Philips Azurion system, Interventional fluoroscopic x-ray… recallZ-0835-2024 · Philips Medical Systems Nederland B.V. · initiated Dec 12, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the procedure.

FDA root cause category: Under Investigation by firm

What the company is doing

On December 12, 2023, the firm notified customers via certified mailing of URGENT Medical Device Correction letters. Customers may identify affected products by the system product name and model number, which can be found on the System Identification Label located on the system stand. The software version of the Philips Azurion system can be identified during start-up. Customers were advised to keep the number of studies in the patient database as low as possible, but at least below 500. Customers should keep the Urgent Medical Device Correction letter with the documentation of the system until Philips corrects the system. Ensure that all users of the system are notified of the issue. Philips is working on a software release that will correct this issue. Philips representatives will contact affected customers to schedule the correction. Please contact your local Philips representative for further information or support concerning this issue.

Recall details

Recall number
Z-0835-2024
Status
Open, Classified
Initiated by the company
Dec 12, 2023
Posted by the FDA
Jan 26, 2024
Recalling firm
Philips Medical Systems Nederland B.V.
Quantity
1,799 US; 5,747 OUS
Product code
OWB
FDA submissions
K200917
Distribution
Worldwide - US Nationwide distribution.

Product description

Philips Azurion system, Interventional fluoroscopic x-ray system. System Product Name - Model Numbers Azurion 3M12 -722063, 722221; Azurion 3M15 - 722064, 722222; Azurion 5M12 - 722227; Azurion 5M20 - 722228; Azurion 7B12/12 - 722067, 722225; Azurion 7B20/15 - 722068, 722226; Azurion 7M12 - 722078, 722223; Azurion 7M20 - 722079, 722224; Azurion 3 M15 (China) - 722280;

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

System Product Name - UDI-DI Azurion 3M12 - 00884838085275, 00884838099203, 00884838101890; Azurion 3M15 - 00884838085282, 00884838099210; Azurion 5M12 - 00884838099227; Azurion 5M20 - 00884838099234; Azurion 7B12/12 - 00884838059290, 00884838064775, 00884838085350, 00884838099265; Azurion 7B20/15 - 00884838059207, 00884838064775, 00884838085367, 00884838099272, 00884838101883; Azurion 7M12 - 00884838085251, 00884838099241, 00884838101890; Azurion 7M20 - 00884838059184, 00884838059191, 00884838059207, 00884838059221, 00884838059290, 00884838064775, 00884838085268, 00884838086777, 00884838099258, 00884838101883, 00884838101890; Azurion 3 M15 (China) - 00884838085282, 00884838099210, 00884838103276; Software Version Numbers R1.0, R1.1, R1.2, R2.0, R2.1, and R2.2

Devices named in this recall

23 devices in the FDA device registry whose identifier appears in the recall's code information.