Skip to content

Azurion recallZ-1350-2026 · Philips Medical Systems Nederland B.V. · initiated Jan 26, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

FDA root cause category: Device Design

What the company is doing

On January 26, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: " Circulate this URGENT Medical Device Correction letter to all users of the system so that they are aware of the issue. " Read and follow the additional instructions in Appendix A regarding the use and configuration of the Reset Geometry function. " Keep Appendix A together with the documentation of the system. Ensure that the letter is in a place likely to be seen/viewed. " Confirm the Reset Geometry function that best suits your clinical workflow. If required, you may request Philips to change the settings. When using a third-party stand-alone device or a stationary accessory. You can request a configuration change free of charge within six (6) months from the date of this Urgent Medical Device Correction letter by contacting your local Philips representative. " In case the affected system has been transferred to another organization, please send a copy of this URGENT Medical Device Correction letter to that organization and inform Philips.

Recall details

Recall number
Z-1350-2026
Status
Open, Classified
Initiated by the company
Jan 26, 2026
Posted by the FDA
Feb 11, 2026
Recalling firm
Philips Medical Systems Nederland B.V.
Quantity
326 units (10 US, 316 OUS)
Product code
OWB
FDA submissions
K172822, K181830, K200917
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Fiji, Finland, France, FrenchPolynesia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.

Product description

Azurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All versions: R1.X, R2.X, R3.X

Code information (models, lots, serial numbers)

System Model Numbers: (1)722227, (2)722231; UDI-DIs: (1)884838099227, (2)884838099258, (3)884838116740; Serial Numbers: All;

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98326).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page