Skip to content

Datascope System 98/98xt, CS 100/CS100i and CS 300… recallZ-1523-2014 · Maquet Datascope Corp - Cardiac Assist Division · initiated Mar 21, 2014

Class I Terminated All recalls posted in May 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Datascope Corp - Cardiac Assist Division or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and cause the IABP to shut down without warning.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Maquet Inc. sent an Urgent Medical Device Field Correction letter dated March 21, 2014, to all affected customers. The letter identified the affected product, problem, and actions to be taken.. Customers were instructed that their Service Representatiave would contact them to schedule the replacement of the fan assembly at no cost to the customer. Upon completion customers were requested to sign a service repair order to verify satisfactory completion of the work. Customers with questions within Puerto Rico, were instructed to contact Technical Support in Mexico at +52-55-9000-8970, ext 104. For questions regarding this recall call 973-709-7652. A field correction letter was sent out on April 15, 2014 to additional consignees that did not receive the March 15, 2014 letter. Maquet issued press on May 9, 2014.

Recall details

Recall number
Z-1523-2014
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Mar 21, 2014
Posted by the FDA
May 21, 2014
Recalling firm
Maquet Datascope Corp - Cardiac Assist Division
Quantity
11,881 units (US 5183 and OUS 6698)
Product code
DSP
FDA submissions
K031636, K063525, K965236
Distribution
Worldwide Distribution - USA (nationwide) and approximately 104 countries.

Product description

Datascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump. The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloons in cardiovascular procedures.

Code information (models, lots, serial numbers)

IABP Model Numbers System 98 0998-00-0446-xx 0998-UC-0446-xx System 98XT 0998-00-0479-xx 0998-UC-0479-xx CS100i 0998-UC-0446Hxx 0998-UC-0479Hxx CS100 0998-00-3013-xx 0998-UC-3013-xx CS300 0998-00-3023-xx 0998-UC-3023-xx

Other devices under the same FDA submission

21 devices cleared under K031636, K063525, K965236 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page