Skip to content

V 40 Femoral Head, Sterile, 22 mm recallZ-1822-2009 · Stryker Howmedica Osteonics Corp. · initiated Jun 18, 2009

Terminated All recalls posted in August 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sterilization issue: During post-sterilization inspection of the outer blister packaging of the affected lots, it was determined that the outer blister did not meet the required minimum seal width.

FDA root cause category: Process control

What the company is doing

Urgent Product Recall letters were sent via Federal Express on June 18, 2009 to all Stryker Branches/Agencies, hospital Risk Management, hospital Chief of Orthopaedics and surgeons who may have used the recalled products. The letters stated the issue, potential hazards, and the risk mitigation. Customers may contact Colleen O'Meara at 201-831-5970 with any questions.

Recall details

Recall number
Z-1822-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jun 18, 2009
Posted by the FDA
Aug 18, 2009
Recalling firm
Stryker Howmedica Osteonics Corp.
Quantity
13 total, both products.
Product code
LWJ
FDA submissions
K022077, K993601
Distribution
Nationwide Distribution

Product description

V 40 Femoral Head, Sterile, 22 mm; Catalog number 6260-4-122, Catalog number 6260-5-226; Howmedica Osteonics Corp., Stryker Ireland, Carrigtwohill County Cork, Ireland; Intended for use with femoral stems and acetabular components in primary or revision total hip arthroplasty.

Code information (models, lots, serial numbers)

lot # 29213701;Exp. 1/28/2014 and lot # 29228301; Exp 1/28/2014

Other devices under the same FDA submission

17 devices cleared under K022077, K993601 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page