Custom Sterile and Non-Sterile Procedural Kits containing… recallZ-2321-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated May 15, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Medline notified customers beginning on 05/15/2023, via mail and email with letter. Customers with Medline Probe Cover Kits were instructed to check inventory for affected kits, destroy any affected kits, and complete and return the response form. Distributors were instructed to notify their customers and have them document and destroy affected kits. Customers with the non-sterile and sterile procedural kits were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits. Distributors were instructed to notify customers of the recall and have customers return any affected product for return to Medline. Medline identified additional units of Medline Probe Cover Kits and notified customers on 06/22/2023. Customers were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits.
Recall details
- Recall number
- Z-2321-2023
- Status
- Open, Classified
- Initiated by the company
- May 15, 2023
- Posted by the FDA
- Aug 3, 2023
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 233,638 kits
- Product code
- LRO
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, China, Costa Rica, India, Lebanon, Panama, Qatar, Saudi Arabia, Singapore, United Arab Emirates.
Product description
Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. PEDIATRIC CATH ANGIO PACK, Model Number: DYNJ54755B; b. PEDIATRIC MAX BARRIER KIT-LF, Model Number: DYNJ40306A; c. PERC PACK-LF, Model Number: DYNJ0774759P; d. PERCUTANEOUS VASCULAR CDS, Model Number: DYNJ902261D; e. PERIPHERAL NERVE BLOCK KIT/P, Model Number: DYNJRA1747; f. PERIPHERAL PACK, Model Number: DYNJ44960D; g. PERIPHERAL ULTRASOUND CATHETER, Model Number: DYNDV2285, MNS9865; h. PERIPHERAL VASCULAR GRAFT PACK, Model Number: DYNJ904401B; i. PERMACATH-LF, Model Number: DYNJ42262D; j. PHLEBECTOMY PACK, Model Number: DYNJ47502B; k. PHYSIATRY PACK, Model Number: DYNJ67414; l. PICC INSERT KIT-LESS THAN 3 FR, Model Number: PHS41739A; m. PICC INSERTION PACK, Model Number: DYNJ60038; n. PICC LINE DOUBLE LUMEN CDS, Model Number: CDS982858F; o. PICC LINE INSERTION, Model Number: DYNJ39885A; p. PICC LINE INSERTION WRO-LF, Model Number: DYNJ41873B; q. PICC LINE TRAY, Model Number: DYNJ52688A; r. PICC LINE TRIPLE LUMEN CDS, Model Number: CDS982859F; s. PICC LINEPACK (PCLUI)642-LF, Model Number: DYNJ47717; t. PICC PACK, Model Number: DYNJ55723A, DYNJ60990; u. PICC PROCEDURE PACK, Model Number: DYNJ59067; v. PICC TRAY, Model Number: DYNJ49177F; w. PIV WITH ULTRASOUND KIT, Model Number: DYNDC2399A; x. PK GEN-RECIPIENT, Model Number: DYNJ41114C; y. PK GEN-TRANSPLANT, Model Number: DYNJ54894A; z. PK, GEN-TRANSPLANT, Model Number: DYNJ54894A; aa. PK, GEN-TRANSPLANT-DONOR, Model Number: DYNJ54982A; bb. PORT INSERTION PACK, Model Number: CVI4765; cc. PORT PACK, Model Number: DYNJ56857B, DYNJ68990; dd. POWERGLIDE PLACEMENT KIT, Model Number: DYNDA1468C; ee. PREMIER PROCEDURE PACK, Model Number: DYNJ64393; ff. PREP KIT, Model Number: DYNJRA1155B, DYNJRA1200B; gg. PREP TRAY, Model Number: DYNJRA1498; hh. PROCEDURE PACK, Model Number: DYNJ55063A, DYNJ57674A, DYNJ59795A, DYNJ62588, DYNJ64372, DYNJ64372A, DYNJ68968; ii. PUH AV FISTULA CDS, Model Number: CDS984329J; jj. PURPLE PACK, Model Number: DYNJ62197; kk. PV INTERVENTIONAL PACK-LF, Model Number: DYNJ45757F; ll. PV PACK, Model Number: DYNJ46549J; mm. RAD CVC TRAY, Model Number: DYNJ35574A, DYNJ54131; nn. RAD ONC LITHOTOMY PACK, Model Number: DYNJ60698B; oo. RADIOFREQUENCY ABLATION PACK, Model Number: DYNJ63155A; pp. RADIOLOGY PACK, Model Number: DYNJ42172A; qq. RADIOLOGY PACK-LF #108402, Model Number: DYNJ58225B; rr. RADIOLOGY SPECIALS PACK, Model Number: PHS479840M; ss. RADIOLOGY TRAY, Model Number: MNS11625; tt. RECIPIENT PACK, Model Number: DYNJ41114B; uu. REGIONAL SINGLE SHOT KIT, Model Number: DYNJRA1256; vv. RFA PACK, Model Number: DYNJ57149, DYNJ67899; ww. RIGHT HEART TRAY, Model Number: DYNJ40379C; xx. ROBOT, Model Number: DYNJ906112; yy. ROBOTIC GYN SSC, Model Number: DYNJ60009A; zz. ROBOTIC PROSTATECTOMY, Model Number: DYNJ905826B, DYNJ905826C, DYNJ905826D, DYNJ905826F; aaa. RRMC AORTIC GRAFT PACK-2, Model Number: DYNJ33546J; bbb. RR-SPECIAL PROCEDURE PACK-LF, Model Number: DYNJ0785821M, DYNJ0785821N, DYNJ0785821O; ccc. SCLEROTHERAPY KIT, Model Number: DYNJ62434A; ddd. SLUH LIVER TRANSPLANT PACK-LF, Model Number: DYNJ0369434M; eee. SMH VASCULAR CDS, Model Number: CDS984584L; fff. SPECIAL PROCEDURE PACK, Model Number: DYNJ53135B, DYNJ53135C; ggg. SPECIAL PROCEDURE PACK-LF, Model Number: DYNJ0416535U, DYNJ0416535Y; hhh. SPECIALS PACK, Model Number: DYNJ67504A; iii. SPECIALS TRACE PACK, Model Number: DYNJ57303A, DYNJ57303B; jjj. SSC GU ROBOTIC PACK, Model Number: DYNJ64027; kkk. ST CHARLES PORT PACK, Model Number: DYNJ49833J; lll. STERILE INJECTION PACK, Model Number: DYNDA2805; mmm. SUPPORT TRAY W/CHLORAPREP, Model Number: DYNJRA0958; nnn. SVMMC AV FISTULA PACK, Model Number: DYNJ48153C; ooo. SVMMC VASCULAR MAJOR PA
Code information (models, lots, serial numbers)
Show the full code information (10k characters)
All lot numbers for the following models and corresponding UPCs distributed from Medline between 04/01/2020 and 04/28/2023: a. Model Number: DYNJ54755B, UPC Number: 088994292536. b. Model Number: DYNJ40306A, UPC Number: 088994260192. c. Model Number: DYNJ0774759P, UPC Number: 019532724613. d. Model Number: DYNJ902261D, UPC Number: 088994275524. e. Model Number: DYNJRA1747, UPC Number: 019348946400. f. Model Number: DYNJ44960D, UPC Number: 019348983324. g. Model Number: DYNDV2285, UPC Number: 019348919454; Model Number: MNS9865, UPC Number: 065316029443. h. Model Number: DYNJ904401B, UPC Number: 019348968343. i. Model Number: DYNJ42262D, UPC Number: 088994269795. j. Model Number: DYNJ47502B, UPC Number: 019348964922. k. Model Number: DYNJ67414, UPC Number: 019348996464. l. Model Number: PHS41739A, UPC Number: 088994290680. m. Model Number: DYNJ60038, UPC Number: 088827779327. n. Model Number: CDS982858F, UPC Number: 088827773921. o. Model Number: DYNJ39885A, UPC Number: 088994253993. p. Model Number: DYNJ41873B, UPC Number: 019348956306. q. Model Number: DYNJ52688A, UPC Number: 088994284356. r. Model Number: CDS982859F, UPC Number: 088827773920. s. Model Number: DYNJ47717, UPC Number: 088827751458. t. Model Number: DYNJ55723A, UPC Number: 088994286553; Model Number: DYNJ60990, UPC Number: 019348953847. u. Model Number: DYNJ59067, UPC Number: 088994293004. v. Model Number: DYNJ49177F, UPC Number: 088827777434. w. Model Number: DYNDC2399A, UPC Number: 088994244693. x. Model Number: DYNJ41114C, UPC Number: 019532726173. y. Model Number: DYNJ54894A, UPC Number: 019348921355. z. Model Number: DYNJ54894A, UPC Number: 019348921355. aa. Model Number: DYNJ54982A, UPC Number: 019348980895. bb. Model Number: CVI4765, UPC Number: 019348995318. cc. Model Number: DYNJ56857B, UPC Number: 019348989314; Model Number: DYNJ68990, UPC Number: 019348987261. dd. Model Number: DYNDA1468C, UPC Number: 088994284317; Model Number: DYNDA1468C, UPC Number: 10889942843176. ee. Model Number: DYNJ64393, UPC Number: 019348980963. ff. Model Number: DYNJRA1155B, UPC Number: 019348958371; Model Number: DYNJRA1200B, UPC Number: 019348956827. gg. Model Number: DYNJRA1498, UPC Number: 019348979574. hh. Model Number: DYNJ55063A, UPC Number: 019348968054; Model Number: DYNJ57674A, UPC Number: 088827777174; Model Number: DYNJ59795A, UPC Number: 019348956698; Model Number: DYNJ62588, UPC Number: 019348966983; Model Number: DYNJ64372, UPC Number: 019348980722; Model Number: DYNJ64372A, UPC Number: 019348988468; Model Number: DYNJ68968, UPC Number: 019348987039. ii. Model Number: CDS984329J, UPC Number: 088994286535. jj. Model Number: DYNJ62197, UPC Number: 019348963300. kk. Model Number: DYNJ45757F, UPC Number: 088827779258. ll. Model Number: DYNJ46549J, UPC Number: 019348933778. mm. Model Number: DYNJ35574A, UPC Number: 088438953806; Model Number: DYNJ54131, UPC Number: 088994246680. nn. Model Number: DYNJ60698B, UPC Number: 019532706789. oo. Model Number: DYNJ63155A, UPC Number: 019348931402. pp. Model Number: DYNJ42172A, UPC Number: 019348950271. qq. Model Number: DYNJ58225B, UPC Number: 088827777898. rr. Model Number: PHS479840M, UPC Number: 019348968272. ss. Model Number: MNS11625, UPC Number: 065316032213. tt. Model Number: DYNJ41114B, UPC Number: 088994269269. uu. Model Number: DYNJRA1256, UPC Number: 019348951155. vv. Model Number: DYNJ57149, UPC Number: 088994272160; Model Number: DYNJ67899, UPC Number: 019348945449. ww. Model Number: DYNJ40379C, UPC Number: 088994281455. xx. Model Number: DYNJ906112, UPC Number: 019348965663. yy. Model Number: DYNJ60009A, UPC Number: 019348964887. zz. Model Number: DYNJ905826B, UPC Number: 019348978095; Model Number: DYNJ905826C, UPC Number: 019348940758; Model Number: DYNJ905826D, UPC Number: 019532702352; Model Number: DYNJ905826F, UPC Number: 019532721705. aaa. Model Number: DYNJ33546J, UPC Number: 019348956435. bbb. Model Number: DYNJ0785821M, UPC Number: 019348956435; Model Number: DYNJ0785821N, UPC Number: 019348956435; Model Number: DYNJ0785821O, UPC Number: 019532726373. ccc. Model Number: DYNJ62434A, UPC Number: 019348936627. ddd. Model Number: DYNJ0369434M, UPC Number: 019532729542. eee. Model Number: CDS984584L, UPC Number: 088994270130. fff. Model Number: DYNJ53135B, UPC Number: 019348966082; Model Number: DYNJ53135C, UPC Number: 019348946843. ggg. Model Number: DYNJ0416535U, UPC Number: 019348956435; Model Number: DYNJ0416535Y, UPC Number: 019348956435. hhh. Model Number: DYNJ67504A, UPC Number: 019532700961. iii. Model Number: DYNJ57303A, UPC Number: 088827778321; Model Number: DYNJ57303B, UPC Number: 019348936064. jjj. Model Number: DYNJ64027, UPC Number: 019348977993. kkk. Model Number: DYNJ49833J, UPC Number: 019532705946. lll. Model Number: DYNDA2805, UPC Number: 019348984614. mmm. Model Number: DYNJRA0958, UPC Number: 088994261890. nnn. Model Number: DYNJ48153C, UPC Number: 088994276562. ooo. Model Number: DYNJ43777G, UPC Number: 088994291573. ppp. Model Number: DYNJ63339A, UPC Number: 019348929531; Model Number: DYNJ63339B, UPC Number: 019348984061. qqq. Model Number: DYNJ53319C, UPC Number: 019348963449. rrr. Model Number: DYNJ63340, UPC Number: 019348974116. sss. Model Number: DYNJ906391B, UPC Number: 019348981688. ttt. Model Number: DYNDH1190, UPC Number: 088994249424. uuu. Model Number: DYKM2125, UPC Number: 019532704487. vvv. Model Number: DYNJ58424, UPC Number: 088994285183. www. Model Number: DYNJ37099C, UPC Number: 019348956435. xxx. Model Number: DYNJ59062, UPC Number: 088994294165. yyy. Model Number: TVS4300A, UPC Number: 081698202335. zzz. Model Number: TVS4500A, UPC Number: 081698202337; Model Number: TVS4500B, UPC Number: 019532722356. aaaa. Model Number: ULT2175B, UPC Number: 019348958303; Model Number: ULT2175C, UPC Number: 10653160344353. bbbb. Model Number: DYNDV2537, UPC Number: 019532718408. cccc. Model Number: DYNDC2099, UPC Number: 088827729567. dddd. Model Number: DYNJ56535, UPC Number: 088994265334. eeee. Model Number: DYNDH1319, UPC Number: 088994279547. ffff. Model Number: DYNDV2385A, UPC Number: 019532729876. gggg. Model Number: IVS3485A, UPC Number: 065316099015; Model Number: IVS3840A, UPC Number: 019348947673. hhhh. Model Number: DYNDV2038B, UPC Number: 019348991983; Model Number: IVS3300, UPC Number: 065316030766; Model Number: IVS3495A, UPC Number: 065316099305. iiii. Model Number: IVS3820, UPC Number: 065316034231. jjjj. Model Number: DYNDV2371, UPC Number: 019348943086. kkkk. Model Number: DYNJ56507B, UPC Number: 019348944347; Model Number: DYNJ64930, UPC Number: 019348923168. llll. Model Number: DYNDH1112, UPC Number: 10889942277360. mmmm. Model Number: IV8635D, UPC Number: 065316099026. nnnn. Model Number: DYNDV2502, UPC Number: 019532701409. oooo. Model Number: DYNDH1204, UPC Number: 088994251983; Model Number: DYNDH1468A, UPC Number: 019348944003; Model Number: DYNJRA0389, UPC Number: 088438972552; Model Number: DYNDH1204, UPC Number: 10889942519835. pppp. Model Number: DYNJ61433, UPC Number: 019348956517. qqqq. Model Number: DYNJ60006A, UPC Number: 019348981007. rrrr. Model Number: DYNJ51638B, UPC Number: 019348994641; Model Number: DYNJ51638C, UPC Number: 019532703956. ssss. Model Number: DYNDH1492, UPC Number: 019348943446. tttt. Model Number: DYNJ904401B, UPC Number: 019348968343. uuuu. Model Number: DYNJ902259A, UPC Number: 088994225654. vvvv. Model Number: CDS982895I, UPC Number: 088994220768. wwww. Model Number: DYNJ902261D, UPC Number: 088994275524. xxxx. Model Number: DYNJ43593A, UPC Number: 088994225559; Model Number: DYNJ53475D, UPC Number: 019348954775. yyyy. Model Number: VC-PK, UPC Number: 019532717689. zzzz. Model Number: DYNJ63268B, UPC Number: 019348930511; Model Number: DYNJ66079, UPC Number: 019348931744. aaaaa. Model Number: DYNJ61148A, UPC Number: 019348954925. bbbbb. Model Number: DYNJ32397B, UPC Number: 088438991665; Model Number: DYNJ54139A, UPC Number: 088994256499; Model Number: DYNJ59094, UPC Number: 088994293350; Model Number: DYNJ65189C, UPC Number: 019532707200. ccccc. Model Number: DYNJS3005, UPC Number: 088438909956. ddddd. Model Number: DYNJ55841B, UPC Number: 088827775768. eeeee. Model Number: DYNJ64190A, UPC Number: 019348943563. fffff. Model Number: DYNJ0309084D, UPC Number: 088994288526; Model Number: DYNJ38682C, UPC Number: 088994289719; Model Number: DYNJ51248B, UPC Number: 088994287340; Model Number: DYNJ59662A, UPC Number: 019348954814; Model Number: DYNJ59681, UPC Number: 088827777552; Model Number: DYNJ61264, UPC Number: 019348955447. ggggg. Model Number: DYNJ59631, UPC Number: 088827776402. hhhhh. Model Number: DYNDA2076, UPC Number: 088994270850. iiiii. Model Number: DYNJ38264C, UPC Number: 019348956435. jjjjj. Model Number: DYNJ57736A, UPC Number: 088827777970. kkkkk. Model Number: DYNJ59751A, UPC Number: 019348959856. lllll. Model Number: DYNJ20094J, UPC Number: 019348956435; Model Number: DYNJ20094K, UPC Number: 019348956435. mmmmm. Model Number: DYNJ56260B, UPC Number: 019348956982. nnnnn. Model Number: DYNJ64666A, UPC Number: 019348931726. ooooo. Model Number: DYNJ901392D, UPC Number: 019348953951. ppppp. Model Number: DYNJ59246, UPC Number: 088827776729. qqqqq. Model Number: DYNJ53349A, UPC Number: 088994257453. rrrrr. Model Number: DYNJ35988G, UPC Number: 019348956435; Model Number: DYNJ35988I, UPC Number: 019348956435; Model Number: DYNJ35988J, UPC Number: 019532702466. sssss. Model Number: DYNJ56392, UPC Number: 088994264067; Model Number: DYNJ56392A, UPC Number: 019532726140.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNDH1204Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Medline Industries, Inc. DYNDH1112Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNDA1468CMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Centurion ULT2175CCenturion Medical Products Corporation · General surgery tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92399).
- Z-2290-2023 · Open, ClassifiedMedline Probe Cover Kits containing Turkuaz Ultrasound Gel, Labeled as the following: a. COVER, PROBE, W/GEL AND BANDS, 4X48 - Model Number DYNJE5900; b. COVER, PROBE, W/GEL AN…
- Z-2292-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. ART LINE KIT ICH OR, Model Number: ART900; b. ART LINE SECUREMEN…
- Z-2297-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. (13) STC IVC FILTER/ANGIOGRAM, Model Number: DYNJ908249A; b. 16…
- Z-2298-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. CRMC MINOR IR PROCEDURE PK-LF, Model Numbers: DYNJ61099B, DYNJ6109…
- Z-2299-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. MULTI-LUMEN CVC BUNDLE KIT, Model Number: ECVC475; b. MULTI-LUM…
- Z-2304-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. 16CM CENTRAL LINE BUNDLE, Model Number: STCVC2020; b. ARTERIAL…
- Z-2305-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. A V FISTULA, Model Number: DYNJ65711; b. ACCESS CATHETER, Mode…
- Z-2307-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. ABLATION/DRAINAGE PACK, Model Number: DYNJ53658B; b. BASIC BIO…
- Z-2313-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. DIALYSIS ACCESSORY PACK W/O CATHETER, Model Number: MNS10750A; …
- Z-2319-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. 7F, 20CM TRIPLE LUMEN CVC BUNDLE, Model Number: ECVC4740; b. A…
- Z-2320-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. DR. ROSEN VEIN PACK, Model Number: DYNJ67517; b. DR. VANBOSSE …
- Z-2322-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. 18GX6" FEMORAL ART LINE KIT, Model Number: ART960; b. 20Gx3" R…
- Z-2323-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. ADULT EPIV INSERTION TRAY, Model Number: DYNDA2482; b. ARTERIAL …
- Z-2324-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. CENTRAL LINE INSERT KIT W/O CA, Model Number: CVI3755; b. CENT…
- Z-2325-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. CVC INFANT INSERTION PACK, Model Number: CVI4770A; b. INSERTIO…
- Z-2329-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. PROBE COVER PACK, Model Number: DYNDA1219
- Z-2339-2023 · Open, ClassifiedCustom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following: a. BREAST BIOPSY KIT, Model Number: DYNDA1224A, DYNDA1224B; b. BR…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page