Medtronic Midas Rex Legend tapered dissecting tool, Model… recallZ-2156-2008 · Medtronic Powered Surgical Solutions · initiated May 30, 2008
Terminated All recalls posted in August 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Powered Surgical Solutions or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Cutting flute geometry of surgical dissecting tools may be outside of established tolerance. This may lead to reduced cutting performance, inability to perform desired dissection, increased risk of overheating, and potential need for revision surgery.
FDA root cause category: Process control
What the company is doing
Firm notified consignees by Urgent: Medical Device Recall letter sent May 30, 2008. The letter requires consignees to immediately compare their inventory of products against the list provided and quarantine any unused dissecting tools. The letter also states that Medtronic would provide replacement product to match the quantity and type returned by the consignees' institution.
Recall details
- Recall number
- Z-2156-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- May 30, 2008
- Posted by the FDA
- Aug 23, 2008
- Recalling firm
- Medtronic Powered Surgical Solutions
- Quantity
- 1,306 units
- Product code
- HBB
- FDA submissions
- K020069
- Distribution
- Nationwide Distribution.
Product description
Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General Surgical and Plastic Surgery applications for transection, osteotomy, graft harvesting, bone shaping, laminectomy, entry hole, suture hole, and midface advancement.
Code information (models, lots, serial numbers)
Lot numbers: 0000710253, 0000710254, 0000710256, 0000710257, 0000710258, 0000710266, 0000710267, 0000710271, 0000710272, 0000710274, 0000710275, 0000710276, 0000710277, 0000710278, 0000710279, 0000710280, 0000710281, and 0000710288
Other devices under the same FDA submission
109 devices cleared under K020069 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Midas Rex® AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC163 reports
- MIDAS REX LEGEND AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC40 reports
- Midas Rex® AF02RMedtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC34 reports
- Midas Rex® AS09SMedtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC31 reports
- Midas Rex® AS09Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC30 reports
- MIDAS REX LEGEND AA10Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC29 reports
- Midas Rex® AA10Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC27 reports
- Midas Rex® AF01Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC25 reports
- Midas Rex® AS08Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC21 reports
- Midas Rex™ F2/8TA23SMedtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC19 reports
- Midas Rex™ F2/8TA23Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC18 reports
- Midas Rex® AA10Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC14 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page