AGENT™ H7493960812300Recalls & reported problems

Boston Scientific Corporation · Paclitaxel-Coated Balloon Catheter · Class 3

All models of AGENT (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00191506014681), received between Apr 21, 2025 and Aug 19, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Adverse Event Without Identified Device or Use Problem2
  • Deformation Due to Compressive Stress1
  • Entrapment of Device1
  • Failure to Advance1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 19, 2026Break; Failure to Advance

    It was reported that a shaft break within a patient occurred. A 3.00 mm x 12mm AGENT US MR was selected for a Percutaneous Coronary Intervention to evaluate the target lesion. During the procedure, the device could not advance in the lesion and the shaft subsequently broke. The procedure was completed as planned and the device was successfully removed in one piece from the patient. No patient complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50334624 ↗
  • MalfunctionMay 4, 2026Break; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT THE BALLOON KINKED AND THE SHAFT BROKE. A 3.00 MM X 12.00 MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR TREATMENT. DURING ADVANCEMENT OF THE DEVICE INTO THE PATIENT, THE SHAFT KINKED AND SUBSEQUENTLY BROKE. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING AN ALTERNATIVE DEVICE. THERE WERE NO PATIENT COMPLICATIONS AND THE PATIENT FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25070664 ↗
  • MalfunctionJul 2, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 3.00 X 12MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR PERCUTANEOUS CORONARY INTERVENTION PROCEDURE. DURING PROCEDURE, BALLOON SHAFT BROKE. DEVICE WAS REMOVED OUT OF PATIENT AND NOTHING WAS LEFT IN THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ALTERNATE DEVICE AND THERE WAS NO PATIENT COMPLICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22383075 ↗
  • MalfunctionApr 21, 2025Break; Entrapment of Device; Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Break; Entrapment of Device; Use of Device Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT CATHETER FROZE ON WIRE OCCURRED. A 3.00 X 12MM AGENT DCB BALLOON WAS SELECTED FOR PERCUTANEOUS CORONARY INTERVENTION PROCEDURE TO TREAT CORONARY ARTERY DISEASE AND IN-STENT RESTENOSIS. DURING THE PROCEDURE, WHEN THE PHYSICIAN ADVANCED THE BALLOON, IT WAS NOT NOTED THAT THE BALLOON PROTECTOR WAS STILL ON THE DEVICE. THE COVER WOULD NOT ADVANCE THROUGH THE TOUHY. THE PHYSICIAN TRIED TO PEEL AWAY THE COVER WHILE THE DEVICE WAS ON THE GUIDEWIRE BUT WAS NOT SUCCESSFUL. THE COVER HAD BEEN PUSHED TO THE MONORAIL EXIT PORT OF THE BALLOON AND THE BALLOON WAS STUCK ON THE GUIDEWIRE. DURING ATTEMPTS TO REMOVE THE BALLOON FROM THE GUIDEWIRE OUTSIDE THE PATIENT, THE BALLOON MONORAIL/HYPOTUBE BROKE AWAY FROM THE SHAFT AND THE GUIDEWIRE ALSO SNAPPED. THE PROCEDURE WAS COMPLETED USING A NEW GUIDEWIRE AND A NEW AGENT BALLOON. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876972 ↗

Other Drug-eluting percutaneous transluminal coronary angioplasty catheter devices