Azurion Azurion 5 M12 (REF 722231)Recalls & reported problems

Philips Medical Systems Nederland B.V. · Monoplane Interventional X-Ray System. A ceiling monoplane system with a 12-inch flat detector. · Class 2

All models of Azurion (22)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

8 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
8

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00884838116740), received between Dec 31, 2024 and Apr 30, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (4 reports).

8 FDA recall records name this device's identifier, 8 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 16, 2026Class IIOpen, Classified

    Two software related issues affecting Philips Azurion R3.0 and R3.1 systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications.

    Z-2932-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Apr 28, 2026Class IIOpen, Classified

    Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.

    Z-2225-2026 · Root cause: Device Design · Recall details · FDA record ↗

  3. Mar 31, 2026Class IIOpen, Classified

    Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.

    Z-2170-2026 · Root cause: Device Design · Recall details · FDA record ↗

  4. Jan 26, 2026Class IIOpen, Classified

    Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

    Z-1350-2026 · Root cause: Device Design · Recall details · FDA record ↗

  5. Dec 12, 2025Class IIOpen, Classified

    Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.

    Z-1113-2026 · Root cause: Software design · Recall details · FDA record ↗

  6. Dec 8, 2025Class IIOpen, Classified

    Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.

    Z-1079-2026 · Root cause: Software design · Recall details · FDA record ↗

  7. Jan 13, 2025Class IOpen, Classified

    Possibility of the patient falling from the table related to the mattress used on the patient table.

    Z-1155-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  8. Jan 13, 2025Class IIOpen, Classified

    Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

    Z-1089-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Failure4
  • Positioning Failure2
  • Radiation Output Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 30, 2026Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO PRODUCE X-RAYS. THERE WAS NO HARM REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THE COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25041134 ↗
  • MalfunctionApr 15, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT. THE SYSTEM WAS OUTSIDE OF CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24887030 ↗
  • MalfunctionMar 24, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS OUTSIDE CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24673592 ↗
  • MalfunctionMar 13, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT TO THE CLINICAL APPLICATION. NO HARM WAS REPORTED TO PHILIPS. THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME OF REPORTED EVENT. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24592433 ↗
  • MalfunctionJun 30, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS INTERMITTENTLY GIVING A COMMUNICATION ERROR DURING BOOT, WHICH CAN IMPACT BOOTING. NO HARM WAS REPORTED TO PHILIPS. WE ARE CONSERVATIVELY REPORTING THIS EVENT AS THE INVESTIGATION IS ONGOING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22358281 ↗
  • MalfunctionMay 15, 2025Radiation Output Failure; Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THERE WAS AN ISSUE WITH THE SYSTEM'S COLLIMATOR, WHICH COULD IMPACT IMAGING. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22025327 ↗
  • MalfunctionMar 5, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED TO PHILIPS THAT THE GEOMETRY MOVEMENTS WERE UNAVAILABLE. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21522096 ↗
  • MalfunctionDec 31, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED TO PHILIPS THAT THE GEOMETRY MOVEMENTS WERE UNAVAILABLE. THERE WAS NO REPORTED PATIENT OR USER HARM. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21048210 ↗