MONOCRYL recallZ-2149-2024 · Ethicon, Inc. · initiated May 10, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ethicon, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
FDA root cause category: Under Investigation by firm
What the company is doing
Erhicon issued Urgent: Medical Device Recall (Removal) letter on May 10, 2024 to US Distributors, primary and secondary consignees via UPS second day. Ethicon Johnson & Johnson affiliates in EMEA, LATAM, APAC and Canada will be contacting all their impacted primary and secondary consignees and/or end customers. Letter states reason for recall, health risk and action to take: 1. Determine whether you have inventory of the lots listed in Attachment 1. If so, you may " Option 1: Remove the first package of suture from a complete box, i. Quarantine and return it for credit. You may use the remaining 35 sutures in accordance with the IFU. (See Figure 2 below for location of first package) OR ii. Examine the first package of a complete box prior to use. Quarantine and return this unit if a packaging defect is identified. (See Figure 2 below for location of first package). If no defect is found, the product and remaining 35 sutures in the box may be used accordance with the IFU. OR "Option 2: Quarantine all product in scope and return all inventory for credit. Please maintain a copy of this notice with the quarantined product and keep a copy for your records. 2. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 3. Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice and fax or email to Sedgwick at 800-901-0335 or [email protected] within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 4. Customers are required to return unused sutures with holes in the primary packaging subject to this recall that are in inventory immediately. To receive credit for items returned, customers must return product no
Recall details
- Recall number
- Z-2149-2024
- Status
- Open, Classified
- Initiated by the company
- May 10, 2024
- Posted by the FDA
- Jun 14, 2024
- Recalling firm
- Ethicon, Inc.
- Quantity
- 287892 units
- Product code
- GAM
- FDA submissions
- K201996
- Distribution
- Worldwide distribution - US Nationwide including PR and the countries of AR, AU, AU, BE, BR, CA, CA, CL, CN, CO, CR, EC, IL, JP, JP, KW, MX, NZ, PA, PR, PR, SA, SG, UY, ZA.
Product description
MONOCRYL Plus Antibacterial (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated. PRODUCT CODE: MCP266H13 MCP3200H16 MCP3212H16 MCP3213H16 MCP340H13 MCP345H13 MCP416H14 MCP417H14 MCP426H13 MCP427H13 MCP4423H16 MCP4424H14 MCP442H14 MCP496H16 MCP497H16 MCP604H13 MCP936H13 MCP945H14 MCPB346H13
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
GTIN: 10705031046337 10705031121393 10705031121423 10705031121430 10705031046405 10705031046429 10705031046481 10705031046498 10705031046504 10705031046511 10705031121997 10705031148116 10705031122000 10705031122314 10705031122338 10705031046634 10705031046757 10705031046795 10705031046863 Product Code Lot Code MCP266H UAMMCX MCP340H UAMQUZ MCP416H UBMHLB MCP417H UBMLKK MCP426H UAMLDQ UAMRSL UAMRUM UAMRUU UAMRMR UAMRUK UAMRMB MCP427H UAMLRK UAMMCH UAMLTR UAMKTS UAMKQL UAMREE MCP604H UAMLQE MCP936H UBMDQM UBMDPZ UBMEXS UBMQAR UBMEQX UBMKZA UBMLDR MCP945H UAMRKQ MCPB346H UAMQZB
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- MONOCRYL MCP426HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall139 reports
- MONOCRYL MCP936HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall73 reports
- MONOCRYL MCP340HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall1 report
- MONOCRYL MCP266HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall0 reports
- MONOCRYL MCP416HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall0 reports
- MONOCRYL MCPB346HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall0 reports
Product lines
Other devices under the same FDA submission
22 devices cleared under K201996 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- MONOCRYL MCP493GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID54 reports
- MONOCRYL MCP316HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID26 reports
- MONOCRYL MCP317HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID26 reports
- MONOCRYL MCP923GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID26 reports
- MONOCRYL MCP943GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID17 reports
- MONOCRYL MCP490GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID14 reports
- MONOCRYL MCP942GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID13 reports
- MONOCRYL MCP315HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID11 reports
- MONOCRYL MCP922HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID10 reports
- MONOCRYL MCP333HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID8 reports
- MONOCRYL MCP358HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID8 reports
- MONOCRYL MCP347HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID5 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94643).
- Z-2144-2024 · Open, ClassifiedVICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not for use in cardiovascular and…
- Z-2145-2024 · Open, ClassifiedVICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological …
- Z-2146-2024 · Open, ClassifiedPDS II (polydioxanone) Suture - Intended for use in general soft tissue approximation, including use in paediatric cardiovascular tissue, in microsurgery and in ophthalmic surgery.…
- Z-2147-2024 · Open, ClassifiedPDS Plus Antibacterial (polydioxanone) Suture - Intended for use in general soft tissue approximation, including use in paediatric cardiovascular tissue, in microsurgery and in oph…
- Z-2148-2024 · Open, ClassifiedMONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated PRODUCT CODE: Y315H Y345…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page