Impella Heart Pump recallZ-3148-2026 · Abiomed, Inc. · initiated Aug 12, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abiomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
FDA root cause category: Under Investigation by firm
What the company is doing
On August 12, 2026 URGENT MEDICAL DEVICE RECALL (REMOVAL) letters were sent to customers. Actions to be Taken: 1. Hospital inventory can continue to be used while awaiting service from Abiomed to ensure continuity of care. 2. Abiomed s field service team will contact you to schedule the specified replacement. 3. In the event of these purge flag component failures and associated alarms, as per IFU, first reinsert purge disc to address the alarm. If that action does not clear the alarm, switch to a backup Automated Impella Controller. 4. Ensure cleaning of the AIC housing with mild detergent as specified in the IFU. 5. Ensure future routine preventive maintenance is performed annually on your AIC(s) following the guidance in the IFU 6. Forward this notice to anyone in your facility that needs to be informed. 7. If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. 8. Post a copy of this notice in a visible area for awareness.
Recall details
- Recall number
- Z-3148-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 12, 2026
- Posted by the FDA
- Sep 16, 2026
- Recalling firm
- Abiomed, Inc.
- Quantity
- 5232 units (3733 US, 1499 OUS) 36 3
- Product code
- OZD
- FDA submissions
- P140003, P170011
- Distribution
- Domestic: Nationwide Distribution. International: United Arab Emirates, Austria, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czechia, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Croatia, Hungary, Ireland, Israel, India, Italy, Japan, Kuwait, Luxembourg, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Pakistan, Poland, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan;
Product description
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN. 4. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP. 5. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 6. Impella Controller, Packaged, US; Product Code: 0042-0000-US.
Code information (models, lots, serial numbers)
Show the full code information (43k characters)
1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA; UDI-DI: 00813502010022; Serial Numbers: IC1324, IC1346, IC1480, IC1481, IC1543, IC1544, IC1572, IC1595, IC1598, IC1664, IC1665, IC1715, IC1769, IC1790, IC1791, IC1911, IC1912, IC1913, IC2032, IC2033, IC2034, IC2035, IC2369, IC2370, IC2686, IC2687, IC3661, IC3662, IC3802, IC3803, IC4944, IC4945, IC5034, IC5050, IC5680, IC5821, IC5824, IC6082, IC6083, IC6633, IC6634, IC6635, IC6636, IC7694, IC7695, IC7716, IC8283, IC9413, IC9414, IC9415, IC9416; 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU; UDI-DI: 00813502011289; Serial Numbers: IC1347, IC1349, IC6845, IC1285, IC1416, IC2765, IC2766, IC3021, IC3022, IC3718, IC3719, IC3720, IC3776, IC5558, IC5559, IC1578, IC1585, IC1902, IC1903, IC1905, IC2678, IC2679, IC2855, IC2856, IC3053, IC3054, IC3329, IC3671, IC3672, IC3830, IC3831, IC4033, IC4837, IC5086, IC5169, IC5171, IC5560, IC5681, IC5682, IC5683, IC6064, IC6081, IC6149, IC6270, IC6272, IC6437, IC6439, IC6441, IC6443, IC6810, IC6812, IC6813, IC6814, IC7261, IC7262, IC7393, IC7394, IC7576, IC7577, IC7646, IC1825, IC1826, IC1870, IC1876, IC4153, IC4175, IC4657, IC4658, IC5476, IC5477, IC5518, IC5520, IC7263, IC7264, IC1011, IC1490, IC1584, IC1590, IC1592, IC2175, IC2176, IC2857, IC2996, IC2997, IC3291, IC3296, IC3317, IC3318, IC3486, IC3488, IC3698, IC3702, IC3724, IC3778, IC3864, IC3870, IC3890, IC3891, IC4151, IC4152, IC4323, IC4324, IC4471, IC4501, IC4681, IC4682, IC4851, IC4892, IC4933, IC4934, IC4973, IC4984, IC4986, IC4987, IC4988, IC5007, IC5048, IC5175, IC5176, IC5195, IC5421, IC5422, IC5433, IC5434, IC5441, IC5442, IC5460, IC5461, IC5463, IC5464, IC5494, IC5495, IC5496, IC5519, IC5611, IC5684, IC5687, IC5688, IC5714, IC5722, IC5818, IC6065, IC6534, IC6535, IC6536, IC6582, IC6587, IC6588, IC6589, IC6590, IC6688, IC6689, IC6690, IC6691, IC6692, IC6693, IC6694, IC6695, IC6706, IC6707, IC6728, IC6738, IC6739, IC6740, IC6741, IC6742, IC6876, IC6877, IC6878, IC6879, IC7388, IC7389, IC7390, IC7392, IC7647, IC7648, IC7649, IC7650, IC7651, IC7652, IC1284, IC2676, IC2677, IC2710, IC2827, IC2828, IC2989, IC3018, IC3020, IC3471, IC5775, IC5777, IC5779, IC5780, IC2099, IC2205, IC2220, IC2654, IC2655, IC2823, IC2824, IC2825, IC2826, IC3668, IC3670, IC3868, IC3869, IC4099, IC4100, IC4209, IC4218, IC4325, IC4326, IC4569, IC4611, IC4624, IC4627, IC4931, IC4932, IC5490, IC5491, IC5492, IC5493, IC6205, IC6275, IC6331, IC6432, IC6434, IC6435, IC6880, IC7407, IC7421, IC7455, IC7675, IC7686, IC7691, IC7692, IC7693, IC7717, IC7720, IC7721, IC7748, IC7809, IC3131, IC3132, IC1549, IC1580, IC2234, IC2236, IC2469, IC2652, IC2653, IC3069, IC3070, IC3311, IC3331, IC3332, IC3657, IC3669, IC3777, IC3828, IC3829, IC4150, IC1396, IC1397, IC1413, IC1415, IC1546, IC1548, IC1724, IC1725, IC1767, IC1768, IC2235, IC2239, IC2483, IC2486, IC2836, IC2838, IC3050, IC3673, IC3696, IC3697, IC4565, IC4839, IC5755, IC5756, IC6487, IC6488, IC7667, IC7674, IC7689, IC7690, IC3301, IC3302, IC1301, IC1302, IC1304, IC1305, IC1217, IC1276, IC1384, IC1386, IC1395, IC1522, IC1523, IC1547, IC1577, IC1579, IC1594, IC1676, IC1684, IC1699, IC1700, IC1722, IC1723, IC1764, IC1821, IC1824, IC1886, IC1974, IC2101, IC2102, IC2329, IC2862, IC2873, IC2874, IC2875, IC2876, IC2877, IC4002, IC4012, IC4013, IC4014, IC4015, IC4464, IC4465, IC4470, IC4566, IC4597, IC4598, IC4609, IC4623, IC4634, IC4635, IC4636, IC4645, IC4656, IC4833, IC4834, IC4835, IC4836, IC4850, IC4852, IC4928, IC4929, IC4930, IC5051, IC5053, IC5057, IC5075, IC5685, IC6579, IC6591, IC6592, IC6593, IC6594, IC6595, IC6708, IC6709, IC6711, IC6715, IC6716, IC6717, IC6804, IC6805, IC6806, IC6807, IC6808, IC6809, IC7098, IC7099, IC7100, IC7258, IC7259, IC7260, IC7580, IC7613, IC7614, IC7615, IC7687, IC7688, IC3817, IC3818, IC5008, IC5009, IC5046, IC5047, IC6223, IC6224, IC7286, IC7378, IC7087, IC7092, IC1265, IC1275, IC1280, IC1283, IC1288, IC1344, IC1345, IC1350, IC1383, IC1411, IC1653, IC1726, IC1761, IC1775, IC1780, IC1881, IC1901, IC1904, IC1907, IC1908, IC1971, IC1972, IC1973, IC1975, IC1976, IC1977, IC1978, IC2005, IC2006, IC2038, IC2039, IC2095, IC2096, IC2098, IC2171, IC2172, IC2173, IC2174, IC2204, IC2207, IC2231, IC2232, IC2233, IC2240, IC2241, IC2242, IC2243, IC2266, IC2267, IC2268, IC2325, IC2326, IC2327, IC2330, IC2332, IC2411, IC2412, IC2413, IC2414, IC2441, IC2473, IC2474, IC2475, IC2481, IC2484, IC2485, IC2490, IC2491, IC2494, IC2512, IC2534, IC2556, IC2557, IC2607, IC2609, IC2610, IC2658, IC2659, IC2670, IC2708, IC2712, IC2752, IC2757, IC2758, IC2763, IC2767, IC2829, IC2830, IC2858, IC2859, IC2878, IC2960, IC2962, IC2964, IC2984, IC2985, IC2987, IC2988, IC3017, IC3019, IC3051, IC3052, IC3067, IC3068, IC3095, IC3096, IC3099, IC3100, IC3101, IC3102, IC3133, IC3134, IC3142, IC3143, IC3144, IC3145, IC3146, IC3147, IC3148, IC3149, IC3178, IC3179, IC3180, IC3181, IC3286, IC3287, IC3288, IC3289, IC3290, IC3292, IC3297, IC3298, IC3299, IC3300, IC3303, IC3304, IC3305, IC3310, IC3313, IC3314, IC3315, IC3316, IC3319, IC3321, IC3322, IC3323, IC3324, IC3327, IC3328, IC3330, IC3458, IC3459, IC3460, IC3461, IC3462, IC3463, IC3466, IC3467, IC3468, IC3472, IC3477, IC3484, IC3485, IC3490, IC3492, IC3493, IC3495, IC3659, IC3660, IC3667, IC3701, IC3703, IC3706, IC3707, IC3708, IC3709, IC3717, IC3781, IC3782, IC3813, IC3814, IC3815, IC3816, IC3832, IC3833, IC3834, IC3859, IC3860, IC3861, IC3862, IC3863, IC3871, IC3872, IC3873, IC3884, IC3885, IC3886, IC3889, IC4034, IC4035, IC4036, IC4037, IC4083, IC4096, IC4097, IC4098, IC4128, IC4129, IC4146, IC4147, IC4148, IC4149, IC4207, IC4208, IC4210, IC4211, IC4212, IC4217, IC4227, IC4228, IC4229, IC4230, IC4234, IC4235, IC4236, IC4238, IC4240, IC4241, IC4242, IC4251, IC4252, IC4253, IC4258, IC4322, IC4328, IC4356, IC4357, IC4359, IC4360, IC4361, IC4362, IC4363, IC4365, IC4366, IC4367, IC4370, IC4371, IC4442, IC4443, IC4444, IC4445, IC4446, IC4466, IC4467, IC4468, IC4469, IC4500, IC4505, IC4506, IC4557, IC4559, IC4560, IC4561, IC4564, IC4567, IC4568, IC4585, IC4594, IC4595, IC4610, IC4612, IC4613, IC4614, IC4626, IC4629, IC4630, IC4631, IC4632, IC4633, IC4646, IC4647, IC4677, IC4678, IC4725, IC4727, IC4728, IC4729, IC4733, IC4734, IC4735, IC4736, IC4737, IC4738, IC4748, IC4757, IC4758, IC4759, IC4760, IC4761, IC4762, IC4763, IC4781, IC4792, IC4793, IC4794, IC4795, IC4796, IC4797, IC4815, IC4816, IC4817, IC4818, IC4819, IC4820, IC4821, IC4822, IC4823, IC4824, IC4840, IC4845, IC4848, IC4869, IC4875, IC5177, IC5517, IC1340, IC1488, IC1586, IC1588, IC1591, IC1677, IC2097, IC2100, IC2178, IC2179, IC2206, IC2244, IC2560, IC2606, IC2656, IC2668, IC2669, IC2709, IC2751, IC2837, IC3015, IC3048, IC3049, IC3097, IC3098, IC3182, IC3183, IC3487, IC3489, IC3491, IC3494, IC3780, IC3866, IC3867, IC4081, IC4329, IC4358, IC4562, IC4563, IC4648, IC4649, IC4723, IC4724, IC4790, IC4791, IC4912, IC4913, IC4915, IC4917, IC5174, IC5356, IC5443, IC5462, IC5686, IC5757, IC5774, IC6078, IC6079, IC6080, IC6147, IC6273, IC6274, IC6431, IC6580, IC6581, IC7077, IC7078, IC7718, IC7719, IC1872, IC1874, IC3704, IC3705, IC5049, IC5055, IC1341, IC1343, IC1381, IC1393, IC1685, IC1720, IC1727, IC1763, IC1765, IC1766, IC2470, IC2472, IC5178, IC5193, IC5194, IC5435, IC5436, IC5440, IC5715, IC5716, IC5717, IC5754, IC6225, IC6236, IC7653, IC7654, IC2442, IC2443, IC2444, IC2471, IC2476, IC2477, IC2478, IC2479, IC3127, IC3128, IC3129, IC3130, IC3306, IC3307, IC3308, IC3309, IC3656, IC3658, IC3663, IC3664, IC3665, IC3666, IC3783, IC3812, IC3887, IC3888, IC4068, IC4069, IC4254, IC4255, IC4256, IC4257, IC4507, IC4508, IC5087, IC5088, IC5210, IC5212, IC5251, IC5252, IC5357, IC5391, IC5392, IC5465, IC5466, IC5467, IC6164, IC6165, IC6214, IC6216, IC6293, IC6294, IC6370, IC6371, IC6372, IC6373, IC6374, IC6375, IC6386, IC6387, IC6510, IC7452, IC7453, IC7454, IC7581, IC7582, IC7583, IC7584, IC3469, IC3470, IC5487, IC5488, IC6449, IC6469, IC6577, IC6578, IC6710, IC6712, IC7093, IC7094, IC7095, IC7096, IC7097, IC7265, IC7395, IC7406, IC7578, IC7579, IC7655, IC7666, IC2961, IC2963, IC3699, IC3700, IC4679, IC4680, IC4914, IC4916, IC5781, IC5817, IC6150, IC6151, IC6436, IC6440, IC6713, IC6714, IC6843, IC7075, IC7076, IC7810, IC1398, IC1524, IC1587, IC1593, IC1678, IC1721, IC1822, IC2021, IC2040, IC2221, IC2222, IC2245, IC2331, IC2380, IC2657, IC2835, IC3016, IC3874, IC3875, IC4055, IC4231, IC4232, IC4233, IC4237, IC4502, IC4503, IC4628, IC4846, IC4911, IC5173, IC5331, IC5489, IC6076, IC6148, IC6474, IC6844; 3. 1. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN; UDI-DI: 00813502011418; Serial Numbers: IC6139, IC6252, IC6253, IC6254, IC6475, IC6476, IC6477, IC6478, IC7079, IC7080, IC7408, IC7409, IC7410, IC7411; 4. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP; UDI-DI: 00813502010428; Serial Numbers: IC4732, IC5027, IC5241, IC5243, IC5245, IC5259, IC5260, IC5261, IC5444, IC5445, IC5447, IC5478, IC5505, IC5506, IC5507, IC5510, IC5511, IC5512, IC5515, IC5516, IC5556, IC5563, IC5651, IC5652, IC5653, IC5654, IC5656, IC5657, IC5658, IC5659, IC5660, IC5665, IC5700, IC5727, IC5728, IC5729, IC5731, IC5732, IC5733, IC5734, IC5735, IC5819, IC5820, IC5860, IC5861, IC5862, IC5863, IC5864, IC5865, IC5866, IC5867, IC5868, IC5930, IC5935, IC5936, IC5937, IC5938, IC6046, IC6047, IC6048, IC6049, IC6084, IC6085, IC6086, IC6087, IC6088, IC6089, IC6090, IC6203, IC6204, IC6210, IC6211, IC6212, IC6306, IC6307, IC6308, IC6309, IC6310, IC6311, IC6312, IC6313, IC6314, IC6315, IC6316, IC6388, IC6389, IC6405, IC6406, IC6407, IC6429, IC6430, IC6442, IC6451, IC6452, IC6466, IC6467, IC6468, IC6490, IC6491, IC6492, IC6519, IC6533, IC6548, IC6549, IC6550, IC6551, IC6552, IC6553, IC6554, IC6555, IC6556, IC6605, IC6606, IC6607, IC6610, IC6642, IC6650, IC6651, IC6652, IC6653, IC6654, IC6656, IC6657, IC6658, IC6659, IC6660, IC6661, IC6662, IC6745, IC6746, IC6747, IC6748, IC6749, IC6750, IC6751, IC6752, IC6855, IC6856, IC6857, IC6858, IC6859, IC6860, IC6861, IC6862, IC6863, IC6882, IC6883, IC6885, IC6886, IC6887, IC6888, IC6985, IC7029, IC7031, IC7033, IC7034, IC7035, IC7058, IC7059, IC7060, IC7061, IC7062, IC7063, IC7064, IC7065, IC7066, IC7067, IC7068, IC7081, IC7082, IC7083, IC7084, IC7085, IC7089, IC7090, IC7091, IC7102, IC7103, IC7104, IC7105, IC7256, IC7257, IC7291, IC7293, IC7294, IC7295, IC7376, IC7377, IC7379, IC7380, IC7381, IC7382, IC7383, IC7384, IC7385, IC7386, IC7387, IC7412, IC7413, IC7414, IC7415, IC7416, IC7417, IC7420, IC7422, IC7423, IC7424, IC7425, IC7426, IC7427, IC7429, IC7430, IC7431, IC7433, IC7434, IC7435, IC7439, IC7440, IC7450, IC7451, IC7593, IC7594, IC7606, IC7607, IC7608, IC7609, IC7610, IC7611, IC7612, IC7672, IC7673, IC7749, IC7750, IC7751, IC7752, IC7753, IC7754, IC7755, IC7779, IC7782, IC7783, IC7784, IC7785, IC7806, IC7808, IC7811, IC7812, IC7813, IC7814, IC7815, IC7846, IC7852, IC7853, IC7896, IC7897, IC7898, IC7899, IC7900, IC7901, IC7902, IC7903, IC7905, IC7926, IC7927, IC7929, IC7933, IC7934, IC7935, IC7976, IC7977, IC7978, IC7979, IC7980, IC7981, IC7982, IC7983, IC7984, IC7985, IC8046, IC8047, IC8048, IC8049, IC8050, IC8051, IC8053, IC8054, IC8076, IC8077, IC8078, IC8079, IC8080, IC8081, IC8082, IC8083, IC8084, IC8276, IC8277, IC8278, IC8279, IC8281; 5. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK; UDI-DI: 00813502011296; Serial Numbers: IC1387, IC1389, IC1519, IC1542, IC1573, IC1792, IC1868, IC1869, IC1909, IC2277, IC2278, IC2493, IC2497, IC2871, IC2872, IC3293, IC3294, IC3295, IC3674, IC3675, IC3784, IC3785, IC4439, IC4440, IC4721, IC4722, IC4739, IC4741, IC4841, IC4842, IC5179, IC5181, IC5219, IC5220, IC5247, IC5553, IC5554, IC5555, IC5831, IC5832, IC5911, IC5912, IC5926, IC5927, IC6286, IC6287, IC6289, IC6290, IC6537, IC6538, IC6540, IC6583, IC6584, IC6585, IC6586, IC6925, IC7056, IC7057, IC7069, IC7070, IC7071, IC7073, IC7074, IC7441, IC7591, IC3326, IC3333, IC3334, IC3464, IC3465, IC3474, IC4599, IC4600, IC5914, IC5915, IC6542, IC6543, IC7776, IC7777, IC5254, IC5833, IC7444, IC7445, IC7446, IC7447, IC1388, IC1545, IC1581, IC1582, IC1583, IC2367, IC2368, IC2492, IC2496, IC4166, IC4330, IC4331, IC4438, IC4843, IC4844, IC4856, IC4857, IC5180, IC5211, IC1910, IC2023, IC2704, IC2705, IC2706, IC2707, IC3312, IC3325, IC4675, IC4676, IC6924, IC6934, IC2024, IC2025, IC5182, IC5184, IC5213, IC5218, IC6291, IC6292, IC7443, IC7585, IC5995, IC5996, IC5997, IC5998, IC5999, IC6001, IC3174, IC3475, IC4441, IC4525, IC5214, IC5215, IC5253, IC5557, IC5830, IC5910, IC6638, IC6669, IC6670, IC6671, IC7668, IC7669, IC7670, IC7671, IC7847, IC7848, IC7778, IC4926, IC4927, IC1494, IC4082, IC7442; 6. Impella Controller, Packaged, US; Product Code: 0042-0000-US; UDI-DI: 00813502010022; Serial Numbers: IC1004, IC1006, IC1007, IC1008, IC1014, IC1015, IC1017, IC1019, IC1024, IC1025, IC1026, IC1027, IC1028, IC1029, IC1030, IC1031, IC1032, IC1033, IC1035, IC1036, IC1037, IC1038, IC1039, IC1040, IC1041, IC1042, IC1043, IC1044, IC1046, IC1047, IC1049, IC1050, IC1051, IC1052, IC1053, IC1055, IC1056, IC1058, IC1060, IC1061, IC1062, IC1064, IC1065, IC1066, IC1067, IC1068, IC1069, IC1070, IC1071, IC1072, IC1073, IC1074, IC1075, IC1076, IC1077, IC1078, IC1079, IC1080, IC1081, IC1082, IC1083, IC1084, IC1085, IC1086, IC1087, IC1088, IC1089, IC1093, IC1094, IC1095, IC1096, IC1097, IC1100, IC1102, IC1103, IC1104, IC1105, IC1106, IC1108, IC1110, IC1111, IC1112, IC1114, IC1115, IC1116, IC1117, IC1119, IC1120, IC1121, IC1122, IC1123, IC1124, IC1132, IC1133, IC1134, IC1135, IC1136, IC1137, IC1138, IC1139, IC1141, IC1142, IC1143, IC1144, IC1145, IC1146, IC1147, IC1148, IC1149, IC1150, IC1151, IC1152, IC1153, IC1154, IC1155, IC1156, IC1157, IC1158, IC1159, IC1160, IC1161, IC1162, IC1163, IC1164, IC1165, IC1166, IC1167, IC1168, IC1169, IC1170, IC1171, IC1172, IC1173, IC1174, IC1175, IC1176, IC1177, IC1178, IC1180, IC1181, IC1182, IC1183, IC1184, IC1185, IC1186, IC1187, IC1188, IC1189, IC1190, IC1191, IC1192, IC1193, IC1194, IC1196, IC1197, IC1198, IC1199, IC1200, IC1201, IC1202, IC1203, IC1204, IC1205, IC1206, IC1207, IC1209, IC1211, IC1212, IC1213, IC1214, IC1215, IC1218, IC1219, IC1220, IC1221, IC1222, IC1223, IC1224, IC1225, IC1226, IC1227, IC1228, IC1229, IC1230, IC1231, IC1233, IC1234, IC1235, IC1236, IC1237, IC1238, IC1239, IC1240, IC1242, IC1243, IC1244, IC1245, IC1246, IC1247, IC1250, IC1251, IC1252, IC1253, IC1254, IC1255, IC1256, IC1257, IC1258, IC1259, IC1261, IC1262, IC1263, IC1264, IC1269, IC1270, IC1271, IC1272, IC1273, IC1274, IC1289, IC1290, IC1291, IC1292, IC1293, IC1294, IC1295, IC1296, IC1297, IC1298, IC1299, IC1300, IC1306, IC1307, IC1309, IC1310, IC1311, IC1312, IC1313, IC1314, IC1315, IC1316, IC1317, IC1319, IC1320, IC1321, IC1322, IC1323, IC1325, IC1326, IC1327, IC1328, IC1329, IC1330, IC1331, IC1332, IC1333, IC1334, IC1335, IC1336, IC1337, IC1338, IC1339, IC1342, IC1351, IC1352, IC1353, IC1354, IC1355, IC1356, IC1357, IC1358, IC1359, IC1360, IC1361, IC1362, IC1363, IC1364, IC1365, IC1368, IC1372, IC1373, IC1374, IC1375, IC1376, IC1377, IC1378, IC1379, IC1380, IC1382, IC1390, IC1394, IC1400, IC1401, IC1402, IC1403, IC1404, IC1405, IC1406, IC1407, IC1408, IC1409, IC1410, IC1412, IC1414, IC1417, IC1418, IC1419, IC1420, IC1421, IC1422, IC1423, IC1424, IC1425, IC1426, IC1427, IC1428, IC1429, IC1430, IC1431, IC1432, IC1433, IC1434, IC1435, IC1436, IC1437, IC1438, IC1439, IC1440, IC1441, IC1442, IC1443, IC1444, IC1445, IC1446, IC1448, IC1449, IC1450, IC1451, IC1452, IC1453, IC1454, IC1455, IC1456, IC1457, IC1458, IC1459, IC1460, IC1461, IC1462, IC1463, IC1464, IC1466, IC1467, IC1469, IC1470, IC1471, IC1472, IC1473, IC1474, IC1475, IC1476, IC1477, IC1478, IC1479, IC1483, IC1484, IC1485, IC1486, IC1489, IC1491, IC1492, IC1495, IC1496, IC1497, IC1498, IC1499, IC1500, IC1502, IC1503, IC1504, IC1505, IC1506, IC1507, IC1508, IC1509, IC1510, IC1512, IC1513, IC1514, IC1515, IC1516, IC1517, IC1518, IC1526, IC1527, IC1528, IC1529, IC1530, IC1531, IC1532, IC1533, IC1534, IC1535, IC1536, IC1537, IC1538, IC1539, IC1540, IC1541, IC1550, IC1551, IC1552, IC1553, IC1554, IC1555, IC1556, IC1557, IC1558, IC1559, IC1560, IC1561, IC1562, IC1563, IC1564, IC1565, IC1566, IC1567, IC1568, IC1569, IC1570, IC1571, IC1574, IC1575, IC1596, IC1599, IC1600, IC1601, IC1602, IC1603, IC1604, IC1605, IC1606, IC1607, IC1608, IC1609, IC1610, IC1612, IC1613, IC1614, IC1615, IC1616, IC1617, IC1618, IC1621, IC1622, IC1623, IC1624, IC1625, IC1626, IC1627, IC1628, IC1629, IC1630, IC1631, IC1632, IC1633, IC1634, IC1635, IC1636, IC1637, IC1638, IC1639, IC1640, IC1641, IC1642, IC1643, IC1644, IC1645, IC1646, IC1648, IC1650, IC1651, IC1652, IC1654, IC1655, IC1656, IC1657, IC1658, IC1659, IC1660, IC1661, IC1662, IC1663, IC1666, IC1668, IC1669, IC1670, IC1671, IC1672, IC1673, IC1674, IC1675, IC1679, IC1680, IC1681, IC1682, IC1683, IC1686, IC1687, IC1688, IC1689, IC1690, IC1691, IC1692, IC1693, IC1694, IC1695, IC1696, IC1697, IC1698, IC1701, IC1702, IC1703, IC1704, IC1705, IC1707, IC1708, IC1709, IC1710, IC1711, IC1712, IC1713, IC1714, IC1716, IC1717, IC1718, IC1719, IC1728, IC1729, IC1730, IC1731, IC1732, IC1733, IC1734, IC1735, IC1736, IC1737, IC1738, IC1739, IC1740, IC1741, IC1742, IC1743, IC1744, IC1746, IC1747, IC1748, IC1749, IC1750, IC1751, IC1752, IC1753, IC1754, IC1755, IC1756, IC1757, IC1758, IC1759, IC1760, IC1770, IC1771, IC1772, IC1773, IC1774, IC1776, IC1777, IC1778, IC1779, IC1781, IC1783, IC1784, IC1785, IC1786, IC1787, IC1788, IC1789, IC1793, IC1794, IC1795, IC1797, IC1799, IC1800, IC1801, IC1802, IC1803, IC1804, IC1805, IC1806, IC1807, IC1808, IC1809, IC1810, IC1811, IC1812, IC1813, IC1814, IC1815, IC1816, IC1817, IC1818, IC1819, IC1820, IC1827, IC1828, IC1829, IC1830, IC1831, IC1832, IC1833, IC1834, IC1835, IC1836, IC1837, IC1838, IC1839, IC1840, IC1841, IC1842, IC1843, IC1844, IC1845, IC1846, IC1847, IC1848, IC1849, IC1850, IC1851, IC1852, IC1853, IC1854, IC1855, IC1857, IC1858, IC1859, IC1860, IC1861, IC1862, IC1863, IC1864, IC1865, IC1866, IC1867, IC1875, IC1877, IC1878, IC1879, IC1880, IC1882, IC1883, IC1884, IC1885, IC1887, IC1888, IC1889, IC1890, IC1891, IC1892, IC1893, IC1894, IC1895, IC1896, IC1897, IC1898, IC1899, IC1900, IC1914, IC1915, IC1916, IC1917, IC1918, IC1919, IC1920, IC1921, IC1922, IC1923, IC1924, IC1925, IC1926, IC1928, IC1929, IC1930, IC1931, IC1932, IC1933, IC1934, IC1935, IC1936, IC1937, IC1938, IC1939, IC1940, IC1941, IC1942, IC1943, IC1944, IC1945, IC1946, IC1947, IC1948, IC1949, IC1950, IC1951, IC1952, IC1953, IC1954, IC1955, IC1956, IC1957, IC1958, IC1959, IC1960, IC1961, IC1962, IC1963, IC1964, IC1965, IC1966, IC1967, IC1968, IC1969, IC1970, IC1979, IC1980, IC1981, IC1982, IC1983, IC1984, IC1985, IC1986, IC1987, IC1988, IC1989, IC1990, IC1991, IC1992, IC1993, IC1994, IC1995, IC1996, IC1997, IC1998, IC1999, IC2000, IC2001, IC2002, IC2003, IC2004, IC2007, IC2008, IC2010, IC2011, IC2012, IC2013, IC2014, IC2015, IC2016, IC2017, IC2018, IC2019, IC2020, IC2022, IC2026, IC2027, IC2028, IC2029, IC2030, IC2031, IC2036, IC2037, IC2041, IC2042, IC2043, IC2044, IC2045, IC2046, IC2047, IC2048, IC2049, IC2050, IC2051, IC2052, IC2053, IC2054, IC2055, IC2056, IC2057, IC2058, IC2059, IC2060, IC2061, IC2062, IC2063, IC2064, IC2065, IC2066, IC2067, IC2068, IC2069, IC2070, IC2071, IC2072, IC2073, IC2074, IC2075, IC2076, IC2077, IC2078, IC2079, IC2080, IC2081, IC2082, IC2083, IC2084, IC2085, IC2086, IC2087, IC2088, IC2089, IC2090, IC2091, IC2092, IC2093, IC2094, IC2103, IC2104, IC2105, IC2106, IC2107, IC2108, IC2109, IC2110, IC2111, IC2112, IC2113, IC2114, IC2115, IC2116, IC2117, IC2118, IC2119, IC2120, IC2121, IC2122, IC2124, IC2125, IC2126, IC2127, IC2128, IC2129, IC2130, IC2131, IC2132, IC2133, IC2134, IC2135, IC2136, IC2137, IC2138, IC2139, IC2140, IC2141, IC2142, IC2146, IC2147, IC2148, IC2149, IC2150, IC2152, IC2153, IC2154, IC2155, IC2157, IC2158, IC2159, IC2160, IC2161, IC2162, IC2163, IC2164, IC2165, IC2166, IC2167, IC2168, IC2169, IC2170, IC2177, IC2180, IC2181, IC2182, IC2183, IC2184, IC2185, IC2186, IC2187, IC2188, IC2189, IC2190, IC2191, IC2192, IC2193, IC2194, IC2195, IC2196, IC2197, IC2198, IC2200, IC2201, IC2202, IC2203, IC2208, IC2209, IC2210, IC2211, IC2212, IC2213, IC2214, IC2215, IC2216, IC2217, IC2218, IC2223, IC2224, IC2225, IC2227, IC2228, IC2229, IC2230, IC2237, IC2238, IC2246, IC2247, IC2248, IC2249, IC2250, IC2251, IC2252, IC2253, IC2254, IC2255, IC2256, IC2257, IC2258, IC2259, IC2260, IC2261, IC2262, IC2263, IC2264, IC2265, IC2269, IC2270, IC2271, IC2272, IC2273, IC2274, IC2275, IC2276, IC2279, IC2280, IC2281, IC2282, IC2283, IC2284, IC2285, IC2286, IC2287, IC2288, IC2289, IC2290, IC2291, IC2292, IC2293, IC2294, IC2295, IC2296, IC2297, IC2298, IC2299, IC2300, IC2301, IC2302, IC2303, IC2304, IC2305, IC2306, IC2307, IC2308, IC2309, IC2310, IC2311, IC2312, IC2313, IC2314, IC2315, IC2316, IC2317, IC2318, IC2319, IC2320, IC2321, IC2322, IC2323, IC2324, IC2333, IC2334, IC2335, IC2336, IC2337, IC2338, IC2339, IC2340, IC2341, IC2342, IC2343, IC2344, IC2345, IC2347, IC2348, IC2349, IC2350, IC2351, IC2352, IC2353, IC2354, IC2355, IC2356, IC2357, IC2358, IC2359, IC2360, IC2361, IC2362, IC2363, IC2364, IC2365, IC2366, IC2371, IC2372, IC2373, IC2374, IC2375, IC2376, IC2377, IC2378, IC2379, IC2381, IC2382, IC2383, IC2384, IC2385, IC2386, IC2387, IC2388, IC2389, IC2390, IC2391, IC2392, IC2393, IC2394, IC2395, IC2396, IC2397, IC2398, IC2399, IC2401, IC2402, IC2403, IC2404, IC2405, IC2406, IC2407, IC2408, IC2409, IC2410, IC2415, IC2416, IC2417, IC2418, IC2419, IC2420, IC2421, IC2422, IC2423, IC2424, IC2425, IC2426, IC2427, IC2428, IC2429, IC2430, IC2431, IC2432, IC2433, IC2434, IC2435, IC2436, IC2437, IC2438, IC2439, IC2440, IC2445, IC2446, IC2447, IC2448, IC2450, IC2454, IC2456, IC2457, IC2458, IC2462, IC2463, IC2464, IC2465, IC2466, IC2467, IC2468, IC2480, IC2482, IC2487, IC2488, IC2489, IC2495, IC2498, IC2499, IC2500, IC2501, IC2502, IC2503, IC2504, IC2505, IC2506, IC2507, IC2508, IC2509, IC2510, IC2511, IC2513, IC2514, IC2515, IC2516, IC2517, IC2518, IC2519, IC2520, IC2521, IC2522, IC2523, IC2524, IC2525, IC2526, IC2527, IC2528, IC2529, IC2530, IC2531, IC2532, IC2533, IC2535, IC2536, IC2537, IC2538, IC2539, IC2540, IC2541, IC2542, IC2543, IC2544, IC2545, IC2546, IC2547, IC2548, IC2549, IC2550, IC2551, IC2552, IC2553, IC2554, IC2555, IC2561, IC2562, IC2563, IC2564, IC2565, IC2566, IC2567, IC2568, IC2569, IC2570, IC2571, IC2572, IC2573, IC2574, IC2575, IC2576, IC2577, IC2578, IC2579, IC2580, IC2581, IC2582, IC2583, IC2584, IC2585, IC2586, IC2587, IC2588, IC2589, IC2590, IC2591, IC2592, IC2593, IC2594, IC2595, IC2596, IC2597, IC2598, IC2599, IC2600, IC2601, IC2602, IC2603, IC2604, IC2605, IC2608, IC2611, IC2612, IC2613, IC2614, IC2615, IC2616, IC2617, IC2618, IC2619, IC2620, IC2621, IC2622, IC2623, IC2624, IC2625, IC2626, IC2627, IC2628, IC2629, IC2630, IC2631, IC2632, IC2633, IC2634, IC2635, IC2636, IC2637, IC2638, IC2639, IC2640, IC2641, IC2643, IC2644, IC2645, IC2646, IC2647, IC2648, IC2649, IC2650, IC2651, IC2660, IC2661, IC2662, IC2663, IC2664, IC2665, IC2666, IC2667, IC2671, IC2672, IC2674, IC2675, IC2680, IC2681, IC2682, IC2684, IC2689, IC2690, IC2691, IC2692, IC2693, IC2694, IC2695, IC2696, IC2697, IC2698, IC2699, IC2700, IC2701, IC2702, IC2703, IC2711, IC2713, IC2714, IC2715, IC2716, IC2717, IC2718, IC2719, IC2721, IC2722, IC2723, IC2724, IC2725, IC2726, IC2727, IC2728, IC2729, IC2730, IC2731, IC2732, IC2733, IC2734, IC2735, IC2736, IC2737, IC2738, IC2739, IC2740, IC2741, IC2742, IC2743, IC2744, IC2745, IC2746, IC2747, IC2748, IC2749, IC2759, IC2760, IC2761, IC2762, IC2764, IC2768, IC2769, IC2770, IC2771, IC2772, IC2773, IC2774, IC2775, IC2776, IC2777, IC2778, IC2779, IC2780, IC2781, IC2782, IC2783, IC2784, IC2785, IC2786, IC2787, IC2788, IC2789, IC2790, IC2791, IC2792, IC2793, IC2794, IC2795, IC2796, IC2797, IC2798, IC2799, IC2800, IC2801, IC2802, IC2803, IC2804, IC2805, IC2806, IC2807, IC2808, IC2809, IC2810, IC2811, IC2812, IC2813, IC2814, IC2815, IC2816, IC2817, IC2818, IC2819, IC2820, IC2821, IC2822, IC2831, IC2832, IC2833, IC2834, IC2840, IC2841, IC2842, IC2843, IC2844, IC2845, IC2846, IC2847, IC2848, IC2849, IC2850, IC2851, IC2852, IC2853, IC2854, IC2863, IC2865, IC2866, IC2867, IC2868, IC2869, IC2870, IC2879, IC2880, IC2881, IC2882, IC2883, IC2884, IC2885, IC2886, IC2887, IC2888, IC2889, IC2890, IC2891, IC2892, IC2893, IC2895, IC2896, IC2897, IC2898, IC2899, IC2901, IC2907, IC2908, IC2909, IC2910, IC2913, IC2914, IC2915, IC2916, IC2917, IC2918, IC2919, IC2920, IC2921, IC2922, IC2923, IC2924, IC2925, IC2926, IC2927, IC2928, IC2929, IC2930, IC2931, IC2932, IC2933, IC2934, IC2935, IC2936, IC2937, IC2938, IC2939, IC2940, IC2941, IC2942, IC2943, IC2944, IC2945, IC2946, IC2947, IC2948, IC2949, IC2950, IC2951, IC2952, IC2953, IC2954, IC2955, IC2956, IC2965, IC2966, IC2969, IC2970, IC2971, IC2972, IC2973, IC2974, IC2975, IC2976, IC2977, IC2978, IC2979, IC2980, IC2981, IC2982, IC2983, IC2990, IC2991, IC2992, IC2993, IC2994, IC2995, IC2998, IC2999, IC3000, IC3001, IC3002, IC3003, IC3004, IC3005, IC3006, IC3007, IC3008, IC3009, IC3010, IC3011, IC3012, IC3013, IC3023, IC3024, IC3025, IC3026, IC3027, IC3028, IC3029, IC3030, IC3031, IC3032, IC3033, IC3034, IC3035, IC3036, IC3037, IC3038, IC3039, IC3040, IC3041, IC3042, IC3043, IC3044, IC3045, IC3046, IC3055, IC3056, IC3057, IC3058, IC3059, IC3060, IC3061, IC3062, IC3063, IC3064, IC3065, IC3066, IC3071, IC3072, IC3073, IC3074, IC3075, IC3076, IC3077, IC3078, IC3079, IC3080, IC3081, IC3082, IC3083, IC3084, IC3085, IC3086, IC3087, IC3088, IC3089, IC3090, IC3091, IC3092, IC3093, IC3094, IC3103, IC3104, IC3105, IC3106, IC3107, IC3108, IC3109, IC3110, IC3111, IC3113, IC3114, IC3115, IC3116, IC3117, IC3118, IC3119, IC3120, IC3121, IC3122, IC3123, IC3124, IC3125, IC3126, IC3135, IC3136, IC3137, IC3138, IC3139, IC3140, IC3141, IC3150, IC3151, IC3152, IC3153, IC3154, IC3155, IC3156, IC3157, IC3158, IC3159, IC3160, IC3161, IC3162, IC3163, IC3164, IC3165, IC3168, IC3169, IC3170, IC3171, IC3172, IC3173, IC3176, IC3177, IC3185, IC3186, IC3187, IC3188, IC3189, IC3191, IC3192, IC3193, IC3194, IC3196, IC3197, IC3198, IC3199, IC3200, IC3201, IC3202, IC3203, IC3204, IC3205, IC3206, IC3207, IC3208, IC3209, IC3210, IC3211, IC3212, IC3214, IC3215, IC3216, IC3217, IC3218, IC3219, IC3220, IC3221, IC3222, IC3224, IC3225, IC3226, IC3227, IC3228, IC3229, IC3230, IC3231, IC3232, IC3233, IC3234, IC3235, IC3236, IC3237, IC3238, IC3239, IC3242, IC3243, IC3244, IC3245, IC3246, IC3247, IC3248, IC3249, IC3250, IC3251, IC3252, IC3253, IC3254, IC3255, IC3257, IC3258, IC3259, IC3260, IC3261, IC3262, IC3263, IC3266, IC3267, IC3268, IC3269, IC3270, IC3271, IC3272, IC3273, IC3274, IC3275, IC3276, IC3277, IC3278, IC3279, IC3281, IC3282, IC3283, IC3284, IC3285, IC3336, IC3337, IC3338, IC3339, IC3340, IC3341, IC3342, IC3343, IC3345, IC3346, IC3347, IC3348, IC3349, IC3350, IC3351, IC3352, IC3353, IC3354, IC3355, IC3356, IC3357, IC3358, IC3359, IC3360, IC3361, IC3362, IC3363, IC3364, IC3365, IC3366, IC3367, IC3368, IC3369, IC3370, IC3371, IC3372, IC3373, IC3374, IC3375, IC3376, IC3377, IC3378, IC3379, IC3380, IC3381, IC3382, IC3383, IC3384, IC3385, IC3386, IC3388, IC3390, IC3394, IC3396, IC3397, IC3398, IC3399, IC3400, IC3401, IC3404, IC3405, IC3406, IC3407, IC3408, IC3409, IC3410, IC3411, IC3412, IC3413, IC3414, IC3415, IC3416, IC3417, IC3418, IC3419, IC3420, IC3421, IC3422, IC3423, IC3424, IC3425, IC3426, IC3427, IC3428, IC3429, IC3432, IC3433, IC3434, IC3436, IC3437, IC3438, IC3439, IC3440, IC3441, IC3443, IC3444, IC3445, IC3446, IC3447, IC3448, IC3449, IC3450, IC3451, IC3452, IC3453, IC3454, IC3455, IC3496, IC3497, IC3498, IC3499, IC3500, IC3501, IC3502, IC3503, IC3504, IC3505, IC3506, IC3507, IC3508, IC3509, IC3510, IC3511, IC3512, IC3513, IC3514, IC3515, IC3516, IC3517, IC3518, IC3519, IC3520, IC3521, IC3522, IC3523, IC3524, IC3525, IC3526, IC3527, IC3528, IC3529, IC3530, IC3542, IC3543, IC3544, IC3545, IC3546, IC3547, IC3548, IC3549, IC3550, IC3551, IC3552, IC3553, IC3554, IC3555, IC3556, IC3557, IC3561, IC3562, IC3563, IC3564, IC3565, IC3566, IC3567, IC3568, IC3569, IC3570, IC3571, IC3572, IC3573, IC3574, IC3577, IC3578, IC3579, IC3580, IC3581, IC3582, IC3589, IC3590, IC3591, IC3592, IC3594, IC3595, IC3596, IC3599, IC3600, IC3601, IC3602, IC3603, IC3604, IC3609, IC3610, IC3622, IC3623, IC3624, IC3625, IC3626, IC3627, IC3628, IC3629, IC3630, IC3631, IC3632, IC3633, IC3634, IC3635, IC3636, IC3637, IC3638, IC3640, IC3641, IC3642, IC3643, IC3644, IC3645, IC3646, IC3647, IC3648, IC3649, IC3650, IC3679, IC3680, IC3681, IC3684, IC3685, IC3688, IC3689, IC3690, IC3691, IC3693, IC3694, IC3695, IC3710, IC3711, IC3712, IC3713, IC3714, IC3715, IC3725, IC3726, IC3727, IC3728, IC3729, IC3730, IC3731, IC3732, IC3733, IC3734, IC3735, IC3736, IC3737, IC3738, IC3739, IC3740, IC3741, IC3742, IC3743, IC3744, IC3745, IC3746, IC3747, IC3748, IC3749, IC3750, IC3751, IC3752, IC3753, IC3754, IC3766, IC3768, IC3769, IC3770, IC3771, IC3772, IC3773, IC3774, IC3775, IC3786, IC3787, IC3788, IC3789, IC3790, IC3791, IC3792, IC3793, IC3794, IC3795, IC3796, IC3797, IC3798, IC3799, IC3800, IC3801, IC3804, IC3805, IC3806, IC3807, IC3808, IC3809, IC3810, IC3811, IC3819, IC3820, IC3821, IC3822, IC3823, IC3825, IC3826, IC3835, IC3838, IC3839, IC3840, IC3841, IC3842, IC3843, IC3844, IC3845, IC3846, IC3847, IC3848, IC3849, IC3850, IC3851, IC3852, IC3853, IC3854, IC3855, IC3856, IC3857, IC3858, IC3865, IC3876, IC3877, IC3878, IC3879, IC3880, IC3881, IC3882, IC3883, IC3892, IC3893, IC3894, IC3895, IC3896, IC3897, IC3898, IC3899, IC3900, IC3901, IC3902, IC3903, IC3904, IC3906, IC3907, IC3908, IC3909, IC3910, IC3911, IC3912, IC3913, IC3914, IC3915, IC3916, IC3917, IC3918, IC3919, IC3920, IC3921, IC3922, IC3923, IC3924, IC3925, IC3926, IC3927, IC3928, IC3929, IC3930, IC3931, IC3932, IC3933, IC3934, IC3935, IC3937, IC3938, IC3939, IC3940, IC3941, IC3942, IC3943, IC3944, IC3945, IC3946, IC3947, IC3948, IC3949, IC3950, IC3951, IC3952, IC3953, IC3954, IC3955, IC3956, IC3957, IC3958, IC3959, IC3960, IC3961, IC3962, IC3963, IC3964, IC3965, IC3966, IC3967, IC3968, IC3969, IC3970, IC3971, IC3972, IC3973, IC3974, IC3975, IC3976, IC3979, IC3980, IC3981, IC3982, IC3983, IC3984, IC3985, IC3986, IC3987, IC3988, IC3989, IC3990, IC3991, IC3992, IC3993, IC3994, IC3995, IC3996, IC3997, IC3998, IC3999, IC4000, IC4001, IC4003, IC4004, IC4005, IC4007, IC4008, IC4009, IC4010, IC4011, IC4016, IC4018, IC4019, IC4020, IC4021, IC4022, IC4023, IC4024, IC4025, IC4026, IC4027, IC4028, IC4029, IC4030, IC4031, IC4038, IC4039, IC4040, IC4041, IC4042, IC4043, IC4044, IC4045, IC4060, IC4061, IC4062, IC4063, IC4064, IC4067, IC4070, IC4071, IC4072, IC4073, IC4074, IC4075, IC4076, IC4077, IC4084, IC4085, IC4087, IC4088, IC4089, IC4090, IC4091, IC4101, IC4102, IC4103, IC4104, IC4105, IC4106, IC4107, IC4108, IC4109, IC4110, IC4111, IC4112, IC4113, IC4114, IC4115, IC4116, IC4130, IC4131, IC4132, IC4133, IC4134, IC4135, IC4136, IC4137, IC4158, IC4159, IC4160, IC4161, IC4162, IC4163, IC4164, IC4165, IC4167, IC4168, IC4169, IC4170, IC4171, IC4172, IC4173, IC4174, IC4191, IC4192, IC4194, IC4195, IC4196, IC4197, IC4198, IC4199, IC4200, IC4201, IC4202, IC4203, IC4204, IC4205, IC4206, IC4219, IC4220, IC4221, IC4222, IC4223, IC4224, IC4225, IC4226, IC4243, IC4244, IC4245, IC4246, IC4247, IC4248, IC4249, IC4250, IC4259, IC4260, IC4261, IC4262, IC4263, IC4264, IC4265, IC4266, IC4267, IC4268, IC4269, IC4270, IC4271, IC4272, IC4273, IC4274, IC4275, IC4276, IC4277, IC4278, IC4279, IC4280, IC4281, IC4282, IC4283, IC4284, IC4285, IC4290, IC4291, IC4292, IC4293, IC4294, IC4295, IC4297, IC4298, IC4299, IC4300, IC4301, IC4302, IC4303, IC4304, IC4305, IC4306, IC4307, IC4308, IC4309, IC4310, IC4311, IC4312, IC4313, IC4314, IC4315, IC4316, IC4317, IC4318, IC4320, IC4321, IC4333, IC4336, IC4337, IC4339, IC4340, IC4341, IC4342, IC4343, IC4344, IC4345, IC4346, IC4347, IC4348, IC4349, IC4350, IC4351, IC4352, IC4353, IC4354, IC4372, IC4373, IC4374, IC4375, IC4376, IC4377, IC4378, IC4379, IC4380, IC4381, IC4382, IC4383, IC4386, IC4387, IC4389, IC4390, IC4391, IC4392, IC4393, IC4394, IC4395, IC4396, IC4397, IC4398, IC4399, IC4400, IC4401, IC4402, IC4403, IC4404, IC4405, IC4406, IC4407, IC4408, IC4409, IC4410, IC4411, IC4412, IC4413, IC4414, IC4415, IC4416, IC4417, IC4418, IC4419, IC4420, IC4422, IC4423, IC4424, IC4425, IC4426, IC4427, IC4428, IC4429, IC4430, IC4431, IC4432, IC4433, IC4434, IC4435, IC4436, IC4437, IC4453, IC4454, IC4455, IC4456, IC4457, IC4458, IC4459, IC4460, IC4461, IC4462, IC4463, IC4472, IC4473, IC4474, IC4475, IC4476, IC4477, IC4478, IC4479, IC4480, IC4481, IC4482, IC4483, IC4484, IC4485, IC4486, IC4487, IC4488, IC4489, IC4490, IC4491, IC4492, IC4493, IC4494, IC4495, IC4496, IC4497, IC4498, IC4499, IC4504, IC4509, IC4510, IC4511, IC4512, IC4513, IC4514, IC4515, IC4516, IC4517, IC4518, IC4519, IC4520, IC4521, IC4522, IC4523, IC4524, IC4526, IC4527, IC4528, IC4529, IC4530, IC4531, IC4532, IC4533, IC4534, IC4535, IC4536, IC4537, IC4538, IC4539, IC4540, IC4541, IC4542, IC4543, IC4544, IC4545, IC4546, IC4547, IC4548, IC4549, IC4550, IC4551, IC4552, IC4554, IC4555, IC4556, IC4558, IC4570, IC4571, IC4572, IC4573, IC4574, IC4575, IC4576, IC4577, IC4578, IC4579, IC4580, IC4581, IC4582, IC4583, IC4584, IC4588, IC4591, IC4593, IC4596, IC4601, IC4602, IC4603, IC4604, IC4605, IC4606, IC4607, IC4608, IC4615, IC4616, IC4617, IC4618, IC4619, IC4620, IC4621, IC4637, IC4638, IC4639, IC4640, IC4641, IC4642, IC4643, IC4644, IC4659, IC4660, IC4661, IC4662, IC4663, IC4664, IC4665, IC4666, IC4667, IC4668, IC4669, IC4670, IC4671, IC4672, IC4673, IC4674, IC4683, IC4684, IC4685, IC4686, IC4687, IC4688, IC4689, IC4690, IC4691, IC4692, IC4693, IC4694, IC4695, IC4696, IC4697, IC4698, IC4699, IC4700, IC4701, IC4702, IC4703, IC4704, IC4705, IC4706, IC4707, IC4708, IC4709, IC4710, IC4711, IC4712, IC4713, IC4714, IC4715, IC4716, IC4717, IC4718, IC4719, IC4720, IC4744, IC4749, IC4750, IC4751, IC4752, IC4753, IC4754, IC4755, IC4756, IC4764, IC4765, IC4766, IC4767, IC4768, IC4769, IC4770, IC4771, IC4772, IC4773, IC4774, IC4775, IC4776, IC4777, IC4778, IC4779, IC4782, IC4783, IC4784, IC4785, IC4786, IC4787, IC4788, IC4789, IC4798, IC4799, IC4800, IC4801, IC4802, IC4803, IC4804, IC4805, IC4806, IC4807, IC4808, IC4809, IC4810, IC4811, IC4812, IC4813, IC4814, IC4825, IC4826, IC4827, IC4828, IC4829, IC4830, IC4831, IC4832, IC4854, IC4858, IC4859, IC4860, IC4861, IC4862, IC4863, IC4864, IC4865, IC4884, IC4885, IC4886, IC4887, IC4888, IC4889, IC4890, IC4893, IC4894, IC4895, IC4896, IC4897, IC4898, IC4899, IC4900, IC4902, IC4903, IC4904, IC4905, IC4906, IC4907, IC4908, IC4909, IC4918, IC4919, IC4920, IC4921, IC4922, IC4923, IC4924, IC4925, IC4935, IC4936, IC4937, IC4938, IC4939, IC4940, IC4941, IC4942, IC4943, IC4956, IC4962, IC4963, IC4964, IC4965, IC4966, IC4967, IC4968, IC4969, IC4970, IC4971, IC4972, IC4974, IC4975, IC4976, IC4977, IC4978, IC4979, IC4980, IC4981, IC4982, IC4983, IC4989, IC4990, IC4991, IC4992, IC4993, IC4994, IC4995, IC4996, IC4997, IC4998, IC4999, IC5001, IC5002, IC5003, IC5004, IC5005, IC5006, IC5017, IC5018, IC5019, IC5024, IC5052, IC5054, IC5056, IC5058, IC5059, IC5060, IC5061, IC5062, IC5063, IC5064, IC5066, IC5067, IC5068, IC5069, IC5070, IC5071, IC5072, IC5073, IC5074, IC5089, IC5090, IC5091, IC5092, IC5093, IC5094, IC5095, IC5116, IC5117, IC5118, IC5119, IC5120, IC5122, IC5123, IC5124, IC5125, IC5146, IC5147, IC5148, IC5149, IC5150, IC5151, IC5152, IC5153, IC5154, IC5155, IC5166, IC5167, IC5168, IC5170, IC5172, IC5183, IC5185, IC5188, IC5189, IC5190, IC5191, IC5192, IC5196, IC5197, IC5198, IC5199, IC5200, IC5201, IC5202, IC5203, IC5204, IC5205, IC5206, IC5207, IC5208, IC5209, IC5221, IC5222, IC5223, IC5224, IC5225, IC5242, IC5244, IC5262, IC5263, IC5264, IC5265, IC5266, IC5267, IC5268, IC5269, IC5270, IC5271, IC5272, IC5273, IC5274, IC5275, IC5286, IC5287, IC5288, IC5289, IC5290, IC5291, IC5292, IC5293, IC5294, IC5295, IC5296, IC5297, IC5298, IC5299, IC5300, IC5301, IC5302, IC5303, IC5304, IC5305, IC5306, IC5307, IC5308, IC5309, IC5310, IC5311, IC5312, IC5313, IC5314, IC5315, IC5316, IC5317, IC5318, IC5319, IC5320, IC5321, IC5322, IC5323, IC5324, IC5325, IC5326, IC5327, IC5328, IC5329, IC5330, IC5332, IC5333, IC5334, IC5335, IC5336, IC5337, IC5338, IC5339, IC5340, IC5342, IC5343, IC5344, IC5345, IC5358, IC5359, IC5360, IC5361, IC5362, IC5363, IC5364, IC5365, IC5366, IC5367, IC5368, IC5369, IC5370, IC5371, IC5372, IC5373, IC5374, IC5375, IC5386, IC5387, IC5388, IC5389, IC5390, IC5393, IC5394, IC5395, IC5416, IC5417, IC5418, IC5419, IC5420, IC5425, IC5453, IC5454, IC5455, IC5456, IC5457, IC5458, IC5459, IC5468, IC5469, IC5470, IC5471, IC5472, IC5473, IC5474, IC5475, IC5479, IC5480, IC5481, IC5482, IC5483, IC5484, IC5485, IC5486, IC5498, IC5499, IC5500, IC5501, IC5502, IC5503, IC5504, IC5536, IC5537, IC5538, IC5539, IC5540, IC5541, IC5542, IC5543, IC5544, IC5545, IC5546, IC5547, IC5548, IC5549, IC5550, IC5551, IC5552, IC5561, IC5562, IC5564, IC5565, IC5596, IC5597, IC5598, IC5599, IC5600, IC5601, IC5602, IC5603, IC5604, IC5605, IC5606, IC5607, IC5608, IC5609, IC5610, IC5612, IC5613, IC5614, IC5615, IC5616, IC5617, IC5619, IC5620, IC5621, IC5623, IC5624, IC5625, IC5647, IC5648, IC5649, IC5650, IC5655, IC5661, IC5662, IC5663, IC5664, IC5676, IC5678, IC5690, IC5691, IC5692, IC5693, IC5694, IC5695, IC5696, IC5697, IC5698, IC5699, IC5701, IC5702, IC5703, IC5704, IC5705, IC5706, IC5707, IC5708, IC5709, IC5710, IC5711, IC5712, IC5713, IC5718, IC5719, IC5720, IC5721, IC5723, IC5724, IC5725, IC5726, IC5736, IC5737, IC5738, IC5739, IC5740, IC5741, IC5742, IC5743, IC5745, IC5746, IC5747, IC5748, IC5749, IC5750, IC5758, IC5759, IC5760, IC5761, IC5762, IC5763, IC5764, IC5765, IC5766, IC5767, IC5768, IC5769, IC5770, IC5772, IC5773, IC5776, IC5778, IC5787, IC5788, IC5789, IC5790, IC5791, IC5792, IC5793, IC5794, IC5795, IC5816, IC5822, IC5823, IC5825, IC5826, IC5827, IC5828, IC5829, IC5838, IC5839, IC5840, IC5841, IC5842, IC5843, IC5844, IC5845, IC5846, IC5847, IC5848, IC5849, IC5850, IC5851, IC5852, IC5853, IC5854, IC5855, IC5856, IC5857, IC5858, IC5859, IC5870, IC5871, IC5872, IC5873, IC5874, IC5875, IC5876, IC5877, IC5878, IC5879, IC5880, IC5881, IC5882, IC5883, IC5884, IC5885, IC5886, IC5887, IC5888, IC5889, IC5890, IC5891, IC5892, IC5893, IC5894, IC5895, IC5906, IC5907, IC5908, IC5909, IC5933, IC5939, IC5940, IC5941, IC5942, IC5943, IC5944, IC5945, IC5946, IC5947, IC5948, IC5949, IC5950, IC5951, IC5952, IC5953, IC5954, IC5955, IC5976, IC5977, IC5978, IC5979, IC5980, IC5981, IC5982, IC5983, IC5984, IC5985, IC5986, IC5987, IC5988, IC5989, IC5990, IC5991, IC5992, IC5993, IC5994, IC6000, IC6002, IC6003, IC6004, IC6005, IC6026, IC6027, IC6028, IC6029, IC6030, IC6031, IC6032, IC6033, IC6034, IC6035, IC6036, IC6037, IC6038, IC6039, IC6040, IC6041, IC6042, IC6043, IC6044, IC6045, IC6050, IC6051, IC6052, IC6053, IC6054, IC6055, IC6056, IC6057, IC6058, IC6059, IC6060, IC6061, IC6062, IC6063, IC6077, IC6092, IC6093, IC6094, IC6095, IC6096, IC6097, IC6098, IC6099, IC6100, IC6101, IC6102, IC6103, IC6104, IC6105, IC6106, IC6107, IC6115, IC6128, IC6129, IC6130, IC6131, IC6132, IC6133, IC6134, IC6135, IC6136, IC6137, IC6138, IC6140, IC6141, IC6142, IC6143, IC6144, IC6145, IC6146, IC6152, IC6153, IC6154, IC6155, IC6156, IC6157, IC6158, IC6159, IC6160, IC6161, IC6176, IC6177, IC6178, IC6179, IC6180, IC6181, IC6182, IC6183, IC6184, IC6185, IC6186, IC6187, IC6188, IC6189, IC6190, IC6191, IC6192, IC6193, IC6194, IC6195, IC6196, IC6197, IC6198, IC6199, IC6200, IC6201, IC6206, IC6208, IC6213, IC6215, IC6217, IC6218, IC6219, IC6220, IC6221, IC6222, IC6237, IC6238, IC6239, IC6240, IC6241, IC6242, IC6243, IC6244, IC6245, IC6246, IC6247, IC6248, IC6249, IC6250, IC6251, IC6255, IC6266, IC6267, IC6268, IC6269, IC6271, IC6288, IC6295, IC6317, IC6318, IC6319, IC6320, IC6321, IC6322, IC6323, IC6324, IC6325, IC6326, IC6327, IC6328, IC6329, IC6333, IC6334, IC6335, IC6346, IC6347, IC6348, IC6349, IC6350, IC6351, IC6352, IC6353, IC6354, IC6355, IC6356, IC6357, IC6358, IC6359, IC6360, IC6361, IC6362, IC6363, IC6364, IC6365, IC6366, IC6367, IC6368, IC6369, IC6390, IC6391, IC6392, IC6393, IC6394, IC6395, IC6396, IC6397, IC6398, IC6399, IC6400, IC6401, IC6402, IC6403, IC6404, IC6408, IC6409, IC6410, IC6411, IC6412, IC6413, IC6414, IC6415, IC6426, IC6427, IC6428, IC6433, IC6438, IC6444, IC6445, IC6446, IC6447, IC6448, IC6450, IC6453, IC6454, IC6455, IC6470, IC6471, IC6472, IC6473, IC6479, IC6480, IC6481, IC6482, IC6483, IC6484, IC6485, IC6486, IC6495, IC6496, IC6497, IC6498, IC6499, IC6500, IC6501, IC6502, IC6503, IC6504, IC6505, IC6506, IC6507, IC6508, IC6509, IC6511, IC6512, IC6513, IC6514, IC6515, IC6516, IC6517, IC6518, IC6520, IC6521, IC6522, IC6523, IC6524, IC6525, IC6544, IC6545, IC6546, IC6547, IC6557, IC6558, IC6559, IC6560, IC6561, IC6562, IC6563, IC6564, IC6565, IC6566, IC6567, IC6568, IC6569, IC6570, IC6571, IC6572, IC6573, IC6574, IC6575, IC6576, IC6596, IC6597, IC6598, IC6599, IC6600, IC6601, IC6602, IC6603, IC6604, IC6611, IC6612, IC6613, IC6614, IC6615, IC6616, IC6617, IC6618, IC6619, IC6620, IC6621, IC6622, IC6623, IC6624, IC6625, IC6626, IC6627, IC6628, IC6629, IC6630, IC6631, IC6632, IC6637, IC6639, IC6640, IC6641, IC6643, IC6644, IC6645, IC6646, IC6647, IC6648, IC6649, IC6663, IC6664, IC6665, IC6666, IC6667, IC6668, IC6672, IC6673, IC6674, IC6675, IC6676, IC6677, IC6678, IC6679, IC6680, IC6681, IC6682, IC6683, IC6684, IC6685, IC6686, IC6687, IC6718, IC6719, IC6720, IC6721, IC6722, IC6723, IC6724, IC6725, IC6726, IC6727, IC6729, IC6730, IC6731, IC6732, IC6733, IC6734, IC6735, IC6736, IC6737, IC6743, IC6744, IC6753, IC6754, IC6755, IC6766, IC6767, IC6768, IC6769, IC6770, IC6771, IC6772, IC6773, IC6774, IC6775, IC6796, IC6797, IC6798, IC6799, IC6800, IC6801, IC6802, IC6803, IC6811, IC6815, IC6816, IC6817, IC6818, IC6819, IC6820, IC6822, IC6823, IC6824, IC6825, IC6826, IC6827, IC6828, IC6829, IC6830, IC6831, IC6832, IC6833, IC6834, IC6835, IC6836, IC6837, IC6838, IC6839, IC6840, IC6841, IC6842, IC6846, IC6847, IC6848, IC6849, IC6850, IC6851, IC6852, IC6853, IC6854, IC6864, IC6865, IC6889, IC6893, IC6894, IC6895, IC6896, IC6897, IC6898, IC6899, IC6900, IC6901, IC6904, IC6905, IC6906, IC6907, IC6908, IC6909, IC6910, IC6911, IC6912, IC6913, IC6914, IC6915, IC6916, IC6917, IC6918, IC6919, IC6920, IC6921, IC6922, IC6923, IC6926, IC6927, IC6928, IC6929, IC6930, IC6931, IC6932, IC6933, IC6935, IC6976, IC6977, IC6978, IC6979, IC6980, IC6981, IC6982, IC6983, IC6984, IC7026, IC7027, IC7028, IC7030, IC7032, IC7919, IC7922, IC3575, IC3576, IC3836, IC3837, IC4065, IC4066, IC4451, IC4452, IC6330, IC6332, IC3166, IC3167, IC3213, IC3223, IC2449, IC2451, IC2452, IC2453, IC2459, IC2460, IC2461, IC1001, IC1002, IC1098, IC1099, IC1140, IC1179;
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Impella 0042-0000-USAbiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump12 recalls120 deaths2,456 reports
- Impella Impella CP Pump SetAbiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump9 recalls9 deaths272 reports
Product lines
Other devices under the same FDA submission
14 devices cleared under P140003, P170011 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Impella 0048-0045Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump5 recalls1,269 deaths11,112 reports
- Impella 1000100Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump5 recalls848 deaths6,093 reports
- Impella 1000080Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump7 recalls987 deaths5,084 reports
- Impella 0042-0040-USAbiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump8 recalls41 deaths1,305 reports
- Impella 1000323Abiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump6 recalls189 deaths943 reports
- Impella 0550-0008Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump4 recalls87 deaths1,005 reports
- Impella 004334Abiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump3 recalls61 deaths408 reports
- Impella 0046-0035Abiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump4 recalls37 deaths258 reports
- Impella 0048-0024Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump2 recalls40 deaths296 reports
- Impella 0048-0032Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump3 recalls3 deaths42 reports
- Impella 0048-0003Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump6 deaths124 reports
- Impella 1000200Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump9 deaths84 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99671).
- Z-3149-2026 · Open, ClassifiedImpella Optical Controller with the below product descriptions and corresponding Product Codes. 1. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 2. Impella …
- Z-3150-2026 · Open, ClassifiedImpella Optical Controller, Automated Impella Controller, Impella Connect Package with the below product descriptions and corresponding Product Codes. 1. Optical, AIC, Impella Conn…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page