Skip to content

AGENT™ H7493960830300Recalls & reported problems

Boston Scientific Corporation · Paclitaxel-Coated Balloon Catheter · Class 3

All models of AGENT (23)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00191506014896), received between Aug 28, 2025 and Sep 30, 2026 — classified as 3 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Adverse Event Without Identified Device or Use Problem2
  • Device-Device Incompatibility1
  • Difficult to Advance1
  • Difficult to Remove1
  • Failure to Advance1
  • Inflation Problem1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Vascular Dissection2
  • Chest Pain1
  • Dyspnea1
  • Myocardial Infarction1
  • Restenosis1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that a grade C dissection occurred. The 2.75 mm x 20 mm target lesion located at the 2nd diagonal branch was pretreated with two devices with the largest balloon size of 3 mm x 15 mm. The non-complaint balloon had a maximum inflation pressure of 8 and two balloon inflations. Final pretreatment stenosis was 40% with TIMI flow 3. IVUS with Opticross 6 catheter was used to confirmed pretreatment results. The target lesion was then successfully treated with a 3.0 mm x 30 mm Agent DCB device. The device achieved an inflation pressure of 6 with an inflation time of 80. However, a grade C dissection was discovered, and an interventional procedure was completed to place a non-Boston Scientific drug eluting stent. The patient was discharged.

    Patient problems: Vascular Dissection

    MAUDE report 50944990 ↗
  • InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that a NSTEMI requiring intervention occurred. In early (b)(6) 2026, the 2.75 mm x 25 mm target lesion located at the right posterior descending vessel was pretreated with a semi-compliant and non-compliant balloon with the largest balloon being 2.5 mm x 25 mm. Maximum inflation pressure was 16 with a total of 2 balloon inflations. Pretreatment resulted in 20% stenosis and TIMI flow 3. The target lesion was then successfully treated with a 3.0 mm x 30 mm AGENT DCB balloon, maximum inflation pressure was 8 with a total of 1 balloon inflation. Prior to the target lesion treatment, the non-target lesion located at the mid left anterior descending artery was also successfully treated with a 3.5 mm x 16 mm drug eluting stent and 3.75 mm x 12 mm balloon angioplasty catheter. Patient discharged on the same day of the procedure.Shortly after the procedure in (b)(6) 2026, the patient experienced dyspnea and had a NSTEMI that required an in-patient hospitalization, stent placement, target vessel revascularization and medication administration (Heparin). Patient discharged.It was later reported: That shortly after the procedure in (b)(6) 2026, the patient also experienced ches

    Patient problems: Chest Pain; Dyspnea; Myocardial Infarction; Restenosis; Vascular Dissection

    MAUDE report 50619349 ↗
  • InjuryJul 8, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THE BALLOON FAILED TO EXIT THE GUIDE CATHETER REQUIRING ADDITIONAL INTERVENTION. AN AGENT (DCB) WAS SELECTED FOR A PERCUTANEOUS CORONARY INTERVENTION IN THE EXAMINATION OF THE RIGHT CORONARY ARTERY. DURING THE PROCEDURE THE LESION WAS PRETREATED WITH BALLOONS AND THEN A SYNERGY XD STENT WAS IMPLANTED. THE AGENT DCB WAS ATTEMPTED BUT WOULD BARLEY COME OUT OF THE GUIDE CATHETER AND A GUIDEZILLA WAS NEEDED TO REMOVE IT. NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25749333 ↗
  • MalfunctionApr 23, 2026Break

    IT WAS REPORTED THAT A SHAFT BREAK OCCURRED. A 3.00 MM X 30 MM AGENT DCB WAS SELECTED FOR USE. HOWEVER, THE BALLOON SHAFT BROKE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING ANOTHER SAME DEVICE AND THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24968979 ↗
  • MalfunctionApr 6, 2026Break; Nonstandard Device; Failure to Advance; Difficult to Advance

    IT WAS REPORTED THAT A SHAFT BREAK OCCURRED. THE IN-STENT RESTENOSED TARGET LESION WAS LOCATED IN THE LEFT CIRCUMFLEX ARTERY. FOLLOWING LESION PREPARATION, THE 3.00 MM X 30.00 MM AGENT DRUG-COATED BALLOON (DCB) WAS INTRODUCED. DURING THE DELIVERY PROCESS, RESISTANCE WAS ENCOUNTERED PROXIMAL TO THE LESION AND THE BALLOON SHAFT BROKE. THE PROCEDURE WAS COMPLETED WITH A SHORTER AGENT DCB. THERE WERE NO PATIENT COMPLICATIONS, AND THE PATIENT WAS EXPECTED TO FULLY RECOVER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24793877 ↗
  • MalfunctionNov 17, 2025Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE DEVICE WAS DIFFICULT TO REMOVE. A 3.00 MM X 30.00 MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR PERCUTANEOUS CORONARY INTERVENTION (PCI). DURING PROCEDURE, AFTER INFLATION, THE BALLON FELT STICKY DURING WITHDRAWAL AND TIGHTNESS WAS EXPERIENCED DURING REMOVAL. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AND THE DEVICE WAS ABLE TO BE REMOVED. THERE WAS NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23569755 ↗
  • MalfunctionOct 16, 2025Break; Break

    IT WAS REPORTED THAT THE SHAFT BREAK OCCURRED. THE PATIENT PRESENTED WITH CORONARY ARTERY DISEASE (CAD). A 3.00 X 30MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR PERCUTANEOUS CORONARY INTERVENTION (PCI). IT WAS NOTED THAT THE DELIVERY SHAFT BROKE AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23316983 ↗
  • MalfunctionAug 28, 2025Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash

    IT WAS REPORTED THAT THE BALLOON HAD A HOLE. THE 60% STENOSED LESION WAS LOCATED IN A MILDLY TORTUOUS LESION. A 3.00 MM X 30.00 MM AGENT DRUG-COATED BALLOON (DCB) WAS SELECTED FOR PERCUTANEOUS CORONARY INTERVENTION (PCI), ALONG WITH VEIN GRAFT ANASTOMOSIS, TO TREAT IN-STENT RESTENOSIS (ISR). NO DAMAGE WAS OBSERVED DURING DEVICE UNPACKING OR PREPARATION, AND THE DEVICE ADVANCED SMOOTHLY OVER THE WIRE AND THROUGH THE GUIDE CATHETER WITHOUT DIFFICULTY. THE BALLOON WAS PREPARED AND ADVANCED OVER THE GUIDEWIRE; HOWEVER, IT FAILED TO MAINTAIN PRESSURE DURING INFLATION DUE TO A HOLE IN THE BALLOON. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER BALLOON AND THERE WAS NO PATIENT COMPLICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22916488 ↗

Other Drug-eluting percutaneous transluminal coronary angioplasty catheter devices

All Drug-eluting percutaneous transluminal coronary angioplasty catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page