Custom Sterile and Non-Sterile Procedural Kits containing… recallZ-2323-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated May 15, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Medline notified customers beginning on 05/15/2023, via mail and email with letter. Customers with Medline Probe Cover Kits were instructed to check inventory for affected kits, destroy any affected kits, and complete and return the response form. Distributors were instructed to notify their customers and have them document and destroy affected kits. Customers with the non-sterile and sterile procedural kits were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits. Distributors were instructed to notify customers of the recall and have customers return any affected product for return to Medline. Medline identified additional units of Medline Probe Cover Kits and notified customers on 06/22/2023. Customers were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits.

Recall details

Recall number
Z-2323-2023
Status
Open, Classified
Initiated by the company
May 15, 2023
Posted by the FDA
Aug 3, 2023
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
218,163 kits
Product code
LRS
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Costa Rica, India, Lebanon, Panama, Qatar, Saudi Arabia, Singapore, United Arab Emirates.

Product description

Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. ADULT EPIV INSERTION TRAY, Model Number: DYNDA2482; b. ARTERIAL LINE START KIT, Model Number: ART920; c. CENTRAL LINE INSERTION TRAY, Model Number: DT19810; d. ECMO INSERTION BUNDLE W/O CANNULA, Model Number: CVI4915; e. FLOOR ULTRASOUND GUIDED IV SECUREMENT, Model Number: IVS3770; f. IV SECUREMENT KIT, Model Number: DYNDV2407A; g. MIDLINE BUNDLE 4FR SL, Model Number: MID0005; h. MIDLINE INSERTION TRAY, Model Number: CVI4340; i. MIDLINE KIT 4F SL MAX BARRIER STD, Model Number: MIDST04011; j. MIDLINE KIT 4F SL MB, Model Number: MID0001; k. MIDLINE KIT 4F SL PL MAX BARR, Model Number: DYNJ80141MB, MID0003; l. MIDLINE KIT 4F SL PL MB, Model Number: PICC0009A; m. MIDLINE KIT 5F DL PL MAX BARR, Model Number: DYNJ80152MB; n. NEONATAL PICC INSERTION TRAY, Model Number: DYNDA2509A, MNS8765; o. NEONATAL PICC LINE INSERTION TRAY, Model Number: DYNDC2120B; p. NICU PICC TRAY, Model Number: CVI3740A; q. PED ULTRASOUND GUIDED PIV, Model Number: IVS3990; r. PEDIATRIC PICC INSERTION TRAY, Model Number: CVI4450B; s. PERIPHERAL ULTRASOUND CATHETER INSERTION, Model Number: DYNDV2285, MNS9865; t. PICC BUNDLE 5FR DL, Model Number: PICC0013; u. PICC INSERTION BUNDLE, Model Number: DT14560; v. PICC INSERTION TRAY, Model Number: CVI3805, CVI3820; w. PICC KIT 4F SL PL MAX BARRIER, Model Number: DYNJ70141MB, PICC0010, PICC0011, PICC0012; x. PICC KIT 5F DL PL MB, Model Number: DYNJ70152MB; y. PIV BSI KIT, Model Number: BSIPIV1010; z. PIV KIT, Model Number: BSIPIV1004; aa. POLY MIDLINE KIT 4F SL, Model Number: DYNJ68937A; bb. QUINTEN CATH ACCESSORY KIT, Model Number: CVI3700; cc. ULTRASOUND GUIDED IV KIT, Model Number: IVS3705, IVS3705A, IVS3905A; dd. ULTRASOUND GUIDED IV SECUREMENT SET, Model Number: IVS3350; ee. ULTRASOUND GUIDED IV START KIT, Model Number: DYNDV2465, DYNDV2518, DYNDV2547; ff. ULTRASOUND GUIDED IV START KIT W/SHIELD, Model Number: BSIPIV1006; gg. ULTRASOUND GUIDED IV/MIDLINE KIT, Model Number: DYNDV2385; hh. ULTRASOUND GUIDED PIV, Model Number: IVS4075; ii. ULTRASOUND GUIDED PIV BUNDLE W/CHG DRSG, Model Number: BSIPIV1011; jj. ULTRASOUND GUIDED PIV INSERTION KIT, Model Number: IVS3485A, IVS3840, IVS3840A, IVS3940, IVS3940A, IVS3940B; kk. ULTRASOUND GUIDED PIV KIT, Model Number: BSIPIV1005, DYNDA2669, IVS3300, IVS3420A, IVS3420B, IVS3495, IVS3505, IVS3655B, IVS3900, IVS3900A, IVS4005; ll. ULTRASOUND GUIDED PIV START KIT, Model Number: IVS3715; mm. ULTRASOUND IV START KIT, Model Number: IV8790; nn. ULTRASOUND PIV BUNDLE, Model Number: IV8635A, IV8635B, IV8635C; oo. ULTRASOUND PIV KIT, Model Number: IVS3970; pp. UNIVERSAL BUNDLE-CENTRAL LINE PACK, Model Number: CVI3910; qq. UNIVERSAL LINE INSERTION TRAY, Model Number: DT16125; rr. USG IV START KIT, Model Number: DYNDV2435; ss. USG PIV KIT, Model Number: DYNDV2536, DYNDV2567; tt. USGPIV Insertion Kit, Model Number: MBSIULSGD01; uu. USGPIV KIT, Model Number: DYNDV2578; vv. USPIV KIT, Model Number: IVS3910; ww. VAMC PICC PREP BUNDLE, Model Number: CVI3880; xx. VANTEX 7FR 3L 16CM CVC INSERTION BNDL, Model Number: STCVC2020A; yy. VANTEX 7FR 3L 20CM CVC INSERTION BNDL, Model Number: STCVC2015A; zz. VI PACK, Model Number: DYNJ64871

Code information (models, lots, serial numbers)

Show the full code information (5k characters)

All lot numbers for the following models and corresponding UPCs distributed from Medline between 04/01/2020 and 04/28/2023: a. Model Number: DYNDA2482, UPC Number: 10193489198188. b. Model Number: ART920, UPC Number: 10653160327899. c. Model Number: DT19810, UPC Number: 10653160285656. d. Model Number: CVI4915, UPC Number: 10653160355489. e. Model Number: IVS3770, UPC Number: 10653160328230. f. Model Number: DYNDV2407A, UPC Number: 019348981701. g. Model Number: MID0005, UPC Number: 10653160349181. h. Model Number: CVI4340, UPC Number: 10653160309475. i. Model Number: MIDST04011, UPC Number: 10653160350699. j. Model Number: MID0001, UPC Number: 10193489381627. k. Model Number: DYNJ80141MB, UPC Number: 10889942733415; Model Number: MID0003, UPC Number: 10653160349334. l. Model Number: PICC0009A, UPC Number: 10193489760231. m. Model Number: DYNJ80152MB, UPC Number: 10889942733422. n. Model Number: DYNDA2509A, UPC Number: 10193489196986; Model Number: MNS8765, UPC Number: 10653160270515. o. Model Number: DYNDC2120B, UPC Number: 10193489198270. p. Model Number: CVI3740A, UPC Number: 10653160324416. q. Model Number: IVS3990, UPC Number: 10653160345282. r. Model Number: CVI4450B, UPC Number: 10653160358428. s. Model Number: DYNDV2285, UPC Number: 10193489194548; Model Number: MNS9865, UPC Number: 10653160294436. t. Model Number: PICC0013, UPC Number: 10653160351818. u. Model Number: DT14560, UPC Number: 10653160236887. v. Model Number: CVI3805, UPC Number: 10653160265085; Model Number: CVI3820, UPC Number: 10653160273509. w. Model Number: DYNJ70141MB, UPC Number: 10889942722808; Model Number: PICC0010, UPC Number: 10653160349112; Model Number: PICC0011, UPC Number: 10653160350842; Model Number: PICC0012, UPC Number: 10653160350491. x. Model Number: DYNJ70152MB, UPC Number: 10889942722822. y. Model Number: BSIPIV1010, UPC Number: 10653160356523. z. Model Number: BSIPIV1004, UPC Number: 10193489198843. aa. Model Number: DYNJ68937A, UPC Number: 10193489111545. bb. Model Number: CVI3700, UPC Number: 10653160259763. cc. Model Number: IVS3705, UPC Number: 10653160325819; Model Number: IVS3705A, UPC Number: 10653160342885; Model Number: IVS3905A, UPC Number: 10653160344391. dd. Model Number: IVS3350, UPC Number: 10653160313045. ee. Model Number: DYNDV2465, UPC Number: 10193489110166; Model Number: DYNDV2518, UPC Number: 019532708907; Model Number: DYNDV2547, UPC Number: 10653160354246. ff. Model Number: BSIPIV1006, UPC Number: 10653160356936. gg. Model Number: DYNDV2385, UPC Number: 10193489195217. hh. Model Number: IVS4075, UPC Number: 10653160350545. ii. Model Number: BSIPIV1011, UPC Number: 10653160353317. jj. Model Number: IVS3485A, UPC Number: 10653160328612; Model Number: IVS3840, UPC Number: 10653160341994; Model Number: IVS3840A, UPC Number: 10653160344612; Model Number: IVS3940, UPC Number: 10653160342304; Model Number: IVS3940A, UPC Number: 10653160345664; Model Number: IVS3940B, UPC Number: 10653160355335. kk. Model Number: BSIPIV1005, UPC Number: 10193489198867; Model Number: DYNDA2669, UPC Number: 10193489191912; Model Number: IVS3300, UPC Number: 10653160307662; Model Number: IVS3420A, UPC Number: 10653160342205; Model Number: IVS3420B, UPC Number: 10653160359067; Model Number: IVS3495, UPC Number: 10653160316596; Model Number: IVS3505, UPC Number: 10653160323761; Model Number: IVS3655B, UPC Number: 10653160358107; Model Number: IVS3900, UPC Number: 10653160344810; Model Number: IVS3900A, UPC Number: 10653160347774; Model Number: IVS4005, UPC Number: 10653160346142. ll. Model Number: IVS3715, UPC Number: 10653160326298. mm. Model Number: IV8790, UPC Number: 019348936055. nn. Model Number: IV8635A, UPC Number: 10653160324010; Model Number: IV8635B, UPC Number: 10653160327622; Model Number: IV8635C, UPC Number: 10653160344858. oo. Model Number: IVS3970, UPC Number: 10653160343431. pp. Model Number: CVI3910, UPC Number: 10653160272311. qq. Model Number: DT16125, UPC Number: 10653160254751. rr. Model Number: DYNDV2435, UPC Number: 019532704366. ss. Model Number: DYNDV2536, UPC Number: 10653160353676; Model Number: DYNDV2567, UPC Number: 10653160356455. tt. Model Number: MBSIULSGD01, UPC Number: 10193489199482. uu. Model Number: DYNDV2578, UPC Number: 10653160358374. vv. Model Number: IVS3910, UPC Number: 10653160338871. ww. Model Number: CVI3880, UPC Number: 10653160272755. xx. Model Number: STCVC2020A, UPC Number: 10653160350569. yy. Model Number: STCVC2015A, UPC Number: 10653160350552. zz. Model Number: DYNJ64871, UPC Number: 019348922536.

Devices named in this recall

76 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 92399).