VICRYL (polyglactin 910) Suture - Indicated for use in… recallZ-2144-2024 · Ethicon, Inc. · initiated May 10, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ethicon, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
FDA root cause category: Under Investigation by firm
What the company is doing
Erhicon issued Urgent: Medical Device Recall (Removal) letter on May 10, 2024 to US Distributors, primary and secondary consignees via UPS second day. Ethicon Johnson & Johnson affiliates in EMEA, LATAM, APAC and Canada will be contacting all their impacted primary and secondary consignees and/or end customers. Letter states reason for recall, health risk and action to take: 1. Determine whether you have inventory of the lots listed in Attachment 1. If so, you may " Option 1: Remove the first package of suture from a complete box, i. Quarantine and return it for credit. You may use the remaining 35 sutures in accordance with the IFU. (See Figure 2 below for location of first package) OR ii. Examine the first package of a complete box prior to use. Quarantine and return this unit if a packaging defect is identified. (See Figure 2 below for location of first package). If no defect is found, the product and remaining 35 sutures in the box may be used accordance with the IFU. OR "Option 2: Quarantine all product in scope and return all inventory for credit. Please maintain a copy of this notice with the quarantined product and keep a copy for your records. 2. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 3. Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice and fax or email to Sedgwick at 800-901-0335 or [email protected] within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 4. Customers are required to return unused sutures with holes in the primary packaging subject to this recall that are in inventory immediately. To receive credit for items returned, customers must return product no
Recall details
- Recall number
- Z-2144-2024
- Status
- Open, Classified
- Initiated by the company
- May 10, 2024
- Posted by the FDA
- Jun 14, 2024
- Recalling firm
- Ethicon, Inc.
- Quantity
- 341532 units
- Product code
- GAM
- FDA submissions
- K022269
- Distribution
- Worldwide distribution - US Nationwide including PR and the countries of AR, AU, AU, BE, BR, CA, CA, CL, CN, CO, CR, EC, IL, JP, JP, KW, MX, NZ, PA, PR, PR, SA, SG, UY, ZA.
Product description
VICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not for use in cardiovascular and neurological tissues. PRODUCT CODE: J214H J218H J258H J268H J269H J327H J344H J345H J370H J416H J417H J426H J427H J458H J493H J603H J699H J936H J978H
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
UDI-DI: N/A PRODUCT CODE PRODUCT LOTS J214H UAMPJQ J218H UBMJSK J258H UBMHSJ J268H UBMLPP J269H UAMMAH UAMRMU J327H UAMPEP J344H UBMHRU J345H UAMJPK UAMPKD J370H UAMMJT J416H UAMRSJ UAMRHB J417H UAMLMX J426H UAMQRQ UBMCAC UBMCML UBMCRD J427H UAMRPR J458H UAMQRT J493H UAMMKE J603H TPMBSX UAMLXC UAMMBQ J699H UAMQME J936H UAMMSL J978H UAMKBD UAMHTE
Other devices under the same FDA submission
145 devices cleared under K022269 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Coated VICRYL J433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID255 reports
- Coated VICRYL J570GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID170 reports
- Coated VICRYL J258HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID155 reports
- VICRYL V960GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall55 reports
- Coated VICRYL J494GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID86 reports
- Coated VICRYL JB840Ethicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID81 reports
- Coated VICRYL J603HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID80 reports
- Coated VICRYL J602HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID76 reports
- VICRYL V449GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID74 reports
- Coated VICRYL J311HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID71 reports
- Coated VICRYL J490GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID67 reports
- Coated VICRYL J493GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID63 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94643).
- Z-2145-2024 · Open, ClassifiedVICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological …
- Z-2146-2024 · Open, ClassifiedPDS II (polydioxanone) Suture - Intended for use in general soft tissue approximation, including use in paediatric cardiovascular tissue, in microsurgery and in ophthalmic surgery.…
- Z-2147-2024 · Open, ClassifiedPDS Plus Antibacterial (polydioxanone) Suture - Intended for use in general soft tissue approximation, including use in paediatric cardiovascular tissue, in microsurgery and in oph…
- Z-2148-2024 · Open, ClassifiedMONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated PRODUCT CODE: Y315H Y345…
- Z-2149-2024 · Open, ClassifiedMONOCRYL Plus Antibacterial (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated. PRODUCT CO…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page