RESTORIS recallZ-1385-2023 · Howmedica Osteonics Corp. · initiated Mar 10, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Ccertain lots of the Restoris MultiCompartmental Knee (MCK) Tibial Baseplate Trials contain product with incorrect size laser marking
FDA root cause category: Process change control
What the company is doing
Stryker issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 03/10/2023 by letter. The notice explained the issue and the hazard and requested the following actions be taken: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the affected products in Table 1. 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and email to [email protected]/ fax (888) 266-7910. 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. 6. Please return ALL affected product to: Stryker Orthopaedics/PFA Product Return. Stryker issued an URGENT MEDICAL DEVICE RECALL UPDATE notice to its consignees on 04/12/2023. The letter contained the same information, but added a section titled Recommendations for patients already treated with an affected device, and advised patients treated with an affected product should continue to be followed per the normal protocol established by his or her surgeon(s).
Recall details
- Recall number
- Z-1385-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 10, 2023
- Posted by the FDA
- Apr 12, 2023
- Recalling firm
- Howmedica Osteonics Corp.
- Quantity
- 52 units
- Product code
- HSX
- FDA submissions
- K082172, K090763
- Distribution
- U.S., Netherlands, Japan, China, India, Korea, and Australia.
Product description
stryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 5 RM/LL), REF 170615, intended for use in a Partial Knee Arthroplasty surgical procedure
Code information (models, lots, serial numbers)
GTIN 00848486003746, Lot Number 26270421
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- RESTORIS 170615Mako Surgical Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
Product lines
Other devices under the same FDA submission
67 devices cleared under K082172, K090763 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- RESTORIS 170616Mako Surgical Corp. · PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- RESTORIS 180606Mako Surgical Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER9 reports
- RESTORIS 180614Mako Surgical Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER9 reports
- RESTORIS 180615Mako Surgical Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER8 reports
- RESTORIS 180704-1Mako Surgical Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER8 reports
- RESTORIS 160405Mako Surgical Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER7 reports
- RESTORIS 180604Mako Surgical Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER6 reports
- RESTORIS 180605Mako Surgical Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER6 reports
- RESTORIS 180613Mako Surgical Corp. · PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER6 reports
- RESTORIS 180602Mako Surgical Corp. · PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER5 reports
- RESTORIS 180603Mako Surgical Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER5 reports
- RESTORIS 160430Mako Surgical Corp. · PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER5 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91913).
- Z-1386-2023 · Open, Classifiedstryker MAKO RESTORIS MCK BASEPLATE TRIAL (SIZE 6 RM/LL), REF 170616, intended for use in a Partial Knee Arthroplasty surgical procedure
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.