Medtronic Recharger Kits, RS6230, RS7230, and RS5220A… recallZ-0372-2025 · Medtronic Neuromodulation · initiated Sep 18, 2024
Class II Open, Classified All recalls posted in November 2024
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.
FDA root cause category: Under Investigation by firm
What the company is doing
The recalling firm issued letters dated September 2024 on 9/18/2024 via UPS 2-Day Air. The letter explained the issue, product scope, and the customer actions. The actions included: (1) Identify and quarantine all unused affected WR Kits. Attachment A of the letter provides guidance on identifying potentially affected devices; (2) Return all unused affected WR Kits using the instructions in the enclosed customer confirmation form and contact Medtronic Customer Service to obtain a replacement WR Kit; (3) Direct patients who cannot establish a recharge session using their WR to promptly contact Medtronic Customer Service to obtain a replacement WR Kit; (4) Complete the enclosed Customer Confirmation Form; (5) Share this notice with all those who need to be aware of this issue within or outside your organization or to any organization where the potentially affected product has been transferred or distributed and maintain a copy of this notice in your records. Medtronic initiated communication of the potential issue via physical mailing to one (1) customer beginning on 08-July-2025 in the U.S.
Recall details
- Recall number
- Z-0372-2025
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Sep 18, 2024
- Posted by the FDA
- Nov 13, 2024
- Recalling firm
- Medtronic Neuromodulation
- Quantity
- 118 wireless rechargers
- Product code
- LGW
- FDA submissions
- P840001, P960009, P970004, P080025
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AZ, CA, FL, GA, ID, MD, MI, MN, MO, NC, NJ, OH, PA, SC, TX, VA, and WA. The countries of Austria, Belgium, Denmark, Finland, Germany, Ireland, Italy, Japan, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland, and United Kingdom.
Product description
Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used with Neurostimulator Inceptiv INS Model 977119 and Percept RC INS Model B35300 as part of the Deep Brain Stimulation (DBS) system. WR9220A is used with InterStim Micro, Model 97810, as part of Pelvic Health system
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
RS6230 - GTIN 00763000564353, Serial numbers: NRA010637N; NRA011389N; NRA013260N; NRA013664N; NRA013924N; NRA014318N; NRA014347N; NRA015823N; NRA015861N; NRA015864N; NRA016334N; NRA016413N; NRA016430N; NRA016432N; NRA016778N; RA017002N; NRA017048N; NRA017227N; NRA017232N; NRA017685N; NRA018697N; NRA018731N; NRA018906N; NRA018936N; NRA018992N; NRA019019N; NRA019026N; NRA019963N; NRA019967N; NRA019968N; NRA019970N; NRA019983N; NRA010657N; NRA012367N; NRA012898N; NRA012915N; NRA013490N; NRA013643N; NRA013877N; NRA013878N; NRA013880N; NRA014248N; NRA014741N; NRA015127N; NRA015142N, NRA010659N; NRA010661N; NRA010663N NRA016424N; NRA016425N; NRA016426N; NRA016431N; RA016433N; NRA021324N. RS7230 - GTIN 00763000540289, Serial numbers: NRA010640N; NRA010650N; NRA010656N: NRA011591N; NRA012896N; NRA012910N; NRA012925N; NRA013884N; NRA010633N; NRA010635N; NRA010641N; NRA010642N; NRA010646N; NRA010652N; NRA010653N; NRA010654N; NRA010655N; NRA010658N; NRA010660N; NRA010664N; NRA010666N; NRA010667N; NRA010668N; NRA010669N; NRA010881N; NRA011216N; NRA011293N; NRA011393N; NRA011713N; NRA011853N; NRA012184N; NRA012897N; NRA012909N; NRA012911N; NRA012912N; NRA012913N; NRA012914N; NRA013502N; NRA013602N; NRA013603N; NRA013634N; NRA013635N; NRA013638N; NRA013639N; NRA013879N; NRA013883N; NRA013899N; NRA013952N; NRA014259N; NRA014944N; NRA015858N; NRA015904N; NRA011189N; NRA011238N; NRA012130N; NRA012185N; NRA012366N; NRA013543N; NRA014132N; NRA014222N; GTIN 0763000827533, Serial numbers: NRA010558N; NRA010627N; NRA015360N; NRA013633N.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- N/A RS7230Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for pain relief1 recall0 reports
- N/A RS6230Medtronic, Inc. · Stimulator, electrical, implanted, for parkinsonian symptoms1 recall0 reports
Other devices under the same FDA submission
262 devices cleared under P840001, P960009, P970004, P080025 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- INTERSTIM® 3058Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence6 deaths23,065 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- InterStim™ X 97800Medtronic, Inc. · Implanted electrical device intended for treatment of fecal incontinence5,930 reports
- RestoreSensor 37714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 death4,116 reports
- ACTIVA RC 37612Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS2,839 reports
- InterStim™ Micro SureScan™ 97810Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence2,757 reports
- PrimeAdvanced® SureScan® 97702Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 death2,263 reports
- RestoreUltra® SureScan® 97712Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1,887 reports
- Activa® RC 37612Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR2 deaths1,849 reports
- InterStim™ X 97800Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence1,822 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page