AGENT™ H7493960830400Recalls & reported problems

Boston Scientific Corporation · Paclitaxel-Coated Balloon Catheter · Class 3

All models of AGENT (23)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00191506014919), received between Jan 8, 2025 and Jul 9, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture4
  • Nonstandard Device2
  • Break1
  • Detachment of Device or Device Component1
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 9, 2026Nonstandard Device; Material Rupture

    It was reported that a balloon rupture occurred.A 4.00 mm x 30.00 mm Agent drug-coated balloon (DCB) was selected for an in-stent restenosis (ISR) treatment. The balloon ruptured immediately upon launch without inflation. The guide catheter was also positioned within the vessel at the time of the event. The device was removed from patient; the procedure was successfully completed with an alternative device. There was no patient complication reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25774276 ↗
  • MalfunctionMay 15, 2026Nonstandard Device; Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE TARGET LESION LOCATED WAS LOCATED IN THE MID RIGHT CORONARY ARTERY (RCA). A 4.00 MM X 30 MM AGENT DCB WAS SELECTED FOR USE AND ADVANCED TO THE LESION SITE. DURING THE FIRST INFLATION AT 12 ATM, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED FROM THE PATIENT, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING TWO SEPARATE 4.00 MM X 15 MM AGENT DCB BALLOONS. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25186906 ↗
  • MalfunctionDec 30, 2025Break; Break

    IT WAS REPORTED THAT A SHAFT BREAK OCCURRED. A 4.00 X 30MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR TREATMENT. DURING THE PROCEDURE, THE SHAFT BROKE. ALTERNATIVE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE DEVICE WAS REMOVED FROM THE PATIENT; THERE WAS NO FURTHER PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23926596 ↗
  • InjuryOct 10, 2025Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE BALLON DETACHED. THE PATIENT PRESENTED WITH IN-STENT RESTENOSIS (ISR). THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY (LAD). A 4.00 MM X 30.00 MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR PERCUTANEOUS CORONARY INTERVENTION (PCI). NO DAMAGE WAS OBSERVED DURING DEVICE UNPACKING OR PREPARATION, AND THE DEVICE ADVANCED SMOOTHLY OVER THE WIRE AND THROUGH THE GUIDE CATHETER WITHOUT DIFFICULTY. THE BALLOON WAS INFLATED ONCE AT NORMAL PRESSURE FOR 60 SECONDS. DURING DEVICE WITHDRAWAL, THE BALLOON DETACHED FROM THE SHAFT AND BECAME STUCK WITHIN A NON-BOSTON SCIENTIFIC GUIDING CATHETER. THE DISLODGED BALLOON WAS SUCCESSFULLY REMOVED USING HEMOSTATS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THIS DEVICE. THERE WAS NO PATIENT COMPLICATION REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23267291 ↗
  • MalfunctionMar 26, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE PROXIMAL RIGHT CORONARY ARTERY. THE 4.00 X 30MM AGENT DCB WAS INTRODUCED TO TREAT THE TARGET LESION BUT THE BALLOON MATERIAL RUPTURED. THE PROCEDURE WAS COMPLETED WITH A BALLOON ANGIOPLASTY CATHETER RESULTING IN 25% RESIDUAL STENOSIS AND TIMI 3 FLOW. IT WAS ADDITIONALLY REPORTED THAT THE TARGET LESION WAS 50% STENOSED, AND THE AGENT DCB WAS INFLATED TO 14 ATM FOR 25 TO 30 SECONDS WHEN IT RUPTURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21700336 ↗
  • MalfunctionJan 8, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE BALLOON BURST. AN AGENT US MR 4.00 X 30MM WAS SELECTED FOR TREATMENT OF THE TARGET LESION, WHICH WAS LOCATED IN THE RIGHT CORONARY ARTERY (RCA). DURING THE PROCEDURE, THE PHYSICIAN INFLATED THE BALLOON TO 11ATM WHEN A LOSS OF PRESSURE WAS NOTED IN THE INFLATION DEVICE. THE DEVICE WAS PULLED, AND THE BALLOON WAS NOTED TO HAVE RUPTURED. THE BALLOON WAS REMOVED FROM THE PATIENT AND ANOTHER SIMILAR DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21105117 ↗

Other Drug-eluting percutaneous transluminal coronary angioplasty catheter devices

All Drug-eluting percutaneous transluminal coronary angioplasty catheter devices →