Azurion recallZ-1349-2026 · Philips Medical Systems Nederland B.V. · initiated Jan 26, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
FDA root cause category: Device Design
What the company is doing
On January 26, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: " Circulate this URGENT Medical Device Correction letter to all users of the system so that they are aware of the issue. " Read and follow the additional instructions in Appendix A regarding the use and configuration of the Reset Geometry function. " Keep Appendix A together with the documentation of the system. Ensure that the letter is in a place likely to be seen/viewed. " Confirm the Reset Geometry function that best suits your clinical workflow. If required, you may request Philips to change the settings. When using a third-party stand-alone device or a stationary accessory. You can request a configuration change free of charge within six (6) months from the date of this Urgent Medical Device Correction letter by contacting your local Philips representative. " In case the affected system has been transferred to another organization, please send a copy of this URGENT Medical Device Correction letter to that organization and inform Philips.
Recall details
- Recall number
- Z-1349-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 26, 2026
- Posted by the FDA
- Feb 11, 2026
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 5136 units (1743 US, 3393 OUS)
- Product code
- OWB
- FDA submissions
- K172822, K181830, K200917
- Distribution
- Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Fiji, Finland, France, FrenchPolynesia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.
Product description
Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Code information (models, lots, serial numbers)
System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); UDI-DIs: (1)884838085268, (2)884838099258, (3)884838116771, (4)N/A; Serial Numbers: All;
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System32 recalls2 deaths2,308 reports
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System28 recalls1 death2,176 reports
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional fluoroscopic x-ray system11 recalls1 death108 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98326).
- Z-1344-2026 · Open, ClassifiedAzurion 3 M12 System Model Numbers: (1) 722063, (2) 722221, (3) 722229; Software Version Number - All versions: R1.X, R2.X, R3.X
- Z-1345-2026 · Open, ClassifiedAzurion 3 M15 System Model Numbers: (1) 722064, (2) 722222, (3) 722230, (4)722280(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
- Z-1346-2026 · Open, ClassifiedAzurion 7 B12 System Model Numbers:(1) 722067, (2) 722225, (3) 722235; Software Version Number - All versions: R1.X, R2.X, R3.X
- Z-1347-2026 · Open, ClassifiedAzurion 7 B20 System Model Numbers: (1)722068, (2)722226, (3)722236; Software Version Number - All versions: R1.X, R2.X, R3.X
- Z-1348-2026 · Open, ClassifiedAzurion 7 M12 System Model Numbers: (1)722078, (2)722223, (3)722233; Software Version Number - All versions: R1.X, R2.X, R3.X
- Z-1350-2026 · Open, ClassifiedAzurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All versions: R1.X, R2.X, R3.X
- Z-1351-2026 · Open, ClassifiedAzurion 5 M20 System Model Numbers: (1)722228, (2)722232, (3)722281(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page