Azurion recallZ-1079-2026 · Philips Medical Systems Nederland B.V. · initiated Dec 8, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
FDA root cause category: Software design
What the company is doing
Philips notified consignees on about 12/08/2025 via emailed URGENT Medical Device Correction letter. The letter informed consignees that they have identified nine (9) software issues affecting Azurion R3.0 systems that may result in loss of imaging (X-ray) functionality, and/or loss of motorized movement, and/or incorrect image content, and/or loss of data. Consignees were instructed to identify all affected units on hand, ensure all users of the affected units are aware of the issue and notify any customers if the unit has been transferred or distributed. Consignees were also instructed to keep the notification with documentation of the system until the issue is resolved, and complete and return the response form. If any of the nine issues are encountered, consignees were asked to contact their local Philips representative and the letter provided recommended actions. ***Updated May 8, 2026*** Previously, Philips planned to address issues 1-7 in software update 3.1.5, while issues 8 and 9 would be addressed separately. Update 3.1.5 has been canceled. Issues 1-9 will now be addressed through software release 3/1/15. Anticipated release date for software version 3.1.15 is Q3 2026. Local Philips representatives will contact each consignee to schedule visits to install the software updates once available.
Recall details
- Recall number
- Z-1079-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 8, 2025
- Posted by the FDA
- Jan 14, 2026
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 990 units
- Product code
- OWB
- FDA submissions
- K200917
- Distribution
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Chile, Colombia, Cote de'Ivore, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Kosovo, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Palestine, Philippines, Poland, Portugal, R¿union, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.
Product description
Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion 5 M12. Model Number: 722231. 4. Azurion 5 M20. Model Number: 722232. 5. Azurion 7 B12. Model Number: 722235. 6. Azurion 7 B20. Model Number: 722236. 7. Azurion 7 M12. Model Number: 722233. 8. Azurion 7 M20. Model Number: 722234.
Code information (models, lots, serial numbers)
Show the full code information (27k characters)
Software Version: R3.0. 1. Model Number: 722229. UDI Numbers: (01)00884838116726(21)22, (01)00884838116726(21)15, (01)00884838116726(21)16, (01)00884838116726(21)6, (01)00884838116726(21)8, (01)00884838116726(21)4, (01)00884838116726(21)24, (01)00884838116726(21)26, (01)00884838116726(21)20, (01)00884838116726(21)25, (01)00884838116726(21)11, (01)00884838116726(21)21, (01)00884838116726(21)13, (01)00884838116726(21)7, (01)00884838116726(21)10, (01)00884838116726(21)2, (01)00884838116726(21)3, (01)00884838116726(21)29. 2. Model Number: 722230. UDI Numbers: (01)00884838116733(21)2, (01)00884838116733(21)40, (01)00884838116733(21)9, (01)00884838116733(21)1, (01)00884838116733(21)38, (01)00884838116733(21)10, (01)00884838116733(21)45, (01)00884838116733(21)7, (01)00884838116733(21)27, (01)00884838116733(21)28, (01)00884838116733(21)39, (01)00884838116733(21)4, (01)00884838116733(21)5, (01)00884838116733(21)23, (01)00884838116733(21)41, (01)00884838116733(21)20, (01)00884838116733(21)19, (01)00884838116733(21)13, (01)00884838116733(21)15, (01)00884838116733(21)12, (01)00884838116733(21)22. 3. Model Number: 722231. UDI Numbers: (01)00884838116740(21)13, (01)00884838116740(21)29, (01)00884838116740(21)50, (01)00884838116740(21)21, (01)00884838116740(21)19, (01)00884838116740(21)12, (01)00884838116740(21)25, (01)00884838116740(21)4, (01)00884838116740(21)1, (01)00884838116740(21)5, (01)00884838116740(21)2, (01)00884838116740(21)73, (01)00884838116740(21)72, (01)00884838116740(21)75, (01)00884838116740(21)33, (01)00884838116740(21)47, (01)00884838116740(21)60, (01)00884838116740(21)53, (01)00884838116740(21)45, (01)00884838116740(21)61, (01)00884838116740(21)34, (01)00884838116740(21)54, (01)00884838116740(21)35, (01)00884838116740(21)49, (01)00884838116740(21)32, (01)00884838116740(21)28, (01)00884838116740(21)27, (01)00884838116740(21)26, (01)00884838116740(21)51, (01)00884838116740(21)11, (01)00884838116740(21)20, (01)00884838116740(21)41, (01)00884838116740(21)52, (01)00884838116740(21)3. 4. Model Number: 722232. UDI Numbers: (01)00884838116757(21)61, (01)00884838116757(21)60, (01)00884838116757(21)5, (01)00884838116757(21)49, (01)00884838116757(21)2, (01)00884838116757(21)91, (01)00884838116757(21)33, (01)00884838116757(21)34, (01)00884838116757(21)48, (01)00884838116757(21)46, (01)00884838116757(21)23, (01)00884838116757(21)43, (01)00884838116757(21)36, (01)00884838116757(21)96, (01)00884838116757(21)95, (01)00884838116757(21)10, (01)00884838116757(21)76, (01)00884838116757(21)42, (01)00884838116757(21)6, (01)00884838116757(21)77, (01)00884838116757(21)3, (01)00884838116757(21)1, (01)00884838116757(21)56, (01)00884838116757(21)51, (01)00884838116757(21)44, (01)00884838116757(21)30, (01)00884838116757(21)29, (01)00884838116757(21)28, (01)00884838116757(21)102, (01)00884838116757(21)19, (01)00884838116757(21)8, (01)00884838116757(21)81, (01)00884838116757(21)20, (01)00884838116757(21)16, (01)00884838116757(21)11, (01)00884838116757(21)112, (01)00884838116757(21)63, (01)00884838116757(21)68, (01)00884838116757(21)32, (01)00884838116757(21)35, (01)00884838116757(21)122, (01)00884838116757(21)14, (01)00884838116757(21)82, (01)00884838116757(21)13, (01)00884838116757(21)50, (01)00884838116757(21)57, (01)00884838116757(21)58, (01)00884838116757(21)4, (01)00884838116757(21)105, (01)00884838116757(21)75, (01)00884838116757(21)121, (01)00884838116757(21)120, (01)00884838116757(21)71, (01)00884838116757(21)72, (01)00884838116757(21)73, (01)00884838116757(21)70, (01)00884838116757(21)69, (01)00884838116757(21)37. 5. Model Number: 722235. UDI Numbers: (01)00884838116788(21)88, (01)00884838116788(21)43, (01)00884838116788(21)12, (01)00884838116788(21)102, (01)00884838116788(21)127, (01)00884838116788(21)57, (01)00884838116788(21)55, (01)00884838116788(21)11, (01)00884838116788(21)76, (01)00884838116788(21)71, (01)00884838116788(21)59, (01)00884838116788(21)77, (01)00884838116788(21)89, (01)00884838116788(21)56, (01)00884838116788(21)99, (01)00884838116788(21)30, (01)00884838116788(21)84, (01)00884838116788(21)82, (01)00884838116788(21)83, (01)00884838116788(21)93, (01)00884838116788(21)31, (01)00884838116788(21)47, (01)00884838116788(21)3, (01)00884838116788(21)13, (01)00884838116788(21)10, (01)00884838116788(21)74, (01)00884838116788(21)98, (01)00884838116788(21)40, (01)00884838116788(21)81, (01)00884838116788(21)66, (01)00884838116788(21)63, (01)00884838116788(21)67, (01)00884838116788(21)64, (01)00884838116788(21)42, (01)00884838116788(21)65, (01)00884838116788(21)58, (01)00884838116788(21)53, (01)00884838116788(21)52, (01)00884838116788(21)49, (01)00884838116788(21)44, (01)00884838116788(21)35, (01)00884838116788(21)45, (01)00884838116788(21)39, (01)00884838116788(21)25, (01)00884838116788(21)34, (01)00884838116788(21)17, (01)00884838116788(21)24, (01)00884838116788(21)19, (01)00884838116788(21)33, (01)00884838116788(21)18, (01)00884838116788(21)9, (01)00884838116788(21)28, (01)00884838116788(21)8, (01)00884838116788(21)23, (01)00884838116788(21)21, (01)00884838116788(21)6, (01)00884838116788(21)7, (01)00884838116788(21)5, (01)00884838116788(21)108, (01)00884838116788(21)109, (01)00884838116788(21)107, (01)00884838116788(21)106, (01)00884838116788(21)4, (01)00884838116788(21)41, (01)00884838116788(21)72, (01)00884838116788(21)70, (01)00884838116788(21)54, (01)00884838116788(21)20, (01)00884838116788(21)14, (01)00884838116788(21)126, (01)00884838116788(21)105, (01)00884838116788(21)94. 6. Model Number: 722236. UDI Numbers: (01)00884838116801(21)43, (01)00884838116801(21)42, (01)00884838116801(21)121, (01)00884838116801(21)50, (01)00884838116801(21)183, (01)00884838116801(21)32, (01)00884838116801(21)38, (01)00884838116801(21)197, (01)00884838116801(21)335, (01)00884838116801(21)325, (01)00884838116801(21)116, (01)00884838116801(21)191, (01)00884838116801(21)235, (01)00884838116801(21)22, (01)00884838116801(21)204, (01)00884838116801(21)249, (01)00884838116801(21)159, (01)00884838116801(21)40, (01)00884838116801(21)233, (01)00884838116801(21)194, (01)00884838116801(21)205, (01)00884838116801(21)276, (01)00884838116801(21)378, (01)00884838116801(21)20, (01)00884838116801(21)81, (01)00884838116801(21)51, (01)00884838116801(21)288, (01)00884838116801(21)25, (01)00884838116801(21)54, (01)00884838116801(21)186, (01)00884838116801(21)21, (01)00884838116801(21)68, (01)00884838116801(21)99, (01)00884838116801(21)208, (01)00884838116801(21)31, (01)00884838116801(21)230, (01)00884838116801(21)203, (01)00884838116801(21)180, (01)00884838116801(21)37, (01)00884838116801(21)342, (01)00884838116801(21)24, (01)00884838116801(21)221, (01)00884838116801(21)17, (01)00884838116801(21)252, (01)00884838116801(21)150, (01)00884838116801(21)60, (01)00884838116801(21)200, (01)00884838116801(21)226, (01)00884838116801(21)118, (01)00884838116801(21)127, (01)00884838116801(21)26, (01)00884838116801(21)36, (01)00884838116801(21)133, (01)00884838116801(21)345, (01)00884838116801(21)134, (01)00884838116801(21)131, (01)00884838116801(21)39, (01)00884838116801(21)231, (01)00884838116801(21)158, (01)00884838116801(21)48, (01)00884838116801(21)227, (01)00884838116801(21)148, (01)00884838116801(21)147, (01)00884838116801(21)272, (01)00884838116801(21)273, (01)00884838116801(21)100, (01)00884838116801(21)224, (01)00884838116801(21)164, (01)00884838116801(21)163, (01)00884838116801(21)170, (01)00884838116801(21)44, (01)00884838116801(21)46, (01)00884838116801(21)113, (01)00884838116801(21)216, (01)00884838116801(21)278, (01)00884838116801(21)102, (01)00884838116801(21)8, (01)00884838116801(21)169, (01)00884838116801(21)338, (01)00884838116801(21)117, (01)00884838116801(21)29, (01)00884838116801(21)34, (01)00884838116801(21)10, (01)00884838116801(21)126, (01)00884838116801(21)237, (01)00884838116801(21)125, (01)00884838116801(21)154, (01)00884838116801(21)157, (01)00884838116801(21)30, (01)00884838116801(21)67, (01)00884838116801(21)292, (01)00884838116801(21)289, (01)00884838116801(21)277.(01)00884838116801(21)244, (01)00884838116801(21)181, (01)00884838116801(21)264, (01)00884838116801(21)232, (01)00884838116801(21)207, (01)00884838116801(21)192, (01)00884838116801(21)132, (01)00884838116801(21)128, (01)00884838116801(21)160, (01)00884838116801(21)94, (01)00884838116801(21)86, (01)00884838116801(21)41, (01)00884838116801(21)27, (01)00884838116801(21)259, (01)00884838116801(21)251, (01)00884838116801(21)261, (01)00884838116801(21)57, (01)00884838116801(21)240, (01)00884838116801(21)107, (01)00884838116801(21)172, (01)00884838116801(21)142, (01)00884838116801(21)109, (01)00884838116801(21)55, (01)00884838116801(21)171, (01)00884838116801(21)155, (01)00884838116801(21)175, (01)00884838116801(21)111, (01)00884838116801(21)88, (01)00884838116801(21)104, (01)00884838116801(21)152, (01)00884838116801(21)124, (01)00884838116801(21)103, (01)00884838116801(21)90, (01)00884838116801(21)70, (01)00884838116801(21)153, (01)00884838116801(21)106, (01)00884838116801(21)151, (01)00884838116801(21)144, (01)00884838116801(21)141, (01)00884838116801(21)76, (01)00884838116801(21)112, (01)00884838116801(21)105, (01)00884838116801(21)85, (01)00884838116801(21)62, (01)00884838116801(21)52, (01)00884838116801(21)75, (01)00884838116801(21)92, (01)00884838116801(21)89, (01)00884838116801(21)108, (01)00884838116801(21)59, (01)00884838116801(21)73, (01)00884838116801(21)84, (01)00884838116801(21)69, (01)00884838116801(21)83, (01)00884838116801(21)80, (01)00884838116801(21)64, (01)00884838116801(21)74, (01)00884838116801(21)53, (01)00884838116801(21)49, (01)00884838116801(21)329, (01)00884838116801(21)130, (01)00884838116801(21)326, (01)00884838116801(21)168, (01)00884838116801(21)145, (01)00884838116801(21)146, (01)00884838116801(21)98, (01)00884838116801(21)96, (01)00884838116801(21)93, (01)00884838116801(21)258, (01)00884838116801(21)56, (01)00884838116801(21)91, (01)00884838116801(21)271, (01)00884838116801(21)201, (01)00884838116801(21)123, (01)00884838116801(21)162, (01)00884838116801(21)122, (01)00884838116801(21)114, (01)00884838116801(21)63, (01)00884838116801(21)374, (01)00884838116801(21)357, (01)00884838116801(21)348, (01)00884838116801(21)347, (01)00884838116801(21)352, (01)00884838116801(21)322, (01)00884838116801(21)330, (01)00884838116801(21)321, (01)00884838116801(21)188, (01)00884838116801(21)283, (01)00884838116801(21)285, (01)00884838116801(21)284, (01)00884838116801(21)269, (01)00884838116801(21)193, (01)00884838116801(21)173, (01)00884838116801(21)311, (01)00884838116801(21)156, (01)00884838116801(21)219, (01)00884838116801(21)167, (01)00884838116801(21)33, (01)00884838116801(21)268, (01)00884838116801(21)23, (01)00884838116801(21)361, (01)00884838116801(21)242, (01)00884838116801(21)223, (01)00884838116801(21)234, (01)00884838116801(21)182, (01)00884838116801(21)189, (01)00884838116801(21)165, (01)00884838116801(21)82, (01)00884838116801(21)66, (01)00884838116801(21)13. 7. Model Number: 722233. UDI Numbers: (01)00884838116764(21)118, (01)00884838116764(21)96, (01)00884838116764(21)259, (01)00884838116764(21)169, (01)00884838116764(21)108, (01)00884838116764(21)212, (01)00884838116764(21)267, (01)00884838116764(21)214, (01)00884838116764(21)147, (01)00884838116764(21)153, (01)00884838116764(21)111, (01)00884838116764(21)122, (01)00884838116764(21)124, (01)00884838116764(21)127, (01)00884838116764(21)59, (01)00884838116764(21)172, (01)00884838116764(21)64, (01)00884838116764(21)170, (01)00884838116764(21)41, (01)00884838116764(21)103, (01)00884838116764(21)144, (01)00884838116764(21)219, (01)00884838116764(21)218, (01)00884838116764(21)192, (01)00884838116764(21)30, (01)00884838116764(21)189, (01)00884838116764(21)186, (01)00884838116764(21)17, (01)00884838116764(21)9, (01)00884838116764(21)245, (01)00884838116764(21)57, (01)00884838116764(21)32, (01)00884838116764(21)203, (01)00884838116764(21)205, (01)00884838116764(21)204, (01)00884838116764(21)184, (01)00884838116764(21)162, (01)00884838116764(21)18, (01)00884838116764(21)83, (01)00884838116764(21)84, (01)00884838116764(21)88, (01)00884838116764(21)76, (01)00884838116764(21)137, (01)00884838116764(21)85, (01)00884838116764(21)86, (01)00884838116764(21)87, (01)00884838116764(21)46, (01)00884838116764(21)47, (01)00884838116764(21)26, (01)00884838116764(21)234, (01)00884838116764(21)200, (01)00884838116764(21)250, (01)00884838116764(21)201, (01)00884838116764(21)123, (01)00884838116764(21)235, (01)00884838116764(21)68, (01)00884838116764(21)14, (01)00884838116764(21)82, (01)00884838116764(21)27, (01)00884838116764(21)38, (01)00884838116764(21)37, (01)00884838116764(21)8, (01)00884838116764(21)145, (01)00884838116764(21)112, (01)00884838116764(21)95, (01)00884838116764(21)173, (01)00884838116764(21)202, (01)00884838116764(21)101, (01)00884838116764(21)165, (01)00884838116764(21)166, (01)00884838116764(21)97, (01)00884838116764(21)81, (01)00884838116764(21)61, (01)00884838116764(21)31, (01)00884838116764(21)29, (01)00884838116764(21)109, (01)00884838116764(21)279, (01)00884838116764(21)260, (01)00884838116764(21)221, (01)00884838116764(21)178, (01)00884838116764(21)181, (01)00884838116764(21)159, (01)00884838116764(21)117, (01)00884838116764(21)141, (01)00884838116764(21)120, (01)00884838116764(21)273, (01)00884838116764(21)163, (01)00884838116764(21)272, (01)00884838116764(21)174, (01)00884838116764(21)249, (01)00884838116764(21)78, (01)00884838116764(21)133, (01)00884838116764(21)102, (01)00884838116764(21)106, (01)00884838116764(21)104, (01)00884838116764(21)80, (01)00884838116764(21)77, (01)00884838116764(21)206, (01)00884838116764(21)67, (01)00884838116764(21)39, (01)00884838116764(21)15, (01)00884838116764(21)13, (01)00884838116764(21)179, (01)00884838116764(21)180, (01)00884838116764(21)136, (01)00884838116764(21)129, (01)00884838116764(21)222, (01)00884838116764(21)149, (01)00884838116764(21)148, (01)00884838116764(21)146, (01)00884838116764(21)5, (01)00884838116764(21)51, (01)00884838116764(21)19, (01)00884838116764(21)7, (01)00884838116764(21)6, (01)00884838116764(21)231, (01)00884838116764(21)107, (01)00884838116764(21)91, (01)00884838116764(21)89, (01)00884838116764(21)90, (01)00884838116764(21)71, (01)00884838116764(21)48, (01)00884838116764(21)74, (01)00884838116764(21)62, (01)00884838116764(21)55, (01)00884838116764(21)50, (01)00884838116764(21)43, (01)00884838116764(21)40, (01)00884838116764(21)11, (01)00884838116764(21)35, (01)00884838116764(21)60, (01)00884838116764(21)28, (01)00884838116764(21)20, (01)00884838116764(21)12, (01)00884838116764(21)138, (01)00884838116764(21)139, (01)00884838116764(21)58, (01)00884838116764(21)257, (01)00884838116764(21)160, (01)00884838116764(21)24, (01)00884838116764(21)268, (01)00884838116764(21)284, (01)00884838116764(21)291, (01)00884838116764(21)238, (01)00884838116764(21)225, (01)00884838116764(21)217, (01)00884838116764(21)211, (01)00884838116764(21)16, (01)00884838116764(21)110, (01)00884838116764(21)183, (01)00884838116764(21)143, (01)00884838116764(21)134, (01)00884838116764(21)44, (01)00884838116764(21)197. 8. Model Number: 722234. UDI Numbers: (01)00884838116771(21)626, (01)00884838116771(21)320, (01)00884838116771(21)256, (01)00884838116771(21)321, (01)00884838116771(21)437, (01)00884838116771(21)101, (01)00884838116771(21)616, (01)00884838116771(21)418, (01)00884838116771(21)426, (01)00884838116771(21)87, (01)00884838116771(21)420, (01)00884838116771(21)827, (01)00884838116771(21)716, (01)00884838116771(21)607, (01)00884838116771(21)183, (01)00884838116771(21)345, (01)00884838116771(21)520, (01)00884838116771(21)662, (01)00884838116771(21)327, (01)00884838116771(21)524, (01)00884838116771(21)191, (01)00884838116771(21)201, (01)00884838116771(21)449, (01)00884838116771(21)435, (01)00884838116771(21)657, (01)00884838116771(21)171, (01)00884838116771(21)124, (01)00884838116771(21)123, (01)00884838116771(21)79, (01)00884838116771(21)409, (01)00884838116771(21)70, (01)00884838116771(21)55, (01)00884838116771(21)68, (01)00884838116771(21)780, (01)00884838116771(21)794, (01)00884838116771(21)500, (01)00884838116771(21)450, (01)00884838116771(21)471, (01)00884838116771(21)459, (01)00884838116771(21)489, (01)00884838116771(21)617, (01)00884838116771(21)874, (01)00884838116771(21)92, (01)00884838116771(21)698, (01)00884838116771(21)758, (01)00884838116771(21)585, (01)00884838116771(21)300, (01)00884838116771(21)502, (01)00884838116771(21)216, (01)00884838116771(21)493, (01)00884838116771(21)358, (01)00884838116771(21)85, (01)00884838116771(21)274, (01)00884838116771(21)804, (01)00884838116771(21)484, (01)00884838116771(21)519, (01)00884838116771(21)346, (01)00884838116771(21)469, (01)00884838116771(21)853, (01)00884838116771(21)852, (01)00884838116771(21)60, (01)00884838116771(21)59, (01)00884838116771(21)94, (01)00884838116771(21)485, (01)00884838116771(21)813, (01)00884838116771(21)433, (01)00884838116771(21)432, (01)00884838116771(21)168, (01)00884838116771(21)223, (01)00884838116771(21)67, (01)00884838116771(21)492, (01)00884838116771(21)71, (01)00884838116771(21)535, (01)00884838116771(21)77, (01)00884838116771(21)288, (01)00884838116771(21)383, (01)00884838116771(21)332, (01)00884838116771(21)762, (01)00884838116771(21)284, (01)00884838116771(21)620, (01)00884838116771(21)76, (01)00884838116771(21)232, (01)00884838116771(21)286, (01)00884838116771(21)267, (01)00884838116771(21)269, (01)00884838116771(21)619, (01)00884838116771(21)260, (01)00884838116771(21)162, (01)00884838116771(21)84, (01)00884838116771(21)161, (01)00884838116771(21)165, (01)00884838116771(21)107, (01)00884838116771(21)233, (01)00884838116771(21)419, (01)00884838116771(21)515, (01)00884838116771(21)212, (01)00884838116771(21)248, (01)00884838116771(21)655, (01)00884838116771(21)674, (01)00884838116771(21)139, (01)00884838116771(21)348, (01)00884838116771(21)586, (01)00884838116771(21)252, (01)00884838116771(21)633, (01)00884838116771(21)689, (01)00884838116771(21)395, (01)00884838116771(21)560, (01)00884838116771(21)263, (01)00884838116771(21)257, (01)00884838116771(21)194, (01)00884838116771(21)403, (01)00884838116771(21)331, (01)00884838116771(21)262, (01)00884838116771(21)362, (01)00884838116771(21)175, (01)00884838116771(21)580, (01)00884838116771(21)877, (01)00884838116771(21)186, (01)00884838116771(21)329, (01)00884838116771(21)266, (01)00884838116771(21)372, (01)00884838116771(21)155, (01)00884838116771(21)779, (01)00884838116771(21)696, (01)00884838116771(21)352, (01)00884838116771(21)667, (01)00884838116771(21)797, (01)00884838116771(21)217, (01)00884838116771(21)509, (01)00884838116771(21)675, (01)00884838116771(21)335, (01)00884838116771(21)347, (01)00884838116771(21)330, (01)00884838116771(21)292, (01)00884838116771(21)43, (01)00884838116771(21)225, (01)00884838116771(21)442, (01)00884838116771(21)322, (01)00884838116771(21)726, (01)00884838116771(21)209, (01)00884838116771(21)333, (01)00884838116771(21)475, (01)00884838116771(21)163, (01)00884838116771(21)624, (01)00884838116771(21)628, (01)00884838116771(21)629, (01)00884838116771(21)108, (01)00884838116771(21)198, (01)00884838116771(21)462, (01)00884838116771(21)768, (01)00884838116771(21)222, (01)00884838116771(21)730, (01)00884838116771(21)287, (01)00884838116771(21)174, (01)00884838116771(21)606, (01)00884838116771(21)97, (01)00884838116771(21)270, (01)00884838116771(21)255, (01)00884838116771(21)88, (01)00884838116771(21)359, (01)00884838116771(21)208, (01)00884838116771(21)102, (01)00884838116771(21)207, (01)00884838116771(21)463, (01)00884838116771(21)711, (01)00884838116771(21)340, (01)00884838116771(21)465, (01)00884838116771(21)140, (01)00884838116771(21)774, (01)00884838116771(21)119, (01)00884838116771(21)565, (01)00884838116771(21)413, (01)00884838116771(21)772, (01)00884838116771(21)54, (01)00884838116771(21)511, (01)00884838116771(21)510, (01)00884838116771(21)483, (01)00884838116771(21)547, (01)00884838116771(21)516, (01)00884838116771(21)447, (01)00884838116771(21)638, (01)00884838116771(21)714, (01)00884838116771(21)185, (01)00884838116771(21)253, (01)00884838116771(21)501, (01)00884838116771(21)178, (01)00884838116771(21)467, (01)00884838116771(21)98, (01)00884838116771(21)404, (01)00884838116771(21)554, (01)00884838116771(21)411, (01)00884838116771(21)344, (01)00884838116771(21)460, (01)00884838116771(21)309, (01)00884838116771(21)133, (01)00884838116771(21)122, (01)00884838116771(21)159, (01)00884838116771(21)461, (01)00884838116771(21)640, (01)00884838116771(21)639, (01)00884838116771(21)99, (01)00884838116771(21)803, (01)00884838116771(21)641, (01)00884838116771(21)513, (01)00884838116771(21)416, (01)00884838116771(21)430, (01)00884838116771(21)514, (01)00884838116771(21)319, (01)00884838116771(21)334, (01)00884838116771(21)181, (01)00884838116771(21)82, (01)00884838116771(21)281, (01)00884838116771(21)508, (01)00884838116771(21)497, (01)00884838116771(21)775, (01)00884838116771(21)873, (01)00884838116771(21)490, (01)00884838116771(21)498, (01)00884838116771(21)782, (01)00884838116771(21)792, (01)00884838116771(21)826, (01)00884838116771(21)295, (01)00884838116771(21)294, (01)00884838116771(21)243, (01)00884838116771(21)114, (01)00884838116771(21)42, (01)00884838116771(21)21, (01)00884838116771(21)399, (01)00884838116771(21)192, (01)00884838116771(21)12, (01)00884838116771(21)13, (01)00884838116771(21)567, (01)00884838116771(21)572, (01)00884838116771(21)566, (01)00884838116771(21)264, (01)00884838116771(21)307, (01)00884838116771(21)308, (01)00884838116771(21)443, (01)00884838116771(21)305, (01)00884838116771(21)304, (01)00884838116771(21)296, (01)00884838116771(21)446, (01)00884838116771(21)187, (01)00884838116771(21)249, (01)00884838116771(21)190, (01)00884838116771(21)200, (01)00884838116771(21)341, (01)00884838116771(21)448, (01)00884838116771(21)558, (01)00884838116771(21)679, (01)00884838116771(21)598, (01)00884838116771(21)385, (01)00884838116771(21)247, (01)00884838116771(21)755, (01)00884838116771(21)205, (01)00884838116771(21)737, (01)00884838116771(21)757, (01)00884838116771(21)582, (01)00884838116771(21)625, (01)00884838116771(21)687, (01)00884838116771(21)401, (01)00884838116771(21)486, (01)00884838116771(21)36, (01)00884838116771(21)37, (01)00884838116771(21)46, (01)00884838116771(21)478, (01)00884838116771(21)479, (01)00884838116771(21)632, (01)00884838116771(21)480, (01)00884838116771(21)242, (01)00884838116771(21)316, (01)00884838116771(21)195, (01)00884838116771(21)250, (01)00884838116771(21)93, (01)00884838116771(21)95, (01)00884838116771(21)18, (01)00884838116771(21)111, (01)00884838116771(21)840, (01)00884838116771(21)650, (01)00884838116771(21)795, (01)00884838116771(21)589, (01)00884838116771(21)754, (01)00884838116771(21)634, (01)00884838116771(21)424, (01)00884838116771(21)635, (01)00884838116771(21)717, (01)00884838116771(21)665, (01)00884838116771(21)491, (01)00884838116771(21)523, (01)00884838116771(21)614, (01)00884838116771(21)439, (01)00884838116771(21)412, (01)00884838116771(21)438, (01)00884838116771(21)445, (01)00884838116771(21)408, (01)00884838116771(21)324, (01)00884838116771(21)645, (01)00884838116771(21)534, (01)00884838116771(21)821, (01)00884838116771(21)482, (01)00884838116771(21)339, (01)00884838116771(21)360, (01)00884838116771(21)677, (01)00884838116771(21)142, (01)00884838116771(21)724, (01)00884838116771(21)505, (01)00884838116771(21)604, (01)00884838116771(21)303, (01)00884838116771(21)364, (01)00884838116771(21)73, (01)00884838116771(21)313, (01)00884838116771(21)374, (01)00884838116771(21)365, (01)00884838116771(21)406, (01)00884838116771(21)388, (01)00884838116771(21)387, (01)00884838116771(21)378, (01)00884838116771(21)239, (01)00884838116771(21)241, (01)00884838116771(21)245, (01)00884838116771(21)229, (01)00884838116771(21)135, (01)00884838116771(21)127, (01)00884838116771(21)47, (01)00884838116771(21)58, (01)00884838116771(21)57, (01)00884838116771(21)41, (01)00884838116771(21)19, (01)00884838116771(21)279, (01)00884838116771(21)152, (01)00884838116771(21)301, (01)00884838116771(21)33, (01)00884838116771(21)156, (01)00884838116771(21)197, (01)00884838116771(21)26, (01)00884838116771(21)31, (01)00884838116771(21)32, (01)00884838116771(21)20, (01)00884838116771(21)29, (01)00884838116771(21)158, (01)00884838116771(21)481, (01)00884838116771(21)275, (01)00884838116771(21)45, (01)00884838116771(21)468, (01)00884838116771(21)425, (01)00884838116771(21)361, (01)00884838116771(21)219, (01)00884838116771(21)39, (01)00884838116771(21)27, (01)00884838116771(21)16, (01)00884838116771(21)721, (01)00884838116771(21)78, (01)00884838116771(21)338, (01)00884838116771(21)235, (01)00884838116771(21)236, (01)00884838116771(21)237, (01)00884838116771(21)115, (01)00884838116771(21)160, (01)00884838116771(21)881, (01)00884838116771(21)690, (01)00884838116771(21)23, (01)00884838116771(21)132, (01)00884838116771(21)189, (01)00884838116771(21)44, (01)00884838116771(21)52, (01)00884838116771(21)51, (01)00884838116771(21)38, (01)00884838116771(21)148, (01)00884838116771(21)49, (01)00884838116771(21)91, (01)00884838116771(21)611, (01)00884838116771(21)120, (01)00884838116771(21)177, (01)00884838116771(21)678, (01)00884838116771(21)596, (01)00884838116771(21)850, (01)00884838116771(21)839, (01)00884838116771(21)504, (01)00884838116771(21)767, (01)00884838116771(21)649, (01)00884838116771(21)588, (01)00884838116771(21)506, (01)00884838116771(21)503, (01)00884838116771(21)688, (01)00884838116771(21)733, (01)00884838116771(21)464, (01)00884838116771(21)739, (01)00884838116771(21)720, (01)00884838116771(21)65, (01)00884838116771(21)499, (01)00884838116771(21)474, (01)00884838116771(21)238, (01)00884838116771(21)282, (01)00884838116771(21)143, (01)00884838116771(21)180, (01)00884838116771(21)134, (01)00884838116771(21)224, (01)00884838116771(21)138, (01)00884838116771(21)149, (01)00884838116771(21)179, (01)00884838116771(21)137, (01)00884838116771(21)131, (01)00884838116771(21)106, (01)00884838116771(21)53, (01)00884838116771(21)30, (01)00884838116771(21)17, (01)00884838116771(21)531, (01)00884838116771(21)259, (01)00884838116771(21)254, (01)00884838116771(21)64, (01)00884838116771(21)83, (01)00884838116771(21)90, (01)00884838116771(21)66, (01)00884838116771(21)452, (01)00884838116771(21)682, (01)00884838116771(21)719, (01)00884838116771(21)494, (01)00884838116771(21)578, (01)00884838116771(21)691, (01)00884838116771(21)544, (01)00884838116771(21)302, (01)00884838116771(21)56, (01)00884838116771(21)145, (01)00884838116771(21)103, (01)00884838116771(21)40, (01)00884838116771(21)627, (01)00884838116771(21)552, (01)00884838116771(21)328, (01)00884838116771(21)658, (01)00884838116771(21)631.
Devices named in this recall
8 devices in the FDA device registry whose identifier appears in the recall's code information.
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional fluoroscopic x-ray system11 recalls1 death108 reports
- Azurion Azurion 7 B12Philips Medical Systems Nederland B.V. · Interventional fluoroscopic x-ray system11 recalls5 reports
- Azurion Azurion 7 B20Philips Medical Systems Nederland B.V. · Interventional fluoroscopic x-ray system10 recalls1 death50 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional fluoroscopic x-ray system10 recalls27 reports
- Azurion Azurion 5 M20Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified9 recalls7 reports
- Azurion Azurion 5 M12Philips Medical Systems Nederland B.V. · Interventional fluoroscopic x-ray system8 recalls8 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified7 recalls1 report
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified6 recalls4 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.