RELIANCE 4-FRONT, endocardial cardioversion/defibrillation… recallZ-2327-2025 · Boston Scientific Corporation · initiated Jul 24, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

FDA root cause category: Component design/selection

What the company is doing

On July 24, 2025, firm notified physicians via Important Medical Device Information letters. The letter provides information on the management of gradually rising daily subthreshold, low-voltage shock impedance (LVSI) pattern associated with calcification of expanded polytetrafluoroethylene (ePTFE) coated single coil (SC) and dual coil (DC) RELIANCE defibrillation leads manufactured by Boston Scientific Corporation between 2002 to 2021. Letter includes recommendations for routine follow-ups of affected leads, and considerations to take if lead replacement is planned. The firm is not recommending any changes to the scheduled follow-up interval for patients with ePTFE lead models.

Recall details

Recall number
Z-2327-2025
Status
Open, Classified
Initiated by the company
Jul 24, 2025
Posted by the FDA
Aug 20, 2025
Recalling firm
Boston Scientific Corporation
Quantity
247,702 US; 344,670 OUS
Product code
NVY
FDA submissions
P910073
Distribution
Worldwide Distribution. US nationwide. International distribution worldwide.

Product description

RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696

Code information (models, lots, serial numbers)

Model 0657 UDI-DI 00802526560934; Model 0682 UDI-DI 00802526560958; Model 0683 UDI-DI 00802526560965; Model 0685 UDI-DI 00802526560972; Model 0686 UDI-DI 00802526560989; Model 0692 UDI-DI 00802526519017 00802526560996 00802526568664 00802526568688; Model 0693 UDI-DI 00802526518850 00802526519055 00802526537769 00802526561009 00802526568701; Model 0695 UDI-DI 00802526519093 00802526561016; Model 0696 UDI-DI 00802526519987 00802526561207 00802526568831

Other devices under the same FDA submission

48 devices cleared under P910073 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97318).