Azurion Azurion 7 B12 (REF 722235)Recalls & reported problems
Philips Medical Systems Nederland B.V. · Biplane Interventional X-Ray System with floor-mounted frontal stand (12-inch flat detector), ceiling-mounted lateral stand (12-inch flat detector) and optional interface for an OR table. · Class 2
All models of Azurion (22)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
11 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 11
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00884838116788), received between Mar 4, 2025 and May 28, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (2 reports).
11 FDA recall records name this device's identifier, 11 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 16, 2026Class IIOpen, Classified
Two software related issues affecting Philips Azurion R3.0 and R3.1 systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications.
Z-2932-2026 · Root cause: Software design · Recall details · FDA record ↗
- Apr 28, 2026Class IIOpen, Classified
Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
Z-2225-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Mar 31, 2026Class IIOpen, Classified
Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
Z-2170-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Mar 3, 2026Class IIOpen, Classified
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Z-1675-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Mar 3, 2026Class IIOpen, Classified
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Z-1676-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Jan 26, 2026Class IIOpen, Classified
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Z-1346-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Dec 12, 2025Class IIOpen, Classified
Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.
Z-1113-2026 · Root cause: Software design · Recall details · FDA record ↗
- Dec 8, 2025Class IIOpen, Classified
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
Z-1079-2026 · Root cause: Software design · Recall details · FDA record ↗
- Jan 13, 2025Class IIOpen, Classified
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
Z-1091-2025 · Root cause: Device Design · Recall details · FDA record ↗
- Jan 13, 2025Class IOpen, Classified
Possibility of the patient falling from the table related to the mattress used on the patient table.
Z-1157-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Aug 22, 2024Class IIOpen, Classified
Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
Z-3165-2024 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Activation Failure2
- Radiation Output Failure2
- Poor Quality Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 28, 2026Poor Quality Image
IT WAS REPORTED TO PHILIPS THAT THERE WAS ARTIFACT IN THE IMAGE. THE SYSTEM WAS IN CLINICAL USE AT THE TIME OF THE REPORTED EVENT. THE PROCEDURE WAS CANCELLED AND RESCHEDULED. THERE WAS NO ALLEGATION OF INJURY TO THE PATIENT OR USER. AN INVESTIGATION INTO THIS COMPLAINT HAS BEEN INITIATED BY PHILIPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25319778 ↗MalfunctionFeb 9, 2026Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S POWER DISTRIBUTION UNIT HAD MALFUNCTIONED, WHICH CAN IMPACT A FULL SYSTEM BOOT. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24289440 ↗MalfunctionSep 18, 2025Activation Failure; Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM SHUT DOWN AND TOOK AN EXTENDED TIME TO BOOT BACK UP. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23085040 ↗MalfunctionJul 31, 2025Radiation Output Failure; Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S FOOTSWITCH WAS UNABLE TO CONNECT WIRELESSLY, PREVENTING IMAGING. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22676983 ↗MalfunctionMar 4, 2025Radiation Output Failure; Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE FLEXVISION MONITOR FAILED TO DISPLAY IMAGES. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21511978 ↗
Other Interventional fluoroscopic x-ray system devices
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System25 recalls2 deaths3,984 reports
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths3,614 reports
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System32 recalls2 deaths2,308 reports
- O-ARM® BI70002000Medtronic Navigation, Inc. · Interventional fluoroscopic x-ray system2 recalls1 death3,667 reports
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System28 recalls1 death2,176 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths2,015 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths1,837 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths1,087 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page