Azurion recallZ-1647-2023 · Philips Medical Systems Nederland B.V. · initiated Mar 21, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure

FDA root cause category: Under Investigation by firm

What the company is doing

Philips issued Urgent Medical Device Letter to US consignees via e-mail on March 21, 2023 and certified email on March 29, 2023. Letters outside of the U.S.A. will be distributed through the local Philips Organizations. Letter states reason for recall, health risk and action to take: Keep this Medical Device Correction with the documentation of the system until Philips corrects your system. Circulate this notice to all users of the system so that they are aware of the issue. Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Medical Device Correction. US: Please return this signed form via email to: [email protected] or alternatively via regular mail to Philips at 222 Jacobs Street, Cambridge, MA 02141 for the attention of Mr. Roland Telson / 3rd Floor Philips is working on a software release that will correct this issue (reference FCO72200528). In the interim, until this software is available and installed in your affected system(s), Philips will be removing the Log Trace Files from the affected systems to free up disk capacity (reference FCO72200529). Philips will be prioritizing these activities based on the time the affected software release has been installed in the Philips Azurion system. You will be contacted by your local Philips representative to schedule these activities. If you need any further information or support concerning this issue, contact your local Philips representative.

Recall details

Recall number
Z-1647-2023
Status
Open, Classified
Initiated by the company
Mar 21, 2023
Posted by the FDA
Sep 30, 2024
Recalling firm
Philips Medical Systems Nederland B.V.
Quantity
USA: 1063 R.O.W. 2136
Product code
OWB
FDA submissions
K181830
Distribution
Nationwide Foreign: Algeria Argentina Armenia Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Cambodia Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cyprus Czech Republic Denmark Dominican Rep Ecuador Egypt Estonia Finland France French Polynesia Germany Greece Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Latvia Lebanon Lithuania Macao Macedonia Malaysia Mexico Morocco Myanmar Nepal Netherlands New Zealand Nicaragua Nigeria Norway Oman Pakistan Palestine Palestine, State of Panama Peru Philippines Poland Portugal Puerto Rico Qatar Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand Tunisia Turkey Turkmenistan Ukraine United Kingdom Utd.Arab.Emir. Uzbekistan Vietnam Yemen

Product description

Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1) Azurion 3M12: 722063, 722221; (2) Azurion 3M15: 722064, 722222; (3) Azurion SM12: 722227; (4) Azurion SM20: 722228; (5) Azurion 7B12/12: 722067, 722225; (6) Azurion 7820/15: 722068, 722226; (7) Azurion 7M12: 722078, 722223; (8) Azurion 7M20 : 722079, 722224

Code information (models, lots, serial numbers)

UDI-DI: System product name Model number UDI (1)Azurion 3M12: 722063 (01)00884838085275(21) 722221 (01)00884838099203(21); (2) Azurion 3M15: 722064 (01)00884838085282(21) 722222 (01)00884838099210(21); (3) Azurion SM12: 722227 (01)00884838099227(21); (4) Azurion SM20: 722228 (01)00884838099234(21); (5) Azurion 7B12/12: 722067 (01)00884838085350(21) 722225 (01)00884838099265(21); (6) Azurion 7820/15: 722068 (01)00884838085367(21) 722226 (01)00884838099272(21); (7) Azurion 7M12: 722078 (01)00884838085251(21) 722223 (01)00884838099241(21); (8) Azurion 7M20: 722079 (01)00884838085268(21) 722224 (01)00884838099258(21), Software: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6