Azurion Azurion 3 M15 (REF 722230)Recalls & reported problems

Philips Medical Systems Nederland B.V. · Monoplane Interventional X-Ray System. A floor-mounted monoplane system with a 15-inch flat detector · Class 2

7 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
7

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838116733), received between Mar 17, 2026 and Mar 17, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Activation Failure” (1 report).

7 FDA recall records name this device's identifier, 7 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 16, 2026Open, Classified

    Two software related issues affecting Philips Azurion R3.0 and R3.1 systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications.

    Z-2932-2026 · Root cause: Software design · FDA record ↗

  2. Apr 28, 2026Open, Classified

    Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.

    Z-2225-2026 · Root cause: Device Design · FDA record ↗

  3. Mar 31, 2026Open, Classified

    Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.

    Z-2170-2026 · Root cause: Device Design · FDA record ↗

  4. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1674-2026 · Root cause: Device Design · FDA record ↗

  5. Jan 26, 2026Open, Classified

    Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

    Z-1345-2026 · Root cause: Device Design · FDA record ↗

  6. Dec 12, 2025Open, Classified

    Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.

    Z-1113-2026 · Root cause: Software design · FDA record ↗

  7. Dec 8, 2025Open, Classified

    Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.

    Z-1079-2026 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Activation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 17, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP PROPERLY. THE DEVICE WAS OUTSIDE CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24618100 ↗

Other System, x-ray, fluoroscopic, image-intensified devices

All System, x-ray, fluoroscopic, image-intensified devices →