Azurion recallZ-1176-2024 · Philips Medical Systems Nederland B.V. · initiated Feb 6, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
FDA root cause category: Under Investigation by firm
What the company is doing
Philips notified U.S customers by Urgent Medical Device Correction letter on 06-Feb-2024 via certified mailing via United States Postal Service. " Letters outside of the U.S.A. will be distributed through the Philips Markets Organizations. Letter states reason for recall, health risk, and action to take: Circulate this Letter to all clinical staff and any service personnel that may service the system so that they are aware of the issue. " Keep this Urgent Medical Device Correction Letter with the documentation of the system until Philips implements the correction in your system. " If the Certeray generator fails, please stop using the system and contact the Customer Care Solutions Center. " Place a copy of this Letter in a visible place in the Control and Technical rooms (e.g., posted on the door). " Please complete and return the attached response form (on page 3) to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction and understanding of the issue and required actions to be taken. 5. Actions planned by Philips IGTS to correct the problem As a remedy, Philips will replace the Power Inverter in the Azurion generator cabinet in all affected Azurion systems through the implementation of Field Change Order (FCO72200544). If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Recall details
- Recall number
- Z-1176-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 6, 2024
- Posted by the FDA
- Feb 22, 2024
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 14 units
- Product code
- OWB
- FDA submissions
- K200917
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of NY, PA, PR, TN, TX, VA, WA and the countries of Algeria, Armenia, Austria, Brazil, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Republic of Kuwait, Lithuania, Macao, Macedonia, Malaysia, Morocco, Netherlands, Norway, Oman, Palestine, State of Panama, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, Ukraine, United Kingdom.
Product description
Azurion 3 with a Certeray generator -To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722221 (2) 722222 (3) 722280
Code information (models, lots, serial numbers)
OUS: (1) Model: 722221 SN: 66 (01)00884838099203(21)66; SN: 91 (01)00884838099203(21)91; SN: 104 (01)00884838099258(21)104, (2) Model: 722222 SN: UDI: 165 (01)00884838099210(21)165 183 (01)00884838099210(21)183 189 (01)00884838099210(21)189 182 (01)00884838099210(21)182 185 (01)00884838099210(21)185 159 (01)00884838099210(21)159 184 (01)00884838099210(21)184 190 (01)00884838099210(21)190 174 (01)00884838099210(21)174 177 (01)00884838099210(21)177 (3)Model: 722280 20 (01)00884838103276(21)20
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System32 recalls2 deaths2,308 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls104 reports
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls121 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93966).
- Z-1177-2024 · Open, ClassifiedAzurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722227 (2) 722228
- Z-1178-2024 · Open, ClassifiedAzurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722223, (2) 722224, (3) 72222…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗