Skip to content

Alaris PCA Module 8120 recallZ-0186-2024 · CareFusion 303, Inc. · initiated Sep 15, 2023

Class I Open, Classified All recalls posted in November 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.

FDA root cause category: Under Investigation by firm

What the company is doing

On September 15, 2023, CareFusion a subsidiary of Becton Dickinson (BD) posted an "Urgent Medical Device Labeling Correction" on their website. On October 4, 2023 CareFusion mailed a copy of the correction notification via FedEx. CareFusion asked consignees to take the following actions: 1. You should cease use of Cardinal Health branded Monoject" syringes with the BD Alaris" Syringe and PCA Modules; except for SKU 8881135609, which is not impacted by this notification. 2. Please refer to Attachments 1 and 2 of this notice for a full selection of syringes that are compatible with the BD Alaris" Syringe and PCA Modules. 3. Replace your current BD Alaris" Syringe and PCA Module Compatibility Lists with Attachments 1 and 2 of this notice. 4. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred BD Alaris" Infusion System devices to. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787

Recall details

Recall number
Z-0186-2024
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Sep 15, 2023
Posted by the FDA
Nov 7, 2023
Recalling firm
CareFusion 303, Inc.
Quantity
86,393 units
Product code
FRN
FDA submissions
K072105, K133532, K211218, K051641
Distribution
Worldwide Distribution: US (nationwide): AK, AL, AR, AS, AZ, CA, CO, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; OUS (foreign): Canada, Germany, Cuba, Italy, Japan, South Korea, Turkey, United Arab Emirates, Australia, Bahrain, Botswana, Kuwait, Mexico, Netherlands, Philippines, Pakistan, Qatar, South Africa, Singapore, Belgium, Israel, India, New Zealand, Malaysia, Taiwan, Switzerland

Product description

Alaris PCA Module 8120

Code information (models, lots, serial numbers)

All Lots/ UDI-DI: 10885403516023,10885403515231,10885403801549,10885403515248, 10885403801532,10885403515224,10885403801518

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

10 devices cleared under K072105, K133532, K211218, K051641 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 93106).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page