EVIS EXERA III Video System Center Model Number CV-190.… recallZ-0056-2025 · Olympus Corporation of the Americas · initiated Sep 14, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Certain serial numbers of the CV-190 do not start up properly because parts that deviated from the specification were assembled into the power supply unit.
FDA root cause category: Nonconforming Material/Component
What the company is doing
An URGENT: MEDICAL DEVICE FIELD CORRECTIVE ACTION notification letter dated 9/14/23 was sent to customers. Action steps to be taken by the end user: Olympus has determined based upon our distribution records that your facility is in possession of one or more affected CV-190 with a serial number shown in Attachment 1. Olympus requires that you take the following actions: 1. Inspect your inventory for the referenced products and identify any devices with the model number and serial number specified in Attachment 1. Cease use of products and quarantine any affected products. Please check all areas of the hospital to determine if any of these products remain in inventory. 2. Please contact our Olympus Repair Center at 1-800-848-9024, option 3, to arrange for a service to exchange the power supply unit in affected devices. Olympus regrets any inconvenience caused and fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at 647-999-3203 or [email protected].
Recall details
- Recall number
- Z-0056-2025
- Status
- Open, Classified
- Initiated by the company
- Sep 14, 2023
- Posted by the FDA
- Oct 11, 2024
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 45 units
- Product code
- FDF
- FDA submissions
- K131780
- Distribution
- US Nationwide distribution.
Product description
EVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera heads, endoscopes, light sources, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation
Code information (models, lots, serial numbers)
UDI-DI: 04953170298622; Serial Numbers: 7384484 7384422 7384487 7384478 7384479 7384442 7384441 7384485 7384451 7384455 7384481 7384470 7384424 7384423 7384489 7384531 7384516 7384492 7384512 7384543 7384483 7384624 7384623 7384497 7384495 7384494 7384533 7384622 7384660 7384654 7384656 7384647 7384658 7384675 7384655 7384704 7384709 7384687 7384706 7384661 7384662 7384703 7384670 7384679 7384678
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EVIS EXERA III VIDEO SYSTEM CENTER OLYMPUS CV-190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall7,688 reports
Other devices under the same FDA submission
21 devices cleared under K131780 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls1,315 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1 death1,194 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall460 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid250 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-PH190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall139 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190LOlympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid99 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-PH190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall47 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid86 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page