Skip to content

Medtronic InterStim INS (Model 3023) with N'Vision Inter… recallZ-0040-2008 · Medtronic Neuromodulation · initiated Jun 25, 2007

Terminated All recalls posted in October 2007

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software issue may cause false low battery alert to be displayed when interrogating an Interstim Implantable Neurostimulator (Model 3023) with an N'Vision Clinician Programmer (Model 8840), or with an Inter Stim iCon Patient Programmer (Model 3037).

FDA root cause category: Software design

What the company is doing

Begining on July 11, 2007 a letter to the physicians was sent informing them of the issue and patient management for the two issues. The letter includes a Cover Letter, A Reply Form, and two Recall letters explaining the issue in detail.

Recall details

Recall number
Z-0040-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jun 25, 2007
Posted by the FDA
Oct 12, 2007
Recalling firm
Medtronic Neuromodulation
Quantity
1885
Product code
EZW
FDA submissions
P970004
Distribution
Nationwide and the Netherlands.

Product description

Medtronic InterStim INS (Model 3023) with N'Vision Inter Stim-B software supplied on the Model 8870 Version MMB_01/NNB_01 with N'Vision Clinician Programmer Model 8840 or InterStim iCon Patient Programmer Model 3037. Implantable Neurostimulator for Urinary Control.

Code information (models, lots, serial numbers)

All serial numbers within these catalog/model numbers are affected (Model 3023, Catalog Number 8870MMB01 and 8870NNB01)

Other devices under the same FDA submission

60 devices cleared under P970004 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page