Boston Scientific Pacemaker, Implantable Pulse Generator… recallZ-1770-2026 · Boston Scientific Corporation · initiated Mar 19, 2026
Class I Open, Classified All recalls posted in April 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia
FDA root cause category: Nonconforming Material/Component
What the company is doing
Boston Scientific issued an Urgent Medical Device Correction notice to physicians, hospitals, and patients on 3/19/2026 via Federal Express or USPS certified mail. The notice explained the availability of the updated software, Brady Software maintenance release 6 (SMR6) for the ACCOLADE family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps), clinical and device longevity of the update, and recommendations for recommendations on the upgrade. In a separate notice, patients were advised to keep and attend all scheduled medical and device check appointments and talk to their doctor about the best steps for managing their pacemaker system. To confirm if the pacemaker or CRT-P is affected, please visit this website: www.BostonScientific.com/lookup. The language, as well as the model and serial numbers of the device, are required for the lookup which can be found on your patient ID card. Our Product Performance Resource Center at www.bostonscientific.com/ppr, includes information on this topic, a device lookup tool, and instructions for returning explanted products. If you have additional questions, or would like to report on a clinical event, please contact your Boston Scientific representative or Technical Services.
Recall details
- Recall number
- Z-1770-2026
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- Mar 19, 2026
- Posted by the FDA
- Apr 24, 2026
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 718,456 units
- Product code
- LWP
- FDA submissions
- P150012
- Distribution
- worldwide
Product description
Boston Scientific Pacemaker, Implantable Pulse Generator, Pacemaker (non- Cardiac Re-Synchronization Therapy) labeled as: 1. ESSENTIO SR SL, Model Number L100; 2. ESSENTIO DR SL Pacemaker, Model Number L101; 3. ESSENTIO SR SL MRI Pacemaker, Model Number L110; 4. ESSENTIO DR SL MRI Pacemaker, Model Number L111; 5. ESSENTIO DR EL Pacemaker, Model Number L121; 6. ESSENTIO DR EL MRI Pacemaker, Model Number L131; 7. PROPONENT SR SL Pacemaker, Model Number L200; 8. PROPONENT DR SL Pacemaker, Model Number L201; 9. PROPONENT DR (VDD) SL Pacemaker, Model Number L209; 10. PROPONENT SR SL Pacemaker, Model Number L210; 11. PROPONENT DR SL MRI Pacemaker, Model Number L211; 12. PROPONENT DR EL Pacemaker, Model Number L221; 13. PROPONENT DR EL MRI Pacemaker, Model Number L231*; 14. ACCOLADE SR SL Pacemaker, Model Number L300; 15. ACCOLADE DR SL Pacemaker, Model Number L301; 16. ACCOLADE SR SL MRI Pacemaker, Model Number L310; 17. ACCOLADE DR SL MRI Pacemaker, Model Number L311; 18. ACCOLADE DR EL Pacemaker, Model Number L321; 19. ACCOLADE DR EL MRI Pacemaker, Model Number L331*; 20. ALTRUA 2 SR SL Pacemaker, Model Number S701; 21. ALTRUA 2 DR SL Pacemaker, Model Number S702; 22. ALTRUA 2 DR EL Pacemaker, Model Number S722.
Code information (models, lots, serial numbers)
The advisory population includes all models listed; however, the bounding differs by battery type: " All serialized DR-EL pacemakers and CRT-Ps from the ACCOLADE family are included in the advisory population. " ACCOLADE DR-SL and SR-SLs with a use-by-date (UBD) on or before 30 June 2025 are included in the advisory population. Model number alone will not precisely identify individual DR-SL or SR-SL devices in the advisory population. To determine if a device is affected, enter a model/serial into the device lookup tool at www.BostonScientific.com/lookup.
Other devices under the same FDA submission
16 devices cleared under P150012 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ACCOLADE™ MRI EL DR L331Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls22 deaths7,984 reports
- INGEVITY™ MRI 7741Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes36 deaths7,874 reports
- INGEVITY™+ 7841Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes15 deaths7,419 reports
- ACCOLADE™ MRI DR L311Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls59 deaths7,283 reports
- INGEVITY™+ 7842Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes21 deaths5,995 reports
- INGEVITY™ MRI 7742Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes24 deaths3,967 reports
- INGEVITY™ MRI 7740Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes11 deaths3,257 reports
- INGEVITY™+ 7840Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes14 deaths2,640 reports
- ESSENTIO™ MRI DR L111Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls9 deaths1,790 reports
- ACCOLADE™ MRI SR L310Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)1 recall10 deaths1,130 reports
- ESSENTIO™ MRI EL DR L131Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls1 death774 reports
- ESSENTIO™ MRI SR L110Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)1 recall2 deaths467 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98615).
- Z-1771-2026 · Open, ClassifiedBoston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.