Skip to content

Libre App 78745-01Recalls & reported problems

Abbott Diabetes Care Inc · Libre Android App US · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00357599840006), received between Sep 12, 2025 and Sep 12, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Application Program Problem” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K233537); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Problem1

Patient problems

  • Confusion/ Disorientation1
  • Hypoglycemia1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 12, 2025Application Program Problem; Application Program Problem

    AN INCOMPATIBLE OPERATING SYSTEM/DEVICE ISSUE WAS REPORTED WITH THE ABBOTT DIABETES CARE (ADC) DEVICE IN USE WITH PIXEL CHROME PRO WITH ANDROID OPERATING SYSTEM VERSION 16. AS A RESULT, THE CUSTOMER WAS UNABLE TO "REGISTER THE SENSOR" AND RECEIVE LOW GLUCOSE ALARMS AND SUBSEQUENTLY RECEIVED A SENSOR ERROR MESSAGE. THE CUSTOMER EXPERIENCED SYMPTOMS DESCRIBED AS LETHARGY, "COMPLETE SATURATION", AND CONFUSION. AFTER THE DEVICE STARTED WORKING, THE CUSTOMER OBTAINED A READING OF 42 MG/DL ON THE ADC DEVICE AND RECEIVED THIRD-PARTY ADMINISTRATION OF COKE AND GLUCOSE TABLETS FOR TREATMENT. THE CUSTOMER DID NOT NOTICE A TREND ARROW ON THE DEVICE. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: Confusion/ Disorientation; Hypoglycemia

    MAUDE report 23044178 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K233537. One submission often covers many models, so check the product and lot details in each record.

  • Z-1022-2026 · Nov 24, 2025 · Open, Classified
    FreeStyle Libre 3 Plus Sensor REF 78769-01 UDI-DI code: 00357599843014 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring Syst
  • Z-1021-2026 · Nov 24, 2025 · Open, Classified
    FreeStyle Libre 3 Plus Sensor REF 78768-01 UDI-DI code: 00357599844011 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring Syst

Other Integrated Continuous Glucose Monitoring system, factory calibrated devices

All Integrated Continuous Glucose Monitoring system, factory calibrated devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page