Why it was recalled
Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
FDA root cause category: Process control
What the company is doing
On October 16, 2023, the firm notified affected customers through URGENT MEDICAL DEVICE REMOVAL letters. Customers were instructed to inspect their inventory and identify affected devices. Affected devices should be quarantined and not used. Olympus Customer Solutions will contact customers by phone and email to arrange the return and repair of the affected product. If you require additional information, please do not hesitate to contact Olympus at 647-999-3203 or [email protected].
Recall details
- Recall number
- Z-0415-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 16, 2023
- Posted by the FDA
- Nov 29, 2023
- Recalling firm
- Aizu Olympus Co., Ltd.
- Quantity
- 24 incorrectly repaired units
- Product code
- FDF
- FDA submissions
- K131780
- Distribution
- Domestic distribution to FL, KY, MA, NJ, NY, and PA
Product description
CF-HQ190L EVIS EXERA III COLONOVIDEOSCOPE
Code information (models, lots, serial numbers)
UDI-DI 04953170305115 Serial Numbers 2635812, 2416423
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
Product lines
Other devices under the same FDA submission
21 devices cleared under K131780 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- EVIS EXERA III VIDEO SYSTEM CENTER OLYMPUS CV-190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall7,688 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls1,315 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1 death1,194 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall460 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid250 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-PH190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall139 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190LOlympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid99 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-PH190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall47 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid86 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93335).
- Z-0416-2024 · Open, ClassifiedGIF-H190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
- Z-0417-2024 · Open, ClassifiedGIF-HQ190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
- Z-0418-2024 · Open, ClassifiedPCF-H190DL, EVIS EXERA III COLONOVIDEOSCOPE
- Z-0419-2024 · Open, ClassifiedPCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page