Azurion Azurion 5 M20 (REF 722228)Recalls & reported problems

Philips Medical Systems Nederland B.V. · Monoplane Interventional X-Ray System. A ceiling or floor-mounted monoplane system with a 20-inch flat detector. · Class 2

All models of Azurion (22)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

22 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Philips Azurion 5 M20 X-ray system has 202 adverse event reports, with 3 injuries and no deaths. Many describe the system failing to activate or to output radiation, and some describe positioning failures. It has 22 open recalls and a few closed ones, covering problems that can stop imaging or table movement, such as software bugs, aging hard drives and worn table controls. Reports fell to 34 in the last 12 months from 102 the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
201
Deaths reported
0
Injuries reported
3
Recalls naming it
24

Summary

The FDA's MAUDE database contains 201 adverse event reports that list this exact device (DI 00884838099234), received between Feb 9, 2022 and Aug 4, 2026 — classified as 3 injuries, 198 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (87 reports).

24 FDA recall records name this device's identifier, 22 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 7 reportsJan 2024: 2 reportsFeb 2024: 5 reportsMar 2024: 2 reportsApr 2024: 3 reportsMay 2024: 3 reportsJun 2024: 2 reportsJul 2024: 3 reportsAug 2024: 6 reportsSep 2024: 6 reportsOct 2024: 3 reportsNov 2024: 5 reportsDec 2024: 8 reportsJan 2025: 5 reportsFeb 2025: 7 reportsMar 2025: 4 reportsApr 2025: 6 reportsMay 2025: 10 reportsJun 2025: 10 reportsJul 2025: 12 reportsAug 2025: 9 reportsSep 2025: 19 reportsOct 2025: 7 reportsNov 2025: 9 reportsDec 2025: 7 reportsJan 2026: 3 reportsFeb 2026: 6 reportsMar 2026: 2 reportsApr 2026: 3 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202619
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 16, 2026Class IIOpen, Classified

    Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications

    Z-2931-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Jun 17, 2026Class IIOpen, Classified

    Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.

    Z-2892-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Apr 30, 2026Class IIOpen, Classified

    It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

    Z-2258-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Apr 28, 2026Class IIOpen, Classified

    Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.

    Z-2225-2026 · Root cause: Device Design · Recall details · FDA record ↗

  5. Mar 31, 2026Class IIOpen, Classified

    Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.

    Z-2170-2026 · Root cause: Device Design · Recall details · FDA record ↗

  6. Mar 19, 2026Class IIOpen, Classified

    Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.

    Z-1805-2026 · Root cause: Labeling design · Recall details · FDA record ↗

  7. Mar 3, 2026Class IIOpen, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1680-2026 · Root cause: Device Design · Recall details · FDA record ↗

  8. Jan 26, 2026Class IIOpen, Classified

    Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

    Z-1351-2026 · Root cause: Device Design · Recall details · FDA record ↗

  9. Dec 12, 2025Class IIOpen, Classified

    Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.

    Z-1113-2026 · Root cause: Software design · Recall details · FDA record ↗

  10. Dec 12, 2025Class IIOpen, Classified

    Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.

    Z-1091-2026 · Root cause: Software design · Recall details · FDA record ↗

  11. Aug 1, 2025Class IIOpen, Classified

    Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.

    Z-2502-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  12. May 28, 2025Class IIOpen, Classified

    A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).

    Z-1969-2025 · Root cause: Software design · Recall details · FDA record ↗

  13. Jan 13, 2025Class IIOpen, Classified

    Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

    Z-1090-2025 · Root cause: Device Design · Recall details · FDA record ↗

  14. Jan 13, 2025Class IOpen, Classified

    Possibility of the patient falling from the table related to the mattress used on the patient table.

    Z-1156-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  15. Aug 22, 2024Class IIOpen, Classified

    Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.

    Z-3165-2024 · Root cause: Device Design · Recall details · FDA record ↗

  16. Feb 6, 2024Class IIOpen, Classified

    Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.

    Z-1177-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  17. Jan 2, 2024Class IIOpen, Classified

    Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.

    Z-1152-2024 · Root cause: Process design · Recall details · FDA record ↗

  18. Dec 12, 2023Class IIOpen, Classified

    System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the procedure.

    Z-0835-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  19. Dec 7, 2023Class IIOpen, Classified

    During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.

    Z-2287-2026 · Root cause: Software design · Recall details · FDA record ↗

  20. Nov 3, 2023Class IIOpen, Classified

    Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose

    Z-0559-2024 · Root cause: Device Design · Recall details · FDA record ↗

  21. Mar 21, 2023Class IIOpen, Classified

    Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure

    Z-1647-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  22. Feb 2, 2022Class IIOpen, Classified

    Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.

    Z-0874-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  23. Nov 12, 2021Class IITerminated

    (1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not executed correctly and movement functionality is not available

    Z-0471-2022 · Root cause: Software design · Recall details · FDA record ↗

  24. Sep 24, 2021Class IITerminated

    When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.

    Z-0213-2022 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Failure87
  • Radiation Output Failure64
  • Positioning Failure31
  • No Display/Image17
  • Failure to Power Up3
  • Fire1
  • Radiation Output Problem1
  • Unexpected/Unintended Radiation Output1

Patient problems

  • No Clinical Signs, Symptoms or Conditions199
  • Insufficient Information4
  • Hemodynamic instability1
  • Needle Stick/Puncture1

Recent adverse event reports

The 25 most recent of 201 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 4, 2026Radiation Output Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50125212 ↗
  • MalfunctionMay 28, 2026Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT DURING A VASCULAR PROCEDURE, THE FOOTSWITCH DID NOT ENABLE FLUOROSCOPY OR EXPOSURE. THE SYSTEM WAS IN CLINICAL USE DURING THE EVENT, AND THE PROCEDURE WAS ABORTED AND RESCHEDULED. THERE WERE NO REPORTS OF INJURY TO THE PATIENT OR USER. AN INVESTIGATION INTO THIS COMPLAINT HAS BEEN INITIATED BY PHILIPS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25319051 ↗
  • MalfunctionMay 12, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25147383 ↗
  • MalfunctionApr 16, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT. THE SYSTEM WAS NOT IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24903755 ↗
  • MalfunctionApr 14, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24879088 ↗
  • MalfunctionApr 13, 2026Positioning Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S C-ARM WAS UNABLE TO PERFORM GEOMETRY MOVEMENTS. NO HARM WAS REPORTED TO PHILIPS. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24861129 ↗
  • MalfunctionMar 31, 2026No Display/Image

    IT WAS REPORTED TO PHILIPS THAT A SYSTEM MONITOR WAS DEFECTIVE, WHICH CAN IMPACT IMAGING. NO HARM WAS REPORTED TO PHILIPS. WE ARE CONSERVATIVELY REPORTING THIS EVENT AS THE INVESTIGATION IS ONGOING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24746657 ↗
  • MalfunctionMar 12, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM CRASHED DURING BOOT, WHICH MAY INDICATE A BOOTING ISSUE. NO HARM WAS REPORTED TO PHILIPS. THE DEVICE WAS OUTSIDE CLINICAL USE AT THE TIME OF REPORTED EVENT. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24581001 ↗
  • MalfunctionFeb 24, 2026Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM GAVE A REMOTE ALERT INDICATING THE FLAT DETECTOR HAD AN AURORA LINK ERROR, WHICH CAN IMPACT IMAGING. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24428439 ↗
  • MalfunctionFeb 20, 2026Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S FLAT DETECTOR CONTROLLER AND FLAT DETECTOR WERE NOT WORKING, WHICH CAN IMPACT IMAGING. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24401849 ↗
  • MalfunctionFeb 9, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24300018 ↗
  • MalfunctionFeb 4, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24256075 ↗
  • MalfunctionFeb 3, 2026Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM GAVE AN ALERT INDICATING "TUBE CURRENT OUT OF RANGE", WHICH CAN IMPACT IMAGING. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24249227 ↗
  • MalfunctionFeb 2, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S FLEXVISION COMPUTER WAS NOT AVAILABLE, WHICH CAN IMPACT A FULL SYSTEM BOOT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24232185 ↗
  • MalfunctionJan 29, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S FLEXVISION COMPUTER WAS UNABLE TO BOOT UP, PREVENTING IMAGING. THE PATIENT WAS MOVED TO ANOTHER SYSTEM. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24218191 ↗
  • MalfunctionJan 23, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO FULLY BOOT, STALLING AT PHILIPS LOGO SCREEN. THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24164539 ↗
  • MalfunctionJan 23, 2026Positioning Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM GEOMETRY MOVEMENTS WERE NOT AVAILABLE.THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS.PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24159245 ↗
  • MalfunctionDec 22, 2025Radiation Output Failure; Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM GAVE A REMOTE ALERT INDICATING THE COLLIMATOR FILTER HARDWARE OR MECHANICS WERE DEFECTIVE, WHICH CAN IMPACT IMAGING. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23871785 ↗
  • MalfunctionDec 17, 2025Radiation Output Failure; Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM GAVE AN ALERT INDICATING X-RAY TUBE PROBLEM, ONLY LOW-LOAD FLUOROSCOPY POSSIBLE, IMPACTING IMAGING. NO HARM WAS REPORTED TO PHILIPS. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23830571 ↗
  • MalfunctionDec 11, 2025Radiation Output Failure; Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S WIRELESS FOOTSWITCH WAS INOPERATIVE, WHICH CAN IMPACT IMAGING. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23775531 ↗
  • MalfunctionDec 11, 2025Radiation Output Failure; Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S WIRED FOOTSWITCH WAS DAMAGED, WHICH CAN IMPACT IMAGING. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23775460 ↗
  • MalfunctionDec 9, 2025Radiation Output Failure; Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE COLLIMATOR SHUTTERS WERE CLOSED, WHICH CAN IMPACT IMAGING. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23754034 ↗
  • MalfunctionDec 5, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23727570 ↗
  • MalfunctionDec 1, 2025Failure to Power Up; Failure to Power Up

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM STARTUP WAS STUCK AND COULD NOT ENTER THE APPLICATION INTERFACE. THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23678406 ↗
  • MalfunctionNov 27, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM GEOMETRY MOVEMENTS WERE NOT AVAILABLE. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23664271 ↗

Other Interventional Fluoroscopic X-Ray System devices

All Interventional Fluoroscopic X-Ray System devices →