AGENT™ H7493960820250Recalls & reported problems
Boston Scientific Corporation · Paclitaxel-Coated Balloon Catheter · Class 3
All models of AGENT (23)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00191506014803), received between Jun 20, 2025 and Sep 1, 2026 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Advance3
- Break2
- Deformation Due to Compressive Stress2
- Adverse Event Without Identified Device or Use Problem1
- Contamination1
- Difficult to Advance1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Cardiac Enzyme Elevation1
- Ischemia1
- Low Blood Pressure/ Hypotension1
- Myocardial Infarction1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjurySep 1, 2026Adverse Event Without Identified Device or Use Problem
It was reported that elevated cardiac enzymes and myocardial infarction occurred. The patient was on a prior regimen of antiplatelet other than Aspirin (>= 72 hours). At the time of the procedure, Heparin or other antithrombotic medication and a loading dose of 325 mg of Aspirin and 300 mg of Clopidogrel were given. The 3.0 mm x 33 mm target lesion located at the distal circumflex (CX) vessel in the main branch was pretreated with 3.0 mm x 15 mm non-compliant balloon angioplasty catheter resulting in 0% residual stenosis and TIMI flow 3. The lesion was then successfully treated with 3.0 mm x 33 mm drug eluting stent (non-Boston Scientific device) achieving 0% residual stenosis with TIMI flow 3. The other 2.5 mm x 15 mm target lesion located at the 2nd obtuse marginal vessel in the side branch was pretreated with a 2.5 mm x 15 mm non-compliant balloon angioplasty catheter resulting in 0% residual stenosis and TIMI flow 3. The lesion was then successfully treated with the 2.5 mm x 20 mm AGENT DCB achieving 0% residual stenosis with TIMI flow 3. One day post procedure, the 18-24-hour post procedure lab draw demonstrated elevated levels of Troponin T HS and Troponin-I HS which met the
Patient problems: Cardiac Enzyme Elevation; Myocardial Infarction
MAUDE report 50550330 ↗MalfunctionMar 31, 2026Material Rupture; Failure to Advance; Difficult to Advance
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL, IN THE OBTUSE MARGINAL (OM) ARTERY. A 2.50 MM X 20.00 MM AGENT DRUG-COATED BALLOON (DCB) WAS ADVANCED WITH RESISTANCE FOR TREATMENT. UPON INITIAL INFLATION OF THE BALLOON AT 6 ATM FOR LESS THAN 20 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24746687 ↗MalfunctionFeb 27, 2026Break; Failure to Advance; Deformation Due to Compressive Stress
IT WAS REPORTED THAT A SHAFT BREAK OCCURRED. THE PATIENT PRESENTED WITH CORONARY ARTERY DISEASE (CAD). A 2.50 MM X 20.00 MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR PERCUTANEOUS CORONARY INTERVENTION (PCI). DURING DELIVERY, THE AGENT BALLOON WAS UNABLE TO CROSS THE LESION, AND THE HYPOTUBE BENT AND SUBSEQUENTLY FRACTURED. THE DEVICE WAS REMOVED WITHOUT DIFFICULTY. A SECOND 3.00 MM X 30 MM AGENT DCB WAS THEN ADVANCED TO A MORE PROXIMAL SEGMENT, BUT THIS DEVICE ALSO FAILED TO CROSS THE LESION AND THE HYPOTUBE BECOME SIGNIFICANTLY BENT. IT WAS ALSO REMOVED WITHOUT INCIDENT. NO PATIENT COMPLICATIONS, AND THE PATIENT FULLY RECOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24474527 ↗MalfunctionDec 16, 2025Break; Deformation Due to Compressive Stress; Break; Deformation Due to Compressive Stress
IT WAS REPORTED THAT THE DEVICE BROKE OFF TOWARDS THE HUB. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS, MODERATELY CALCIFIED DISTAL RIGHT CORONARY ARTERY (RCA). DURING THE PROCEDURE, A 2.50 MM X 20.00 MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR TREATMENT. DURING THE PROCEDURE, WHILE ADVANCING THE BALLOON AND TRYING TO CROSS THE LESION, THE SHAFT BROKE OFF AND WAS KINKED TOWARDS THE HUB. THE DEVICE WAS REMOVED FROM THE PATIENT WITHOUT COMPLICATION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DIFFERENT DEVICE. THERE WAS NO PATIENT COMPLICATION, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23821677 ↗MalfunctionSep 4, 2025Contamination; Contamination
IT WAS REPORTED THAT DEVICE CONTAMINATION OCCURRED. THE PATIENT PRESENTED WITH IN-STENT RESTENOSIS. A 2.50 X 20 MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR USE. IT WAS NOTED THAT THE DEVICE APPEARED DIRTY UPON OPENING THE BOX. THE PROCEDURE SUCCESSFULLY COMPLETED WITH ANOTHER AGENT DCB. THERE WAS NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22965121 ↗InjuryJun 20, 2025Failure to Advance; Failure to Advance
IT WAS REPORTED THAT ISCHEMIA AND HYPOTENSION OCCURRED. A 2.50 X 20 MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR USE. AGGRESSIVE PRE-DILATATION WAS PERFORMED FOR IN-STENT RESTENOSIS (ISR) AT THE OSTIAL CIRCUMFLEX, LEFT ANTERIOR DESCENDING ARTERY (LAD), AND LEFT MAIN BIFURCATION USING NC EMERGE AND NON-BOSTON SCIENTIFIC BALLOONS. THE DEVICE WAS PREPARED ACCORDING TO STANDARD PROTOCOL. DURING ATTEMPTED DELIVERY TO THE OSTIAL CIRCUMFLEX, THE AGENT BALLOON ENCOUNTERED RESISTANCE AND FAILED TO CROSS EASILY. MULTIPLE ATTEMPTS WITH DEEP GUIDE ENGAGEMENT LED TO ISCHEMIA. THE BALLOON WAS EVENTUALLY SUCCESSFULLY ADVANCED AND INFLATED; HOWEVER, AFTER 30 SECONDS OF INFLATION, THE PATIENT DEVELOPED SEVERE HYPOTENSION, REQUIRING ADMINISTRATION OF NEO-SYNEPHRINE (NEO), DEFLATION OF THE AGENT BALLOON, AND DISENGAGEMENT OF THE GUIDE CATHETER. FURTHER NEO-SYNEPHRINE WAS ADMINISTERED, AND THE PATIENT WAS INSTRUCTED TO COUGH TO MITIGATE THE RISK OF CARDIAC ARREST. AFTER 20 SECONDS, THE HYPOTENSION RESOLVED, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT FURTHER COMPLICATIONS. THE PATIENT FULLY RECOVERED.
Patient problems: Ischemia; Low Blood Pressure/ Hypotension
MAUDE report 22299034 ↗
Other Drug-eluting percutaneous transluminal coronary angioplasty catheter devices
- AGENT™ H7493960830250Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter10 reports
- AGENT™ H7493960830300Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter8 reports
- AGENT™ H7493960820300Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter8 reports
- AGENT™ H7493960830400Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter6 reports
- AGENT™ H7493960830350Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter5 reports
- AGENT™ H7493960820400Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter5 reports
- AGENT™ H7493960815300Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter5 reports
- AGENT™ H7493960820350Boston Scientific Corporation · Drug-eluting percutaneous transluminal coronary angioplasty catheter4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page