Medtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive… recallZ-1440-2010 · Medtronic Inc. Cardiac Rhythm Disease Managment · initiated Feb 11, 2010
Terminated All recalls posted in April 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc. Cardiac Rhythm Disease Managment or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic is informing doctors of two specific battery issues with EnRhythm pacemakers that will be addressed by a Medtronic software update available mid-2010. If the software update referenced above is not implemented, there will be a potential risk of loss of device functionality or risk for loss of therapy at or near ERI in a small number of devices. The software will eliminate this issue, fo
FDA root cause category: Nonconforming Material/Component
What the company is doing
Consignees were given a Medtronic "Important : Medical Device Correction" letter addressed to "Dear Doctor" and dated February 2010. In the letter, Medtronic is informing doctors of two specific battery issues with EnRhythm pacemakers that will be addressed by a Medtronic software update available mid-2010. The letter included recommendations including that Medtronic recommends physicians continue to use the ERI notification to determine time for device replacement. At this time, no other action, reprogramming or change in the frequency of follow-up is recommended. For additional information or assistance, please contact your local Medtronic Representative or Medtronic Technical Services at 800-505-4636.
Recall details
- Recall number
- Z-1440-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 11, 2010
- Posted by the FDA
- Apr 26, 2010
- Recalling firm
- Medtronic Inc. Cardiac Rhythm Disease Managment
- Quantity
- approximately 120,061 worldwide
- Product code
- DXY
- FDA submissions
- P980035
- Distribution
- All States in the USA including Guam, Puerto Rico and DC. OUS: Customers in the following countries: Andorra, Argentina, Aruba, Australia, Austria, Bangladesh, Belarus, Belgium, Brazil, Brunei, Darussalam, Canada, Cayman Islands, Chile, China, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guam, Hong Kong, Hungary, India, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Vatican City, Venezuela, Viet Nam, Virgin Islands
Product description
Medtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects, and treats atrial tachyarrhythmia episodes. It also provides bradycardia pacing and monitoring of ventricular tachycardia (VT) episodes. Manufactured at: Juncos, Puerto Rico, USA. Medtronic, Inc., Minneapolis, MN 55432 USA.
Code information (models, lots, serial numbers)
All EnRhythm pacemakers
Other devices under the same FDA submission
71 devices cleared under P980035 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)187 deaths7,655 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)107 deaths5,971 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Pulse generator, permanent, implantable51 deaths2,298 reports
- Adapta™ ADDR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)39 deaths1,801 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)41 deaths1,758 reports
- Adapta™ L ADDRL1Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)20 deaths1,072 reports
- Azure™ S DR MRI SureScan™ W3DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)12 deaths715 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)13 deaths654 reports
- Azure™ XT SR MRI SureScan™ W1SR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)22 deaths469 reports
- Azure™ S DR MRI SureScan™ W3DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)2 deaths443 reports
- Sensia™ SEDR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)10 deaths434 reports
- Azure™ XT SR MRI SureScan™ W1SR01Medtronic, Inc. · Pulse generator, permanent, implantable11 deaths425 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page