Datascope System 98/98xt, CS 100/CS100i and CS 300… recallZ-1523-2014 · Maquet Datascope Corp - Cardiac Assist Division · initiated Mar 21, 2014
Class I Terminated All recalls posted in May 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Maquet Datascope Corp - Cardiac Assist Division or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and cause the IABP to shut down without warning.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Maquet Inc. sent an Urgent Medical Device Field Correction letter dated March 21, 2014, to all affected customers. The letter identified the affected product, problem, and actions to be taken.. Customers were instructed that their Service Representatiave would contact them to schedule the replacement of the fan assembly at no cost to the customer. Upon completion customers were requested to sign a service repair order to verify satisfactory completion of the work. Customers with questions within Puerto Rico, were instructed to contact Technical Support in Mexico at +52-55-9000-8970, ext 104. For questions regarding this recall call 973-709-7652. A field correction letter was sent out on April 15, 2014 to additional consignees that did not receive the March 15, 2014 letter. Maquet issued press on May 9, 2014.
Recall details
- Recall number
- Z-1523-2014
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Mar 21, 2014
- Posted by the FDA
- May 21, 2014
- Recalling firm
- Maquet Datascope Corp - Cardiac Assist Division
- Quantity
- 11,881 units (US 5183 and OUS 6698)
- Product code
- DSP
- FDA submissions
- K031636, K063525, K965236
- Distribution
- Worldwide Distribution - USA (nationwide) and approximately 104 countries.
Product description
Datascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump. The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloons in cardiovascular procedures.
Code information (models, lots, serial numbers)
IABP Model Numbers System 98 0998-00-0446-xx 0998-UC-0446-xx System 98XT 0998-00-0479-xx 0998-UC-0479-xx CS100i 0998-UC-0446Hxx 0998-UC-0479Hxx CS100 0998-00-3013-xx 0998-UC-3013-xx CS300 0998-00-3023-xx 0998-UC-3023-xx
Other devices under the same FDA submission
21 devices cleared under K031636, K063525, K965236 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CS100 Intra-Aortic Balloon Pump, English, 220V 0998-00-3013-55Datascope Corp. · System, Balloon, Intra-Aortic And Control4 deaths933 reports
- CS100 Intra-Aortic Balloon Pump, Chinese, 220V 0998-00-3013-45Datascope Corp. · System, Balloon, Intra-Aortic And Control2 deaths684 reports
- CS100 Intra-Aortic Balloon Pump, Refurbished, ENGLISH,NON-UTS,INTL 0998-UC-3013-55Datascope Corp. · System, Balloon, Intra-Aortic And Control1 death129 reports
- CS100 Intra-Aortic Balloon Pump, Japanese, 110V 0998-00-3013-65Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall1 death60 reports
- CS100 Intra-Aortic Balloon Pump, Spanish, 220V 0998-00-3013-58Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall32 reports
- CS100 Intra-Aortic Balloon Pump, Portuguese, 110V 0998-00-3013-69Datascope Corp. · System, Balloon, Intra-Aortic And Control66 reports
- CS100 Intra-Aortic Balloon Pump, Italian, 220V 0998-00-3013-67Datascope Corp. · System, Balloon, Intra-Aortic And Control65 reports
- CS100 Accessory Kit, Spanish, 220 V 0020-00-0463-05Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall0 reports
- CS100 Intra-Aortic Balloon Pump, Portuguese, 220V 0998-00-3013-68Datascope Corp. · System, Balloon, Intra-Aortic And Control49 reports
- CS100 Intra-Aortic Balloon Pump, French, 220V 0998-00-3013-64Datascope Corp. · System, Balloon, Intra-Aortic And Control16 reports
- CS100 Intra-Aortic Balloon Pump, Refurbished, CHINESE,NON-UTS,INTL 0998-UC-3013-45Datascope Corp. · System, Balloon, Intra-Aortic And Control13 reports
- CS100 Intra-Aortic Balloon Pump, Refurbished, JAPANESE,NON-UTS,DOM 0998-UC-3013-65Datascope Corp. · System, Balloon, Intra-Aortic And Control11 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page