Skip to content

Monitron II Waveform Analyzer, REF: S00015-1, used with… recallZ-0259-2024 · Percussionaire Corporation · initiated Sep 14, 2020

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Percussionaire Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Note this recall occurred in 2020 and 2021. Reports of screen freeze on visual displays of waveform analyzers that are paired with ventilators. Subsequently, waveform analyzers were serviced to install PAL microprocessor chips and firmware.

FDA root cause category: Device Design

What the company is doing

Waveform analyzers were serviced from September 2020 through September 2021.

Recall details

Recall number
Z-0259-2024
Status
Completed
Initiated by the company
Sep 14, 2020
Posted by the FDA
Nov 7, 2023
Recalling firm
Percussionaire Corporation
Quantity
212
Product code
CBK
FDA submissions
K905235
Distribution
US: CA, ND, MS, MD

Product description

Monitron II Waveform Analyzer, REF: S00015-1, used with ventilators.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

UDI-DI: 00849436000068, Serial Numbers: WF21150, WF21159, WF21163, WF21164, WF21167, WF21171, WF21371, WF21372, WF21377, WF21378, WF21158, WF21391, WF21165, WF21166, WF21168, WF21169, WF21170, WF21172, WF21173, WF21174, WF21175, WF21176, WF21177, WF21178, WF21179, WF21182, WF21183, WF21184, WF21185, WF21186, WF21187, WF21188, WF21189, WF21190, WF21192, WF21193, WF21194, WF21195, WF21196, WF21197, WF21198, WF21199, WF21200, WF21201, WF21202, WF21203, WF21204, WF21205, WF21206, WF21207, WF21208, WF21209, WF21210, WF21211, WF21212, WF21213, WF21214, WF21215, WF21217, WF21218, WF21219, WF21220, WF21221, WF21222, WF21223, WF21224, WF21225, WF21226, WF21227, WF21228, WF21229, WF21230, WF21231, WF21232, WF21233, WF21234, WF21235, WF21236, WF21237, WF21238, WF21239, WF21240, WF21241, WF21242, WF21243, WF21244, WF21245, WF21246, WF21247, WF21248, WF21249, WF21250, WF21251, WF21252, WF21253, WF21254, WF21255, WF21256, WF21257, WF21258, WF21259, WF21260, WF21261, WF21262, WF21263, WF21264, WF21265, WF21266, WF21267, WF21268, WF21269, WF21270, WF21271, WF21272, WF21273, WF21274, WF21275, WF21276, WF21277, WF21278, WF21279, WF21280, WF21281, WF21282, WF21283, WF21284, WF21285, WF21286, WF21287, WF21288, WF21289, WF21290, WF21291, WF21292, WF21293, WF21294, WF21295, WF21296, WF21297, WF21298, WF21299, WF21300, WF21301, WF21302, WF21303, WF21304, WF21305, WF21306, WF21307, WF21308, WF21309, WF21310, WF21311, WF21312, WF21313, WF21314, WF21315, WF21316, WF21317, WF21318, WF21319, WF21320, WF21321, WF21322, WF21323, WF21324, WF21325, WF21326, WF21327, WF21328, WF21329, WF21330, WF21331, WF21332, WF21333, WF21334, WF21335, WF21336, WF21337, WF21338, WF21339, WF21340, WF21341, WF21342, WF21343, WF21344, WF21345, WF21346, WF21347, WF21348, WF21349, WF21350, WF21351, WF21352, WF21353, WF21354, WF21355, WF21356, WF21357, WF21358, WF21359, WF21360, WF21361, WF21362, WF21363, WF21364, WF21365, WF21366, WF21367, WF21368, WF21369, WF21370

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

6 devices cleared under K905235 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates