Boston Scientific Corporation · Paclitaxel-Coated Balloon Catheter · Class 3
All models of AGENT (23)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00191506014827), received between Mar 24, 2025 and Aug 12, 2026 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Failure to Advance2
- Nonstandard Device2
- Adverse Event Without Identified Device or Use Problem1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Vascular Dissection1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryAug 12, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Vascular Dissection
MAUDE report 50229874 ↗MalfunctionJun 8, 2026Break; Nonstandard Device
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 3.00 X 20MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR PERCUTANEOUS CORONARY INTERVENTION (PCI). DURING PROCEDURE, WHILE TRYING TO DELIVER BALLOON, THE SHAFT BROKE. PHYSICIAN WAS ABLE TO RETRIEVE BROKEN SHAFT FROM GUIDE CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25429747 ↗MalfunctionApr 17, 2026Nonstandard Device; Failure to Advance
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 26MARCH2026. IT WAS REPORTED THAT THE DEVICE WAS UNABLE TO CROSS. THE PATIENT PRESENTED WITH IN-STENT RESTENOSIS (ISR) FOR A PERCUTANEOUS CORONARY INTERVENTION (PCI). A 3.00 MM X 20.00 MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR USE. FOLLOWING PRE-TREATMENT WITH BALLOON ANGIOPLASTY AND CONFIRMED WITH INTRAVASCULAR ULTRASOUND, THE AGENT DCB WAS INSERTED BUT WAS UNABLE TO CROSS THE TARGET LESION. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS, AND THE PATIENT FULLY RECOVERED. HOWEVER, THE RETURN DEVICE ANALYSIS REVEALED A TEAR WAS NOTED TO THE OUTER SHAFT POLYMER EXTRUSION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24913672 ↗MalfunctionMar 20, 2026Break
IT WAS REPORTED THAT THE SHAFT BROKE. AN AGENT US MR 3.00 X 20MM WAS SELECTED FOR TREATMENT OF THE TARGET LESION. DURING THE PROCEDURE WHILE INSIDE THE GUIDE CATHETER, THE AGENT BALLOON SHAFT BROKE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24650604 ↗MalfunctionDec 22, 2025Break; Break
IT WAS REPORTED THAT A SHAFT BREAK OCCURRED. A 3.00 MM X 20.00 MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR TREATMENT. DURING THE PROCEDURE, WHILE INSERTING THE BALLOON THE SHAFT BROKE. THE DEVICE WAS REMOVED FROM PATIENT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WAS NO PATIENT COMPLICATION, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23865385 ↗MalfunctionAug 4, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT THE BALLOON RUPTURED. THE PATIENT PRESENTED WITH IN-STENT RESTENOSIS (ISR) FOR A PERCUTANEOUS CORONARY INTERVENTION (PCI). A 3.00 MM X 20.00 MM AGENT DRUG-COATED BALLOON (DCB) WAS USED AT THE RIGHT CORONARY ARTERY (RCA). VESSEL PREPARATION WAS INITIATED WITH PLAIN OLD BALLOON ANGIOPLASTY (POBA) USING AN EMERGE BALLOON, FOLLOWED BY INTRAVASCULAR LITHOTRIPSY (IVL). THE AGENT DEVICE WAS FOUND TO BE UNDAMAGED DURING UNPACKING AND PREPARATION FOR USE, AND IT ADVANCED SMOOTHLY OVER THE WIRE AND THROUGH THE GUIDEWIRE WITHOUT ANY DIFFICULTY. THE AGENT DCB WAS ADVANCED; HOWEVER, THE BALLOON RUPTURED AT 6 ATM. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER AGENT DEVICE. THERE WAS NO PATIENT COMPLICATION REPORTED, AND THE PATIENT WAS DOING WELL POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22701416 ↗MalfunctionJun 9, 2025Break; Break
IT WAS REPORTED THAT SHAFT BROKE. A 3.00 X 20MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR USE. DURING PROCEDURE, SHAFT BROKE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22169737 ↗MalfunctionMar 24, 2025Break; Failure to Advance; Break; Failure to Advance
IT WAS REPORTED THAT THE SHAFT SEPARATION OCCURRED. THE 60% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED PROXIMAL LEFT ANTERIOR DESCENDING (LAD). A 3.00 X 20MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR LEFT HEART CATHETERIZATION (LHC). THE TARGET LESION PROPERLY PREPARED WITH NON-BOSTON SCIENTIFIC INTRAVASCULAR LITHOTRIPSY (IVL) AND INTRAVASCULAR ULTRASOUND (IVUS). WHEN THE CATHETER WAS ADVANCED AND POSITIONED PROXIMAL TO THE LESION, THE CATHETER UNABLE TO CROSS THE LESION, AND THE SHAFT BROKE. THE PHYSICIAN WAS REMOVED THE DEVICE IN ONE PIECE WITHOUT ANY PROBLEMS. THE PROCEDURE COMPLETED WITH ANOTHER AGENT DCB AND THERE WAS NO PATIENT COMPLICATION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21677962 ↗
Other Drug-eluting percutaneous transluminal coronary angioplasty catheter devices
- AGENT™ H7493960830250Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter10 reports
- AGENT™ H7493960830300Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter8 reports
- AGENT™ H7493960830400Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter6 reports
- AGENT™ H7493960820250Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter6 reports
- AGENT™ H7493960830350Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter5 reports
- AGENT™ H7493960815300Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter5 reports
- AGENT™ H7493960820400Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter5 reports
- AGENT™ H7493960820350Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page