Custom Sterile and Non-Sterile Procedural Kits containing… recallZ-2324-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated May 15, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Medline notified customers beginning on 05/15/2023, via mail and email with letter. Customers with Medline Probe Cover Kits were instructed to check inventory for affected kits, destroy any affected kits, and complete and return the response form. Distributors were instructed to notify their customers and have them document and destroy affected kits. Customers with the non-sterile and sterile procedural kits were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits. Distributors were instructed to notify customers of the recall and have customers return any affected product for return to Medline. Medline identified additional units of Medline Probe Cover Kits and notified customers on 06/22/2023. Customers were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits.
Recall details
- Recall number
- Z-2324-2023
- Status
- Open, Classified
- Initiated by the company
- May 15, 2023
- Posted by the FDA
- Aug 3, 2023
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 19,907 kits
- Product code
- MCY
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, China, Costa Rica, India, Lebanon, Panama, Qatar, Saudi Arabia, Singapore, United Arab Emirates.
Product description
Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. CENTRAL LINE INSERT KIT W/O CA, Model Number: CVI3755; b. CENTRAL LINE KIT, Model Number: DYNDC2582, DYNDC2582A; c. CENTRAL LINE PACK-COMPANION, Model Number: DYNJ47215D; d. CENTRAL LINE TRAY, Model Number: DYNJ42902B; e. CENTRAL LINE UNIVERSAL INSERTION, Model Number: CVI3780; f. CLOSURE PACK, Model Number: DYNJ83054A; g. DBD-PICC LINEPACK PCLUI642-LF, Model Number: DYNJ47717A; h. ER CENTRAL LINE KIT, Model Number: P155508B, P155508C; i. HEMODIALYSIS CENTRAL LINE, Model Number: DYNJ63347; j. LINE KIT, Model Number: DYNJ47987A; k. PEDIATRIC CENTRAL LINE KIT, Model Number: DYNJ38885B; l. PK CUST CENTRAL LINE PACK, Model Number: DYNJ60586; m. TRAY LACERATION, Model Number: DYNDL1134A; n. TUNNEL LINE PACK, Model Number: DYNJ67990; o. ULTRASOUND GUIDED IV/MIDLINE K, Model Number: DYNDV2385;
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
All lot numbers for the following models and corresponding UPCs distributed from Medline between 04/01/2020 and 04/28/2023: a. Model Number: CVI3755, UPC Number: 019348981943. b. Model Number: DYNDC2582, UPC Number: 10889942670444; Model Number: DYNDC2582A, UPC Number: 10193489198164. c. Model Number: DYNJ47215D, UPC Number: 088827774302. d. Model Number: DYNJ42902B, UPC Number: 088827779800. e. Model Number: CVI3780, UPC Number: 019348943486. f. Model Number: DYNJ83054A, UPC Number: 019532731051. g. Model Number: DYNJ47717A, UPC Number: 019348945452. h. Model Number: P155508B, UPC Number: 019532709134; Model Number: P155508C, UPC Number: 019532709134. i. Model Number: DYNJ63347, UPC Number: 019348981031. j. Model Number: DYNJ47987A, UPC Number: 088994268310. k. Model Number: DYNJ38885B, UPC Number: 019348946233. l. Model Number: DYNJ60586, UPC Number: 019348952804. m. Model Number: DYNDL1134A, UPC Number: 008019699476. n. Model Number: DYNJ67990, UPC Number: 019348946253. o. Model Number: DYNDV2385, UPC Number: 019348947559.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- MEDLINE DYNDC2582AMedline Industries, Inc. · Wound dressing kit1 recall0 reports
- Medline Industries, Inc. DYNDC2582Medline Industries, Inc. · Wound dressing kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92399).
- Z-2290-2023 · Open, ClassifiedMedline Probe Cover Kits containing Turkuaz Ultrasound Gel, Labeled as the following: a. COVER, PROBE, W/GEL AND BANDS, 4X48 - Model Number DYNJE5900; b. COVER, PROBE, W/GEL AN…
- Z-2292-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. ART LINE KIT ICH OR, Model Number: ART900; b. ART LINE SECUREMEN…
- Z-2297-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. (13) STC IVC FILTER/ANGIOGRAM, Model Number: DYNJ908249A; b. 16…
- Z-2298-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. CRMC MINOR IR PROCEDURE PK-LF, Model Numbers: DYNJ61099B, DYNJ6109…
- Z-2299-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. MULTI-LUMEN CVC BUNDLE KIT, Model Number: ECVC475; b. MULTI-LUM…
- Z-2304-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. 16CM CENTRAL LINE BUNDLE, Model Number: STCVC2020; b. ARTERIAL…
- Z-2305-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. A V FISTULA, Model Number: DYNJ65711; b. ACCESS CATHETER, Mode…
- Z-2307-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. ABLATION/DRAINAGE PACK, Model Number: DYNJ53658B; b. BASIC BIO…
- Z-2313-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. DIALYSIS ACCESSORY PACK W/O CATHETER, Model Number: MNS10750A; …
- Z-2319-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. 7F, 20CM TRIPLE LUMEN CVC BUNDLE, Model Number: ECVC4740; b. A…
- Z-2320-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. DR. ROSEN VEIN PACK, Model Number: DYNJ67517; b. DR. VANBOSSE …
- Z-2321-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. PEDIATRIC CATH ANGIO PACK, Model Number: DYNJ54755B; b. PEDIAT…
- Z-2322-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. 18GX6" FEMORAL ART LINE KIT, Model Number: ART960; b. 20Gx3" R…
- Z-2323-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. ADULT EPIV INSERTION TRAY, Model Number: DYNDA2482; b. ARTERIAL …
- Z-2325-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. CVC INFANT INSERTION PACK, Model Number: CVI4770A; b. INSERTIO…
- Z-2329-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. PROBE COVER PACK, Model Number: DYNDA1219
- Z-2339-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. BREAST BIOPSY KIT, Model Number: DYNDA1224A, DYNDA1224B; b. BR…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page