Datascope/ Maquet IABPs. System 98/98XT Intra-Aortic… recallZ-2414-2011 · Maquet Datascope Corp - Cardiac Assist Division · initiated Mar 16, 2011
Terminated All recalls posted in June 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Maquet Datascope Corp - Cardiac Assist Division or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Several models of the Intra-Aortic Balloon Pumps are being recalled due to a potential problem with the fan's assembly which may cause the power supply to detect an overheating event and shut down.
FDA root cause category: Device Design
What the company is doing
Maquet sent an Urgent Product Field Correction letter dated March 16, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to note the WARNING in the IABP Operating Instruction. Customers were advised to refer to the IAB Instructions for Use, Section D (Manually inflating and Deflating a Catheter). Section D of the IAB instruction for Use reiterates that an IAB should not remain inactive for more than 30 minutes and also provides detailed instruction on how to keep an IAB active during IABP failure. Customers were advised that their Service Representatiave would arrange to replace their supply which would contain a new fan assembly. Upon completion of the power supply replacement, customers were asked to sign a service repair order to verify satisfactory completion of the work. For technical questions contact Technical Support Department at 1-800-777-4222 and press 4. For scheduling a service visit by a Field Service Representatiave, contact the firm's US Call Center at 1-800-777-4222 and press 3. Maquet issued a press release on June 6, 2011 advising consumers of the issue and that their recall was considered a class one.
Recall details
- Recall number
- Z-2414-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 16, 2011
- Posted by the FDA
- Jun 13, 2011
- Recalling firm
- Maquet Datascope Corp - Cardiac Assist Division
- Quantity
- 885 units
- Product code
- FMZ
- FDA submissions
- K031636, K063525, K965236
- Distribution
- Worldwide Distribution
Product description
Datascope/ Maquet IABPs. System 98/98XT Intra-Aortic Balloon Pump; CS100/CS100i Intra-Aortic Balloon Pump; CS300 Intra-Aortic Balloon Pump; Manufactured by Datascope Corp., 1300 MacArthur Blvd., Mahwah, NJ 07430 The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloon. It provides temporary support to the left ventricle via the principle of counterpulsion. The IABP is placed in the descending aorta, via the principe of counterpulsation. The IABP is placed in the descending aorta, just distal to the left subclavian artery. Once the balloon is positioned, the IABP is adjusted to trigger in synchrony with the ECG or arterial pressure waveform to ensure that inflation and deflation occur at the appropriate points during the cardiac cycle. The CS300 has the additional capability to automatically calibrate a fiber-optic pressure sensor placed in the tip of the intra-aortic balloon. The IABP will recalibrate every 2 hours or sooner should the patient or environmental conditions change. The target populations are adult and pediatric. The IABP is intended for use in the health care facility setting.
Code information (models, lots, serial numbers)
System 98 - Part Numbers 0998-00-0446-xx, 0998-UC-0446-xx; System 98XT - Part numbers 0998-00-0479-xx, 0998-UC-0479-xx ; CS100i- Part numbers 0998-UC-0446Hxx, 0998-UC-0479Hxx; CS100 - Part numbers 0998-00-3013-xx, 0998-UC-3013-xx; CS300 - Part numbers 0998-00-3023-xx, 0998-UC-3023-xx
Other devices under the same FDA submission
21 devices cleared under K031636, K063525, K965236 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CS100 Intra-Aortic Balloon Pump, English, 220V 0998-00-3013-55Datascope Corp. · System, Balloon, Intra-Aortic And Control4 deaths933 reports
- CS100 Intra-Aortic Balloon Pump, Chinese, 220V 0998-00-3013-45Datascope Corp. · System, Balloon, Intra-Aortic And Control2 deaths684 reports
- CS100 Intra-Aortic Balloon Pump, Refurbished, ENGLISH,NON-UTS,INTL 0998-UC-3013-55Datascope Corp. · System, Balloon, Intra-Aortic And Control1 death129 reports
- CS100 Intra-Aortic Balloon Pump, Japanese, 110V 0998-00-3013-65Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall1 death60 reports
- CS100 Intra-Aortic Balloon Pump, Spanish, 220V 0998-00-3013-58Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall32 reports
- CS100 Intra-Aortic Balloon Pump, Portuguese, 110V 0998-00-3013-69Datascope Corp. · System, Balloon, Intra-Aortic And Control66 reports
- CS100 Intra-Aortic Balloon Pump, Italian, 220V 0998-00-3013-67Datascope Corp. · System, Balloon, Intra-Aortic And Control65 reports
- CS100 Accessory Kit, Spanish, 220 V 0020-00-0463-05Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall0 reports
- CS100 Intra-Aortic Balloon Pump, Portuguese, 220V 0998-00-3013-68Datascope Corp. · System, Balloon, Intra-Aortic And Control49 reports
- CS100 Intra-Aortic Balloon Pump, French, 220V 0998-00-3013-64Datascope Corp. · System, Balloon, Intra-Aortic And Control16 reports
- CS100 Intra-Aortic Balloon Pump, Refurbished, CHINESE,NON-UTS,INTL 0998-UC-3013-45Datascope Corp. · System, Balloon, Intra-Aortic And Control13 reports
- CS100 Intra-Aortic Balloon Pump, Refurbished, JAPANESE,NON-UTS,DOM 0998-UC-3013-65Datascope Corp. · System, Balloon, Intra-Aortic And Control11 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page