Skip to content

Medtronic Recharger Kits, RS6230, RS7230, and RS5220A… recallZ-0372-2025 · Medtronic Neuromodulation · initiated Sep 18, 2024

Class II Open, Classified All recalls posted in November 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.

FDA root cause category: Under Investigation by firm

What the company is doing

The recalling firm issued letters dated September 2024 on 9/18/2024 via UPS 2-Day Air. The letter explained the issue, product scope, and the customer actions. The actions included: (1) Identify and quarantine all unused affected WR Kits. Attachment A of the letter provides guidance on identifying potentially affected devices; (2) Return all unused affected WR Kits using the instructions in the enclosed customer confirmation form and contact Medtronic Customer Service to obtain a replacement WR Kit; (3) Direct patients who cannot establish a recharge session using their WR to promptly contact Medtronic Customer Service to obtain a replacement WR Kit; (4) Complete the enclosed Customer Confirmation Form; (5) Share this notice with all those who need to be aware of this issue within or outside your organization or to any organization where the potentially affected product has been transferred or distributed and maintain a copy of this notice in your records. Medtronic initiated communication of the potential issue via physical mailing to one (1) customer beginning on 08-July-2025 in the U.S.

Recall details

Recall number
Z-0372-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Sep 18, 2024
Posted by the FDA
Nov 13, 2024
Recalling firm
Medtronic Neuromodulation
Quantity
118 wireless rechargers
Product code
LGW
FDA submissions
P840001, P960009, P970004, P080025
Distribution
Worldwide - US Nationwide distribution in the states of AL, AZ, CA, FL, GA, ID, MD, MI, MN, MO, NC, NJ, OH, PA, SC, TX, VA, and WA. The countries of Austria, Belgium, Denmark, Finland, Germany, Ireland, Italy, Japan, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland, and United Kingdom.

Product description

Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used with Neurostimulator Inceptiv INS Model 977119 and Percept RC INS Model B35300 as part of the Deep Brain Stimulation (DBS) system. WR9220A is used with InterStim Micro, Model 97810, as part of Pelvic Health system

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

RS6230 - GTIN 00763000564353, Serial numbers: NRA010637N; NRA011389N; NRA013260N; NRA013664N; NRA013924N; NRA014318N; NRA014347N; NRA015823N; NRA015861N; NRA015864N; NRA016334N; NRA016413N; NRA016430N; NRA016432N; NRA016778N; RA017002N; NRA017048N; NRA017227N; NRA017232N; NRA017685N; NRA018697N; NRA018731N; NRA018906N; NRA018936N; NRA018992N; NRA019019N; NRA019026N; NRA019963N; NRA019967N; NRA019968N; NRA019970N; NRA019983N; NRA010657N; NRA012367N; NRA012898N; NRA012915N; NRA013490N; NRA013643N; NRA013877N; NRA013878N; NRA013880N; NRA014248N; NRA014741N; NRA015127N; NRA015142N, NRA010659N; NRA010661N; NRA010663N NRA016424N; NRA016425N; NRA016426N; NRA016431N; RA016433N; NRA021324N. RS7230 - GTIN 00763000540289, Serial numbers: NRA010640N; NRA010650N; NRA010656N: NRA011591N; NRA012896N; NRA012910N; NRA012925N; NRA013884N; NRA010633N; NRA010635N; NRA010641N; NRA010642N; NRA010646N; NRA010652N; NRA010653N; NRA010654N; NRA010655N; NRA010658N; NRA010660N; NRA010664N; NRA010666N; NRA010667N; NRA010668N; NRA010669N; NRA010881N; NRA011216N; NRA011293N; NRA011393N; NRA011713N; NRA011853N; NRA012184N; NRA012897N; NRA012909N; NRA012911N; NRA012912N; NRA012913N; NRA012914N; NRA013502N; NRA013602N; NRA013603N; NRA013634N; NRA013635N; NRA013638N; NRA013639N; NRA013879N; NRA013883N; NRA013899N; NRA013952N; NRA014259N; NRA014944N; NRA015858N; NRA015904N; NRA011189N; NRA011238N; NRA012130N; NRA012185N; NRA012366N; NRA013543N; NRA014132N; NRA014222N; GTIN 0763000827533, Serial numbers: NRA010558N; NRA010627N; NRA015360N; NRA013633N.

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

262 devices cleared under P840001, P960009, P970004, P080025 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates