Brand Name: DJO SURGICAL recallZ-2549-2025 · Encore Medical, LP · initiated Jul 18, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Encore Medical, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Their is a potential that the reamer may kick or bind up during or immediately prior to use.
FDA root cause category: Device Design
What the company is doing
On July 29, 2025, Enovis issued a "Urgent Medical Device Recall Notification via Email to affected consignees. On August 4, 2025, Enovis issued an update to the recall notification via E-Mail requesting consignees rework affected trays and returned affected products. Enovis asked consignees to take the following actions: 1. Continue using the wedge reamer assembly while carefully inspecting and following the specific details in the supplied Quality Bulletin. 2. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 3. Please see the attached quality Bulletin 4. By 9/13/25, the firm will reach out regarding reworking and returning trays. 5. Acknowledge receipt of notification.
Recall details
- Recall number
- Z-2549-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 18, 2025
- Posted by the FDA
- Sep 10, 2025
- Recalling firm
- Encore Medical, LP
- Quantity
- 439 kits
- Product code
- FSM
- FDA submissions
- K233481
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, FL, HI, IA, IL, IN, KS, KY, LA, MD, ME, NE, NJ, NY, OH, PR, RI, SC, TN, TX, UT, VA, WA, WI.
Product description
Brand Name: DJO SURGICAL Product Name: FA S Altivate Reverse Glenoid Tray Model/Catalog Number: None Software Version: NA Product Description: Enovis" shoulder devices are intended for treatment of patients who are candidates for shoulder arthroplasty per the Indication for use. While total shoulder replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients. The AltiVate Reverse Shoulder Prosthesis is indicated as a reverse shoulder replacement for patients with a functional deltoid muscle and a grossly deficient rotator cuff joint suffering from pain and dysfunction due to: " Severe arthropathy with a grossly deficient rotator cuff; " Previously failed joint replacement with a grossly deficient rotator cuff; " Fracture of glenohumeral joint from trauma or pathologic conditions of the shoulder including humeral head fracture, displaced 3- or 4-part fractures of proximal humerus, or reconstruction after tumor resection; " Bone defect in proximal humerus; " Non-inflammatory degenerative disease including osteoarthritis and avascular necrosis of the natural humeral head and/or glenoid; " Inflammatory arthritis including rheumatoid arthritis; " Correction of functional deformity. The glenoid baseplate is intended for cementless application with addition of screws for fixation. This device may also be indicated in the salvage of previously failed surgical attempts for anatomic and hemi procedures. All RSP Monoblock and AltiVate Reverse humeral stems are intended for cemented or cementless use. Enovis" instrumentation consists of devices and their accessories used in surgical procedures. Implantation of Enovis" products should only be performed with Enovis" instrumentation or instrumentation distributed by Enovis". Enovis" instruments and instrument cases are generally composed of titanium, stainless steel, aluminum, and/or polymeric materials. The cases may be multi-layered with various inserts to hold surgical instrumentation in place during handling and storage. The inserts may consist of trays, holders, and silicone mats. The instrument cases will allow sterilization of the contents to occur in a steam autoclave utilizing the cleaning, sterilization, and drying cycle that has been validated and detailed below. Instrument cases do not provide a sterile barrier and must be used in conjunction with FDA cleared sterilization wrap to maintain sterility. Instruments are provided non-sterile and should be stored in their original packaging until cleaned and sterilized according to the recommended guidelines listed below. Component: NA
Code information (models, lots, serial numbers)
Lot Code: All GTIN:
Other devices under the same FDA submission
21 devices cleared under K233481 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- DJO SURGICAL 804-06-310Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall12 reports
- DJO SURGICAL 506-04-118Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall1 report
- DJO SURGICAL 804-06-311Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall1 report
- DJO SURGICAL 506-04-122Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall0 reports
- DJO SURGICAL 506-04-126Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall0 reports
- DJO SURGICAL 506-04-130Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall0 reports
- DJO SURGICAL 804-06-312Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall0 reports
- DJO SURGICAL 508-36-107Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED25 reports
- DJO SURGICAL 508-40-107Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED13 reports
- DJO SURGICAL 508-65-001Encore Medical, L.P. · shoulder prosthesis, reverse configuration13 reports
- DJO SURGICAL 804-06-324Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED12 reports
- DJO SURGICAL 508-32-105Encore Medical, L.P. · shoulder prosthesis, reverse configuration4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97412).
- Z-2546-2025 · Open, ClassifiedBrand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD Model/Catalog Number: 804-06-311 Product Description: Material: S.S/Nitronic, Non-Sterile
- Z-2547-2025 · Open, ClassifiedBrand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER BOSS DRILL, 6.5mm Model/Catalog Number: 804-06-312 Product Description: Material: S.S., Non Sterile
- Z-2548-2025 · Open, ClassifiedBrand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER SLEEVE Model/Catalog Number: 804-06-310 Product Description: Material: S.S/CoCrMo, Non-Sterile
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page