Philips Azurion system, Interventional fluoroscopic x-ray… recallZ-0835-2024 · Philips Medical Systems Nederland B.V. · initiated Dec 12, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the procedure.
FDA root cause category: Under Investigation by firm
What the company is doing
On December 12, 2023, the firm notified customers via certified mailing of URGENT Medical Device Correction letters. Customers may identify affected products by the system product name and model number, which can be found on the System Identification Label located on the system stand. The software version of the Philips Azurion system can be identified during start-up. Customers were advised to keep the number of studies in the patient database as low as possible, but at least below 500. Customers should keep the Urgent Medical Device Correction letter with the documentation of the system until Philips corrects the system. Ensure that all users of the system are notified of the issue. Philips is working on a software release that will correct this issue. Philips representatives will contact affected customers to schedule the correction. Please contact your local Philips representative for further information or support concerning this issue.
Recall details
- Recall number
- Z-0835-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 12, 2023
- Posted by the FDA
- Jan 26, 2024
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 1,799 US; 5,747 OUS
- Product code
- OWB
- FDA submissions
- K200917
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Philips Azurion system, Interventional fluoroscopic x-ray system. System Product Name - Model Numbers Azurion 3M12 -722063, 722221; Azurion 3M15 - 722064, 722222; Azurion 5M12 - 722227; Azurion 5M20 - 722228; Azurion 7B12/12 - 722067, 722225; Azurion 7B20/15 - 722068, 722226; Azurion 7M12 - 722078, 722223; Azurion 7M20 - 722079, 722224; Azurion 3 M15 (China) - 722280;
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
System Product Name - UDI-DI Azurion 3M12 - 00884838085275, 00884838099203, 00884838101890; Azurion 3M15 - 00884838085282, 00884838099210; Azurion 5M12 - 00884838099227; Azurion 5M20 - 00884838099234; Azurion 7B12/12 - 00884838059290, 00884838064775, 00884838085350, 00884838099265; Azurion 7B20/15 - 00884838059207, 00884838064775, 00884838085367, 00884838099272, 00884838101883; Azurion 7M12 - 00884838085251, 00884838099241, 00884838101890; Azurion 7M20 - 00884838059184, 00884838059191, 00884838059207, 00884838059221, 00884838059290, 00884838064775, 00884838085268, 00884838086777, 00884838099258, 00884838101883, 00884838101890; Azurion 3 M15 (China) - 00884838085282, 00884838099210, 00884838103276; Software Version Numbers R1.0, R1.1, R1.2, R2.0, R2.1, and R2.2
Devices named in this recall
23 devices in the FDA device registry whose identifier appears in the recall's code information.
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System32 recalls2 deaths2,308 reports
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System28 recalls1 death2,176 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths1,087 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths835 reports
- Azurion Azurion 7 B20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls1 death634 reports
- Azurion Azurion 7 B20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls594 reports
- Azurion Azurion 5 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls201 reports
- Azurion Azurion 7 B12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls232 reports
- Azurion Azurion 7 B12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls222 reports
- Azurion Azurion 5 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls1 death197 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths205 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls104 reports
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls121 reports
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls73 reports
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- HeartNavigator HeartNavigator - 2Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System6 recalls3 reports
- Dynamic Coronary Roadmap Dynamic Coronary Roadmap - 2Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System4 recalls0 reports
- EmboGuide EmboGuide - 1Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System3 recalls2 reports
- VesselNavigator VesselNavigator - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological3 recalls1 report
- 3D Roadmap 3D Roadmap - 1Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System3 recalls0 reports
- Dynamic Coronary Roadmap Dynamic Coronary Roadmap - 3Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System2 recalls10 reports
- StentBoost Live StentBoost Live - 2Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System2 recalls2 reports
- EP navigator EP navigator - 5Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System2 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page