Skip to content

Medtronic Midas Rex Legend High Speed Pneumatic System recallZ-0278-05 · Power Surgical Solutions · initiated Nov 11, 2004

Terminated All recalls posted in December 2004

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Power Surgical Solutions or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Softer hose increases the propensity for outer exhaust hose to kink or become occluded as a result of clamping. Complaints of hose rupture/burst have been reported.

FDA root cause category: Other

What the company is doing

The firm initiated the recall via letter on November 11, 2004.

Recall details

Recall number
Z-0278-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Nov 11, 2004
Posted by the FDA
Dec 4, 2004
Recalling firm
Power Surgical Solutions
Quantity
1432 units
Product code
HBB
FDA submissions
K020069
Distribution
Nationwide and to the following foreign countries; Australia; Japan; Canada, China; Netherlands; Columbia; Mexico; Argentina; Sweden; Poland; Spain; Germany; Great Britan; Chile; Switzerland; Venezuela

Product description

Medtronic Midas Rex Legend High Speed Pneumatic System

Code information (models, lots, serial numbers)

Show the full code information (17k characters)

Serial Numbers: A0046-V02 ; A0079-V01 ; A0089-V01 ; A0091-V01 ; A0096-V02 ; A0106-V02 ; A0233-V02 ; A0235-V02 ; A0249-V01 ; A0251-V01 ; A0259-V01 ; A0260-V02 ; A0261-V02 ; A0264-V01 ; A0271-V02 ; A0272-V02 ; A0274-V02 ; A0275-V02 ; A0281-V02 ; A0406-V02 ; A0422-V01 ; A0424-V01 ; A0426-V01 ; A0444-V01 ; A0448-V01 ; A0453-V02 ; A0454-V02 ; A0456-V02 ; A0535-V02 ; A0693-V01 ; A0697-V01 ; A0698-V01 ; A0699-V01 ; A0710-V02 ; A0717-V02 ; A0725-V02 ; A0726-V02 ; A0735-V02 ; A0736-V02 ; A0744-V02 ; A0750-V02 ; A0761-V02 ; A0837-V01 ; A0838-V01 ; A0854-V01 ; A0855-V01 ; A0863-V02 ; A1008-V02 ; A1079-V02 ; A1081-V02 ; A1082-V02 ; A1083-V02 ; A1084-V02 ; A1114-V02 ; A1136-V01 ; A1137-V01 ; A1138-V01 ; A1145-V01 ; A1147-V01 ; A1148-V01 ; A1300-V02 ; A1304-V02 ; A1341-V02 ; A1342-V02 ; A1391-V02 ; A1393-V02 ; A1467-V02 ; A1475-V01 ; A1499-V02 ; A1533-V02 ; A1535-V02 ; A1545-V02 ; A1659-V02 ; A1660-V02 ; A2031-V01 ; A2169-V01 ; A2309-V02 ; A2310-V02 ; A2312-V02 ; A2314-V02 ; A2316-V02 ; A2318-V02 ; A2321-V02 ; A2322-V02 ; A2325-V02 ; A2342-V01 ; A2344-V01 ; A2346-V01 ; A2356-V01 ; A2363-V01 ; a2366-v01 ; A3195-V02 ; A3368-V02 ; A3374-V01 ; A3671-V02 ; A3683-V02 ; A3692-V01 ; A3927-V02 ; A3985-V02 ; A3996-V01 ; A4244-V01 ; A4300-V02 ; A4340-V01 ; A4635-V02 ; A4700-V02 ; A4712-V02 ; A4718-V02 ; A4721-V02 ; A4722-V02 ; A4725-V02 ; A4731-V02 ; A5263-V01 ; A5270-V01 ; A5271-V01 ; A5368-V01 ; A5375-V01 ; A5539-V01 ; A5540-V01 ; A5541-V01 ; A5547-V01 ; A5553-V01 ; A5558-V01 ; A5563-V02 ; A5900-V01 ; A5902-V01 ; A5903-V01 ; A5904-V01 ; A5906-V01 ; A5908-V01 ; A5909-V01 ; A5910-V01 ; A5912-V01 ; A5913-V01 ; A5914-V01 ; A5920-V02 ; A5923-V02 ; A5923-VO2 ; A5924-V02 ; A5925-V02 ; A5926-V02 ; A5927-V02 ; A5928-V02 ; A5929-V02 ; A5935-V02 ; A5936-V02 ; A5939-V02 ; A5940-V02 ; A5941-V02 ; A5943-V02 ; A5944-V02 ; A5945-V02 ; A5948-V02 ; A5949-V02 ; A5952-V02 ; A5955-V02 ; A5956-V02 ; A5957-V02 ; A5959-V02 ; A6034-V02 ; A6035-V02 ; A6036-V02 ; A6037-V02 ; A6038-V02 ; A6039-V02 ; A6040-V02 ; A6041-V02 ; A6042-V02 ; A6043-V02 ; A6044-V02 ; A6048-V02 ; A6049-V02 ; A6050-V02 ; A6051-V02 ; A6052-V02 ; A6054-V01 ; A6055-V01 ; A6056-V01 ; A6057-V01 ; A6058-V01 ; A6059-V01 ; A6060-V01 ; A6061-V01 ; A6062-V01 ; A6066-V01 ; A6067-V01 ; A6068-V01 ; A6069-V01 ; A6070-V01 ; A6071-V01 ; A6072-V01 ; A6073-V02 ; A6075-V02 ; A6077-V02 ; A6078-V02 ; A6079-V02 ; A6080-V02 ; A6082-V02 ; A6235-V02 ; A6236-V02 ; A6237-V02 ; A6239-V02 ; A6240-V02 ; A6241-V02 ; A6242-V02 ; A6243-V02 ; A6245-V02 ; A6247-V02 ; A6248-V02 ; A6250-V02 ; A6251-V02 ; A6252-V02 ; A6253-V02 ; A6256-V01 ; A6257-V01 ; A6258-V01 ; A6259-V01 ; A6260-V01 ; A6262-V01 ; A6264-V01 ; A6266-V01 ; A6267-V01 ; A6269-V01 ; A6271-V01 ; A6272-V01 ; A6273-V01 ; A6435-V02 ; A6436-V02 ; A6441-V01 ; A6442-V01 ; A6443-V01 ; A6444-V01 ; A6448-V01 ; A6450-V01 ; A6453-V01 ; A6455-V01 ; A6459-V01 ; A6475-V02 ; A6476-V02 ; A6477-V02 ; A6479-V02 ; A6480-V02 ; A6481-V02 ; A6483-V02 ; A6484-V02 ; A6486-V02 ; A6487-V02 ; A6488-V02 ; A6489-V02 ; A6490-V02 ; A6491-V02 ; A6492-V02 ; A6493-V02 ; A6494-V02 ; A6497-V02 ; A6498-V02 ; A6499-V02 ; A6500-V02 ; A6502-V02 ; A6509-V02 ; A6515-V02 ; A6516-V02 ; A6517-V02 ; A6518-V02 ; A6726-V01 ; A6729-V01 ; A6736-V01 ; A6741-V01 ; A6742-V01 ; A6743-V01 ; A6744-V01 ; A6745-V01 ; A6746-V01 ; A6747-V01 ; A6748-V01 ; A6749-V01 ; A6750-V01 ; A6751-V01 ; A6752-V01 ; A6753-V01 ; A6754-V01 ; A6880-V01 ; A6882-V01 ; A6883-V01 ; A6884-V01 ; A6886-V01 ; A6888-V01 ; A6892-V01 ; A6893-V01 ; A6894-V01 ; A6895-V01 ; A6896-V01 ; A7074-V01 ; A7075-V01 ; A7076-V01 ; A7077-V01 ; A7078-V01 ; A7079-V01 ; A7080-V01 ; A7081-V01 ; A7082-V01 ; A7083-V01 ; A7113-V02 ; A7114-V02 ; A7115-V02 ; A7117-V02 ; A7118-V02 ; A7119-V02 ; A7120-V02 ; A7121-V02 ; A7122-V02 ; A7123-V02 ; A7124-V02 ; A7127-V02 ; A7128-V02 ; A7129-V02 ; A7130-V02 ; A7131-V02 ; A7132-V02 ; A7134-V02 ; A7136-V02 ; A7142-V02 ; A7384-V01 ; A7385-V01 ; A7386-V01 ; A7388-V01 ; A7389-V01 ; A7390-V01 ; A7391-V01 ; A7392-V01 ; A7393-V01 ; A7394-V01 ; A7395-V01 ; A7397-V01 ; A7398-V01 ; A7399-V01 ; A7401-V01 ; A7402-V01 ; A7403-V01 ; A7404-V01 ; A7405-V01 ; A7407-V01 ; A7411-V01 ; A7415-V01 ; A7416-V01 ; A7417-V01 ; A7418-V01 ; A7419-V01 ; A7421-V01 ; A7422-V01 ; A7423-V01 ; A7424-V01 ; A7425-V01 ; A7427-V01 ; A7436-V02 ; A7440-V02 ; A7441-V02 ; A7444-V02 ; A7445-V02 ; A7446-V02 ; A7447-V02 ; A7448-V02 ; A7449-V02 ; A7453-V02 ; A7454-V02 ; A7455-V02 ; A7462-V02 ; A7469-V02 ; A7470-V02 ; A7471-V02 ; A7474-V02 ; A7475-V02 ; A7476-V02 ; A7477-V02 ; A7478-V02 ; A7479-V02 ; A7480-V02 ; A7481-V02 ; A7482-V02 ; A7483-V02 ; A7871-V01 ; A7961-V02 ; A7962-V02 ; A7963-V02 ; A7964-V02 ; A7965-V02 ; A7966-V02 ; A7967-V02 ; A7968-V02 ; A7969-V02 ; A7970-V02 ; A7981-V02 ; A7982-V02 ; A7983-V02 ; A7984-V02 ; A7985-V02 ; A7986-V02 ; A7987-V02 ; A7988-V02 ; A7989-V02 ; A7990-V02 ; B0019-V02 ; B0020-V02 ; B0027-V02 ; B0032-V02 ; B0033-V02 ; B0034-V02 ; B0037-V02 ; B0038-V02 ; B0044-V02 ; B0046-V02 ; B0047-V02 ; B0048-V02 ; B0051-V02 ; B0058-V02 ; B0060-V02 ; B0064-V02 ; B0066-V02 ; B0071-V02 ; B0072-V02 ; B0074-V02 ; B0079-V02 ; B0092-V02 ; B0102-V02 ; B0108-V02 ; B0111-V02 ; B0114-V02 ; B0121-V02 ; B0122-V02 ; B0133-V02 ; B0137-V02 ; B0166-V02 ; B0179-V02 ; B0225-V02 ; B0248-V01 ; B0283-V01 ; B0297-V02 ; B0298-V02 ; B0300-V02 ; B0301-V02 ; B0302-V02 ; B0303-V02 ; B0305-V02 ; B0306-V02 ; B0308-V02 ; B0315-V02 ; B0316-V02 ; B0319 ; B0320-V02 ; B0321-V02 ; B0325-V02 ; R8002-V01 ; T0088-V01 ; T0236-V02 ; T0400 ; T0447-V02 ; T0458 ; T0461-V02 ; T0465-V01 ; T0486 ; T0490-V02 ; T0500-V02 ; T0505-V02 ; T0514-V02 ; T0615-V01 ; T0619-V01 ; T0620 ; T0681-V02 ; T0687-V02 ; T0688-V02 ; T0699-V02 ; t0706 ; T073 ; T0753 ; T0756-V01 ; T0972-V02 ; T0995-V01 ; T1012-V02 ; T1020-V02 ; T1021-V02 ; T1133-V02 ; T1137 ; T1159-V01 ; T1257-V02 ; T1259-V02 ; T1299-V02 ; T1315-V01 ; T1318 V)1 ; T1322-V01 ; T1332-V01 ; T1337-V01 ; T1364-V02 ; T1368-V02 ; T1372-V01 ; T1375-V02 ; T1381-V02 ; T1391-V02 ; T1392-V02 ; T1414-V01 ; T1433-V01 ; T1434-V01 ; T1435-V01 ; T1462-V02 ; T1525-V02 ; T1541-V01 ; T1548-V01 ; T1549-V01 ; T1550-V01 ; T1559-V01 ; T1573-V02 ; T1579-V02 ; T1580 ; T1603-V02 ; T1604-VO2 ; T1606-V02 ; T1625-V02 ; T1636-V01 ; T1639-V01 ; T1642 ; T1651 ; T1654 ; T1669-V01 ; t1696-v02 ; T1697 ; T1699 ; T1702-V02 ; T1705-V02 ; T1715-V02 ; T1717-V02 ; T1719-V02 ; T1727-V02 ; T1729-V02 ; T1737-V02 ; T1738-V02 ; T1755 ; T1760-V02 ; T1766-V02 ; T1772-V02 ; T1818-V01 ; T1827-V01 ; T1832 ; T1835-V01 ; T1918-V02 ; T1924-V02 ; T1947-V02 ; T1955-V02 ; T1957-V02 ; T1961-V02 ; T1963-V02 ; T1965-V02 ; T1966-V02 ; T1969-V02 ; T1971-V02 ; T1973-V02 ; T1975-V02 ; T1979-V02 ; T1999-V01 ; T2004-V01 ; T2016-V01 ; T2019-V01 ; T2027-V01 ; T2031-V01 ; T2056-V02 ; T2057-V02 ; T2060-V02 ; T2063-V02 ; T2069-V01 ; T2070-V02 ; T2071-V01 ; T2077-V01 ; T2079-V01 ; T2083-V01 ; T2087-V01 ; T2090-V01 ; T2095-V01 ; T2100-V01 ; T2164-V02 ; T2166-V02 ; T2167-V02 ; T2168-V02 ; T2170-V02 ; T2175-V02 ; T2181-V02 ; T2182-V02 ; T2183-V02 ; T2185-V02 ; T2187-V02 ; T2205-V02 ; T2208-V02 ; T2211-V02 ; T2276-V02 ; T2301-V02 ; T2343-V02 ; T2362-V01 ; T2368-V01 ; T2385-V01 ; T2395-V01 ; T2401-V01 ; T2408 ; T2415-V02 ; T2418-V02 ; T2421-V02 ; T2424-V02 ; T2431-V02 ; T2437-V02 ; T2447-V02 ; T2457-V02 ; T2461-V02 ; T2464-V02 ; T2476-V01 ; T2480-V01 ; T2484-V01 ; t2485-v01 ; T2487-V01 ; T2522-V01 ; T2523-V01 ; T2528-V01 ; T2529 ; T2530 ; T2533-V01 ; T2534-V01 ; T2540-V01 ; T2541-V01 ; T2545-V01 ; T2674-V02 ; T2675-V02 ; T2678-V02 ; T2686-V02 ; T2690-V02 ; T2698-V02 ; T2703-V02 ; T2710-V02 ; T2723-V02 ; T2726-V02 ; T2774-V02 ; T2794-V02 ; T2830-V02 ; T2833-V02 ; T2885-V02 ; T2887-V02 ; T2889-V02 ; T2892-V02 ; T2920-V01 ; T2939-V01 ; T2940 ; T2941-V01 ; T2947-V01 ; T2960-V01 ; T2963 ; T2968-V01 ; T2969-V02 ; T2970-V01 ; T2971-V01 ; T2972-V01 ; T2977-V01 ; T2983-V01 ; T2984-V01 ; T3017-V02 ; T3028-V02 ; T3033-V02 ; T3036-V02 ; T3042-V02 ; T3047-V02 ; T3050-V02 ; T3051-V02 ; T3058-V02 ; T3059-V02 ; T3061-V02 ; T3068-V02 ; T3073-V02 ; T3075-V02 ; T3076-V02 ; T3081-V02 ; T3086-V02 ; T3089-V02 ; T3310-V01 ; T3324-V01 ; T3328-V01 ; T3352-V02 ; T3357-V02 ; T3360-V02 ; T3363-V02 ; T3366-V02 ; T3376-V02 ; T3384-V02 ; T3387-0-V02 ; T3388-V02 ; T3391-V02 ; T3393-V02 ; T3394-V02 ; T3400-V02 ; T3601-V01 ; T3606-V01 ; T3617-V01 ; T3619-V01 ; T3630-V02 ; T3635-V02 ; T3639-V02 ; T3645-V02 ; T3651-V02 ; T3655-V02 ; T3659-V02 ; T3662-V02 ; T3794-V02 ; T3799-V02 ; T3801-V02 ; T3808-V02 ; T3809-V02 ; T3821-V02 ; T3829-V02 ; T3831-V02 ; T4003 ; T4007-V01 ; T4015-V01 ; T4016-V01 ; T4019-V01 ; T4021-V01 ; T4050-V02 ; t4053-v02 ; T4058-V02 ; T4062-V02 ; T4070-V02 ; T4078-V02 ; T4133-V02 ; T4134-V02 ; T4139-V02 ; T4141-V01 ; T4143-V02 ; T4147-V02 ; T4148-V02 ; T4150-V02 ; T4152-V02 ; T4155-V02 ; t4161 ; T4163-V02 ; T4165-V02 ; T4169-V02 ; T4173-V02 ; T4190-V02 ; T4359-V01 ; T4361-V01 ; T4369-V01 ; T4370 ; T4385-V01 ; T4388-V01 ; T4395-V01 ; T4396-V01 ; T4397-V01 ; T4460-V02 ; T4464-V02 ; T4469-V02 ; T4476-V02 ; T4479-V01 ; t4513 ; T4530-V02 ; T4538 ; T4543-VO2 ; T4547-V02 ; T4551-V02 ; T4552-V02 ; T4554-V02 ; T4561-V02 ; T4567-V02 ; T4575-V02 ; T4599-V02 ; T4604-V02 ; T4607-V02 ; T4609-V02 ; T4681-V01 ; T4683-V01 ; T4692-V01 ; T4693-V01 ; T4694-V01 ; T4697-V01 ; T4700-V01 ; T4704-V01 ; T4773-V01 ; T4776-V01 ; T4778-V01 ; T4779-V01 ; T4782-V01 ; T4793-V01 ; T4794-V01 ; T4802-V01 ; T4805-V01 ; T5220 ; T5224 ; T5227-V02 ; T5228-V02 ; T5241-V02 ; T5244-V02 ; T5253-V02 ; T5261-V02 ; T5276-V02 ; T5277-V02 ; T5283-V02 ; T5287-V02 ; T5289-V02 ; T5290-V02 ; T5293-V02 ; T5295-V02 ; T5300-V02 ; T5302-V02 ; T5303-V02 ; T5317-V02 ; T5324-V02 ; T5327-V02 ; T5330-V02 ; T5332-V02 ; T5334-V02 ; T5335-V02 ; T5337-V02 ; T5350-V02 ; T5359 ; T5376-V02 ; T5378-V02 ; T5380-V02 ; T5384-V02 ; T5389-V02 ; T5396-V01 ; T5397-V01 ; T5398-V01 ; T5400-V01 ; T5401-V01 ; T5402-V01 ; T5406-V01 ; T5427-V01 ; T5433-V01 ; T5440-V01 ; T5441-V01 ; T5443-V01 ; T5445-V01 ; T5447-V01 ; T5448-V01 ; T5453-V01 ; T5547-V02 ; T5556-V02 ; T5558-V02 ; T5564VO2 ; T5704-V02 ; T5709-V02 ; T5710-V02 ; T5711-V02 ; T5713-V02 ; T5714-V02 ; T57152-V02 ; T5718-V02 ; T5729-V02 ; T5730-V02 ; T5733-V02 ; T5737-V02 ; T5747-V02 ; T5749-V02 ; T5750-V02 ; T5752-V02 ; T5753-V02 ; T5756-V02 ; T5757-V02 ; T6007-V01 ; T600-V02 ; T6020-V01 ; T6026 ; T6027-V01 ; T6046-V02 ; T6050-V02 ; T6067-V02 ; T6077-V02 ; T6079-V02 ; T6081-V02 ; T6145-V01 ; T6159-V01 ; T6188-V02 ; T6193-V02 ; T6196-V02 ; T6198-V02 ; T6205-V02 ; T6211-V02 ; T6215-V02 ; T6216-V02 ; T6226-V02 ; T6235 ; T6241-V02 ; T6286-V01 ; T6292-V01 ; T6293-V01 ; T6302-V01 ; T6303-V01 ; T6307-V02 ; T6309-V02 ; T6315-V02 ; T6318-V02 ; T6321-V02 ; T6322-V02 ; T6325-V02 ; T6326-V02 ; T6327-V02 ; T6354-V02 ; T6420-V01 ; T6426-V01 ; T6439-V01 ; T6442-V01 ; T6447-V02 ; T6451-V02 ; T6455-V02 ; T6468-V02 ; T6473-V02 ; T6474-V02 ; T6488-V02 ; T6490-V02 ; T6496-V02 ; T6507-V02 ; T6508-V02 ; T6513-V02 ; T6667-V02 ; T6668-V02 ; T6669-V02 ; T6672-V02 ; T6676-V02 ; T6684-V02 ; T6685-V02 ; T6701-V02 ; T6704-V02 ; T6706-V02 ; T6770-V01 ; T6771-V01 ; T6799-V01 ; T6816-V02 ; T6827-V01 ; T6830-V01 ; T6833-V01 ; T6836-V01 ; T6843-V01 ; T6844-V01 ; T6881-V01 ; T6882-V01 ; T6884-V01 ; T6886 ; T6886-V01 ; T6892-V01 ; T6893-V01 ; T6906-V01 ; T6911-V01 ; T6921-V01 ; T6925-V01 ; T6927-V01 ; T6928-V01 ; T6933-V02 ; T6939-V02 ; T6941-V02 ; T6943-V02 ; T6947-V02 ; T6958-V02 ; T7001-V02 ; T7017-V02 ; T7021-V01 ; T7024 ; T7031-V01 ; T7032-V01 ; T7038-V01 ; T7058-V01 ; T7062-V01 ; T7063-V01 ; T7064-V01 ; T7074-V01 ; T7082-V02 ; T7085-V02 ; T7086-V02 ; T7088-V02 ; T7094-V02 ; T7168-V01 ; T7170-V01 ; T7172-V01 ; T7174-V01 ; T7256-V02 ; T7257-V02 ; T7262-V02 ; T7268-V02 ; T7275-V02 ; T7288-V02 ; T7289-V02 ; T7291-V02 ; T7292-V02 ; T7293-V02 ; T7298-V02 ; T7305-V02 ; T7312-V02 ; T7314-V02 ; T7316-V02 ; T7328-V01 ; T7335-V01 ; T7338-V01 ; T7340-V01 ; T7347-V01 ; T7513-V02 ; T7514-V02 ; T7518-V02 ; T7520-V02 ; T7523-V02 ; T7525-V02 ; T7682-V02 ; T7754-V01 ; T7757-V01 ; T7758-V01 ; T7762-V01 ; T7766-V01 ; T7767-V01 ; T7780-V02 ; T7785-V02 ; T7787-V02 ; T7790-V02 ; T7793-V02 ; T7795-V02 ; T7801-V02 ; T7819-V01 ; T7823-V01 ; T7827-V01 ; T7836-V02 ; T7842 ; T7888-V02 ; T7890-V02 ; T7903-V02 ; T7927-V01 ; T7933-V01 ; T7937-V01 ; T7938-V01 ; T7940-V01 ; T7942-V01 ; T7944-V01 ; T7988-V01 ; T7989-V01 ; T7991-V01 ; T7999-V01 ; T8019-V02 ; T8023-V02 ; T8024-V02 ; T8030-V01 ; T8045-V01 ; T8098-V01 ; T8110-V01 ; T8114-V01 ; T8116-V01 ; T8123-V02 ; T8133-V02 ; T8141-V02 ; T8143-V02 ; T8147-V02 ; T8151-V02 ; t8154-V02 ; T8155-V02 ; T8159-V02 ; T8160-V01 ; T8161-V02 ; T8162-V02 ; T8175-V02 ; T8261-V02 ; T8273-V01 ; T8276-V01 ; T8277-V01 ; T8280-V01 ; T8281-V01 ; T8283-V01 ; T8285-V01 ; T8286-V01 ; T8289-V01 ; T8291-V01 ; T8297-V02 ; T8299-V02 ; T8304-V02 ; T8305-V02 ; T8309-V02 ; T8310-V02 ; T8312-V02 ; T8313-V02 ; T8389-V01 ; T8390-V01 ; T8392-V01 ; T8398-V01 ; T8403-V02 ; T8409-V02 ; T8410-V02 ; T8413-V02 ; T8416-V02 ; T8417-V02 ; T8418-V02 ; T8422-V02 ; T8425-V02 ; T8426-V01 ; T8428-V02 ; T8432-V02 ; T8434-V02 ; T8442-V02 ; T8456-V01 ; T8460-V01 ; T8462-V01 ; T8466-V01 ; T8472-V01 ; T8473-V01 ; T8497-V02 ; T8517-V02 ; T8518-V02 ; T8578-V02 ; T8581-V02 ; T8590-V02 ; T8600-V01 ; T8603-V01 ; T8605-V01 ; T8606-V01 ; T8612-V01 ; T8660-V02 ; T8662-V02 ; T8675-V02 ; T8682-V01 ; T8685-V01 ; T8688-V01 ; T8689-V01 ; T8692-V01 ; T8693-V01 ; T8713-V02 ; T8715-V02 ; T8716-V02 ; T8730-V02 ; T8732-V02 ; T8734-VO2 ; T8736-V02 ; T8738-V02 ; T8746-V01 ; T8758-V01 ; T8763-V01 ; T8764-V01 ; T8770-V01 ; T8798-V01 ; T8869-V01 ; T8870-V01 ; T8876-V02 ; T8881-V02 ; T8933 ; T9193-V01 ; T9279-V01 ; T9291-V02 ; T9293-V02 ; T9304-V01 ; T9315-V01 ; T9324-V02 ; T9340-V02 ; T9343-V02 ; T9350-V02 ; T9354-V02 ; T9549-V02 ; T9562-V02 ; T9575-V02 ; T9578-V02 ; T9585-V02 ; T9588-V02 ; T9593-V01 ; T9600-V01 ; T9604-V01 ; T9726-V02 ; T9730-V02 ; T9731-V02 ; T9736-V02 ; T9739-V02 ; T9743-V02 ; T9744-V02 ; T9773-V02 ; T9778-V02 ; T9882-V02 ; T9885-V02 ; T9910-V01 ; T9917-V01 ; T9953-V02 ; A0087-V01 ; A0236-V01 ; A0401-V02 ; A0408-V02 ; A0425-V01 ; A0463-V02 ; A1305-V02 ; A1345-V02 ; A1649-V02 ; A2032-V01 ; A2462-V02 ; a2465-v01 ; A278-V01 ; A3178-V02 ; A3375-V01 ; A3992-V01 ; A4589-V02 ; A4652-V02 ; A4719-V02 ; A4720-V02 ; A4724-V02 ; A4728-V02 ; A4730-V02 ; A4733-V02 ; A5905-V01 ; A5907-V01 ; A5911-V01 ; A5921-V02 ; A5922-V02 ; A5930-V02 ; A5931-V02 ; A5932-V02 ; A5933-V02 ; A5934-V02 ; A5937-V02 ; A5938-V02 ; A5942-V02 ; A5946-V02 ; A5950-V02 ; A5951-V02 ; A5953-V02 ; A5954-V02 ; A5958-V02 ; A6033-V02 ; A6045-V02 ; A6046-V02 ; A6047-V02 ; A6065-V01 ; A6074-V02 ; A6076-V02 ; A6081-V02 ; A6238-V02 ; A6244-V02 ; A6246-V02 ; A6249-V02 ; A6254-V02 ; A6261-V01 ; A6263-V01 ; A6265-V01 ; A6268-V01 ; A6270-V01 ; A6427-V02 ; A6428-V02 ; A6429-V02 ; A6430-V02 ; A6431-V02 ; A6432-V02 ; A6445-V01 ; A6446-V01 ; A6447-V01 ; A6449-V01 ; A6451-V01 ; A6452-V01 ; A6454-V01 ; A6456-V01 ; A6457-V01 ; A6458-V01 ; A6460-V01 ; A6478-V02 ; A6482-V02 ; A6485-V02 ; A6495-V02 ; A6496-V02 ; A6501-V02 ; A6503-V02 ; A6504-V02 ; A6505-V02 ; A6506-V02 ; A6507-V02 ; A6508-V02 ; A6510-V02 ; A6511-V02 ; A6512-V02 ; A6513-V02 ; A6514-V02 ; A6519-V02 ; A6520-V02 ; A6521-V02 ; A6522-V02 ; A6523-V02 ; A6524-V02 ; A6725-V01 ; A6727-V01 ; A6728-V01 ; A6730-V01 ; A6731-V01 ; A6732-V01 ; A6733-V01 ; A6734-V01 ; A6735-V01 ; A6737-V01 ; A6738-V01 ; A6739-V01 ; A6740-V01 ; A6877-V01 ; A6878-V01 ; A6879-V01 ; A6881-V01 ; A6885-V01 ; A6887-V01 ; A6889-V01 ; A6890-V01 ; A6891-V01 ; A7116-V02 ; A7125-V02 ; A7126-V02 ; A7133-V02 ; A7135-V02 ; A7137-V02 ; A7138-V02 ; A7139-V02 ; A7140-V02 ; A7141-V02 ; A7175-V01 ; A7176-V01 ; A7177-V01 ; A7178-V01 ; A7179-V01 ; A7180-V01 ; A7181-V01 ; A7182-V01 ; A7387-V01 ; A7396-V01 ; A7400-V01 ; A7406-V01 ; A7408-V01 ; A7409-V01 ; A7410-V01 ; A7414-V01 ; A7420-V01 ; A7426-V01 ; A7428-V01 ; A7429-V01 ; A7430-V01 ; A7431-V01 ; A7432-V01 ; A7433-V01 ; A7434-V02 ; A7435-V02 ; A7437-V02 ; A7438-V02 ; A7439-V02 ; A7442-V02 ; A7443-V02 ; A7450-V02 ; A7451-V02 ; A7452-V02 ; A7456-V02 ; A7457-V02 ; A7458-V02 ; A7459-V02 ; A7460-V02 ; A7461-V02 ; A7463-V02 ; A7464-V02 ; A7465-V02 ; A7466-V02 ; A7467-V02 ; A7468-V02 ; A7472-V02 ; A7473-V02 ; A7868-V01 ; A7869-V01 ; A7875-V01 ; A7876-V01 ; A7877-V01 ; A7878-V01 ; A7977-V01 ; B0030-V02 ; B0109-V02 ; B0299-V02 ; B0304-V02 ; B0307-V02 ; B0310-V02 ; B0311-V02 ; B0312-V02 ; B0313-V02 ; B0314-V02 ; T0476-V0-2 ; T0752-V01 ; T0985-V02 ; T0988 ; T1145-V01 ; T1330-V01 ; T1344-V02 ; T1449-V01 ; T1519 ; T1563-V01 ; T1838-V01 ; T1840 ; T1991-V02 ; T2078-V01 ; T2291-V02 ; T2398-V01 ; T2433-VO2 ; T2446-V02 ; T2525-V01 ; T2937 ; T2953 ; T2964-V01 ; T2969-V01 ; T4176-V02 ; t4358-v01 ; T4370-V01 ; T4377-V01 ; T4384 ; T4465-V02 ; T4558-V02 ; T4582-V02 ; T4799-V01 ; T5230-V02 ; T5388-V02 ; T5390 ; T5439-V01 ; T5449-V01 ; T5561-V02 ; T6012-V01 ; T6042-V02 ; T6237-V02 ; T6443-V01 ; T6481-V02 ; T6841-V01 ; T6912 ; T6931-V02 ; T7079-V02 ; T730--V02 ; T7930-V01 ; T8111-V01 ; T8288-V01 ; t8458-v01 ; T8673-V01 ; T8684-V01 ; T8717-V02 ; T8739-V02 ; T8771-V01 ; T8779-V01 ; T8889-V02 ; T9232-V01 ; T9292-V02 ; T9295-V02 ; T9314-V01 ; T9570-V02 ; T9786-V01

Other devices under the same FDA submission

109 devices cleared under K020069 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page