TRIGEN recallZ-2456-2026 · Smith & Nephew, Inc. · initiated Mar 16, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.
FDA root cause category: Process control
What the company is doing
On 3/16/2026, recall notices were emailed to customers who were asked to do the following: 1) Locate and quarantine affected devices immediately. If you have further distributed the product to other organizations, please inform them at once of this Field Action and provide to them a copy of this letter. 3. Please complete the Customer Response form and email or fax it to your national Smith+Nephew agency/distributor. 4. Return quarantined product to your national Smith+Nephew agency/distributor. 5. Please maintain awareness of this notice and resulting action for an appropriate period to ensure the effectiveness of the corrective action. Firm contact information: + 1 901 396 2121, 1 800 821 5700
Recall details
- Recall number
- Z-2456-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 16, 2026
- Posted by the FDA
- Jun 16, 2026
- Recalling firm
- Smith & Nephew, Inc.
- Quantity
- 5
- Product code
- HSB
- FDA submissions
- K210980
- Distribution
- International distribution to the countries of BE, DE, IT.
Product description
TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
Code information (models, lots, serial numbers)
UDI-DI: 03596010514394. Lot/Batch: 25BM03616
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- TRIGEN 71760816Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall2 reports
Product lines
Other devices under the same FDA submission
130 devices cleared under K210980 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- TRIGEN 71647336Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES2 recalls1 report
- TRIGEN 71642690Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall3 reports
- TRIGEN 71676627Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall2 reports
- TRIGEN 71676628Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall0 reports
- TRIGEN 71677100Smith & Nephew, Inc. · NAIL, FIXATION, BONE20 reports
- TRIGEN 71677095Smith & Nephew, Inc. · NAIL, FIXATION, BONE18 reports
- TRIGEN 71675202Smith & Nephew, Inc. · NAIL, FIXATION, BONE15 reports
- INTERTAN 71934479Smith & Nephew, Inc. · NAIL, FIXATION, BONE12 reports
- TRIGEN 71645030Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES12 reports
- TRIGEN 71701016LSmith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES12 reports
- TRIGEN 71645035Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES11 reports
- TRIGEN 71677090Smith & Nephew, Inc. · NAIL, FIXATION, BONE11 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.