PDS recallZ-2146-2024 · Ethicon, Inc. · initiated May 10, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ethicon, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
FDA root cause category: Under Investigation by firm
What the company is doing
Erhicon issued Urgent: Medical Device Recall (Removal) letter on May 10, 2024 to US Distributors, primary and secondary consignees via UPS second day. Ethicon Johnson & Johnson affiliates in EMEA, LATAM, APAC and Canada will be contacting all their impacted primary and secondary consignees and/or end customers. Letter states reason for recall, health risk and action to take: 1. Determine whether you have inventory of the lots listed in Attachment 1. If so, you may " Option 1: Remove the first package of suture from a complete box, i. Quarantine and return it for credit. You may use the remaining 35 sutures in accordance with the IFU. (See Figure 2 below for location of first package) OR ii. Examine the first package of a complete box prior to use. Quarantine and return this unit if a packaging defect is identified. (See Figure 2 below for location of first package). If no defect is found, the product and remaining 35 sutures in the box may be used accordance with the IFU. OR "Option 2: Quarantine all product in scope and return all inventory for credit. Please maintain a copy of this notice with the quarantined product and keep a copy for your records. 2. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 3. Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice and fax or email to Sedgwick at 800-901-0335 or [email protected] within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 4. Customers are required to return unused sutures with holes in the primary packaging subject to this recall that are in inventory immediately. To receive credit for items returned, customers must return product no
Recall details
- Recall number
- Z-2146-2024
- Status
- Open, Classified
- Initiated by the company
- May 10, 2024
- Posted by the FDA
- Jun 14, 2024
- Recalling firm
- Ethicon, Inc.
- Quantity
- 56952 units
- Product code
- GAM
- FDA submissions
- K183183
- Distribution
- Worldwide distribution - US Nationwide including PR and the countries of AR, AU, AU, BE, BR, CA, CA, CL, CN, CO, CR, EC, IL, JP, JP, KW, MX, NZ, PA, PR, PR, SA, SG, UY, ZA.
Product description
PDS II (polydioxanone) Suture - Intended for use in general soft tissue approximation, including use in paediatric cardiovascular tissue, in microsurgery and in ophthalmic surgery. These sutures are particularly useful where the combination of an absorbable suture and extended wound support (up to six weeks) is desirable. PRODUCT CODE: D6261 Z117H Z149H Z259H Z305H Z320H
Code information (models, lots, serial numbers)
GTIN 10705031026216 10705031060166 10705031060289 10705031060395 10705031060517 PRODUCT CODE PRODUCT LOTS VCP196H UBMDDB VCP417H UAMQML VCP602H UAMQQL UAMSCA VCP603H UAMHTC UAMHZZ UAMEDC VCP662H UBMJSH VCPB259H UAMPKR VCPB946H UAMLJA
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
Product lines
Other devices under the same FDA submission
37 devices cleared under K183183 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Coated VICRYL J570GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID170 reports
- Coated VICRYL J416HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID60 reports
- Coated VICRYL J556GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID45 reports
- Coated VICRYL J566GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID39 reports
- PDS Z924HEthicon, LLC · Suture, surgical, absorbable, polydioxanone19 reports
- Coated VICRYL J493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID18 reports
- Coated VICRYL J649GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID17 reports
- Coated VICRYL J328HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID11 reports
- Coated VICRYL J376HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID9 reports
- Coated VICRYL D5890Ethicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID8 reports
- Coated VICRYL J458HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID6 reports
- Coated VICRYL J495HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94643).
- Z-2144-2024 · Open, ClassifiedVICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not for use in cardiovascular and…
- Z-2145-2024 · Open, ClassifiedVICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological …
- Z-2147-2024 · Open, ClassifiedPDS Plus Antibacterial (polydioxanone) Suture - Intended for use in general soft tissue approximation, including use in paediatric cardiovascular tissue, in microsurgery and in oph…
- Z-2148-2024 · Open, ClassifiedMONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated PRODUCT CODE: Y315H Y345…
- Z-2149-2024 · Open, ClassifiedMONOCRYL Plus Antibacterial (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated. PRODUCT CO…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page