Shockwave IVL System with the Shockwave L6 Peripheral IVL Catheter L6IVL090030Recalls & reported problems

Shockwave Medical, Inc. · The IVL Catheter is a proprietary lithotripsy device delivered through the peripheral arterial system of the lower extremities to the site of an otherwise difficult to treat calcified stenosis. Energizing the lithotripsy device will generate pulsatile mechanical energy within the target treatment site, disrupting calcium within the lesion and allowing subsequent dilation of a peripheral artery stenosis using low balloon pressure. The IVL Catheter is comprised of an integrated balloon with an array of lithotripsy emitters for the localized delivery of pulsatile mechanical energy. The system consists of an IVL Catheter, an IVL Connector Cable and an IVL Generator. The IVL Catheter is available in four (4) sizes: 8.0 x 30mm, 9.0 x 30mm, 10.0 x 30mm, and 12.0 x 30mm. The L6 Peripheral IVL Catheter is compatible with a 7 or 8F sheath and has a working length of 110cm. The Shockwave L6 Peripheral IVL Catheter shaft contains an inflation lumen, a guidewire lumen, and the lithotripsy emitters. The inflation lumen is used for inflation and deflation of the balloon with 50/50 saline/contrast medium. The guidewire lumen enables the use of a 0.018” guidewire to facilitate advancement of the catheter to and through the target stenosis. The system is designed as ‘Over-the-wire’ (OTW) with 110cm shaft working length, so an exchange length (300cm) guidewire is indicated. The emitters are positioned along the length of the balloon working length for delivery of pulsatile mechanical energy. The balloon is located near the distal tip of the catheter. Two radiopaque marker bands within the balloon denote the length of the balloon to aid in positioning of the balloon during treatment. The balloon is designed to provide an expandable segment of known length and diameter at a specific pressure. The proximal hub has three ports: one for inflation/deflation of the balloon, one for guidewire lumen, and one for connection to the IVL Connector Cable. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00195451000393), received between Jun 17, 2025 and Mar 9, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Difficult to Advance1
  • Difficult to Remove1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 9, 2026Difficult to Remove; Stretched; Detachment of Device or Device Component; Difficult to Advance

    A SHOCKWAVE L6 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED TO TREAT A LESION IN THE COMMON ILIAC ARTERY AND DISTAL AORTA. IT WAS REPORTED THAT WHILE ATTEMPTING TO ADVANCE THE BALLOON CATHETER OVER THE GUIDEWIRE, THERE WAS SOME RESISTANCE. DURING REMOVAL OF THE DEVICE FROM THE PATIENT, RESISTANCE WAS AGAIN ENCOUNTERED. THE BALLOON APPEARED TO BE FULLY DEFLATED PRIOR TO REMOVAL. HOWEVER, AS IT STRETCHED DURING EXTRACTION, IT LIKELY DREW FLUID BACK IN AND BECAME STUCK ON THE GUIDEWIRE. THE SHEATH WAS WITHDRAWN AND THE GUIDEWIRE WAS CUT TO FACILITATE RETRIEVAL OF THE DEVICE. THE BALLOON SEPARATED AND CAME APART FROM THE CATHETER AS IT WAS BEING REMOVED. THE CATHETER BUT WAS SUCCESSFULLY REMOVED USING SIMPLE TRACTION. THERE WERE NO CATHETER FRAGMENTS RETAINED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT ISSUES, AND NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24552666 ↗
  • MalfunctionJun 17, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    A SHOCKWAVE L6 PERIPHERAL INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER WAS USED TO TREAT A LESION IN THE ILIAC ARTERY. 300 IVL PULSES WERE DELIVERED SUCCESSFULLY. A DETACHMENT OCCURRED DURING IVL CATHETER REMOVAL, RESULTING IN THE CATHETER BREAKING INTO TWO PIECES. THE IVL BALLOON WAS FULLY DEFLATED, AND NO EXCESSIVE FORCE WAS APPLIED DURING REMOVAL. THE DETACHED COMPONENTS WERE RETRIEVED IN THE INTRODUCER SHEATH AND WERE NOT LEFT INSIDE THE PATIENT. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22245366 ↗

Other Percutaneous Catheter, Ultrasound devices