Alaris PC Unit 8015 utilizing 802.11 b/g or 802.11 a/b/g… recallZ-1192-2023 · CareFusion 303, Inc. · initiated Jan 24, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems. Users with interfaces to electronic medication record or admin discharge transfer, may be unable to receive/send data; infusions may need to be manually programmed
FDA root cause category: Software Design Change
What the company is doing
On 01/24/23, notices were mailed to customers, who were told to do the following: 1) Inspect the back of the PC Unit and observe the wireless card. Wireless cards that protrude out of the PC Unit are impacted. 2) Software version can be seen on the PCU using the following steps: a) Press Options Key, b) Press Page Down soft key, c) Press Software Versions soft key to display Software Versions menu, d) Press View soft key for the PCU. To prevent this issue from occurring, do one of the following: 1. Exclude IP address ranges ending in x.x.x.224 to x.x.x.255 from all the DHCP scopes where the affected PC Units are being utilized. 2. Assign the affected PC Units IP addresses utilizing DHCP reservations to prevent the PC Unit from obtaining an IP address in the aforementioned range. 3. Assign the affected PC Units a static IP address outside of the aforementioned range. Actions for Clinical Users: Check if wireless icon is illuminated. If the PC Unit is not connected to the wireless network when using Interoperability, the clinical user can input all infusion parameters manually. In addition, the clinical end user can manually document the infusion in the EMR. Data sets can also manually be programmed or uploaded to the PC unit using a serial cable. Continue to rely on hospital established protocols for wireless downtime procedures. Customers who require additional assistance can contact: Customer Order Management, 1-800-482-4822, [email protected]; Technical Support, 1-866-488-1408, [email protected]; Clinical & Pharmacy Support, 1-858-617-1316, [email protected]; BD Designated Complaint Handling Unit, 1-888-812-3266, [email protected] Complete and return the customer response form to [email protected]
Recall details
- Recall number
- Z-1192-2023
- Status
- Open, Classified
- Initiated by the company
- Jan 24, 2023
- Posted by the FDA
- Mar 1, 2023
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- 83,185
- Product code
- FRN
- FDA submissions
- K072105
- Distribution
- US: FL, CA, OR, IL, WI, NJ, AZ, NV, CO, WY, NE, TX, MI, VA, OH, NC, LA, NM, AL, TN, AR, PA, CT, IN, KS, GA, SC, MD, UT, MA, DC, OK, IA, NY, WA, SD, ND, MN, KY, MO, WV, MS, DE, ME, MT, ID, NH, HI, AK, VT OUS: Canada, Qatar, Saudi Arabia,, Kuwait, United Arab Emirates, Israel, South Africa, Zimbabwe, Botswana, Australia, New Zealand
Product description
Alaris PC Unit 8015 utilizing 802.11 b/g or 802.11 a/b/g wireless network cards, with IP addresses from x.x.x.224 to x.x.x.255
Code information (models, lots, serial numbers)
All PC Units, UDI-DI: 10885403801549, 10885403801532, with software versions 12.1.0, 12.1.2, 12.1.3, utilizing affected network cards and IP addresses
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Alaris 8015Carefusion 303, Inc. · Pump, infusion3 recalls6 deaths13,022 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion4 recalls4,302 reports
Product lines
Other devices under the same FDA submission
6 devices cleared under K072105 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Alaris 8100Carefusion 303, Inc. · Pump, infusion4 recalls78 deaths344,514 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion3 recalls1 death6,859 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion1 recall748 reports
- Alaris 8120Carefusion 303, Inc. · Pump, infusion2 recalls70 reports
- Alaris 9601Carefusion 303, Inc. · Pump, infusion32 reports
- Alaris 8975Carefusion 303, Inc. · Pump, infusion27 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page