CADD 2120 (REF 21-2120-0104-231)Recalls & reported problems
Icu Medical, Inc. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 4
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15019517126617), received between Mar 17, 2021 and Apr 21, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Infusion or Flow Problem” (1 report).
4 FDA recall records name this device's identifier, 4 still open.
3 other recalls cover devices cleared under the same FDA submission (K111275); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Apr 10, 2025Open, Classified
Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
Z-1667-2025 · Root cause: Labeling design · FDA record ↗
- Apr 10, 2025Open, Classified
There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.
Z-1682-2025 · Root cause: Under Investigation by firm · FDA record ↗
- Apr 10, 2025Open, Classified
Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
Z-1668-2025 · Root cause: Labeling design · FDA record ↗
- Feb 27, 2024Open, Classified
The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
Z-2421-2024 · Root cause: Software Manufacturing/Software Deployment · FDA record ↗
Most reported problems
Device problems
- Infusion or Flow Problem1
- Insufficient Flow or Under Infusion1
- Therapeutic or Diagnostic Output Failure1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 21, 2021Infusion or Flow Problem; Infusion or Flow Problem
INFORMATION WAS RECEIVED INDICATING THAT WHILE IN USE, A SMITHS MEDICAL CADD SOLIS VIP PUMP WAS NOT INFUSING MEDICATION (CHEMO). IT WAS ALSO REPORTED THAT THE USER WAS INSTRUCTED TO DO SEVERAL THINGS, BUT THE ISSUE COULD NOT BE RESOLVED. SUBSEQUENTLY, USER ARRIVED AT THE CLINIC WITH THE PUMP. NO PATIENT CONSEQUENCES WERE REPORTED. NO ADVERSE EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11700595 ↗MalfunctionMar 17, 2021Insufficient Flow or Under Infusion; Therapeutic or Diagnostic Output Failure
PATIENT CAME INTO THE CANCER CENTER FOR DISCONNECTION OF D2 EPOCH (CHEMO REGIMEN) AND RECONNECTION WITH D3 EPOCH VIA CADD PUMP. WHEN PATIENT ARRIVED THEY REPORTED THAT DURING THE NIGHT THEY DISCOVERED THE CADD PUMP WAS NOT INFUSING THE CHEMO. THEY CALLED THE 800 NUMBER ON THE PUMP AND THE CENTER'S ON CALL/SICK AFTER HOURS LINE. THEY WERE INSTRUCTED TO DO SEVERAL THINGS BUT COULD NOT GET THE PUMP RUNNING. THEY WERE INSTRUCTED TO COME TO THE CLINIC IN THE MORNING. WHEN THEY ARRIVED THE PUMP WAS ON, BLINKING GREEN BUT THE BAG APPEARED FULL. THE COUNTER ON THE PUMP SHOWED 500CC TO BE INFUSED (THE WIFE REPORTS THAT SHE CLEARED THE VOLUME INFUSED WHILE TRYING TO GET IT RUNNING AGAIN PER THE SUPPORT STAFF).
Patient problems: Insufficient Information
MAUDE report 11504693 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111275. One submission often covers many models, so check the product and lot details in each record.
- Z-0222-2025 · Jul 16, 2024 · Open, ClassifiedCADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-
- Z-0466-2025 · Jun 26, 2024 · Open, Classifiedsmiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis infusion pumps
- Z-0107-2022 · Mar 24, 2020 · TerminatedCADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-0105-03L c. 21-2120-0105-06L d. 21-2120-0105-07L e. 21-2120-0105-08L f. 21-2120-0105-12
Other Pump, infusion devices
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion4 recalls78 deaths344,514 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion2 recalls45 deaths148,093 reports
- BD® Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion5 recalls30 deaths51,256 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls4 deaths21,956 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths15,547 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion3 recalls6 deaths13,022 reports
- CADD 21-6500-51Icu Medical, Inc. · Pump, infusion2 deaths8,231 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.