Custom Sterile and Non-Sterile Procedural Kits containing… recallZ-2297-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated May 15, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Medline notified customers beginning on 05/15/2023, via mail and email with letter. Customers with Medline Probe Cover Kits were instructed to check inventory for affected kits, destroy any affected kits, and complete and return the response form. Distributors were instructed to notify their customers and have them document and destroy affected kits. Customers with the non-sterile and sterile procedural kits were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits. Distributors were instructed to notify customers of the recall and have customers return any affected product for return to Medline. Medline identified additional units of Medline Probe Cover Kits and notified customers on 06/22/2023. Customers were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits.
Recall details
- Recall number
- Z-2297-2023
- Status
- Open, Classified
- Initiated by the company
- May 15, 2023
- Posted by the FDA
- Aug 3, 2023
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 97,677 kits
- Product code
- OES
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, China, Costa Rica, India, Lebanon, Panama, Qatar, Saudi Arabia, Singapore, United Arab Emirates.
Product description
Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. (13) STC IVC FILTER/ANGIOGRAM, Model Number: DYNJ908249A; b. 16CM 3L 7FR CVC INSERTION BUNDLE, Model Numbers: ECVC7410A; c. 16CM 7F 3L CVC MAX BARRIER INSERTION, Model Number: ECVC2205A; d. 16CM 7F CVC INSERTION TRAY, Model Number: ECVC8095; e. 16CM CVC BUNDLE W/MICROPUNCTURE KIT, Model Number: ECVC5010A; f. 16CM CVC INSERTION KIT, Model Number: ECVC6800A; g. 16CM TRIPLE LUMEN 7FR INSERT BUNDLE, Model Number: ECVC2555; h. 16CM TRIPLE LUMEN CVC INSERT BUNDLE, Model Number: ECVC4125; i. 20CM 3L 7FR CVC INSERTION BUNDLE, Model Number: ECVC5135B; j. 20CM 3L CVC INSERTION TRAY, Model Number: ECVC7325; k. 20CM 7F 3L CVC MAX BARRIER INSERTION, Model Number: ECVC1970B; l. 20CM 7F CVC INSERTION TRAY, Model Number: ECVC4335; m. 20CM CVC TRIPLE LUMEN SAFETY BUNDLE, Model Number: ECVC7160; n. 20CM TRIPLE LUMEN 7FR INSERT BUNDLE, Model Number: ECVC2560; o. 20CM, 3L INSERT BUNDLE-CINCINNATI VA, Model Number: ECVC3505; p. 2L 7FR 20CM CVC INSERTION TRAY, Model Number: ECVC7075; q. 2L 8.5FR 16CM CVC INSERTION KIT, Model Number: ECVC8055; r. 3L 7FR 16 CM CVC INSERTION TRAY, Model Number: ECVC8165; s. 3L 7FR 16CM CVC INSERTION BUNDLE, Model Numbers: ECVC4635A, ECVC7040A, ECVC7430, ECVC7475, ECVC7605, ECVC7675, ECVC8130, STCVC2055US; t. 3L 7FR 20CM CVC INSERTION BUNDLE, Model Number: ECVC7600; u. 3L 7FR 20CM CVC INSERTION BUNDLE, Model Numbers: ECVC4640A, ECVC6490A, ECVC6845, ECVC7470; v. 3L 7FR 20CM CVC INSERTION TRAY, Model Number: ECVC6170; w. 7 FR 20 CM CVC TRIPLE LUMEN BUNDLE, Model Number: ECVC3545; x. 7 FR 20CM CVC TRIPLE LUMEN BUNDLE, Model Number: ECVC2440B; y. 7F 20CM 3L CVC VANTEX BUNDLE, Model Number: ECVC7350; z. 7F, 16cm, 3L MAX BARRIER CVC BUNDLE, Model Number: ECVC5480; aa. 7F, 20CM TRIPLE LUMEN CVC BUNDLE, Model Number: ECVC3475; bb. 7FR 16 CM TRIPLE LUMEN CVC INSERTION KIT, Model Number: ECVC2770; cc. 7FR 16CM TRIPLE LUMEN CVC BUNDLE, Model Number: ECVC6370; dd. 7FR 3L 16CM CVC INSERTION BUNDLE, Model Number: ECVC7910; ee. 7FR 3L 20CM CVC INSERTION BUNDLE, Model Numbers: ECVC7235, ECVC7905; ff. 7FR TRIPLE LUMEN CVC INSERTION BUNDLE, Model Number: ECVC2795; gg. 7FR,20CM,3L ANTIMICROBIAL CVC BUNDLE, Model Number: ECVC705; hh. 8.5F 16CM 2L CVC VANTEX BUNDLE, Model Number: ECVC7355; ii. 9F AVA HF RESUSCIT CVC BNDL, Model Number: ECVC6640; jj. ABLATION EP PACK, Model Number: DYNJ43851B; kk. ABLATION PACK, Model Number: DYNJ40629C; ll. ADULT CARDIAC CATH PACK, Model Numbers: DYNJ23455K, DYNJ23455L, DYNJ23455M, DYNJ23455N; mm. ANGIO, Model Numbers: DYNJ902507C, DYNJ905151B; nn. ANGIO CATH COMBINED PACK, Model Numbers: DYNJ43609O, DYNJ43609P, DYNJ43609R, DYNJ43609S; oo. ANGIO CATH LAB PACK-LF, Model Numbers: DYNJ0313233V, DYNJ0313233W, DYNJ0313233X; pp. ANGIO CATH PACK, Model Number: DYNJ52000B; qq. ANGIO DRAPE PACK, Model Numbers: DYNJ44320G, DYNJ44320I, DYNJ44320J; rr. ANGIO DRAPE PACK (ADLUG)642-LF, Model Numbers: DYNJ47665D, DYNJ47665F; ss. ANGIO HOLLAND PACK-LF, Model Numbers: DYNJ0037936C, DYNJ0037936D; tt. ANGIO PACK, Model Numbers: DYNJ42903C, DYNJ42903D, DYNJ43661C, DYNJ44642B, DYNJ58111D, DYNJ58111F, DYNJ69200, DYNJ69951; uu. ANGIO PACK-LF, Model Number: DYNJ0407399P; vv. ANGIO SAVANNAH PACK, Model Number: DYNJ69944; ww. ANGIO TRAY, Model Number: DYNJ23438M, DYNJ52121A, DYNJ69421, DYNJ69792, DYNJ82999, DYNJ82999A; xx. ANGIO/PICC PACK, Model Number: DYNJ36049A; yy. ANGIOGRAM, Model Number: DYNJ0149444N; zz. ANGIOGRAPHIC TRAY, Model Number: DYNJ67320A; aaa. ANGIOGRAPHY DRAPE PACK, Model Numbers: 00-400305R, DYNJ49621C, DYNJ63060B, DYNJ63684G, DYNJ68065B, DYNJ63684G; bbb. ANGIOGRAPHY PACK, Model Numbers: DYNJ44293F, DYNJ45059B, DYNJ55252C, DYNJ65190D, DYNJ65190F, DYNJ66476A, DYNJ67144A; ccc. ANGIOGRAPHY PACK-LF, Model Numbers: DYNJ0290716M, DYNJ0854485V; ddd. ANGIOPLASTY PACK, Model Number: DYNJ44700J; eee
Code information (models, lots, serial numbers)
Show the full code information (10k characters)
All lot numbers for the following models and corresponding UPCs distributed from Medline between 04/01/2020 and 04/28/2023: a. Model Number: DYNJ908249A, UPC Number: 019532724050. b. Model Number: ECVC7410A, UPC Number: 10653160353645. c. Model Number: ECVC2205A, UPC Number: 10653160357148. d. Model Number: ECVC8095, UPC Number: 10653160357759. e. Model Number: ECVC5010A, UPC Number: 10653160335290. f. Model Number: ECVC6800A, UPC Number: 10653160355694. g. Model Number: ECVC2555, UPC Number: 10653160336570. h. Model Number: ECVC4125, UPC Number: 10653160339502. i. Model Number: ECVC5135B, UPC Number: 10653160353652. j. Model Number: ECVC7325, UPC Number: 10653160345589. k. Model Number: ECVC1970B, UPC Number: 10653160357155. l. Model Number: ECVC4335, UPC Number: 10653160357285. m. Model Number: ECVC7160, UPC Number: 10653160335306. n. Model Number: ECVC2560, UPC Number: 10653160334286. o. Model Number: ECVC3505, UPC Number: 10653160334767. p. Model Number: ECVC7075, UPC Number: 10653160335122. q. Model Number: ECVC8055, UPC Number: 10653160355656. r. Model Number: ECVC8165, UPC Number: 10653160358749. s. Model Number: ECVC4635A, UPC Number: 10653160352112; Model Number: ECVC7040A, UPC Number: 10653160344414; Model Number: ECVC7430, UPC Number: 10653160346524; Model Number: ECVC7475, UPC Number: 10653160347569; Model Number: ECVC7605, UPC Number: 10653160347743; Model Number: ECVC7675, UPC Number: 10653160348122; Model Number: ECVC8005, UPC Number: 10653160355137; Model Number: ECVC8130, UPC Number: 10653160358343; Model Number: STCVC2055US, UPC Number: 10653160355564. t. Model Number: ECVC7600, UPC Number: 10653160347736. u. Model Number: ECVC4640A, UPC Number: 10653160352105; Model Number: ECVC6490A, UPC Number: 10653160344360; Model Number: ECVC6845, UPC Number: 10653160339366; Model Number: ECVC7470, UPC Number: 10653160347583. v. Model Number: ECVC6170, UPC Number: 10653160333395. w. Model Number: ECVC3545, UPC Number: 10653160334774. x. Model Number: ECVC2440B, UPC Number: 10653160358930. y. Model Number: ECVC7350, UPC Number: 10653160346265. z. Model Number: ECVC5480, UPC Number: 10653160339458. aa. Model Number: ECVC3475, UPC Number: 10653160334736. bb. Model Number: ECVC2770, UPC Number: 10653160332671. cc. Model Number: ECVC6370, UPC Number: 10653160333425. dd. Model Number: ECVC7910, UPC Number: 10653160357254. ee. Model Number: ECVC7235, UPC Number: 10653160341703; Model Number: ECVC7905, UPC Number: 10653160356912. ff. Model Number: ECVC2795, UPC Number: 10653160334200. gg. Model Number: ECVC705, UPC Number: 10653160331889. hh. Model Number: ECVC7355, UPC Number: 10653160346272. ii. Model Number: ECVC6640, UPC Number: 10653160333494. jj. Model Number: DYNJ43851B, UPC Number: 019532720032. kk. Model Number: DYNJ40629C, UPC Number: 019348999731. ll. Model Number: DYNJ23455K, UPC Number: 019348956435; Model Number: DYNJ23455L, UPC Number: 019348956435; Model Number: DYNJ23455M, UPC Number: 019348956435; Model Number: DYNJ23455N, UPC Number: 019532723775. mm. Model Number: DYNJ902507C, UPC Number: 019532713788; Model Number: DYNJ905151B, UPC Number: 019532721996. nn. Model Number: DYNJ43609O, UPC Number: 019348946871; Model Number: DYNJ43609P, UPC Number: 019348991453; Model Number: DYNJ43609R, UPC Number: 019532722878; Model Number: DYNJ43609S, UPC Number: 019532732298. oo. Model Number: DYNJ0313233V, UPC Number: 019348956435; Model Number: DYNJ0313233W, UPC Number: 019348956435; Model Number: DYNJ0313233X, UPC Number: 019532725098. pp. Model Number: DYNJ52000B, UPC Number: 019348938498. qq. Model Number: DYNJ44320G, UPC Number: 019348996630; Model Number: DYNJ44320I, UPC Number: 019532707035; Model Number: DYNJ44320J, UPC Number: 019532713235. rr. Model Number: DYNJ47665D, UPC Number: 019348995282; Model Number: DYNJ47665F, UPC Number: 019532735521. ss. Model Number: DYNJ0037936C, UPC Number: 019348956435; Model Number: DYNJ0037936D, UPC Number: 019532712953. tt. Model Number: DYNJ42903C, UPC Number: 019532707966; Model Number: DYNJ42903D, UPC Number: 019532717111; Model Number: DYNJ43661C, UPC Number: 019532710435; Model Number: DYNJ44642B, UPC Number: 019532701142; Model Number: DYNJ58111D, UPC Number: 019348993292; Model Number: DYNJ58111F, UPC Number: 019532722351; Model Number: DYNJ69200, UPC Number: 019348990681; Model Number: DYNJ69951, UPC Number: 019532700030. uu. Model Number: DYNJ0407399P, UPC Number: 019532733686. vv. Model Number: DYNJ69944, UPC Number: 019348999845. ww. Model Number: DYNJ23438M, UPC Number: 019348997060; Model Number: DYNJ52121A, UPC Number: 019532726009; Model Number: DYNJ69421, UPC Number: 019348993759; Model Number: DYNJ69792, UPC Number: 019348997928; Model Number: DYNJ82999, UPC Number: 019532722310; Model Number: DYNJ82999A, UPC Number: 019532731027. xx. Model Number: DYNJ36049A, UPC Number: 019532714720. yy. Model Number: DYNJ0149444N, UPC Number: 019348996671. zz. Model Number: DYNJ67320A, UPC Number: 019348993757. aaa. Model Number: 00-400305R, UPC Number: 019348994428; Model Number: DYNJ49621C, UPC Number: 019532700858; Model Number: DYNJ63060B, UPC Number: 019348991282; Model Number: DYNJ63684G, UPC Number: 019532703722; Model Number: DYNJ68065B, UPC Number: 019532701333. bbb. Model Number: DYNJ44293F, UPC Number: 019532726518; Model Number: DYNJ45059B, UPC Number: 088994213563; Model Number: DYNJ55252C, UPC Number: 019532726791; Model Number: DYNJ65190D, UPC Number: 019532706664; Model Number: DYNJ65190F, UPC Number: 019532709265; Model Number: DYNJ66476A, UPC Number: 019532705189; Model Number: DYNJ67144A, UPC Number: 019532717902. ccc. Model Number: DYNJ0290716M, UPC Number: 019532708886; Model Number: DYNJ0854485V, UPC Number: 019532709688. ddd. Model Number: DYNJ44700J, UPC Number: 019532727741. eee. Model Number: DYNJ43203G, UPC Number: 019348995291; Model Number: DYNJ64655B, UPC Number: 019348996101; Model Number: DYNJ64655C, UPC Number: 019532717289. fff. Model Number: DYNJ43261J, UPC Number: 019532728879; Model Number: DYNJ52774, UPC Number: 088994232247; Model Number: DYNJ53095, UPC Number: 088994237177; Model Number: DYNJ55169D, UPC Number: 019348972168; Model Number: DYNJ55169F, UPC Number: 019348984849. ggg. Model Number: DYNJ62733A, UPC Number: 019348997067. hhh. Model Number: ECVC7840A, UPC Number: 10653160355427. iii. Model Number: STCVC12US, UPC Number: 10653160311829. jjj. Model Number: ECVC3440, UPC Number: 10653160332770. kkk. Model Number: ECVC5640A, UPC Number: 10653160339977. lll. Model Number: ECVC7875, UPC Number: 10653160351535. mmm. Model Number: ECVC5860, UPC Number: 10653160341543. nnn. Model Number: ECVC5615, UPC Number: 10653160341512. ooo. Model Number: ECVC5665, UPC Number: 10653160341505. ppp. Model Number: ECVC4300, UPC Number: 10653160334873. qqq. Model Number: ECVC2600, UPC Number: 10653160334279. rrr. Model Number: ECVC3210, UPC Number: 10653160334071. sss. Model Number: ECVC7070B, UPC Number: 10653160346166. ttt. Model Number: DYNJV0177AF, UPC Number: 019348926312. uuu. Model Number: DYNJ41700A, UPC Number: 088994275695. vvv. Model Number: DYNJ907014A, UPC Number: 019348947402. www. Model Number: DYNJ38866D, UPC Number: 019348956435; Model Number: DYNJ38866F, UPC Number: 019348991486; Model Number: DYNJ67174A, UPC Number: 019532704680; Model Number: DYNJ67174B, UPC Number: 019532713108; Model Number: DYNJ82280, UPC Number: 019532718392; Model Number: DYNJ82296, UPC Number: 019532718420. xxx. Model Number: DYNJ0842644F, UPC Number: 019532728860; Model Number: DYNJ0855138P, UPC Number: 019348956435; Model Number: DYNJ0855138Q, UPC Number: 019532709484; Model Number: DYNJ42996C, UPC Number: 019348998843. yyy. Model Number: DYNJ45323F, UPC Number: 019532708236. zzz. Model Number: ECVC2660, UPC Number: 10653160338178. aaaa. Model Number: DYNJ64050A, UPC Number: 019348994382; Model Number: DYNJ63684G, UPC Number: 019532703722. bbbb. Model Number: DYNJ66182, UPC Number: 019348932098. cccc. Model Number: DYNJ33709F, UPC Number: 088994271392. dddd. Model Number: DYNJ33739G, UPC Number: 088994271393. eeee. Model Number: DYNJ55680A, UPC Number: 088994263978. ffff. Model Number: DYNJ68763, UPC Number: 019348983759; Model Number: DYNJ68763A, UPC Number: 019532726478. gggg. Model Number: DYNJ69864, UPC Number: 019348999048. hhhh. Model Number: DYNJ40926A, UPC Number: 019532713721; Model Number: DYNJ55455A, UPC Number: 019348938623; Model Number: DYNJ60763, UPC Number: 019348952796; Model Number: DYNJ66166A, UPC Number: 019348939599. iiii. Model Number: DYNJ67895, UPC Number: 019348945396. jjjj. Model Number: DYNJ57723B, UPC Number: 019348951627. kkkk. Model Number: DYNJ80269, UPC Number: 019532701780. llll. Model Number: DYNJ60329, UPC Number: 088994297098. mmmm. Model Number: DYNJ51338C, UPC Number: 019348924171; Model Number: DYNJ56547C, UPC Number: 019348987983; Model Number: DYNJ59632D, UPC Number: 019532703407; Model Number: DYNJ63181A, UPC Number: 019532704548; Model Number: DYNJ63181B, UPC Number: 019532712721; Model Number: DYNJ63181C, UPC Number: 019532724200; Model Number: DYNJ66473A, UPC Number: 019348991777; Model Number: DYNJ83089, UPC Number: 019532723083; Model Number: DYNJ83105, UPC Number: 019532723151. nnnn. Model Number: ECVC6970, UPC Number: 10653160336693. oooo. Model Number: ECVC360, UPC Number: 10653160332855. pppp. Model Number: ECVC3285, UPC Number: 10653160336105. qqqq. Model Number: ECVC2745, UPC Number: 10653160334224. rrrr. Model Number: ECVC2750, UPC Number: 10653160334217. ssss. Model Number: ECVC1140A, UPC Number: 10653160331964. tttt. Model Number: ECVC2300, UPC Number: 10653160334330. uuuu. Model Number: ECVC2305, UPC Number: 10653160332473. vvvv. Model Number: ECVC1345A, UPC Number: 10653160334606. wwww. Model Number: ECVC1725, UPC Number: 10653160334507. xxxx. Model Number: DYNJ81316, UPC Number: 019532710379. yyyy. Model Number: ECVC5530A, UPC Number: 10653160345596. zzzz. Model Number: DYNJ61082C, UPC Number: 019532711141; Model Number: DYNJ61082D, UPC Number: 019532726199. aaaaa. Model Number: DYNJ0378596I, UPC Number: 10195327188818; Model Number: DYNJ0378596J, UPC Number: 10195327310608.
Devices named in this recall
72 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. ECVC2750Centurion Medical Products Corporation · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. ECVC2745Centurion Medical Products Corporation · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. ECVC2555Centurion Medical Products Corporation · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. ECVC6845Centurion Medical Products Corporation · Cardiac catheterization kit3 recalls0 reports
- MEDLINE ECVC4335Medline Industries, Inc. · Cardiac catheterization kit2 recalls1 report
- Medline Industries, Inc. ECVC2770Centurion Medical Products Corporation · Cardiac catheterization kit2 recalls0 reports
- Medline Industries, Inc. ECVC6370Centurion Medical Products Corporation · General Surgery Tray2 recalls0 reports
- Medline Industries, Inc. ECVC2560Centurion Medical Products Corporation · Cardiac catheterization kit2 recalls0 reports
- Medline Industries, Inc. ECVC1725Centurion Medical Products Corporation · Cardiac catheterization kit2 recalls0 reports
- Medline Industries, Inc. ECVC1345ACenturion Medical Products Corporation · Cardiac catheterization kit2 recalls0 reports
- Medline Industries, Inc. ECVC3505Centurion Medical Products Corporation · Cardiac catheterization kit2 recalls0 reports
- Medline Industries, Inc. ECVC3545Centurion Medical Products Corporation · Cardiac catheterization kit2 recalls0 reports
- Medline Industries, Inc. ECVC4300Centurion Medical Products Corporation · Cardiac catheterization kit2 recalls0 reports
- Medline Industries, Inc. ECVC5010ACenturion Medical Products Corporation · Cardiac catheterization kit2 recalls0 reports
- Medline Industries, Inc. ECVC6970Centurion Medical Products Corporation · Cardiac catheterization kit2 recalls0 reports
- Medline Industries, Inc. ECVC5480Centurion Medical Products Corporation · Cardiac catheterization kit2 recalls0 reports
- MEDLINE INDUSTRIES, INC. ECVC5665Centurion Medical Products Corporation · Cardiac catheterization kit2 recalls0 reports
- MEDLINE ECVC6490AMedline Industries, Inc. · Cardiac catheterization kit2 recalls0 reports
- MEDLINE ECVC5530AMedline Industries, Inc. · Cardiac catheterization kit2 recalls0 reports
- MEDLINE ECVC7675Medline Industries, Inc. · General Surgery Tray2 recalls0 reports
- MEDLINE ECVC4640AMedline Industries, Inc. · Cardiac catheterization kit2 recalls0 reports
- MEDLINE ECVC4635AMedline Industries, Inc. · Cardiac catheterization kit2 recalls0 reports
- MEDLINE ECVC5135BMedline Industries, Inc. · Cardiac catheterization kit2 recalls0 reports
- MEDLINE ECVC6800AMedline Industries, Inc. · Cardiac catheterization kit2 recalls0 reports
- MEDLINE ECVC2205AMedline Industries, Inc. · Cardiac catheterization kit2 recalls0 reports
- MEDLINE ECVC1970BMedline Industries, Inc. · Cardiac catheterization kit2 recalls0 reports
- MEDLINE ECVC8165Medline Industries, Inc. · Cardiac catheterization kit2 recalls0 reports
- MEDLINE ECVC2440BMedline Industries, Inc. · Cardiac catheterization kit2 recalls0 reports
- MEDLINE ECVC8130Medline Industries, Inc. · Cardiac catheterization kit1 recall2 reports
- Medline Industries, Inc. DYNJ0378596IMedline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. DYNJ0378596JMedline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- Centurion STCVC12USCenturion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC705Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC1140ACenturion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC2305Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC3440Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC360Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC6170Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC6640Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC3210Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC2795Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC2600Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC2300Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC3475Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC7075Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7160Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. ECVC3285Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC2660Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC4125Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. ECVC5640ACenturion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- MEDLINE INDUSTRIES, INC. ECVC5615Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- MEDLINE INDUSTRIES, INC. ECVC5860Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7235Centurion Medical Products Corporation · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7040AMedline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7325Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7070BMedline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7350Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7355Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7430Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7475Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7470Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7600Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7605Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7875Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7410AMedline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC8005Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7840AMedline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE STCVC2055USMedline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC8055Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7905Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC7910Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- MEDLINE ECVC8095Medline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92399).
- Z-2290-2023 · Open, ClassifiedMedline Probe Cover Kits containing Turkuaz Ultrasound Gel, Labeled as the following: a. COVER, PROBE, W/GEL AND BANDS, 4X48 - Model Number DYNJE5900; b. COVER, PROBE, W/GEL AN…
- Z-2292-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. ART LINE KIT ICH OR, Model Number: ART900; b. ART LINE SECUREMEN…
- Z-2298-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. CRMC MINOR IR PROCEDURE PK-LF, Model Numbers: DYNJ61099B, DYNJ6109…
- Z-2299-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. MULTI-LUMEN CVC BUNDLE KIT, Model Number: ECVC475; b. MULTI-LUM…
- Z-2304-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. 16CM CENTRAL LINE BUNDLE, Model Number: STCVC2020; b. ARTERIAL…
- Z-2305-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. A V FISTULA, Model Number: DYNJ65711; b. ACCESS CATHETER, Mode…
- Z-2307-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. ABLATION/DRAINAGE PACK, Model Number: DYNJ53658B; b. BASIC BIO…
- Z-2313-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. DIALYSIS ACCESSORY PACK W/O CATHETER, Model Number: MNS10750A; …
- Z-2319-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. 7F, 20CM TRIPLE LUMEN CVC BUNDLE, Model Number: ECVC4740; b. A…
- Z-2320-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. DR. ROSEN VEIN PACK, Model Number: DYNJ67517; b. DR. VANBOSSE …
- Z-2321-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. PEDIATRIC CATH ANGIO PACK, Model Number: DYNJ54755B; b. PEDIAT…
- Z-2322-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. 18GX6" FEMORAL ART LINE KIT, Model Number: ART960; b. 20Gx3" R…
- Z-2323-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. ADULT EPIV INSERTION TRAY, Model Number: DYNDA2482; b. ARTERIAL …
- Z-2324-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. CENTRAL LINE INSERT KIT W/O CA, Model Number: CVI3755; b. CENT…
- Z-2325-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. CVC INFANT INSERTION PACK, Model Number: CVI4770A; b. INSERTIO…
- Z-2329-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. PROBE COVER PACK, Model Number: DYNDA1219
- Z-2339-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. BREAST BIOPSY KIT, Model Number: DYNDA1224A, DYNDA1224B; b. BR…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page