Infinity recallZ-2510-2023 · Abbott Medical · initiated Jul 18, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Medical or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).
FDA root cause category: Under Investigation by firm
What the company is doing
On 07/20/2023, the firm will send an "URGENT MEDICAL DEVICE CORRECTION" Letter via mail with delivery tracking to customers informing them, Abbott has received complaints from patients who are unable to exit MRI mode, as their Patient Controller has lost the ability to connect or communicate with their IPG while in MRI mode. Situations where this has occurred include where the user deleted the Bluetooth pairing, lost or disabled their Patient Controller, or upgraded the iOS! software on their Patient Controller while in MRI mode. Next Steps for Customers: Per Abbott's Instructions for Use, patients are advised to contact their physician before having an MRI to discuss all critical information regarding MRI scans and MRI mode. In alignment with the Clinician Programmer and Patient Controller Instructions for Use, Abbott recommends physicians do the following: " For the Patient Controller, advise the patient not to delete the paired Bluetooth connection between their IPG and the Patient Controller and not to alter, damage or lose their Patient Controller while the IPG is in MRI mode. " For the Patient Controller prior to entering MRI mode, ensure patients have upgraded their Patient Controller to the latest "Patient Controller NR - US" application from the Apple! App Store!. This version of the application provides instructions for the user not to delete the IPG pairing while the system is in MRI mode. " For the Clinician Programmer, maintain the paired Bluetooth connection between the IPG and the Clinician Programmer by avoiding deleting the pairing and disabling automatic iOS! software upgrades from the iOS! settings. " For questions about this issue or to report patients who are unable to exit MRI mode, please contact your local Abbott representative or Abbott Technical Support at 1-800-727-7846 (option 3) for assistance.
Recall details
- Recall number
- Z-2510-2023
- Status
- Open, Classified
- Initiated by the company
- Jul 18, 2023
- Posted by the FDA
- Sep 11, 2023
- Recalling firm
- Abbott Medical
- Quantity
- 5,932 devices (US-4,464; OUS-1,468)
- Product code
- LGW
- FDA submissions
- P010032, P010032S096, P010032S151
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Korean, Kuwait, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.
Product description
Infinity 5 Implantable Pulse Generator REF 6660 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensions or leads with up to 16 electrodes total. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG can deliver stimulation with a single program or with multiple programs. Each program can provide stimulation to a single anatomical area or to multiple areas. The IPG communicates wirelessly with system programmers and controllers, and the IPGs are available in small and large sizes to accommodate different power needs. The Proclaim" XR and Proclaim" Plus IPGs are used for Spinal Cord Stimulation (SCS). The Proclaim" DRG IPG is used for Dorsal Root Ganglion (DRG) stimulation. The Infinity" IPG is used for Deep Brain Stimulation (DBS).
Code information (models, lots, serial numbers)
Show the full code information (53k characters)
Model Number: 6660 UDI-DI Codes: 05415067030016 Serial Numbers: US AAD255.1 AAN997.1 AAP328.1 AAT540.1 AAT579.1 AAT652.1 AAT772.1 AAT783.1 AAT795.1 AAT797.1 AAT845.1 AAT902.1 AAT989.1 AAU009.1 AAU022.1 AAU046.1 AAU078.1 AAU094.1 ABM541.1 ABM592.1 ABM630.1 ABT798.1 ABT829.1 ABT841.1 ABT900.1 ACD800.1 ACE689.1 ACE771.1 ACE786.1 ACE838.1 ACE875.1 ACE936.1 ACE939.1 ACE994.1 ACF048.1 ACK145.1 ACK161.1 ACK241.1 ACK511.1 ACK514.1 ACK549.1 ACK622.1 ACK629.1 ACK637.1 ACK699.1 ACK702.1 ACK739.1 ACK752.1 ACK803.1 ACK805.1 ACK806.1 ACK828.1 ACK855.1 ACP686.1 ACP700.1 ACP712.1 ACP770.1 ACP780.1 ACP835.1 ACP869.1 ACP881.1 ACP883.1 ACT538.1 ACT548.1 ACT565.1 ACT738.1 ACU349.1 ACU381.1 ACU424.1 ADR673.1 ADR679.1 ADR685.1 ADR787.1 ADR794.1 ADR852.1 ADR856.1 ADW762.1 ADW766.1 ADW771.1 AEH704.1 AEH745.1 AEH748.1 AEH761.1 AEH764.1 AEH772.1 AEH826.1 AEH831.1 AEH856.1 AEH862.1 AEH865.1 AEH877.1 AEH883.1 AEH953.1 AEH955.1 AEH962.1 AEH972.1 AEJ054.1 AEJ100.1 AEJ101.1 AEJ111.1 AEJ145.1 AEJ164.1 AEJ167.1 AEJ196.1 AEJ203.1 AEJ241.1 AEJ251.1 AEJ252.1 AEJ257.1 AEM803.1 AEM833.1 AEM857.1 AEM877.1 AEM885.1 AEM887.1 AEM888.1 AEM951.1 AEM964.1 AEM967.1 AEM969.1 AEM988.1 AFB225.1 AFB349.1 AFB376.1 AFB425.1 AFB433.1 AFB434.1 AFB436.1 AFB444.1 AFB454.1 AFB495.1 AFB658.1 AFG344.1 AFG422.1 AFG446.1 AFG471.1 AFG496.1 AFG545.1 AFG668.1 AFG672.1 AFG679.1 AFG704.1 AFG724.1 AFG750.1 AFG810.1 AFG815.1 AFG819.1 AFG851.1 AFG852.1 AFG857.1 AFG871.1 AFG947.1 AFG959.1 AFZ148.1 AFZ174.1 AFZ198.1 AFZ211.1 AFZ224.1 AFZ260.1 AFZ359.1 AFZ364.1 AFZ384.1 AFZ392.1 AFZ409.1 AFZ530.1 AFZ532.1 AFZ590.1 AFZ807.1 AFZ835.1 AFZ851.1 AFZ996.1 AGM596.1 AGM602.1 AGM732.1 AGM764.1 AGM765.1 AGM950.1 AGN231.1 AGN324.1 AGN447.1 AGN552.1 AGN616.1 AHD578.1 AHD589.1 AHD612.1 AHD646.1 AHD648.1 AHD663.1 AHD677.1 AHD686.1 AHD713.1 AHD818.1 AHD821.1 AHK567.1 AHK727.1 AHK768.1 AHK778.1 AHK897.1 AHK906.1 AHK922.1 AHK976.1 AHL054.1 AHL069.1 AHL093.1 AHL095.1 AHL097.1 AHL105.1 AHL122.1 AHL133.1 AHL135.1 AHL138.1 AHL195.1 AHV587.1 AHV591.1 AHV705.1 AHV797.1 AHV801.1 AHV802.1 AHV807.1 AHV810.1 AHV818.1 AHV824.1 AHV843.1 AHV851.1 AHV854.1 AHV855.1 AHV860.1 AHV873.1 AHV887.1 AHV937.1 AHV945.1 AHV947.1 AHV965.1 AHV971.1 AHV975.1 AHV992.1 AHW055.1 AHW080.1 AHW082.1 AHW104.1 AHW496.1 AHW498.1 AHW661.1 AHW663.1 AHW670.1 AHW696.1 AJP994.1 AJP995.1 AJR053.1 AJR058.1 AJR073.1 AJR075.1 AJR076.1 AJR077.1 AJW278.1 AJW280.1 AJW285.1 AJW287.1 AJW290.1 AJW296.1 AJW310.1 AJW328.1 AJW329.1 AJW334.1 AJW336.1 AJW345.1 AJW365.1 AJW384.1 AJW388.1 AJW393.1 AJW412.1 AJW422.1 AJW431.1 AJW487.1 ALE668.1 ALE944.1 ALJ257.1 ALJ402.1 ALJ928.1 ALJ984.1 ALJ992.1 ALN367.1 ALN429.1 ALN438.1 ALN441.1 ALN444.1 ALN448.1 ALN451.1 ALN469.1 ALN477.1 ALN487.1 ALN547.1 ALN593.1 ALN596.1 ALN622.1 ALN658.1 ALX037.1 ALX043.1 ALX053.1 ALX085.1 ALX203.1 ALX313.1 ALY252.1 ALY270.1 ALY272.1 ALY313.1 ALY329.1 ALY367.1 ALY369.1 ALY371.1 ALY400.1 ALY431.1 ALY533.1 ALY541.1 ALY544.1 ALY610.1 ALY624.1 ALY625.1 ALY636.1 ALY653.1 ALY660.1 ALY774.1 ALY799.1 ALY801.1 ALY802.1 ALY810.1 ALZ051.1 ALZ080.1 ALZ174.1 ALZ191.1 ALZ311.1 AMJ759.1 AMK053.1 AMN778.1 AMN786.1 AMN800.1 AMN825.1 AMN826.1 AMN872.1 AMN919.1 AMN963.1 AMP153.1 AMP207.1 AMP263.1 AMP265.1 AMP271.1 AMS002.1 AMS004.1 AMS007.1 AMS022.1 AMS024.1 AMS026.1 AMS038.1 AMS068.1 AMS073.1 AMS126.1 AMS127.1 AMS161.1 AMS169.1 AMS193.1 AMS228.1 AMS241.1 AMS248.1 AMS256.1 AMS258.1 AMS259.1 AMS270.1 AMS400.1 AMS431.1 AMS445.1 AMS452.1 AMS480.1 AMS487.1 AMS488.1 AMS504.1 AMS678.1 AMS705.1 AMS817.1 AMS829.1 AMS854.1 AMS865.1 AMS867.1 AMS879.1 AMS889.1 AMS891.1 AMS892.1 AMS919.1 AMS920.1 AMS925.1 AMS928.1 AMS931.1 AMS939.1 AMS944.1 AMS946.1 AMS978.1 AMS983.1 AMS988.1 AMS989.1 AMS990.1 AMT006.1 AMT007.1 AMT018.1 AMT023.1 AMT027.1 AMT094.1 AMT114.1 AMT122.1 AMT212.1 AMT260.1 AMT275.1 AMT289.1 AMT292.1 AMT351.1 ANG899.1 ANG900.1 ANG969.1 ANG990.1 ANG999.1 ANH005.1 ANH036.1 ANH041.1 ANH056.1 ANH061.1 ANH062.1 ANH066.1 ANH075.1 ANH077.1 ANH094.1 ANH096.1 ANH100.1 ANH104.1 ANH110.1 ANH113.1 ANH123.1 ANH126.1 ANH127.1 ANH131.1 ANH151.1 ANH163.1 ANH166.1 ANH170.1 ANH178.1 ANH185.1 ANH187.1 ANH207.1 ANH215.1 ANH232.1 ANH258.1 ANH269.1 ANH275.1 ANH278.1 ANH335.1 ANH375.1 ANH465.1 ANH468.1 ANH469.1 ANH503.1 ANH544.1 ANH598.1 ANH618.1 ANH636.1 ANL523.1 ANL534.1 ANL602.1 ANL644.1 ANL657.1 ANL660.1 ANL663.1 ANL668.1 ANL760.1 ANL782.1 ANL784.1 ANL792.1 ANL958.1 ANM023.1 ANM056.1 ANM059.1 ANM072.1 ANT198.1 ANT218.1 ANT235.1 ANT659.1 ANT740.1 ANT829.1 ANT950.1 ANU127.1 APA825.1 APA912.1 APA938.1 APB043.1 APB241.1 APB265.1 APB298.1 APB363.1 APB386.1 APB389.1 APB516.1 APB527.1 APB632.1 APB745.1 APB914.1 APB921.1 APC040.1 APE539.1 APE596.1 APE620.1 APE635.1 APE964.1 APF063.1 APL204.1 APL416.1 APS470.1 APS480.1 APZ202.1 APZ232.1 APZ462.1 ARG349.1 ARJ197.1 ARJ337.1 ARK025.1 ARK331.1 ARK552.1 ARK654.1 ARK786.1 ARK864.1 ARK930.1 ARK994.1 ARL206.1 ASG295.1 ASG296.1 ASG333.1 ASG362.1 ASG552.1 ASG564.1 ASG604.1 ASG619.1 ASG681.1 ASG706.1 ASG771.1 ASG933.1 ASJ064.1 ASJ118.1 ASJ122.1 ASJ154.1 ASJ230.1 ASJ282.1 ASM274.1 ASM331.1 ASM350.1 ASM398.1 ASM405.1 ASM461.1 ASM463.1 ASM481.1 ASM490.1 ASM530.1 ASM540.1 ASM666.1 ASM673.1 ASM694.1 ASM697.1 ASM699.1 ASM732.1 ASR446.1 ASR714.1 ASR759.1 ASR770.1 ASR800.1 ASR825.1 ASV344.1 ASV356.1 ASV396.1 ASV409.1 ASV417.1 ASV568.1 ASV616.1 ASV626.1 ASV632.1 ASV680.1 ASV681.1 ASV768.1 ASV814.1 ASV868.1 ASW187.1 ASW190.1 ASW206.1 ATC610.1 ATC612.1 ATC691.1 ATC758.1 ATC761.1 ATC830.1 ATC832.1 ATC837.1 ATC847.1 ATC866.1 ATC939.1 ATC943.1 ATC949.1 ATC951.1 ATC954.1 ATD040.1 ATD078.1 ATD142.1 ATD171.1 ATD175.1 ATD200.1 ATD257.1 ATD271.1 ATD277.1 ATD289.1 ATD361.1 ATD480.1 ATD492.1 ATD533.1 ATD535.1 ATD676.1 ATD686.1 ATD739.1 ATD765.1 ATD886.1 ATD920.1 ATD927.1 ATD945.1 ATD966.1 ATD979.1 ATD996.1 ATE032.1 ATE693.1 ATE702.1 ATE735.1 ATE740.1 ATE777.1 ATE826.1 ATE860.1 ATE938.1 ATE958.1 ATE960.1 ATE965.1 ATE998.1 ATF003.1 ATF042.1 ATF110.1 ATF136.1 ATF147.1 ATF192.1 ATF198.1 ATF202.1 ATF213.1 ATF227.1 ATF233.1 ATF242.1 ATF274.1 ATF363.1 ATF385.1 ATF415.1 ATF463.1 ATJ745.1 ATJ763.1 ATJ777.1 ATJ807.1 ATJ882.1 ATJ936.1 ATJ941.1 ATJ996.1 ATK009.1 ATK062.1 ATK068.1 ATK090.1 ATK103.1 ATK110.1 ATK130.1 ATK173.1 ATK250.1 ATK261.1 ATK262.1 ATK271.1 ATK289.1 ATK313.1 ATK316.1 ATK335.1 ATT994.1 ATU005.1 ATU013.1 ATU075.1 ATU298.1 ATU348.1 ATU383.1 ATU389.1 ATU488.1 ATU491.1 ATU532.1 ATU542.1 ATU738.1 ATV051.1 ATV132.1 ATV156.1 ATV191.1 ATV947.1 ATV963.1 ATW107.1 ATW120.1 ATW127.1 ATW129.1 ATW161.1 ATZ910.1 ATZ947.1 ATZ991.1 AUA079.1 AUA087.1 AUA107.1 AUA119.1 AUA131.1 AUA133.1 AUA137.1 AUA158.1 AUA168.1 AUA269.1 AUA309.1 AUA325.1 AUA386.1 AUA390.1 AUA393.1 AUA402.1 AUA492.1 AUA512.1 AUA580.1 AUA594.1 AUA620.1 AUA625.1 AUA641.1 AUA663.1 AUA667.1 AUA681.1 AUA707.1 AUA730.1 AUB150.1 AUE347.1 AUE384.1 AUE438.1 AUE441.1 AUE465.1 AUE535.1 AUE730.1 AUE741.1 AUE743.1 AUE793.1 AUE797.1 AUE800.1 AUE811.1 AUE815.1 AUE836.1 AUE853.1 AUE861.1 AUE891.1 AUE900.1 AUE970.1 AUH558.1 AUH656.1 AUH679.1 AUH734.1 AUH805.1 AUH870.1 AUH875.1 AUH876.1 AUH919.1 AUH950.1 AUM041.1 AUM139.1 AUM154.1 AUM166.1 AUM380.1 AUM383.1 AUM449.1 AUM590.1 AUM597.1 AUM639.1 AUM686.1 AUM706.1 AUM760.1 AUM789.1 AUM796.1 AUM805.1 AUM841.1 AUM856.1 AUM864.1 AUN363.1 AUN997.1 AUP020.1 AUP041.1 AUP086.1 AUP104.1 AUP129.1 AUP298.1 AUR649.1 AUR687.1 AUV797.1 AUV824.1 AUV837.1 AUV866.1 AUV883.1 AUV904.1 AUV915.1 AUV919.1 AUV924.1 AUW664.1 AUW772.1 AUX496.1 AUX508.1 AUX554.1 AVA019.1 AVA035.1 AVA046.1 AVA081.1 AVA086.1 AVA103.1 AVA209.1 AVA426.1 AVA462.1 AVA488.1 AVA532.1 AVA550.1 AVA581.1 AVA605.1 AVA641.1 AVA653.1 AVA670.1 AVA675.1 AVA679.1 AVA709.1 AVA726.1 AVA728.1 AVA734.1 AVA784.1 AVA814.1 AVA824.1 AVA826.1 AVA858.1 AVA894.1 AVA903.1 AVB020.1 AVB060.1 AVB070.1 AVB196.1 AVD374.1 AVD385.1 AVD450.1 AVD457.1 AVD551.1 AVD794.1 AVD837.1 AVE365.1 AVE421.1 AVE439.1 AVE447.1 AVE557.1 AVE603.1 AVE621.1 AVE645.1 AVE705.1 AVE756.1 AVE763.1 AVE765.1 AVE775.1 AVF268.1 AVF278.1 AVF284.1 AVF368.1 AVF369.1 AVF409.1 AVF454.1 AVF593.1 AVF642.1 AVF661.1 AVF685.1 AVF752.1 AVF756.1 AVF827.1 AVF828.1 AVK650.1 AVK758.1 AVK792.1 AVK815.1 AVK817.1 AVK860.1 AVK901.1 AVK922.1 AVK981.1 AVK991.1 AVL045.1 AVL099.1 AVL100.1 AVL139.1 AVL159.1 AVL195.1 AVL219.1 AVL230.1 AVL938.1 AVL944.1 AVL984.1 AVL988.1 AVM004.1 AVM009.1 AVM015.1 AVM021.1 AVM029.1 AVM222.1 AVM223.1 AVM224.1 AVM256.1 AVM325.1 AVN614.1 AVN642.1 AVN661.1 AVN680.1 AVN700.1 AVN735.1 AVN738.1 AVN754.1 AVN757.1 AVN868.1 AVN893.1 AVN894.1 AVN896.1 AVN908.1 AVN957.1 AVN986.1 AVP108.1 AVP383.1 AVP396.1 AVP407.1 AVP728.1 AVP894.1 AVP932.1 AVP945.1 AVP959.1 AVR218.1 AVR363.1 AVR422.1 AVR437.1 AVR505.1 AVR549.1 AVR711.1 AVR723.1 AVW763.1 AVW875.1 AVW879.1 AVW916.1 AVW943.1 AVW976.1 AVX009.1 AVX036.1 AVX143.1 AVX291.1 AVX368.1 AVX385.1 AVX395.1 AVX424.1 AVX459.1 AVX472.1 AVX546.1 AVX580.1 AVY647.1 AVY728.1 AVY773.1 AVY780.1 AVY787.1 AVY806.1 AVY893.1 AVY922.1 AVY938.1 AVZ038.1 AVZ065.1 AVZ073.1 AVZ074.1 AVZ080.1 AWB300.1 AWB303.1 AWB308.1 AWB324.1 AWB325.1 AWB348.1 AWB385.1 AWB399.1 AWB405.1 AWB490.1 AWB500.1 AWB534.1 AWB544.1 AWB674.1 AWB691.1 AWB748.1 AWB753.1 AWB756.1 AWB788.1 AWB795.1 AWB801.1 AWB814.1 AWB819.1 AWB840.1 AWB847.1 AWB869.1 AWB910.1 AWB914.1 AWB928.1 AWB935.1 AWB954.1 AWC005.1 AWC016.1 AWC025.1 AWC076.1 AWC195.1 AWC300.1 AWC343.1 AWC346.1 AWC349.1 AWC376.1 AWC503.1 AWC519.1 AWC537.1 AWC601.1 AWC678.1 AWC693.1 AWC873.1 AWC897.1 AWC989.1 AWF789.1 AWH715.1 AWL444.1 AWL462.1 AWP731.1 AWP842.1 AWP930.1 AWP944.1 AWR029.1 AWR035.1 AWR070.1 AWR099.1 AWR101.1 AWR108.1 AWR111.1 AWR128.1 AWR137.1 AWR157.1 AWR254.1 AWR299.1 AWR311.1 AWR354.1 AWR357.1 AWR391.1 AWR402.1 AWR406.1 AWR444.1 AWR445.1 AWR495.1 AWR512.1 AWR580.1 AWR581.1 AWR601.1 AWR639.1 AWR643.1 AWR646.1 AWR677.1 AWR680.1 AWR684.1 AWT975.1 AWX023.1 AWX030.1 AWX106.1 AWX204.1 AWX343.1 AWX420.1 AWX490.1 AWX541.1 AWX564.1 AWX719.1 AWX802.1 AWY341.1 AXE634.1 AXE735.1 AXE819.1 AXE827.1 AXE968.1 AXE978.1 AXF059.1 AXF062.1 AXF347.1 AXF385.1 AXF406.1 AXF533.1 AXF535.1 AXF537.1 AXF555.1 AXF561.1 AXF619.1 AXK090.1 AXK114.1 AXK120.1 AXK167.1 AXK223.1 AXK252.1 AXK271.1 AXK274.1 AXK275.1 AXK351.1 AXP789.1 AXP803.1 AXP804.1 AXP841.1 AXP868.1 AXP869.1 AXP871.1 AXP874.1 AXP880.1 AXP886.1 AXP902.1 AXP940.1 AXP979.1 AXP981.1 AXP983.1 AXR036.1 AXR154.1 AXR175.1 AXR181.1 AXT391.1 AXT404.1 AXT485.1 AXU278.1 AXU279.1 AXU283.1 AXU293.1 AXU311.1 AXU317.1 AXU338.1 AXU361.1 AXU451.1 AXU479.1 AXU567.1 AXU609.1 AXU628.1 AXU633.1 AXU668.1 AXU672.1 AXU725.1 AXU776.1 AXU793.1 AXU808.1 AXU820.1 AXU829.1 AXU864.1 AXV790.1 AXV815.1 AXV830.1 AXV850.1 AXV971.1 AXV978.1 AXV984.1 AXV991.1 AXW011.1 AXW058.1 AXX856.1 AXX870.1 AXX873.1 AXX983.1 AXX990.1 AXY013.1 AXY101.1 AXY104.1 AXY135.1 AXY204.1 AXY218.1 AXY223.1 AXY226.1 AXY237.1 AXY244.1 AXY311.1 AXY315.1 AXY332.1 AXY404.1 AXY639.1 AXY640.1 AXY650.1 AXY676.1 AXY681.1 AXY688.1 AXY689.1 AXY703.1 AXY709.1 AXY732.1 AXY736.1 AXY740.1 AXY744.1 AXY760.1 AXY806.1 AXY877.1 AXY878.1 AXY887.1 AXY893.1 AXY926.1 AXY944.1 AXY985.1 AXY986.1 AXY989.1 AXY994.1 AXZ043.1 AYD251.1 AYD263.1 AYD274.1 AYD290.1 AYD339.1 AYD362.1 AYD397.1 AYD454.1 AYD460.1 AYD496.1 AYD506.1 AYD559.1 AYD607.1 AYD625.1 AYD632.1 AYD644.1 AYD651.1 AYD665.1 AYD669.1 AYD676.1 AYD696.1 AYD702.1 AYD703.1 AYD721.1 AYD730.1 AYD776.1 AYD778.1 AYD802.1 AYD803.1 AYD819.1 AYD839.1 AYD855.1 AYD862.1 AYE002.1 AYE024.1 AYE032.1 AYE060.1 AYE079.1 AYE155.1 AYE180.1 AYE203.1 AYE216.1 AYE222.1 AYE236.1 AYE244.1 AYE245.1 AYE274.1 AYE287.1 AYE306.1 AYE340.1 AYE399.1 AYE413.1 AYE417.1 AYE426.1 AYE432.1 AYE437.1 AYE465.1 AYE485.1 AYE507.1 AYE523.1 AYE525.1 AYE543.1 AYE555.1 AYE564.1 AYE573.1 AYE580.1 AYE650.1 AYE659.1 AYE660.1 AYE670.1 AYE683.1 AYE713.1 AYE721.1 AYE741.1 AYE745.1 AYE803.1 AYE810.1 AYE839.1 AYE852.1 AYE899.1 AYM173.1 AYM214.1 AYM217.1 AYM220.1 AYM234.1 AYM277.1 AYM288.1 AYM602.1 AYM915.1 AYM971.1 AYN015.1 AYN029.1 AYN056.1 AYN060.1 AYP863.1 AYP876.1 AYP877.1 AYP878.1 AYP883.1 AYP895.1 AYP900.1 AYP904.1 AYP921.1 AYP963.1 AYP964.1 AYP966.1 AYP972.1 AYP974.1 AYP975.1 AYP976.1 AYP993.1 AYR012.1 AYR053.1 AYR064.1 AYR230.1 AYR272.1 AYR303.1 AYR307.1 AYR334.1 AYR356.1 AYR358.1 AYR359.1 AYR382.1 AYR404.1 AYR405.1 AYR406.1 AYR419.1 AYR431.1 AYR450.1 AYR458.1 AYR487.1 AYR548.1 AYR571.1 AYR589.1 AYR615.1 AYR633.1 AYR640.1 AYR644.1 AYR654.1 AYR658.1 AYR660.1 AYR670.1 AYR672.1 AYR688.1 AYR690.1 AYR702.1 AYR729.1 AYR748.1 AYR749.1 AYR750.1 AYR796.1 AYR810.1 AYR941.1 AYS342.1 AYS625.1 AYS756.1 AYS765.1 AYS851.1 AYS885.1 AYS960.1 AYS987.1 AYT111.1 AYT116.1 AYT279.1 AYT318.1 AYT384.1 AYT393.1 AYX793.1 AYX840.1 AYX913.1 AYX927.1 AYX953.1 AYX958.1 AYX991.1 AYY004.1 AYY107.1 AYY129.1 AYY137.1 AYY170.1 AYY176.1 AYY283.1 AYY294.1 AYY332.1 AYY393.1 AYY394.1 AYY428.1 AYY434.1 AYY440.1 AYY563.1 AYY579.1 AYY679.1 AYY838.1 AYY840.1 AYY926.1 AYY941.1 AYY957.1 BAK166.1 BAK175.1 BAK180.1 BAK185.1 BAK199.1 BAK203.1 BAK315.1 BAK406.1 BAK424.1 BAK560.1 BAK576.1 BAK579.1 BAK593.1 BAK598.1 BAK607.1 BAK724.1 BAK779.1 BAK791.1 BAK794.1 BAK857.1 BAK861.1 BAK880.1 BAK999.1 BAL007.1 BAL013.1 BAL017.1 BAL027.1 BAL030.1 BAL036.1 BAL048.1 BAL069.1 BAL102.1 BAL110.1 BAL116.1 BAL132.1 BAL150.1 BAL571.1 BAL699.1 BAR382.1 BAR415.1 BAR458.1 BAR493.1 BAR499.1 BAR507.1 BAR513.1 BAR521.1 BAR525.1 BAR588.1 BAR639.1 BAR658.1 BAR688.1 BAR692.1 BAR712.1 BAR714.1 BAR717.1 BAR727.1 BAR747.1 BAR753.1 BAR754.1 BAR777.1 BAR787.1 BAR822.1 BAR827.1 BAR912.1 BAR938.1 BAR943.1 BAR951.1 BAR971.1 BAR974.1 BAR987.1 BAR992.1 BAS022.1 BAS023.1 BAS064.1 BAS068.1 BAS070.1 BAS073.1 BAS204.1 BAS212.1 BAS234.1 BAS243.1 BAS266.1 BAS267.1 BAS304.1 BAS305.1 BAS325.1 BAS335.1 BAS337.1 BAS357.1 BAS358.1 BAS363.1 BAS364.1 BAS371.1 BAS384.1 BAS399.1 BAS414.1 BAS428.1 BAS433.1 BAS480.1 BAS481.1 BAS482.1 BAS485.1 BAS496.1 BAS498.1 BAS502.1 BAS521.1 BAS525.1 BAS529.1 BAS544.1 BAS587.1 BAS590.1 BAW199.1 BAW259.1 BAW761.1 BAW771.1 BAW774.1 BAW800.1 BAW813.1 BAW814.1 BAW839.1 BAW842.1 BAW866.1 BAW867.1 BAW883.1 BAW886.1 BAW892.1 BAW893.1 BAW979.1 BAX006.1 BAX033.1 BAX053.1 BAX126.1 BAX214.1 BAX219.1 BAX227.1 BAX244.1 BAX258.1 BAX328.1 BAX339.1 BAX359.1 BAX363.1 BAX383.1 BAX385.1 BAX408.1 BAX453.1 BAX771.1 BAX778.1 BAX819.1 BAX868.1 BAX931.1 BAY206.1 BAY341.1 BAY373.1 BAY391.1 BAY396.1 BAY404.1 BAY606.1 BAY675.1 BAY693.1 BAY890.1 BAY903.1 BAY945.1 BAY947.1 BAY948.1 BAY959.1 BAZ007.1 BAZ304.1 BBG450.1 BBG592.1 BBG975.1 BBH131.1 BBH152.1 BBH186.1 BBH331.1 BBK994.1 BBS102.1 BBS301.1 BBS395.1 BBS415.1 BBS499.1 BBW282.1 BBW319.1 BBW334.1 BBW381.1 BBW673.1 BBW675.1 BBW709.1 BBW714.1 BBW717.1 BBW726.1 BBW742.1 BBW756.1 BBW778.1 BBW780.1 BBW801.1 BBW802.1 BBW804.1 BBW806.1 BBW840.1 BBW844.1 BBW852.1 BBW855.1 BBW876.1 BBW877.1 BBW878.1 BBW915.1 BBW920.1 BBW954.1 BBW961.1 BBW963.1 BBW969.1 BBW990.1 BBX028.1 BBX032.1 BBY533.1 BBY537.1 BBY538.1 BBY549.1 BBY578.1 BBY583.1 BBY585.1 BBY589.1 BBY615.1 BBY619.1 BBY636.1 BBY649.1 BBY670.1 BBY673.1 BBY676.1 BBY964.1 BCA109.1 BCA145.1 BCA618.1 BCA695.1 BCA786.1 BCC777.1 BCC826.1 BCC834.1 BCC858.1 BCD979.1 BCD992.1 BCD995.1 BCE001.1 BCE013.1 BCE017.1 BCE020.1 BCE035.1 BCG196.1 BCG198.1 BCG199.1 BCG200.1 BCG204.1 BCG209.1 BCG212.1 BCG233.1 BCG235.1 BCG237.1 BCG238.1 BCG240.1 BCG261.1 BCG279.1 BCG283.1 BCG308.1 BCG320.1 BCG327.1 BCG331.1 BCG332.1 BCG335.1 BCG339.1 BCG340.1 BCH282.1 BCH285.1 BCH301.1 BCH307.1 BCH311.1 BCH315.1 BCH316.1 BCH318.1 BCH357.1 BCH364.1 BCH380.1 BCH382.1 BCH388.1 BCH398.1 BCH401.1 BCH412.1 BCH441.1 BCH459.1 BCH461.1 BCH467.1 BCH469.1 BCH473.1 BCH475.1 BCH503.1 BCH505.1 BCH509.1 BCH512.1 BCH519.1 BCH527.1 BCH533.1 BCH540.1 BCM573.1 BCN767.1 BCN810.1 BCN811.1 BCN880.1 BCN962.1 BCN977.1 BCP027.1 BCP074.1 BCP116.1 BCS909.1 BCS929.1 BCS942.1 BCS955.1 BCT037.1 BCT038.1 BCT059.1 BCT071.1 BCT092.1 BCT125.1 BCT166.1 BCT191.1 BCT192.1 BCU909.1 BCU967.1 BCU969.1 BCU994.1 BCV023.1 BCV033.1 BCV059.1 BCV246.1 BCV285.1 BCV297.1 BCV307.1 BCV332.1 BCV354.1 BCX776.1 BCX783.1 BCX799.1 BCX818.1 BCX946.1 BCY017.1 BCY040.1 BCY170.1 BCY196.1 BCY211.1 BDA086.1 BDA154.1 BDA169.1 BDA199.1 BDA214.1 BDA230.1 BDA242.1 BDA302.1 BDA330.1 BDA336.1 BDA349.1 BDA371.1 BDA377.1 BDA387.1 BDA396.1 BDA398.1 BDA405.1 BDA418.1 BDA455.1 BDA470.1 BDA486.1 BDA517.1 BDA521.1 BDA524.1 BDA528.1 BDB890.1 BDB947.1 BDB951.1 BDB954.1 BDB974.1 BDB985.1 BDC018.1 BDC039.1 BDC046.1 BDC064.1 BDC103.1 BDC116.1 BDC145.1 BDC184.1 BDC187.1 BDC197.1 BDC202.1 BDC205.1 BDC233.1 BDC239.1 BDC240.1 BDC249.1 BDC259.1 BDC265.1 BDC301.1 BDC304.1 BDC309.1 BDC317.1 BDC320.1 BDC323.1 BDC370.1 BDC379.1 BDC387.1 BDC428.1 BDC443.1 BDC450.1 BDE866.1 BDE875.1 BDE893.1 BDE902.1 BDE908.1 BDE910.1 BDE913.1 BDE915.1 BDE926.1 BDE932.1 BDE946.1 BDE955.1 BDE957.1 BDE959.1 BDE995.1 BDF003.1 BDF013.1 BDF015.1 BDF025.1 BDF052.1 BDF055.1 BDF057.1 BDF065.1 BDF066.1 BDF070.1 BDF071.1 BDF087.1 BDF107.1 BDF108.1 BDF157.1 BDF162.1 BDF166.1 BDF171.1 BDF188.1 BDF231.1 BDF232.1 BDF264.1 BDF268.1 BDF278.1 BDF287.1 BDF292.1 BDF295.1 BDF320.1 BDF339.1 BDJ211.1 BDJ255.1 BDJ258.1 BDJ260.1 BDJ264.1 BDJ281.1 BDJ282.1 BDJ295.1 BDJ336.1 BDJ354.1 BDJ360.1 BDJ361.1 BDJ369.1 BDJ382.1 BDJ388.1 BDJ393.1 BDJ414.1 BDJ438.1 BDJ440.1 BDJ473.1 BDJ488.1 BDJ497.1 BDJ511.1 BDJ514.1 BDJ533.1 BDJ537.1 BDJ548.1 BDJ622.1 BDJ626.1 BDJ639.1 BDJ687.1 BDL933.1 BDL947.1 BDM024.1 BDM497.1 BDM635.1 BDN451.1 BDN463.1 BDN617.1 BDN619.1 BDU819.1 BDU831.1 BDU856.1 BDU977.1 BDU979.1 BDU980.1 BDU982.1 BDV003.1 BDV013.1 BDV014.1 BDV015.1 BDV023.1 BDV028.1 BDV032.1 BDV045.1 BDV052.1 BDV053.1 BDV069.1 BDV070.1 BDV077.1 BDV079.1 BDV083.1 BDV097.1 BDV107.1 BDV120.1 BDV131.1 BDV133.1 BDV139.1 BDV140.1 BDV147.1 BDV152.1 BDV157.1 BDV174.1 BDV178.1 BDV191.1 BDV196.1 BDV203.1 BDV213.1 BDV223.1 BDV225.1 BDV231.1 BDV237.1 BDV240.1 BDV243.1 BDV245.1 BDV249.1 BDV255.1 BDV257.1 BDV258.1 BDV259.1 BDV263.1 BDV265.1 BDV274.1 BDV276.1 BDV279.1 BDV283.1 BDV561.1 BDV610.1 BDY878.1 BDY888.1 BDY898.1 BDY970.1 BDZ106.1 BDZ145.1 BDZ200.1 BDZ255.1 BDZ270.1 BDZ279.1 BDZ299.1 BDZ309.1 BDZ419.1 BDZ454.1 BDZ502.1 BEB416.1 BEB417.1 BEB466.1 BEB468.1 BEB496.1 BEB497.1 BEB511.1 BEB515.1 BEB526.1 BEB534.1 BEB564.1 BEB592.1 BEB629.1 BEB631.1 BEB640.1 BEB664.1 BEB708.1 BEB714.1 BEB768.1 BEB774.1 BEB776.1 BEB781.1 BEB871.1 BEB891.1 BEB899.1 BEB922.1 BEB934.1 BEB938.1 BEB941.1 BEB944.1 BEB967.1 BEB976.1 BEC067.1 BEC088.1 BEC105.1 BEC106.1 BEE449.1 BEE460.1 BEE475.1 BEE685.1 BEF089.1 BEK830.1 BEL045.1 BEU442.1 BEU502.1 BEU519.1 BEU526.1 BEU531.1 BEX653.1 BEX665.1 BEX679.1 BEX683.1 BEX693.1 BEX727.1 BEX731.1 BEX741.1 BEX742.1 BEX746.1 BEX759.1 BEX774.1 BEX839.1 BEX866.1 BEX950.1 BEX965.1 BEX994.1 BEX997.1 BEX998.1 BEY005.1 BEY006.1 BEY010.1 BEY022.1 BEY029.1 BEY117.1 BEY123.1 BEY129.1 BEY137.1 BEY145.1 BEY157.1 BEY160.1 BEY181.1 BEY183.1 BEY205.1 BEY233.1 BEY246.1 BEY341.1 BEY349.1 BEY365.1 BEY376.1 BEY386.1 BEY393.1 BEY428.1 BEY436.1 BEY444.1 BEY572.1 BEY599.1 BEY648.1 BEY659.1 BEY674.1 BEY676.1 BEY692.1 BEY727.1 BEY766.1 BEY778.1 BEY814.1 BEY815.1 BEY838.1 BEY839.1 BEY843.1 BEY857.1 BEY858.1 BFA685.1 BFA717.1 BFA722.1 BFA730.1 BFA797.1 BFA803.1 BFA804.1 BFA951.1 BFA954.1 BFB012.1 BFB041.1 BFB135.1 BFB139.1 BFB198.1 BFB228.1 BFB233.1 BFB279.1 BFB282.1 BFB310.1 BFB373.1 BFB580.1 BFB732.1 BFB738.1 BFB742.1 BFB770.1 BFB779.1 BFE142.1 BFE182.1 BFE195.1 BFE243.1 BFE286.1 BFE315.1 BFE379.1 BFE381.1 BFE414.1 BFE536.1 BFE602.1 BFE635.1 BFE713.1 BFE810.1 BFE818.1 BFE897.1 BFE955.1 BFF016.1 BFF043.1 BFF084.1 BFF107.1 BFJ305.1 BFJ312.1 BFJ442.1 BFJ579.1 BFJ583.1 BFJ607.1 BFJ613.1 BFJ617.1 BFJ647.1 BFJ652.1 BFJ690.1 BFJ692.1 BFJ702.1 BFJ716.1 BFJ740.1 BFJ763.1 BFJ764.1 BFJ766.1 BFJ777.1 BFJ779.1 BFJ789.1 BFJ794.1 BFJ800.1 BFJ808.1 BFJ845.1 BFJ913.1 BFJ924.1 BFJ934.1 BFJ944.1 BFJ994.1 BFK003.1 BFK047.1 BFK055.1 BFK067.1 BFK119.1 BFK175.1 BFK205.1 BFK223.1 BFL191.1 BFL215.1 BFL217.1 BFL219.1 BFL234.1 BFL259.1 BFL262.1 BFL361.1 BFL554.1 BFL559.1 BFL861.1 BFX892.1 BGA075.1 BGA291.1 BGA876.1 BGA961.1 BGB060.1 BGD155.1 BGD272.1 BGD285.1 BGD287.1 BGD345.1 BGD468.1 BGD471.1 BGD496.1 BGD561.1 BGD588.1 BGD657.1 BGD664.1 BGD719.1 BGD798.1 BGD855.1 BGD871.1 BGD878.1 BGD895.1 BGD897.1 BGD913.1 BGF430.1 BGF437.1 BGF451.1 BGF458.1 BGF513.1 BGF519.1 BGF536.1 BGF565.1 BGF570.1 BGF571.1 BGF589.1 BGF601.1 BGF607.1 BGF615.1 BGH507.1 BGH531.1 BGH549.1 BGH567.1 BGH568.1 BGH569.1 BGH574.1 BGH585.1 BGH589.1 BGH606.1 BGH609.1 BGH617.1 BGH639.1 BGH640.1 BGH642.1 BGH664.1 BGH723.1 BGH737.1 BGH794.1 BGH798.1 BGH806.1 BGH844.1 BGH845.1 BGJ971.1 BGJ972.1 BGJ974.1 BGJ992.1 BGK001.1 BGK032.1 BGK116.1 BGK124.1 BGK134.1 BGK142.1 BGK147.1 BGK208.1 BGK237.1 BGK246.1 BGK270.1 BGK616.1 BGK665.1 BGK749.1 BGK756.1 BGK770.1 BGK809.1 BGK836.1 BGK872.1 BGK879.1 BGK896.1 BGK917.1 BGK993.1 BGL023.1 BGL047.1 BGL078.1 BGL082.1 BGL084.1 BGL140.1 BGL154.1 BGL155.1 BGL163.1 BGL186.1 BGL223.1 BGL230.1 BGL234.1 BGL240.1 BGL242.1 BGL245.1 BGL269.1 BGL271.1 BGL275.1 BGL277.1 BGL281.1 BGL292.1 BGN654.1 BGN663.1 BGN692.1 BGN708.1 BGN763.1 BGN773.1 BGN778.1 BGN784.1 BGN807.1 BGN815.1 BGN827.1 BGN828.1 BGN834.1 BGN877.1 BGN881.1 BGN892.1 BGN917.1 BGN973.1 BGN974.1 BGP020.1 BGP028.1 BGP043.1 BGP047.1 BGP058.1 BGP060.1 BGP091.1 BGP126.1 BGP128.1 BGP140.1 BGP147.1 BGP149.1 BGP179.1 BGP184.1 BGP223.1 BGP233.1 BGP273.1 BGP295.1 BGP299.1 BGS120.1 BGS134.1 BGS159.1 BGS277.1 BGS283.1 BGS286.1 BGS295.1 BGS302.1 BGS331.1 BGS332.1 BGS340.1 BGS368.1 BGS483.1 BGS527.1 BGS591.1 BGS611.1 BGS613.1 BGS623.1 BGS624.1 BGS630.1 BGS645.1 BGS655.1 BGS656.1 BGS673.1 BGV287.1 BGV291.1 BGV293.1 BGV328.1 BGV333.1 BGV346.1 BGV347.1 BGV353.1 BGV513.1 BGV601.1 BGV610.1 BGV633.1 BGV653.1 BGV683.1 BGV801.1 BGV834.1 BGV870.1 BGV871.1 BGV890.1 BGV974.1 BGV987.1 BGY914.1 BGY952.1 BGZ032.1 BGZ049.1 BGZ293.1 BGZ364.1 BGZ409.1 BGZ461.1 BGZ477.1 BGZ553.1 BGZ560.1 BGZ607.1 BGZ659.1 BGZ670.1 BHD065.1 BHD066.1 BHD110.1 BHD143.1 BHD146.1 BHD159.1 BHD186.1 BHD188.1 BHD236.1 BHD271.1 BHD273.1 BHD286.1 BHD334.1 BHD346.1 BHD350.1 BHD352.1 BHD357.1 BHD369.1 BHD438.1 BHD448.1 BHD565.1 BHD567.1 BHD652.1 BHD734.1 BHH294.1 BHH300.1 BHH323.1 BHH439.1 BHH480.1 BHH505.1 BHH595.1 BHH638.1 BHH669.1 BHH943.1 BHP742.1 BHP805.1 BHP961.1 BHR015.1 BHR037.1 BHR072.1 BHR400.1 BHR435.1 BHR458.1 BHR493.1 BHU344.1 BHU346.1 BHU407.1 BHU446.1 BHU486.1 BHU495.1 BHU520.1 BHU524.1 BHU530.1 BHU597.1 BHU600.1 BHU608.1 BHU622.1 BHU633.1 BHX118.1 BHX218.1 BHX471.1 BHX535.1 BHX563.1 BHX578.1 BHX596.1 BHX650.1 BHX672.1 BHX840.1 BHX858.1 BHX870.1 BHX882.1 BHX985.1 BJD496.1 BJD556.1 BJD586.1 BJD605.1 BJD689.1 BJD726.1 BJD791.1 BJD798.1 BJD988.1 BJE143.1 BJE211.1 BJE224.1 BJE266.1 BJG600.1 BJG744.1 BJK779.1 BJP593.1 BJW615.1 BJW641.1 BJY965.1 BJZ017.1 BJZ225.1 BJZ340.1 BJZ412.1 BJZ441.1 BJZ554.1 BKC156.1 BKC191.1 BKC245.1 BKC412.1 BKC853.1 BKC868.1 BKD123.1 BKD225.1 BKD258.1 BKF280.1 BKF325.1 BKF349.1 BKF397.1 BKF403.1 BKF467.1 BKF537.1 BKF630.1 BKF649.1 BKF695.1 BKF770.1 BKF806.1 BKF827.1 BKF902.1 BKF937.1 BKF959.1 BKF966.1 BKF972.1 BKG021.1 BKG097.1 BKG120.1 BKG293.1 BKG297.1 BKG308.1 BKG310.1 BKG319.1 BKG328.1 BKH621.1 BKH692.1 BKH698.1 BKH705.1 BKH734.1 BKH767.1 BKH781.1 BKH843.1 BKH899.1 BKH900.1 BKH905.1 BKH982.1 BKJ005.1 BKJ137.1 BKJ326.1 BKJ342.1 BKJ356.1 BKJ357.1 BKS898.1 BKT059.1 BKT095.1 BKT231.1 BKT355.1 BKT369.1 BKT507.1 BKZ174.1 BKZ236.1 BKZ257.1 BKZ282.1 BKZ309.1 BLA857.1 BLA860.1 BLA901.1 BLA928.1 BLC375.1 BLC420.1 BLC443.1 BLC448.1 BLC449.1 BLC467.1 BLC476.1 BLC478.1 BLC484.1 BLC505.1 BLC548.1 BLK473.1 BLK582.1 BLK584.1 BLK834.1 BLK873.1 BLL032.1 BLL122.1 BLL193.1 BLP002.1 BLP009.1 BLP010.1 BLP079.1 BLP084.1 BLP210.1 BLP269.1 BLP302.1 BLP311.1 BLP313.1 BLP323.1 BLP334.1 BLP345.1 BLP578.1 BLP653.1 BLP666.1 BLP672.1 BLR138.1 BLR186.1 BLR228.1 BLT452.1 BLT460.1 BLT480.1 BLT482.1 BLT521.1 BLT550.1 BLT554.1 BLT567.1 BLT574.1 BLT651.1 BLT653.1 BLT685.1 BLT702.1 BLT713.1 BLT747.1 BLT764.1 BLT770.1 BLT776.1 BLT784.1 BLT829.1 BLT880.1 BLT946.1 BLU044.1 BLU068.1 BLU096.1 BLU126.1 BLU189.1 BLY131.1 BLY223.1 BLY264.1 BLY394.1 BLY438.1 BLY538.1 BLY565.1 BLY624.1 BLY701.1 BLY743.1 BLY849.1 BLZ234.1 BLZ270.1 BMA955.1 BMB963.1 BMC148.1 BMC279.1 BMC291.1 BMC314.1 BMC351.1 BMC376.1 BMC379.1 BMD829.1 BMD835.1 BMD875.1 BMD968.1 BMD971.1 BMD989.1 BME007.1 BME029.1 BME069.1 BMF625.1 BMF666.1 BMF697.1 BMF704.1 BMF734.1 BMF833.1 BMF849.1 BMF857.1 BMF870.1 BMF875.1 BMF932.1 BMG003.1 BMG061.1 BMG114.1 BMG123.1 BMG140.1 BMG146.1 BMG188.1 BMG225.1 BMG278.1 BMG315.1 BMG335.1 BMG395.1 BMG406.1 BMG457.1 BMG460.1 BMG468.1 BMG472.1 BMG479.1 BMJ788.1 BMJ854.1 BMJ860.1 BMJ944.1 BMJ966.1 BMJ975.1 BMK029.1 BMK062.1 BMK068.1 BMK076.1 BMK088.1 BMK101.1 BMK161.1 BMK196.1 BMK204.1 BMK206.1 BMK214.1 BMK221.1 BMK225.1 BMK230.1 BMK295.1 BMK311.1 BMK339.1 BMK361.1 BML082.1 BML126.1 BML161.1 BML185.1 BML240.1 BML291.1 BML375.1 BML425.1 BML435.1 BML488.1 BML507.1 BML538.1 BML608.1 BML625.1 BMM040.1 BMM042.1 BMM044.1 BMM079.1 BMM080.1 BMM109.1 BMM149.1 BMM201.1 BMM275.1 BMM303.1 BMM306.1 BMM770.1 BMM810.1 BMM833.1 BMM922.1 BMM969.1 BMN028.1 BMN244.1 BMN247.1 BMN248.1 BMN273.1 BMR321.1 BMR378.1 BMR430.1 BMR488.1 BMT018.1 BMT047.1 BMT126.1 BMT225.1 BMT292.1 BMT301.1 BMT313.1 BMT570.1 BMT583.1 BMW755.1 BMX052.1 BNC197.1 BNC370.1 BNK814.1 BNV969.1 BNW304.1 BNW305.1 BNX029.1 BNX206.1 BNX280.1 BNX362.1 BNX501.1 BPG024.1 BPG232.1 BPK571.1 BPK581.1 BPM586.1 BPM811.1 BPM917.1 BPN077.1 BPS920.1 BPS921.1 BPS932.1 BPS949.1 BPV031.1 BPV034.1 BPV062.1 BPV075.1 BPV121.1 BPV138.1 BPV164.1 BPV177.1 BPV227.1 BPV263.1 BPV274.1 BPV283.1 BPV328.1 BPV332.1 BPV342.1 BPV344.1 BPV346.1 BPV352.1 BPV354.1 BPV357.1 BPV381.1 BPV390.1 BPV398.1 BPV407.1 BPV463.1 BPV473.1 BPV493.1 BPV520.1 BPV521.1 BPV524.1 BPV525.1 BPV527.1 BPV529.1 BPV541.1 BPV545.1 BPV547.1 BPV548.1 BPV558.1 BPV566.1 BPV581.1 BPV591.1 BPV592.1 BPV593.1 BPZ705.1 BRA669.1 BRA836.1 BRB154.1 BRC880.1 BRF100.1 BRG516.1 BRG545.1 BRG614.1 BRG643.1 BRG693.1 BRG737.1 BRG903.1 BRH057.1 BRH092.1 BRH209.1 BRH248.1 BRJ421.1 BRJ475.1 BRJ476.1 BRJ519.1 BRJ533.1 BRJ543.1 BRJ572.1 BRJ602.1 BRJ628.1 BRJ633.1 BRJ638.1 BRJ642.1 BRJ650.1 BRJ682.1 BRJ707.1 BRJ713.1 BRJ715.1 BRJ741.1 BRJ742.1 BRJ749.1 BRJ777.1 BRJ782.1 BRJ974.1 BRK062.1 BRK084.1 BRK135.1 BRK929.1 BRK946.1 BRK956.1 BRK993.1 BRL046.1 BRM908.1 BRM926.1 BRM960.1 BRM974.1 BRM977.1 BRN390.1 BRN457.1 BRN499.1 BRN500.1 BRN603.1 BRN671.1 BRN688.1 BRN723.1 BRN731.1 BRN758.1 BRN764.1 BRN774.1 BRN827.1 BRR131.1 BRR151.1 BRR179.1 BRR223.1 BRR242.1 BRR260.1 BRR273.1 BRR298.1 BRR311.1 BRR379.1 BRR454.1 BRR457.1 BRR542.1 BRR697.1 BRR771.1 BRR778.1 BRR861.1 BRR902.1 BRR915.1 BRT716.1 BRT722.1 BRT730.1 BRT742.1 BRT762.1 BRT783.1 BRT897.1 BRT905.1 BRT907.1 BRT961.1 BRT973.1 BRU013.1 BRU159.1 BRU175.1 BRW919.1 BRW983.1 BRX081.1 BRX102.1 BRX104.1 BRX119.1 BRX570.1 BRX579.1 BRX609.1 BRX717.1 BRX724.1 BRX738.1 BRX764.1 BRZ314.1 BRZ417.1 BRZ433.1 BRZ456.1 BRZ458.1 BRZ539.1 BRZ541.1 BSA249.1 BSA323.1 BSA346.1 BSA423.1 BSA484.1 BSA529.1 BSA540.1 BSA602.1 BSA635.1 BSA698.1 BSA702.1 BSD036.1 BSD114.1 BSD115.1 BSD157.1 BSD164.1 BSD309.1 BSD339.1 BSD344.1 BSD356.1 BSD394.1 BSD399.1 BSE957.1 BSE969.1 BSE981.1 BSE984.1 BSE998.1 BSF057.1 BSF081.1 BSF088.1 BSF131.1 BSF141.1 BSF148.1 BSF165.1 BSF166.1 BSF210.1 BSF218.1 BSF219.1 BSF220.1 BSF282.1 BSN600.1 BSV820.1 BSW089.1 BSY350.1 BSY811.1 BTA891.1 BTA999.1 BTB021.1 BTB067.1 BTB104.1 BTB113.1 BTB201.1 BTB293.1 BTB372.1 BTB447.1 BTB558.1 BTC183.1 BTC581.1 BTC602.1 BTC630.1 BTC634.1 BTC697.1 BTC709.1 BTC715.1 BTC741.1 BTC843.1 BTC851.1 BTC862.1 BTC880.1 BTC981.1 BTD004.1 BTD061.1 BTD150.1 BTD159.1 BTD162.1 BTD166.1 BTD168.1 BTD256.1 BTM811.1 BTN069.1 BTN101.1 BTN426.1 BTR935.1 BTS838.1 BTS874.1 BTW631.1 BTW650.1 BTW651.1 BTW678.1 BTW717.1 BTW724.1 BTW815.1 BTW839.1 BTW842.1 BTW846.1 BTW986.1 BTX047.1 BTX060.1 BTX126.1 BTX131.1 BTX139.1 BTX150.1 BTX209.1 BTX216.1 BTX219.1 BTX273.1 BTX299.1 BTX328.1 BTX337.1 BTX390.1 BTX400.1 BTX413.1 BTX436.1 BTX470.1 BTX477.1 BTX488.1 BTX503.1 BTX509.1 BTX608.1 BTX671.1 BTX690.1 BTX694.1 BTX699.1 BTX747.1 BTX762.1 BTX767.1 BTX836.1 BTX844.1 BTX845.1 BTX851.1 BUA445.1 BUA481.1 BUA487.1 BUA567.1 BUA608.1 BUA654.1 BUA727.1 BUA770.1 BUA793.1 BUA806.1 BUA823.1 BUA827.1 BUA848.1 BUA871.1 BUB039.1 BUB058.1 BUB081.1 BUB129.1 BUB154.1 BUB162.1 BUB753.1 BUB766.1 BUB781.1 BUB806.1 BUB830.1 BUB849.1 BUB864.1 BUB873.1 BUB884.1 BUB896.1 BUB904.1 BUB910.1 BUB911.1 BUB932.1 BUB936.1 BUB955.1 BUB962.1 BUC024.1 BUC965.1 BUD004.1 BUD024.1 BUD046.1 BUD056.1 BUF199.1 BUF226.1 BUF243.1 BUF275.1 BUF318.1 BUF385.1 BUF394.1 BUF400.1 BUF426.1 BUF443.1 BUF496.1 BUF542.1 BUF568.1 BUF604.1 BUF659.1 BUF669.1 BUF686.1 BUF699.1 BUF722.1 BUF757.1 BUJ866.1 BUJ871.1 BUJ873.1 BUJ981.1 BUK540.1 BUK664.1 BUK705.1 BUK715.1 BUL448.1 BUL454.1 BUL602.1 BUL671.1 BUL683.1 BUL695.1 BUP221.1 BUP247.1 BUP291.1 BUP323.1 BUP337.1 BUP366.1 BUP402.1 BUP412.1 BUP496.1 BUP541.1 BUP595.1 BUP622.1 BUP653.1 BUP681.1 BUP688.1 BUP690.1 BUP698.1 BUP704.1 BUP709.1 BUR853.1 BUR864.1 BUR919.1 BUR920.1 BUR938.1 BUR954.1 BUR958.1 BUS028.1 BUS080.1 BUS081.1 BUS098.1 BUS117.1 BUV015.1 BUV050.1 BUV104.1 BUV107.1 BUV123.1 BUV154.1 BUV184.1 BUV195.1 BUV219.1 BUV269.1 BUV282.1 BUV286.1 BUV303.1 BUV307.1 BUV344.1 BUV353.1 BUV387.1 BUV396.1 BUV481.1 BUV520.1 BUV551.1 BUW642.1 BUW675.1 BUW721.1 BUW770.1 BUX368.1 BUX403.1 BUX449.1 BUX501.1 BUX551.1 BUX563.1 BUX573.1 BUX633.1 BUX636.1 BUX637.1 BUX692.1 BUX725.1 BUX775.1 BUY351.1 BUY353.1 BUY362.1 BUY399.1 BUY434.1 BUY469.1 BUY488.1 BUY509.1 BUY514.1 BUY522.1 BUY535.1 BUY545.1 BUY652.1 BUY690.1 BUY706.1 BUY709.1 BUY710.1 BUY718.1 BUY725.1 BUY748.1 BUY750.1 BUY783.1 BUY866.1 BUY874.1 BUY893.1 BUY897.1 BUY953.1 BUY963.1 BUY972.1 BUY981.1 BUY982.1 BUZ044.1 BUZ067.1 BVD707.1 BVD713.1 BVD722.1 BVD724.1 BVD741.1 BVD755.1 BVD787.1 BVD804.1 BVD844.1 BVD855.1 BVD863.1 BVD953.1 BVD988.1 BVE081.1 BVE203.1 BVE272.1 BVE289.1 BVE329.1 BVE331.1 BVE363.1 BVE445.1 BVE448.1 BVE452.1 BVE500.1 BVH241.1 BVH246.1 BVH305.1 BVH316.1 BVH376.1 BVH418.1 BVH460.1 BVH468.1 BVH489.1 BVH548.1 BVK034.1 BVK125.1 BVK168.1 BVK170.1 BVK174.1 BVK298.1 BVL098.1 BVL214.1 BVL570.1 BVP083.1 BVP106.1 BVP118.1 BVP176.1 BVP269.1 BVP327.1 BVP351.1 BVP368.1 BVX141.1 BVX143.1 BVX147.1 BVX232.1 BVX243.1 BVX261.1 BVX310.1 BVX351.1 BVX355.1 BVX381.1 BVX396.1 BVX445.1 BVX493.1 BVX503.1 BVX530.1 BVX531.1 BVX587.1 BVX612.1 BVY103.1 BVY109.1 BVY122.1 BVY155.1 BVY190.1 BVY232.1 BVY248.1 BVY251.1 BVY254.1 BVY257.1 BVY308.1 BVY322.1 BVY366.1 BVY376.1 BVY407.1 BVY460.1 BVY492.1 BVY511.1 BVY527.1 BVY532.1 BVY560.1 BVY572.1 BVZ680.1 BVZ707.1 BVZ803.1 BVZ812.1 BVZ899.1 BVZ946.1 BVZ958.1 BVZ962.1 BVZ963.1 BWA048.1 BWA090.1 BWA110.1 BWA113.1 BWC135.1 BWC196.1 BWC315.1 BWC328.1 BWC377.1 BWC420.1 BWC422.1 BWC453.1 BWC462.1 BWC463.1 BWC472.1 BWC484.1 BWC553.1 BWC559.1 BWE769.1 BWE788.1 BWE842.1 BWE870.1 BWE903.1 BWE949.1 BWE953.1 BWE976.1 BWF046.1 BWF055.1 BWF064.1 BWF090.1 BWF091.1 BWF107.1 BWF128.1 BWF146.1 BWF148.1 BWF151.1 BWF229.1 BWF233.1 BWF257.1 BWF258.1 BWF265.1 BWF841.1 BWF843.1 BWF933.1 BWG022.1 BWG024.1 BWG066.1 BWG091.1 BWG108.1 BWG199.1 BWG355.1 BWG357.1 BWG367.1 BWJ623.1 BWJ659.1 BWJ682.1 BWJ727.1 BWJ777.1 BWJ802.1 BWJ859.1 BWJ904.1 BWJ942.1 BWK043.1 BWP103.1 BWP212.1 BWP274.1 BWP285.1 BWP289.1 BWP291.1 BWP337.1 BWP452.1 BWP559.1 BWP620.1 BWP623.1 BWP626.1 BWP643.1 BWP675.1 BWP688.1 BWP690.1 BWP760.1 BWP762.1 BWP768.1 BWP869.1 BWP877.1 BWP903.1 BWR039.1 BWT946.1 BWU031.1 BWU040.1 BWU057.1 BWV690.1 BWX122.1 BWX158.1 BWX212.1 BWX346.1 BXB034.1 BXB106.1 BXB174.1 BXB256.1 BXB265.1 BXB294.1 BXB377.1 BXB439.1 BXB583.1 BXB672.1 BXB726.1 BXB851.1 BXB902.1 BXC044.1 BXC086.1 BXC110.1 BXC121.1 BXH584.1 BXH594.1 BXH601.1 BXH616.1 BXH640.1 BXH655.1 BXH673.1 BXH744.1 BXH773.1 BXH776.1 BXH855.1 BXH858.1 BXH974.1 BXJ813.1 BXJ844.1 BXJ901.1 BXJ928.1 BXJ948.1 BXK035.1 BXK515.1 BXK573.1 BXK611.1 BXK707.1 BXK821.1 BXK833.1 BXK919.1 BXK926.1 BXL001.1 BXL012.1 BXM509.1 BXM570.1 BXM584.1 BXM671.1 BXM727.1 BXM761.1 BXM807.1 BXM859.1 BXM875.1 BXM920.1 BXM934.1 BXM955.1 BXN569.1 BXN695.1 BXN729.1 BXN774.1 BXN867.1 BXN877.1 BXN972.1 BXP041.1 BXP105.1 BXP113.1 BXR159.1 BXR190.1 BXS586.1 BXS717.1 BXS779.1 BXS846.1 BXS848.1 BXS887.1 BXT707.1 BXT715.1 BXT770.1 BXT787.1 BXT823.1 BXT828.1 BXT840.1 BXT851.1 BXV957.1 BXV969.1 BXW203.1 BXW302.1 BXW478.1 BXY564.1 BXY591.1 BXY626.1 BXY646.1 BXY665.1 BXY806.1 BXY814.1 BXY816.1 BXY897.1 BXY898.1 BXY919.1 BYA223.1 BYA283.1 BYA289.1 BYB676.1 BYB678.1 BYB685.1 BYD752.1 BYD756.1 BYD798.1 BYD820.1 BYD844.1 BYE015.1 BYE077.1 BYE083.1 BYE103.1 BYE134.1 BYE149.1 BYE168.1 BYE181.1 BYE208.1 BYE242.1 BYE260.1 BYE262.1 BYE385.1 BYE410.1 BYE483.1 BYE527.1 BYG537.1 BYG590.1 BYG620.1 BYL765.1 BYL782.1 BYL807.1 BYL848.1 BYL860.1 BYL912.1 BYL924.1 BYL934.1 BYL945.1 BYL958.1 BYL959.1 BYL961.1 BYL968.1 BYM066.1 BYM104.1 BYM126.1 BYM188.1 BYP305.1 BYP334.1 BYP341.1 BYP348.1 BYP355.1 BYP369.1 BYP412.1 BYP425.1 BYP431.1 BYR162.1 BYR259.1 BYR330.1 BYR484.1 BYR499.1 BYR513.1 BYS448.1 BYS459.1 BYS533.1 BYS583.1 BYS591.1 BYS593.1 BYS618.1 BYS650.1 BYS653.1 BYU021.1 BYU132.1 BYU399.1 BYU410.1 BYU450.1 BYU453.1 BYU456.1 BYU461.1 BYU462.1 BYU479.1 BYU485.1 BYU486.1 BYU493.1 BYU500.1 BYU503.1 BYU511.1 BYU516.1 BYU526.1 BYU533.1 BYU552.1 BYU571.1 BYU960.1 BYU999.1 BYV050.1 BYV088.1 BYV176.1 BYV188.1 BYV198.1 BYV217.1 BYV222.1 BYV246.1 BYV249.1 BYV263.1 BYV277.1 BYV305.1 BYV364.1 BYV414.1 BYV431.1 BYX885.1 CAD997.1 CAE047.1 CAE059.1 CAE072.1 CAE075.1 CAE110.1 CAE111.1 CAE112.1 CAE122.1 CAE126.1 CAE130.1 CAE138.1 CAE173.1 CAE183.1 CAE191.1 CAE193.1 CAE199.1 CAE202.1 CAE205.1 CAE214.1 CAE216.1 CAE217.1 CAE226.1 CAE239.1 CAE243.1 CAE244.1 CAE245.1 CAE260.1 CAE275.1 CAE282.1 CAE284.1 CAF736.1 CAF835.1 CAF842.1 CAF891.1 CAF926.1 CAF969.1 CAF993.1 CAG025.1 CAG145.1 CAG383.1 CAG400.1 CAG418.1 CAG440.1 CAG481.1 CAG681.1 CAG705.1 CAG776.1 CAG947.1 CAH015.1 CAH037.1 CAH041.1 CAH078.1 CAH106.1 CAH135.1 CAH143.1 CAH145.1 CAH223.1 CAH234.1 CAH258.1 CAH261.1 CAH868.1 CAH932.1 CAH940.1 CAJ003.1 CAJ313.1 CAJ403.1 CAJ418.1 CAJ453.1 CAK262.1 CAK278.1 CAK300.1 CAK305.1 CAK319.1 CAK358.1 CAK363.1 CAK373.1 CAK413.1 CAK482.1 CAK484.1 CAK587.1 CAK631.1 CAK674.1 CAK690.1 CAK702.1 CAK738.1 CAK755.1 CAK767.1 CAK775.1 CAK798.1 CAM329.1 CAM338.1 CAM444.1 CAM481.1 CAM533.1 CAN466.1 CAN477.1 CAN655.1 CAN796.1 CAN884.1 CAP502.1 CAP552.1 CAP559.1 CAP631.1 CAP655.1 CAP763.1 CAR185.1 CAR203.1 CAS501.1 CAS681.1 CAV515.1 CAV525.1 CAV561.1 CAV781.1 CAW966.1 CAW983.1 CAX001.1 CAX065.1 CAX078.1 CAX258.1 CAX285.1 CAX313.1 CAX320.1 CAX355.1 CAX374.1 CAX401.1 CAX404.1 CAX428.1 CAX449.1 CAX552.1 CAY118.1 CAY173.1 CAY186.1 CBA385.1 CBA397.1 CBA402.1 CBA403.1 CBA408.1 CBB243.1 CBB275.1 CBB306.1 CBB327.1 CBB331.1 CBB346.1 CBB429.1 CBB452.1 CBC216.1 CBC276.1 CBC295.1 CBC371.1 CBC378.1 CBC382.1 CBC393.1 CBC395.1 CBC464.1 CBC541.1 CBC548.1 CBC576.1 CBC591.1 CBC619.1 CBE280.1 CBE550.1 CBF569.1 CBF580.1 CBF611.1 CBF658.1 CBF867.1 CBF913.1 CBG604.1 CBJ980.1 CBJ992.1 CBK357.1 CBK359.1 CBK375.1 CBK533.1 CBK572.1 CBK675.1 CBK679.1 CBK711.1 CBK715.1 CBK723.1 CBK725.1 CBK730.1 CBK776.1 CBM205.1 CBM228.1 CBM284.1 CBM289.1 CBM338.1 CBM340.1 CBM342.1 CBM359.1 CBM373.1 CBM387.1 CBM410.1 CBM417.1 CBM477.1 CBM527.1 CBM532.1 CBM541.1 CBM593.1 CBR704.1 CBR739.1 CBR743.1 CBR766.1 CBR884.1 CBR889.1 CBR927.1 CBR928.1 CBR939.1 CBT998.1 CBU015.1 CBU019.1 CBU029.1 CBU031.1 CBU088.1 CBU140.1 CBU226.1 CBU288.1 CBW115.1 CBW117.1 CBW119.1 CBW120.1 CBW124.1 CBW131.1 CBW133.1 CBW146.1 CBW148.1 CBW162.1 CBW188.1 CBW190.1 CBW207.1 CBW220.1 CBW263.1 CBW292.1 CBW306.1 CBW344.1 CBW350.1 CBW371.1 CBW389.1 CBX222.1 CBY568.1 CBY582.1 CBY639.1 CBY711.1 CBY731.1 CBY735.1 CBY828.1 CBY923.1 CBZ604.1 CBZ607.1 CBZ649.1 CBZ677.1 CBZ756.1 CBZ766.1 CBZ807.1 CBZ866.1 CBZ891.1 CCA099.1 CCA146.1 CCA157.1 CCA169.1 CCA199.1 CCA208.1 CCA271.1 CCA294.1 CCA308.1 CCA330.1 CCC308.1 CCC314.1 CCC352.1 CCC390.1 CCC480.1 CCC498.1 CCC552.1 CCC678.1 CCD137.1 CCD263.1 CCD265.1 CCD306.1 CCD314.1 CCD366.1 CCD376.1 CCD401.1 CCD445.1 CCD459.1 CCD520.1 CCD569.1 CCF570.1 CCF580.1 CCF587.1 CCF588.1 CCF724.1 CCF828.1 CCF829.1 CCF834.1 CCF896.1 CCF948.1 CCG680.1 CCG705.1 CCG859.1 CCG904.1 CCG992.1 CCH043.1 CCH592.1 CCH650.1 CCS136.1 CCS178.1 CCS372.1 CCS560.1 CCS611.1 CCS997.1 CCT206.1 CCT229.1 CCT276.1 CCV844.1 CCV915.1 CCV930.1 CCV984.1 CCV988.1 CCW786.1 CCW796.1 CCW800.1 CCW806.1 CCW987.1 CCW997.1 CCX049.1 CCX075.1 CCX112.1 CCX119.1 CCX628.1 CCX634.1 CCX637.1 CCX667.1 CCX668.1 CDH274.1 CDL624.1 CDL692.1 CDL707.1 CDP689.1 CDR334.1 CDR450.1 CDR476.1 CDR559.1 CDT549.1 CDT627.1 CDV091.1 CDV095.1 CDV177.1 CDV189.1 CDV227.1 CDV239.1 CDW233.1 CDW277.1 CDW519.1 CDW528.1 CDW571.1 CEB846.1 CEB866.1 CEB908.1 CEC605.1 CEC737.1 CEC784.1 CEF807.1 CEF837.1 CEH116.1 CEH885.1 CEH969.1 CEK130.1 CEK140.1 CEK198.1 CEK202.1 CEK272.1 CEK296.1 CEK394.1 CEK423.1 CEK478.1 CEM110.1 CEM170.1 CEN284.1 CEN327.1 CEN328.1 CEN397.1 CEN416.1 CEP521.1 CEP812.1 CER549.1 CER589.1 CER648.1 CET203.1 CET259.1 CET473.1 CET658.1 CFF123.1 CFF179.1 CFF217.1 CFG114.1 CFH121.1 CFL436.1 CFL541.1 CFL703.1 CFL813.1 CFM139.1 CFP990.1 CFR031.1 CFR055.1 CFR120.1 CFR186.1 CFR255.1 CFR300.1 CFR303.1 CFR416.1 CFR509.1 CFR533.1 CFR541.1 CFU095.1 CFU109.1 CFU125.1 CFU249.1 CFU250.1 CFU305.1 CFU318.1 CFU440.1 CFU478.1 CFU479.1 CFU482.1 CFU533.1 CFU616.1 CFU659.1 CFV557.1 CFV573.1 CFV574.1 CFV578.1 CFV591.1 CFV633.1 CFV659.1 CFV692.1 CFV696.1 CFV728.1 CFV740.1 CFV765.1 CFV776.1 CFV807.1 CFV811.1 CFV823.1 CFV830.1 CFX116.1 CFX121.1 CFX145.1 CFX184.1 CFX191.1 CFX193.1 CFX217.1 CFX257.1 CFX261.1 CFX267.1 CFX336.1 CFX382.1 CFX391.1 CFX394.1 CFZ943.1 CFZ970.1 CFZ984.1 CFZ998.1 CGA094.1 CGA178.1 CGA202.1 CGA225.1 CGA251.1 CGA276.1 CGA289.1 CGA349.1 CGA392.1 CGA518.1 CGA565.1 CGA603.1 CGA613.1 CGD078.1 CGD786.1 CGD794.1 CGD961.1 CGG883.1 CGH090.1 CGH343.1 CGH723.1 CGH724.1 CGH730.1 CGH851.1 CGH969.1 CGH980.1 CGH992.1 CGK932.1 CGK948.1 CGL012.1 CGL035.1 CGL131.1 CGL195.1 CGL251.1 CGL275.1 CGL289.1 CGL315.1 CGL323.1 CGL358.1 CGL360.1 CGL685.1 CGL686.1 CGL687.1 CGL696.1 CGL708.1 CGL719.1 CGL776.1 CGL788.1 CGL791.1 CGL820.1 CGP048.1 CGP057.1 CGP067.1 CGP073.1 CGP130.1 CGP144.1 CGP157.1 CGP159.1 CGP175.1 CGP197.1 CGP218.1 CGP223.1 CGP226.1 CGP261.1 CGP275.1 CGP300.1 CGP309.1 CGP338.1 CGP362.1 CGP363.1 CGP385.1 CGP395.1 CGP419.1 CGP423.1 CGP441.1 CGP500.1 CGP589.1 CGP613.1 CGP622.1 CGS697.1 CGS703.1 CGS705.1 CGS739.1 CGS757.1 CGS768.1 CGS831.1 CGS833.1 CGS865.1 CGS937.1 CGU046.1 CGU054.1 CGU056.1 CGU066.1 CGU081.1 CGU100.1 CGU141.1 CGU174.1 CGU185.1 CGU186.1 CGU213.1 CGU243.1 CGU247.1 CGU250.1 CGU253.1 CGW382.1 CGW427.1 CGX832.1 CHB176.1 CHB269.1 CHB332.1 CHE515.1 CHE540.1 CHE585.1 CHE730.1 CHE795.1 CHE811.1 CHE903.1 CHE905.1 CHE969.1 CHF107.1 CHF110.1 CHF216.1 CHF222.1 CHH664.1 CHH677.1 CHH740.1 CHH818.1 CHH824.1 CHH854.1 CHH862.1 CHH936.1 CHH947.1 CHH980.1 CHJ006.1 CHK284.1 CHK290.1 CHK342.1 CHK344.1 CHK384.1 CHK394.1 CHK405.1 CHK417.1 CHK467.1 CHK486.1 CHK499.1 CHK535.1 CHK624.1 CHK631.1 CHK690.1 CHL932.1 CHL935.1 CHL961.1 CHM058.1 CHM071.1 CHM085.1 CHM135.1 CHM137.1 CHM161.1 CHM162.1 CHM324.1 CHM325.1 CHM393.1 CHM480.1 CHM505.1 CHM532.1 CHM580.1 CHM587.1 CHP522.1 CHP572.1 CHP606.1 CHP655.1 CHP668.1 CHP701.1 CHT189.1 CHT199.1 CHT230.1 CHT244.1 CHT274.1 CHT327.1 CHT358.1 CHT379.1 CHT392.1 CHT421.1 CHT545.1 CHT550.1 CHT555.1 CHT775.1 CHT848.1 CHT851.1 CHV639.1 CHV703.1 CHV769.1 CHV774.1 CHV800.1 CHV869.1 CHV909.1 CHV911.1 CHV934.1 CHW238.1 CHW360.1 CHW381.1 CHW550.1 CHW565.1 CHW641.1 CHZ465.1 CHZ496.1 CHZ527.1 CHZ566.1 CHZ617.1 CHZ626.1 CHZ942.1 CHZ945.1 CHZ949.1 CHZ950.1 CHZ966.1 CHZ967.1 CHZ987.1 CHZ991.1 CHZ993.1 CJA045.1 CJA050.1 CJA052.1 CJA077.1 CJA092.1 CJA139.1 CJA178.1 CJA185.1 CJA209.1 CJA212.1 CJA231.1 CJA243.1 CJA272.1 CJA283.1 CJA284.1 CJA290.1 CJA302.1 CJA310.1 CJA314.1 CJA316.1 CJE813.1 CJE845.1 CJE855.1 CJE873.1 CJE923.1 CJE937.1 CJF026.1 CJF105.1 CJF108.1 CJF150.1 CJF177.1 CJF193.1 CJF972.1 CJF997.1 CJG001.1 CJG003.1 CJG016.1 CJG076.1 CJG078.1 CJG084.1 CJG107.1 CJG112.1 CJG187.1 CJJ344.1 CJJ353.1 CJJ355.1 CJJ380.1 CJJ394.1 CJJ422.1 CJJ474.1 CJJ574.1 CJJ577.1 CJJ642.1 CJJ670.1 CJJ677.1 CJJ713.1 CJJ715.1 CJM607.1 CJM660.1 CJM743.1 CJM791.1 CJM846.1 CJM851.1 CJN080.1 CJN157.1 CJN185.1 CJN198.1 CJP010.1 CJW478.1 CJY125.1 CJY153.1 CJY201.1 CJY219.1 CJY353.1 CJZ216.1 CJZ266.1 CJZ268.1 CKB674.1 YS1278.1 YW0010.1 YW0026.1 YW0249.1 YX9047.1 YY2921.1 YY5780.1 YY5876.1 YY6049.1 YY6118.1 YZ1487.1 YZ1509.1 YZ1514.1 YZ1515.1 YZ1531.1 YZ1574.1 YZ9624.1 YZ9636.1 YZ9648.1 YZ9651.1 YZ9654.1 YZ9691.1 YZ9709.1 YZ9712.1 YZ9732.1 YZ9755.1 YZ9768.1 YZ9826.1 YZ9857.1 ZK3232.1 ZK3240.1 ZK3253.1 ZK3370.1 ZV3919.1 ZW0302.1 ZW0587.1 ZW8865.1 ZW8885.1 ZW8943.1 ZW8991.1 ZY8811.1 ZY8821.1 ZY8915.1 ZY9164.1 ZY9264.1 ZY9274.1 ZY9280.1 ZY9296.1 ZY9321.1 ZY9345.1 ZY9349.1 ZY9359.1 ZY9361.1 ZY9369.1 ZY9406.1 OUS ABJ141.1 ABJ148.1 ABJ386.1 ABJ392.1 ABJ501.1 ABM312.1 ABM334.1 ACE780.1 ACE912.1 ACK135.1 ACK224.1 ACK231.1 ACK554.1 ACK632.1 ACP670.1 ACP676.1 ACP684.1 ACP756.1 ACP804.1 ACP885.1 ACP915.1 AFZ274.1 AFZ356.1 AFZ357.1 AFZ383.1 AHK779.1 AHK908.1 AHV609.1 ALZ177.1 ALZ260.1 AMN775.1 AMN787.1 AMN795.1 AMN799.1 AMN911.1 AMP020.1 AMP170.1 AMP201.1 AMP230.1 AMP248.1 AMS442.1 AMS665.1 AMS761.1 AMS948.1 AMT139.1 AMT189.1 AMT361.1 ANH134.1 ANT889.1 ANU305.1 APA824.1 APA826.1 APA834.1 APA854.1 APA859.1 APA885.1 APA891.1 APA903.1 APA908.1 APB200.1 APB325.1 APB515.1 APB667.1 APB675.1 APB923.1 APC031.1 APE431.1 APE526.1 APE600.1 APE629.1 APE698.1 APE700.1 APF147.1 APK939.1 APK950.1 APL209.1 APL299.1 APL300.1 APL310.1 APL312.1 APL342.1 APL392.1 APL561.1 APL608.1 APS149.1 APS254.1 APS280.1 APS367.1 APS384.1 APS395.1 APZ353.1 APZ378.1 APZ380.1 APZ531.1 APZ642.1 ARF911.1 ARF926.1 ARF942.1 ARG080.1 ARG305.1 ARG511.1 ARG563.1 ARG569.1 ARG595.1 ARJ354.1 ARJ999.1 ARK072.1 ARK081.1 ARK082.1 ARK087.1 ARK160.1 ARK291.1 ARK340.1 ARK361.1 ARK372.1 ARK393.1 ARK402.1 ARK753.1 ARL001.1 ARL006.1 ARL126.1 ASG313.1 ASM153.1 ASM503.1 ASM613.1 ASM633.1 ASR557.1 ASW136.1 ATC609.1 ATC664.1 ATC707.1 ATC876.1 ATC877.1 ATC879.1 ATC911.1 ATC934.1 ATD015.1 ATD076.1 ATD186.1 ATD220.1 ATD316.1 ATD354.1 ATD405.1 ATD585.1 ATD645.1 ATD698.1 ATD701.1 ATD909.1 ATE005.1 ATE028.1 ATE659.1 ATE841.1 ATF007.1 ATF009.1 ATF021.1 ATF038.1 ATF065.1 ATF176.1 ATF436.1 ATF453.1 ATJ737.1 ATJ906.1 ATJ958.1 ATK033.1 ATK066.1 ATK082.1 ATK085.1 ATK145.1 ATK159.1 ATK163.1 ATK167.1 ATK258.1 ATK326.1 ATT887.1 ATT888.1 ATT889.1 ATU067.1 ATU119.1 ATU353.1 ATU390.1 ATU519.1 ATU623.1 ATU707.1 ATU709.1 ATU874.1 ATU875.1 ATU963.1 ATV164.1 ATV979.1 ATW052.1 ATZ977.1 AUA027.1 AUA054.1 AUA132.1 AUA175.1 AUA198.1 AUA263.1 AUA272.1 AUA461.1 AUA483.1 AUA575.1 AUA668.1 AUA759.1 AUA907.1 AUB124.1 AUB409.1 AUE592.1 AUE604.1 AUE687.1 AUE950.1 AUH678.1 AUH741.1 AUH843.1 AUM059.1 AUM376.1 AUP128.1 AUV890.1 AUV902.1 AUV929.1 AVA023.1 AVA045.1 AVA051.1 AVA076.1 AVA253.1 AVA297.1 AVA407.1 AVA593.1 AVA604.1 AVA673.1 AVA743.1 AVA947.1 AVB149.1 AVD347.1 AVE355.1 AVE866.1 AVE877.1 AVF366.1 AVF564.1 AVF665.1 AVK762.1 AVL174.1 AVL196.1 AVM031.1 AVN634.1 AVN742.1 AVN758.1 AVR501.1 AVR524.1 AVR825.1 AVY637.1 AWC215.1 AWC239.1 AWC452.1 AWC480.1 AWC499.1 AWC517.1 AWH888.1 AWP704.1 AWP716.1 AWR182.1 AWR296.1 AWR460.1 AWR530.1 AWW952.1 AWX134.1 AWX145.1 AWX481.1 AWX678.1 AWX761.1 AWX839.1 AWX909.1 AWX928.1 AWX935.1 AWX940.1 AWX968.1 AWY014.1 AWY056.1 AWY067.1 AWY141.1 AWY152.1 AWY167.1 AWY208.1 AWY244.1 AWY263.1 AWY335.1 AXE739.1 AXE744.1 AXE871.1 AXF082.1 AXF355.1 AXF383.1 AXK222.1 AXK285.1 AXR199.1 AXU282.1 AXU294.1 AXV810.1 AXV835.1 AXV865.1 AXW005.1 AXW022.1 AXW152.1 AXY183.1 AXY259.1 AXY651.1 AXY702.1 AXY733.1 AXY737.1 AYD281.1 AYD369.1 AYD649.1 AYD670.1 AYD678.1 AYD777.1 AYE119.1 AYE166.1 AYE178.1 AYE349.1 AYE411.1 AYE512.1 AYE677.1 AYE833.1 AYE887.1 AYM149.1 AYM245.1 AYM301.1 AYM333.1 AYM350.1 AYM382.1 AYM425.1 AYM460.1 AYM538.1 AYM573.1 AYM600.1 AYM688.1 AYM719.1 AYM769.1 AYM773.1 AYM805.1 AYM818.1 AYM819.1 AYM848.1 AYM857.1 AYM867.1 AYM893.1 AYN001.1 AYN050.1 AYN058.1 AYP957.1 AYR244.1 AYR293.1 AYR353.1 AYR423.1 AYR466.1 AYR476.1 AYR679.1 AYR722.1 AYR887.1 AYR910.1 AYS372.1 AYS401.1 AYS434.1 AYS469.1 AYS477.1 AYS501.1 AYS528.1 AYS572.1 AYS580.1 AYS635.1 AYS797.1 AYS923.1 AYT004.1 AYT013.1 AYT022.1 AYT035.1 AYT069.1 AYT077.1 AYT091.1 AYT161.1 AYT344.1 AYX829.1 AYX892.1 AYX977.1 AYY079.1 AYY650.1 BAK641.1 BAK657.1 BAK659.1 BAK792.1 BAL029.1 BAL038.1 BAL131.1 BAR652.1 BAR721.1 BAR908.1 BAS005.1 BAS301.1 BAS354.1 BAS466.1 BAS483.1 BAS499.1 BAS548.1 BAS566.1 BAW229.1 BAW242.1 BAW807.1 BAW820.1 BAW835.1 BAW859.1 BAX243.1 BAX282.1 BAX304.1 BAX388.1 BAX392.1 BAX629.1 BAX679.1 BAY334.1 BAY418.1 BAY497.1 BAY916.1 BAY927.1 BAZ107.1 BBG633.1 BBS111.1 BBS118.1 BBS554.1 BBT165.1 BBW274.1 BBW781.1 BBW914.1 BBX586.1 BBY678.1 BBZ010.1 BBZ106.1 BCA154.1 BCA177.1 BCA563.1 BCA603.1 BCC630.1 BCC661.1 BCC679.1 BCC736.1 BCC757.1 BCC776.1 BCC970.1 BCC997.1 BCD238.1 BCP169.1 BCT030.1 BCT160.1 BCU939.1 BCU981.1 BCV044.1 BCV284.1 BCX932.1 BCY185.1 BDA034.1 BDA146.1 BDA161.1 BDA164.1 BDA212.1 BDA257.1 BDA304.1 BDA320.1 BDA472.1 BDB883.1 BDC089.1 BDC101.1 BDC120.1 BDC132.1 BDC303.1 BDC425.1 BDC453.1 BDE991.1 BDF022.1 BDF054.1 BDF131.1 BDF241.1 BDF271.1 BDF291.1 BDF307.1 BDJ205.1 BDJ278.1 BDJ453.1 BDJ482.1 BDJ503.1 BDJ549.1 BDJ689.1 BDL985.1 BDM006.1 BDM587.1 BDN547.1 BDU827.1 BDU891.1 BDU999.1 BDV004.1 BDV017.1 BDV060.1 BDV061.1 BDV186.1 BDV187.1 BDV193.1 BDV219.1 BDV273.1 BDZ229.1 BDZ271.1 BDZ284.1 BDZ405.1 BDZ455.1 BEB604.1 BEB767.1 BEB877.1 BEB915.1 BEL091.1 BEL147.1 BEP086.1 BET114.1 BET204.1 BEU398.1 BEU440.1 BEU494.1 BEU520.1 BEU537.1 BEU559.1 BEU566.1 BEX542.1 BEX564.1 BEX618.1 BEX659.1 BEX666.1 BEX670.1 BEX713.1 BEX849.1 BEX904.1 BEX913.1 BEY043.1 BEY072.1 BEY101.1 BEY128.1 BEY223.1 BEY237.1 BEY342.1 BEY374.1 BEY385.1 BEY397.1 BEY532.1 BEY540.1 BEY542.1 BEY665.1 BFA809.1 BFA831.1 BFB043.1 BFB192.1 BFB227.1 BFB706.1 BFE363.1 BFE509.1 BFE548.1 BFE591.1 BFE689.1 BFE881.1 BFE917.1 BFF048.1 BFF134.1 BFJ988.1 BFK222.1 BGD848.1 BGF487.1 BGH419.1 BGH830.1 BGK033.1 BGK063.1 BGK621.1 BGK745.1 BGK771.1 BGL030.1 BGL227.1 BGL249.1 BGL276.1 BGL302.1 BGL307.1 BGL379.1 BGL384.1 BGN711.1 BGN776.1 BGN794.1 BGN809.1 BGN860.1 BGN863.1 BGN869.1 BGN920.1 BGN946.1 BGP080.1 BGP081.1 BGP090.1 BGP115.1 BGP118.1 BGP177.1 BGP188.1 BGP194.1 BGP247.1 BGP303.1 BGS056.1 BGS057.1 BGS157.1 BGS584.1 BGS641.1 BGV778.1 BGY919.1 BGZ533.1 BHD433.1 BHR028.1 BHR156.1 BHU565.1 BHX191.1 BHX447.1 BHX504.1 BHX567.1 BHX580.1 BHX688.1 BHX717.1 BHX726.1 BHX896.1 BJE280.1 BJP430.1 BJY862.1 BJY992.1 BJZ015.1 BKC315.1 BKC866.1 BKD263.1 BKD361.1 BKF417.1 BKF559.1 BKF598.1 BKF919.1 BKF957.1 BKG028.1 BKG137.1 BKG177.1 BKG195.1 BKG304.1 BKG376.1 BKH672.1 BKH682.1 BKH684.1 BKH783.1 BKH846.1 BKH854.1 BKJ010.1 BKJ048.1 BKJ071.1 BKJ081.1 BKJ104.1 BKJ303.1 BKT336.1 BKT479.1 BKT493.1 BKT511.1 BLA549.1 BLA601.1 BLA623.1 BLA664.1 BLA673.1 BLA746.1 BLA836.1 BLC302.1 BLC314.1 BLC445.1 BLC486.1 BLC492.1 BLC585.1 BLC593.1 BLC601.1 BLC641.1 BLC668.1 BLC680.1 BLC722.1 BLC725.1 BLC760.1 BLC789.1 BLC851.1 BLC852.1 BLK992.1 BLL135.1 BLN995.1 BLP028.1 BLP212.1 BLP321.1 BLP358.1 BLP412.1 BLP418.1 BLP642.1 BLP679.1 BLT671.1 BLT727.1 BLT897.1 BLT955.1 BLU025.1 BLU141.1 BLU160.1 BLU168.1 BLY168.1 BLY203.1 BLY251.1 BLY287.1 BLY365.1 BLY555.1 BLY562.1 BLY580.1 BMC352.1 BMC459.1 BMD880.1 BMF711.1 BMG022.1 BMG043.1 BMG048.1 BMG062.1 BMG164.1 BMG165.1 BMG329.1 BMG452.1 BMJ779.1 BMJ937.1 BMK147.1 BMK285.1 BML069.1 BML131.1 BML345.1 BML363.1 BML501.1 BML513.1 BML584.1 BMM062.1 BMM180.1 BMM206.1 BMM223.1 BMM841.1 BMM875.1 BMM892.1 BMM939.1 BMM947.1 BMN042.1 BMN133.1 BMN144.1 BMN182.1 BMN230.1 BMN284.1 BMR375.1 BMR426.1 BMR501.1 BMT207.1 BMT228.1 BMT230.1 BMT234.1 BMT236.1 BMT247.1 BMT263.1 BMT328.1 BMT489.1 BMT517.1 BMW580.1 BMZ720.1 BMZ931.1 BNA150.1 BNC159.1 BNJ236.1 BNW987.1 BNX009.1 BPD102.1 BPK304.1 BPK633.1 BPM667.1 BPM819.1 BPS731.1 BPV110.1 BRG627.1 BRH073.1 BRK992.1 BRM863.1 BRM877.1 BRN012.1 BRN036.1 BRN054.1 BRN367.1 BRN374.1 BRN381.1 BRN410.1 BRN522.1 BRN644.1 BRN669.1 BRN808.1 BRR237.1 BRR253.1 BRR507.1 BRR901.1 BRT702.1 BRU121.1 BRV339.1 BRX053.1 BRX132.1 BRX532.1 BRX681.1 BRX707.1 BRX762.1 BRY527.1 BRY530.1 BRY563.1 BRZ272.1 BRZ461.1 BRZ476.1 BSA288.1 BSA408.1 BSA557.1 BSA675.1 BSD147.1 BSD183.1 BSD194.1 BSD212.1 BSD244.1 BSD297.1 BSD300.1 BSE948.1 BSF026.1 BSF110.1 BSF151.1 BSF273.1 BSJ903.1 BTB141.1 BTB412.1 BTC486.1 BTC669.1 BTC716.1 BTD069.1 BTD158.1 BTD225.1 BTJ148.1 BTJ197.1 BTM657.1 BTM877.1 BTM971.1 BTN446.1 BTN682.1 BTS107.1 BTS800.1 BTS897.1 BTS957.1 BTS970.1 BTS985.1 BTW681.1 BTW691.1 BTW732.1 BTW817.1 BTW987.1 BTX025.1 BTX055.1 BTX255.1 BTX340.1 BTX391.1 BTX771.1 BTX862.1 BTX866.1 BUA542.1 BUA678.1 BUA717.1 BUA722.1 BUA781.1 BUA911.1 BUA949.1 BUB189.1 BUB821.1 BUF276.1 BUF284.1 BUF355.1 BUF420.1 BUF470.1 BUF574.1 BUF615.1 BUF655.1 BUJ872.1 BUJ924.1 BUJ936.1 BUJ942.1 BUK116.1 BUK709.1 BUL536.1 BUL624.1 BUP193.1 BUP222.1 BUP228.1 BUP309.1 BUP339.1 BUP422.1 BUP452.1 BUP468.1 BUP469.1 BUP511.1 BUP722.1 BUP766.1 BUP768.1 BUP782.1 BUR846.1 BUR930.1 BUR998.1 BUS003.1 BUS143.1 BUV018.1 BUV081.1 BUV272.1 BUV301.1 BUV348.1 BUV425.1 BUV474.1 BUV487.1 BUV591.1 BUW586.1 BUW659.1 BUW720.1 BUX369.1 BUX419.1 BUX702.1 BUX704.1 BUY439.1 BUY467.1 BUY521.1 BUY875.1 BUY996.1 BUZ097.1 BVD868.1 BVD934.1 BVD951.1 BVE082.1 BVE234.1 BVE237.1 BVE284.1 BVE301.1 BVE318.1 BVK023.1 BVK063.1 BVL022.1 BVL275.1 BVL521.1 BVL603.1 BVL641.1 BVL688.1 BVL793.1 BVP221.1 BVP364.1 BVP424.1 BVX198.1 BVX240.1 BVX254.1 BVX332.1 BVX361.1 BVX380.1 BVX417.1 BVX549.1 BVX574.1 BVY386.1 BVY451.1 BVY504.1 BVY547.1 BVZ742.1 BVZ796.1 BVZ915.1 BVZ971.1 BWC163.1 BWC166.1 BWC191.1 BWC200.1 BWC211.1 BWC242.1 BWC397.1 BWC511.1 BWC536.1 BWC574.1 BWE867.1 BWE881.1 BWF132.1 BWG067.1 BWG069.1 BWG081.1 BWG141.1 BWG166.1 BWG241.1 BWG283.1 BWG419.1 BWJ619.1 BWJ651.1 BWJ684.1 BWJ743.1 BWK032.1 BWP561.1 BWV536.1 BWV559.1 BWV561.1 BWV660.1 BXB050.1 BXB433.1 BXB473.1 BXB509.1 BXH686.1 BXH843.1 BXH910.1 BXJ839.1 BXK403.1 BXK457.1 BXK475.1 BXK703.1 BXK751.1 BXK822.1 BXK996.1 BXL051.1 BXM527.1 BXM650.1 BXM707.1 BXM724.1 BXM760.1 BXM763.1 BXM782.1 BXM789.1 BXM805.1 BXM935.1 BXM936.1 BXN562.1 BXN762.1 BXN787.1 BXP083.1 BXP088.1 BXR187.1 BXR226.1 BXR228.1 BXT598.1 BXT623.1 BXW379.1 BXW391.1 BXW505.1 BXW533.1 BXW598.1 BXY645.1 BXY746.1 BXY747.1 BYD744.1 BYD754.1 BYD815.1 BYD841.1 BYD869.1 BYD873.1 BYD999.1 BYE036.1 BYE045.1 BYE051.1 BYE096.1 BYE107.1 BYE116.1 BYE205.1 BYE210.1 BYE235.1 BYE484.1 BYG463.1 BYG468.1 BYG480.1 BYG555.1 BYG564.1 BYG710.1 BYG732.1 BYG834.1 BYL777.1 BYL840.1 BYL854.1 BYL913.1 BYL981.1 BYM005.1 BYM042.1 BYM062.1 BYR453.1 BYR496.1 BYR526.1 BYR528.1 BYR573.1 BYU429.1 BYV311.1 CAE026.1 CAE096.1 CAF758.1 CAF975.1 CAG778.1 CAG966.1 CAG982.1 CAH025.1 CAH031.1 CAH867.1 CAH873.1 CAH948.1 CAH984.1 CAJ299.1 CAJ343.1 CAK520.1 CAK648.1 CAK665.1 CAK669.1 CAK697.1 CAK803.1 CAM524.1 CAM539.1 CAN511.1 CAN514.1 CAN564.1 CAS397.1 CAT192.1 CAV718.1 CAV738.1 CAX043.1 CAX105.1 CAX170.1 CAX273.1 CAX460.1 CAY169.1 CAY238.1 CAY243.1 CBA332.1 CBA370.1 CBB249.1 CBB260.1 CBB278.1 CBB304.1 CBB433.1 CBB475.1 CBB500.1 CBC244.1 CBC254.1 CBC259.1 CBC280.1 CBC282.1 CBC290.1 CBC294.1 CBC296.1 CBC391.1 CBC418.1 CBC425.1 CBC448.1 CBC509.1 CBC514.1 CBC520.1 CBC553.1 CBC556.1 CBC557.1 CBC580.1 CBC610.1 CBC611.1 CBC612.1 CBC616.1 CBC624.1 CBF866.1 CBG708.1 CBG712.1 CBH251.1 CBH274.1 CBH353.1 CBH534.1 CBK361.1 CBK419.1 CBK697.1 CBM236.1 CBM368.1 CBM376.1 CBM435.1 CBM447.1 CBM465.1 CBM583.1 CBM588.1 CBM592.1 CBR921.1 CBR930.1 CBS011.1 CBT987.1 CBU093.1 CBU125.1 CBU277.1 CBW105.1 CBW136.1 CBW210.1 CBW212.1 CBW305.1 CBW380.1 CBX069.1 CBX153.1 CBX242.1 CBX249.1 CBY577.1 CBY589.1 CBY632.1 CBY663.1 CBY671.1 CBY672.1 CBY700.1 CBY757.1 CBY790.1 CBY835.1 CBY873.1 CBY876.1 CBY909.1 CBY911.1 CBY936.1 CBZ721.1 CCA103.1 CCA137.1 CCA232.1 CCC628.1 CCC645.1 CCD143.1 CCD246.1 CCD280.1 CCD299.1 CCD316.1 CCD447.1 CCD526.1 CCD578.1 CCD584.1 CCD588.1 CCD592.1 CCF628.1 CCF655.1 CCF679.1 CCF735.1 CCF899.1 CCF923.1 CCG703.1 CCG733.1 CCG743.1 CCG790.1 CCG842.1 CCG917.1 CCG941.1 CCG966.1 CCG995.1 CCH012.1 CCS156.1 CCS255.1 CCS466.1 CCV843.1 CCV987.1 CCW739.1 CCW742.1 CCW820.1 CCX038.1 CCX062.1 CCX084.1 CCX099.1 CCX101.1 CCZ074.1 CDR535.1 CDV128.1 CDV184.1 CDW533.1 CDW598.1 CEB897.1 CEG029.1 CEH452.1 CEK456.1 CEP478.1 CEP512.1 CEP517.1 CEP570.1 CEP573.1 CER406.1 CER502.1 CET330.1 CET789.1 CFE899.1 CFF102.1 CFF171.1 CFL746.1 CFP967.1 CFU275.1 CFU454.1 CFU558.1 CFU599.1 CFU637.1 CFV512.1 CFV536.1 CFV571.1 CFV705.1 CFV795.1 CFV808.1 CFX138.1 CFX140.1 CFX179.1 CFX201.1 CFX214.1 CFX301.1 CFX401.1 CFZ937.1 CGA006.1 CGA019.1 CGA022.1 CGA122.1 CGA227.1 CGA239.1 CGA343.1 CGA530.1 CGA543.1 CGD806.1 CGD820.1 CGG926.1 CGG939.1 CGH092.1 CGH118.1 CGH313.1 CGH430.1 CGH779.1 CGH785.1 CGH811.1 CGH885.1 CGL026.1 CGL064.1 CGL099.1 CGL263.1 CGL298.1 CGL299.1 CGL341.1 CGL365.1 CGL411.1 CGL427.1 CGL729.1 CGL745.1 CGL761.1 CGL762.1 CGL767.1 CGL771.1 CGL773.1 CGL802.1 CGL813.1 CGP070.1 CGP074.1 CGP115.1 CGP183.1 CGP204.1 CGP211.1 CGP262.1 CGP286.1 CGP303.1 CGP448.1 CGS702.1 CGS704.1 CGS744.1 CGS789.1 CGS806.1 CGS813.1 CGS821.1 CGS848.1 CGS850.1 CGU053.1 CGU059.1 CGU085.1 CGU092.1 CGU129.1 CGU142.1 CGU240.1 CGU252.1 CGU297.1 CGU302.1 CGU303.1 CGW270.1 CGY885.1 CHB090.1 CHB130.1 CHB184.1 CHB228.1 CHE807.1 CHE852.1 CHE859.1 CHF028.1 CHF051.1 CHF189.1 CHH652.1 CHH748.1 CHH796.1 CHH798.1 CHH808.1 CHH834.1 CHH845.1 CHH873.1 CHK299.1 CHK381.1 CHK492.1 CHK574.1 CHK579.1 CHK636.1 CHK662.1 CHL944.1 CHL953.1 CHM024.1 CHM052.1 CHM123.1 CHM201.1 CHM235.1 CHM270.1 CHM350.1 CHM402.1 CHM429.1 CHM463.1 CHM495.1 CHM546.1 CHM569.1 CHM591.1 CHP624.1 CHP729.1 CHR146.1 CHR202.1 CHT294.1 CHT322.1 CHT466.1 CHT533.1 CHT534.1 CHT537.1 CHT687.1 CHT833.1 CHV708.1 CHV746.1 CHW477.1 CHW483.1 CHW520.1 CHW558.1 CHZ470.1 YW0022.1 YW0038.1 YW0089.1 YW0185.1 YW0238.1 YW0282.1 YW0387.1 YW0483.1 YX8973.1 YX8974.1 YX9006.1 YX9010.1 YX9012.1 YX9026.1 YX9031.1 YX9048.1 YX9058.1 YX9063.1 YX9149.1 YY3001.1 YY3014.1 YY6058.1 YZ1455.1 YZ1474.1 YZ1543.1 YZ1550.1 YZ1556.1 YZ1613.1 YZ9626.1 YZ9694.1 YZ9704.1 YZ9730.1 YZ9748.1 YZ9758.1 YZ9771.1 YZ9780.1 YZ9817.1 YZ9897.1 YZ9902.1 ZY8841.1 ZY9281.1 ZY9324.1 ZY9483.1
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Infinity™ 5St. Jude Medical, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR2 recalls28 reports
Product lines
Other devices under the same FDA submission
110 devices cleared under P010032, P010032S096, P010032S151 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
- Penta™ 3228St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11 deaths5,201 reports
- Eon Mini™ 3788St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 death4,666 reports
- Prodigy MRI™ 3772St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths3,162 reports
- Octrode™ 3189St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF8 deaths2,238 reports
- Protégé MRI™ 3771St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1,997 reports
- Eterna™ 32400St. Jude Medical, Inc. · Stimulator, spinal-cord, totally implanted for pain relief1 recall5 deaths1,547 reports
- Octrode® 3186Advanced Neuromodulation Systems, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths1,455 reports
- Swift-Lock™ 1192St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1,400 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92688).
- Z-2505-2023 · Open, ClassifiedProclaim XR 5 Implantable Pulse Genterator REF 3660 (previously known as Proclaim" 5 Elite IPG) Model 3660 Product Description: The implantable pulse generator (IPG) is an electr…
- Z-2506-2023 · Open, ClassifiedProclaim" XR 7 Implantable Pulse Generator REF 3662 (previously known as Proclaim" 7 Elite IPG) Product Description: The implantable pulse generator (IPG) is an electronic device …
- Z-2507-2023 · Open, ClassifiedProclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext…
- Z-2508-2023 · Open, ClassifiedProclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more ext…
- Z-2509-2023 · Open, ClassifiedProclaim DRG Implantable Pulse Generator REF 3664 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extens…
- Z-2511-2023 · Open, ClassifiedInfinity 7 Implantable Pulse Generator REF 6662 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensio…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page