Azurion recallZ-1178-2024 · Philips Medical Systems Nederland B.V. · initiated Feb 6, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
FDA root cause category: Under Investigation by firm
What the company is doing
Philips notified U.S customers by Urgent Medical Device Correction letter on 06-Feb-2024 via certified mailing via United States Postal Service. " Letters outside of the U.S.A. will be distributed through the Philips Markets Organizations. Letter states reason for recall, health risk, and action to take: Circulate this Letter to all clinical staff and any service personnel that may service the system so that they are aware of the issue. " Keep this Urgent Medical Device Correction Letter with the documentation of the system until Philips implements the correction in your system. " If the Certeray generator fails, please stop using the system and contact the Customer Care Solutions Center. " Place a copy of this Letter in a visible place in the Control and Technical rooms (e.g., posted on the door). " Please complete and return the attached response form (on page 3) to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction and understanding of the issue and required actions to be taken. 5. Actions planned by Philips IGTS to correct the problem As a remedy, Philips will replace the Power Inverter in the Azurion generator cabinet in all affected Azurion systems through the implementation of Field Change Order (FCO72200544). If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Recall details
- Recall number
- Z-1178-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 6, 2024
- Posted by the FDA
- Feb 22, 2024
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 10 units US; 124 units OUS
- Product code
- OWB
- FDA submissions
- K200917
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of NY, PA, PR, TN, TX, VA, WA and the countries of Algeria, Armenia, Austria, Brazil, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Republic of Kuwait, Lithuania, Macao, Macedonia, Malaysia, Morocco, Netherlands, Norway, Oman, Palestine, State of Panama, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, Ukraine, United Kingdom.
Product description
Azurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722223, (2) 722224, (3) 722225, (4) 722226
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
US: (1) Model 722223 SN: 580 (01)00884838099241(21)580 (2) Model: 722224 SN: UDI: 1518 (01)00884838099258(21)1518 1317 (01)00884838099258(21)1317 1221 (01)00884838099258(21)1221 845 (01)00884838099258(21)845 974 (01)00884838099258(21)974 905 (01)00884838099258(21)905 (3) Model: 722225 227 (01)00884838099265(21)227 (4) Model: 722226 441 (01)00884838099272(21)441 398 (01)00884838099272(21)398 OUS: 179 (01)00884838099265(21)179 145 (01)00884838099265(21)145 223 (01)00884838099265(21)223 226 (01)00884838099265(21)226 232 (01)00884838099265(21)232 196 (01)00884838099265(21)196 159 (01)00884838099265(21)159 200 (01)00884838099265(21)200 234 (01)00884838099265(21)234 206 (01)00884838099265(21)206. 412 (01)00884838099272(21)412 425 (01)00884838099272(21)425 420 (01)00884838099272(21)420 355 (01)00884838099272(21)355 393 (01)00884838099272(21)393 407 (01)00884838099272(21)407 402 (01)00884838099272(21)402 527 (01)00884838099272(21)527 479 (01)00884838099272(21)479 509 (01)00884838099272(21)509 460 (01)00884838099272(21)460 505 (01)00884838099272(21)505 513 (01)00884838099272(21)513 485 (01)00884838099272(21)485 364 (01)00884838099272(21)364 507 (01)00884838099272(21)507 577 (01)00884838099272(21)577 476 (01)00884838099272(21)476 515 (01)00884838099272(21)515 596 (01)00884838099272(21)596 598 (01)00884838099272(21)598 521 (01)00884838099272(21)521 455 (01)00884838099272(21)455 589 (01)00884838099272(21)589 575 (01)00884838099272(21)575 522 (01)00884838099272(21)522 540 (01)00884838099272(21)540. 506 (01)00884838099241(21)506 550 (01)00884838099241(21)550 613 (01)00884838099241(21)613 121 (01)00884838099241(21)121 630 (01)00884838099241(21)630 474 (01)00884838099241(21)474 614 (01)00884838099241(21)614 612 (01)00884838099241(21)612 523 (01)00884838099241(21)523 517 (01)00884838099241(21)517 554 (01)00884838099241(21)554 450 (01)00884838099241(21)450 542 (01)00884838099241(21)542 623 (01)00884838099241(21)623 475 (01)00884838099241(21)475 547 (01)00884838099241(21)547 397 (01)00884838099241(21)397 586 (01)00884838099241(21)586 42 (01)00884838099241(21)42 492 (01)00884838099241(21)492 427 (01)00884838099241(21)427 520 (01)00884838099241(21)520 708 (01)00884838099241(21)708 524 (01)00884838099241(21)524 497 (01)00884838099241(21)497 519 (01)00884838099241(21)519 394 (01)00884838099241(21)394 600 (01)00884838099241(21)600. 1154 (01)00884838099258(21)1154 1438 (01)00884838099258(21)1438 1351 (01)00884838099258(21)1351 1139 (01)00884838099258(21)1139 1326 (01)00884838099258(21)1326 1391 (01)00884838099258(21)1391 1075 (01)00884838099258(21)1075 1281 (01)00884838099258(21)1281 1497 (01)00884838099258(21)1497 802 (01)00884838099258(21)802 1553 (01)00884838099258(21)1553 1401 (01)00884838099258(21)1401 1348 (01)00884838099258(21)1348 1350 (01)00884838099258(21)1350 1367 (01)00884838099258(21)1367 1407 (01)00884838099258(21)1407 1181 (01)00884838099258(21)1181 1365 (01)00884838099258(21)1365 1503 (01)00884838099258(21)1503 455 (01)00884838099258(21)455 1489 (01)00884838099258(21)1489 1342 (01)00884838099258(21)1342 1666 (01)00884838099258(21)1666 1714 (01)00884838099258(21)1714 1673 (01)00884838099258(21)1673 1658 (01)00884838099258(21)1658 1733 (01)00884838099258(21)1733 1741 (01)00884838099258(21)1741 1173 (01)00884838099258(21)1173 1566 (01)00884838099258(21)1566 915 (01)00884838099258(21)915 1713 (01)00884838099258(21)1713 1688 (01)00884838099258(21)1688 936 (01)00884838099258(21)936 1602 (01)00884838099258(21)1602 1588 (01)00884838099258(21)1588 1193 (01)00884838099258(21)1193 1586 (01)00884838099258(21)1586 1665 (01)00884838099258(21)1665 1609 (01)00884838099258(21)1609 1463 (01)00884838099258(21)1463 969 (01)00884838099258(21)969 1280 (01)00884838099258(21)1280 1402 (01)00884838099258(21)1402 1598 (01)00884838099258(21)1598 1624 (01)00884838099258(21)1624 1537 (01)00884838099258(21)1537 1645 (01)00884838099258(21)1645 1355 (01)00884838099258(21)1355 1349 (01)00884838099258(21)1349 1736 (01)00884838099258(21)1736 1505 (01)00884838099258(21)1505 1664 (01)00884838099258(21)1664 1711 (01)00884838099258(21)1711 1862 (01)00884838099258(21)1862 1708 (01)00884838099258(21)1708 1447 (01)00884838099258(21)1447 1786 (01)00884838099258(21)1786 1428 (01)00884838099258(21)1428
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System32 recalls2 deaths2,308 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths835 reports
- Azurion Azurion 7 B20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls1 death634 reports
- Azurion Azurion 7 B12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls232 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93966).
- Z-1176-2024 · Open, ClassifiedAzurion 3 with a Certeray generator -To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722221 (2) 722222 (3) 722280
- Z-1177-2024 · Open, ClassifiedAzurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722227 (2) 722228
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page