Impella Heart Pump recallZ-2211-2026 · Abiomed, Inc. · initiated Apr 20, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abiomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.

FDA root cause category: Under Investigation by firm

What the company is doing

On April 20, 2026 URGENT MEDICAL DEVICE RECALL (REMOVAL) letters were sent to customers via 1-day direct mail (US). OUS notification process will be commensurate with local requirements. Actions to be taken: 1. For the devices identified, the Abiomed servicing team will contact you to coordinate the return of the device(s) to implement necessary changes. To ensure continuity of care, hospital inventory can continue to be used. 2. Upon contact from Abiomed s field servicing team, please work with them to return the identified device(s) for the change to be implemented. Refer to Attachment 2 for list of impacted AIC serial numbers. 3. Forward to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). 4. If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.

Recall details

Recall number
Z-2211-2026
Status
Open, Classified
Initiated by the company
Apr 20, 2026
Posted by the FDA
May 22, 2026
Recalling firm
Abiomed, Inc.
Quantity
91 units
Product code
OZD
FDA submissions
P140003, P170011
Distribution
Worldwide - US Nationwide distribution in the states of CA, CT, FL, LA, MA, MI, MO, MS, ND, NY, OH, PA, TX, WY and the countries of Brazil, Canada, Colombia, Hong Kong, Italy, Kuwait, Saudi Arabia, UAE, United Kingdom;

Product description

Automated Impella Controller (AIC); Product Code: 0042-0000-US;

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Product Code: 0042-0000-US; UDI-DI: 00813502010022; Serial Numbers: IC2449, IC2451, IC2452, IC2453, IC2459, IC2460, IC1096, IC1142, IC1202, IC1203, IC1224, IC1231, IC1238, IC1284, IC1285, IC1390, IC1545, IC1577, IC1665, IC1708, IC1741, IC1852, IC1853, IC2029, IC2071, IC2154, IC2277, IC2278, IC2314, IC2315, IC2368, IC2369, IC2370, IC2442, IC2443, IC2444, IC2461, IC2471, IC2476, IC2477, IC2478, IC2479, IC2575, IC2989, IC3127, IC3128, IC3129, IC3130, IC3166, IC3167, IC3213, IC3223, IC3306, IC3307, IC3308, IC3309, IC3342, IC3398, IC3475, IC3656, IC3658, IC3663, IC3664, IC3665, IC3666, IC3674, IC3675, IC3719, IC3720, IC3776, IC3812, IC3887, IC3888, IC4045, IC4060, IC1301, IC1544, IC1572, IC2686, IC2687, IC2848, IC3174, IC3269, IC3272, IC3277, IC3278, IC3281, IC3282, IC3284, IC3294, IC3783;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

14 devices cleared under P140003, P170011 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.