Skip to content

Cordis Neurovascular Pre-Shaped PROWLER Infusion Catheters recallZ-1482-06 · Cordis Neurovascular, Inc. · initiated Aug 11, 2006

Terminated All recalls posted in September 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cordis Neurovascular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sterility (package integrity) compromised-Cordis Neurovascular, Inc. discovered, during internal testing, that some catheters within the affected lots of CNV preshaped PROWLER and pre-shaped PROWLER SELECT Infusion Catheters may have a pinhole or tear in the Mylar pouch, which may result in a compromise of the sterility inside the pouch.

FDA root cause category: Other

What the company is doing

All consignees were notified by letter on 8/14/06. All Cordis Neurovascular sales representatives will also participate, to ensure expedited communication.

Recall details

Recall number
Z-1482-06
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 11, 2006
Posted by the FDA
Sep 2, 2006
Recalling firm
Cordis Neurovascular, Inc.
Quantity
15 Lot numbers (46,452)
Product code
KRA
FDA submissions
K021591
Distribution
Class II Recall - Worldwide distribution --- including USA and countries of Asia Pacific, Europe, Latin America, Canada, Japan and Australia.

Product description

Cordis Neurovascular Pre-Shaped PROWLER Infusion Catheters.

Code information (models, lots, serial numbers)

All lot numbers equal to or lower than: CAT 606051FX - LOT 13084977; CAT 606051JX - LOT 13077739; CAT 606051MX - LOT 13100828; CAT 606151FX - LOT 13095822; CAT 606151JX - LOT 13090978; CAT 606151MX - LOT 13090976; CAT 6062310F - LOT 13090973; CAT 6062310J - LOT 13090974; CAT 6062310M - LOT 13082015; CAT 6062510FX - LOT 13075472; CAT 6062510JX - LOT 13090975; CAT 6062510MX - LOT 13082020; CAT 6062511FX - LOT 13090979; CAT 6062511JX - LOT 13090971; CAT 6062511MX - LOT 13090972.

Other devices under the same FDA submission

9 devices cleared under K021591 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 36077).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page