Azurion 5 with a Certeray generator-To perform image… recallZ-1177-2024 · Philips Medical Systems Nederland B.V. · initiated Feb 6, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.

FDA root cause category: Under Investigation by firm

What the company is doing

Philips notified U.S customers by Urgent Medical Device Correction letter on 06-Feb-2024 via certified mailing via United States Postal Service. " Letters outside of the U.S.A. will be distributed through the Philips Markets Organizations. Letter states reason for recall, health risk, and action to take: Circulate this Letter to all clinical staff and any service personnel that may service the system so that they are aware of the issue. " Keep this Urgent Medical Device Correction Letter with the documentation of the system until Philips implements the correction in your system. " If the Certeray generator fails, please stop using the system and contact the Customer Care Solutions Center. " Place a copy of this Letter in a visible place in the Control and Technical rooms (e.g., posted on the door). " Please complete and return the attached response form (on page 3) to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction and understanding of the issue and required actions to be taken. 5. Actions planned by Philips IGTS to correct the problem As a remedy, Philips will replace the Power Inverter in the Azurion generator cabinet in all affected Azurion systems through the implementation of Field Change Order (FCO72200544). If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).

Recall details

Recall number
Z-1177-2024
Status
Open, Classified
Initiated by the company
Feb 6, 2024
Posted by the FDA
Feb 22, 2024
Recalling firm
Philips Medical Systems Nederland B.V.
Quantity
27 units
Product code
OWB
FDA submissions
K200917
Distribution
Worldwide distribution - US Nationwide distribution in the states of NY, PA, PR, TN, TX, VA, WA and the countries of Algeria, Armenia, Austria, Brazil, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Republic of Kuwait, Lithuania, Macao, Macedonia, Malaysia, Morocco, Netherlands, Norway, Oman, Palestine, State of Panama, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, Ukraine, United Kingdom.

Product description

Azurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722227 (2) 722228

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

OUS (1) Model: 722227 SN: UDI: 75 (01)00884838099227(21)75 122 (01)00884838099227(21)122 131 (01)00884838099227(21)131 132 (01)00884838099227(21)132 165 (01)00884838099227(21)165 116 (01)00884838099227(21)116 136 (01)00884838099227(21)136.' (2) Model: 722228 Serial Number UDI 168 (01)00884838099234(21)168 197 (01)00884838099234(21)197 154 (01)00884838099234(21)154 193 (01)00884838099234(21)193 165 (01)00884838099234(21)165 163 (01)00884838099234(21)163 139 (01)00884838099234(21)139 202 (01)00884838099234(21)202 141 (01)00884838099234(21)141 200 (01)00884838099234(21)200 174 (01)00884838099234(21)174 178 (01)00884838099234(21)178 147 (01)00884838099234(21)147 208 (01)00884838099234(21)208 171 (01)00884838099234(21)171 167 (01)00884838099234(21)167 194 (01)00884838099234(21)194 176 (01)00884838099234(21)176 228 (01)00884838086777(10)2.0.0.10015 161 (01)00884838099234(21)161

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93966).