VICRYL (polyglactin 910) Suture - Indicated for use in… recallZ-2144-2024 · Ethicon, Inc. · initiated May 10, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ethicon, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .

FDA root cause category: Under Investigation by firm

What the company is doing

Erhicon issued Urgent: Medical Device Recall (Removal) letter on May 10, 2024 to US Distributors, primary and secondary consignees via UPS second day. Ethicon Johnson & Johnson affiliates in EMEA, LATAM, APAC and Canada will be contacting all their impacted primary and secondary consignees and/or end customers. Letter states reason for recall, health risk and action to take: 1. Determine whether you have inventory of the lots listed in Attachment 1. If so, you may " Option 1: Remove the first package of suture from a complete box, i. Quarantine and return it for credit. You may use the remaining 35 sutures in accordance with the IFU. (See Figure 2 below for location of first package) OR ii. Examine the first package of a complete box prior to use. Quarantine and return this unit if a packaging defect is identified. (See Figure 2 below for location of first package). If no defect is found, the product and remaining 35 sutures in the box may be used accordance with the IFU. OR "Option 2: Quarantine all product in scope and return all inventory for credit. Please maintain a copy of this notice with the quarantined product and keep a copy for your records. 2. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 3. Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice and fax or email to Sedgwick at 800-901-0335 or [email protected] within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 4. Customers are required to return unused sutures with holes in the primary packaging subject to this recall that are in inventory immediately. To receive credit for items returned, customers must return product no

Recall details

Recall number
Z-2144-2024
Status
Open, Classified
Initiated by the company
May 10, 2024
Posted by the FDA
Jun 14, 2024
Recalling firm
Ethicon, Inc.
Quantity
341532 units
Product code
GAM
FDA submissions
K022269
Distribution
Worldwide distribution - US Nationwide including PR and the countries of AR, AU, AU, BE, BR, CA, CA, CL, CN, CO, CR, EC, IL, JP, JP, KW, MX, NZ, PA, PR, PR, SA, SG, UY, ZA.

Product description

VICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not for use in cardiovascular and neurological tissues. PRODUCT CODE: J214H J218H J258H J268H J269H J327H J344H J345H J370H J416H J417H J426H J427H J458H J493H J603H J699H J936H J978H

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

UDI-DI: N/A PRODUCT CODE PRODUCT LOTS J214H UAMPJQ J218H UBMJSK J258H UBMHSJ J268H UBMLPP J269H UAMMAH UAMRMU J327H UAMPEP J344H UBMHRU J345H UAMJPK UAMPKD J370H UAMMJT J416H UAMRSJ UAMRHB J417H UAMLMX J426H UAMQRQ UBMCAC UBMCML UBMCRD J427H UAMRPR J458H UAMQRT J493H UAMMKE J603H TPMBSX UAMLXC UAMMBQ J699H UAMQME J936H UAMMSL J978H UAMKBD UAMHTE

Other devices under the same FDA submission

145 devices cleared under K022269 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 94643).