FreeStyle Libre 3 Plus 78769 (UDI 00357599843007)Recalls & reported problems

Abbott Diabetes Care Inc · FreeStyle Libre 3 PLUS Sensor NFRS · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This is an Abbott FreeStyle Libre 3 Plus sensor, a small factory-calibrated wearable sensor that reads glucose through the skin without a fingerstick. Most of the 34 reports describe the sensor giving inaccurate or unreliable readings, including some readings far off from other meters, plus a few low readings. Three injuries and no deaths were reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
44
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 44 adverse event reports that list this exact device (DI 00357599843007), received between Apr 2, 2025 and Sep 29, 2026 — classified as 3 injuries, 41 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (39 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 3 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 4 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 5 reportsAug 2026: 12 reportsSep 2026: 10 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings39
  • Low Readings4
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions41
  • Hyperglycemia1
  • Insufficient Information1
  • Purulent Discharge1
  • Skin Infection1

Recent adverse event reports

The 25 most recent of 44 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 69.00 mg/dL compared to readings of 202.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 14 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932005 ↗
  • MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 44.00 mg/dL compared to readings of 407.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the D Zone, showing the difference in values to be clinically significant. The length of sensor wear was 12 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911850 ↗
  • MalfunctionSep 21, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 42.00 mg/dL compared to readings of 145.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50839422 ↗
  • MalfunctionSep 15, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 353.00 mg/dL compared to readings of 170.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 11 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50772299 ↗
  • MalfunctionSep 15, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 53 mg/dL compared to readings of 351 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the D Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50770323 ↗
  • MalfunctionSep 15, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 60 mg/dL compared to readings of 300 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50769555 ↗
  • MalfunctionSep 15, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50769076 ↗
  • MalfunctionSep 14, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 51.00 mg/dL compared to readings of 200.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was unknown. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50750671 ↗
  • MalfunctionSep 11, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 50 mg/dL compared to readings of 300 mg/dL respectively obtained on a Competitor Brand Meter and the results, when plotted on a Parkes Error Grid, fall into the D Zone, showing the difference in values to be clinically significant. The length of sensor wear was 10 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50730585 ↗
  • MalfunctionSep 11, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 57.00 mg/dL compared to readings of 192.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50728320 ↗
  • MalfunctionAug 31, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50534079 ↗
  • MalfunctionAug 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50498741 ↗
  • MalfunctionAug 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50498614 ↗
  • MalfunctionAug 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50498601 ↗
  • MalfunctionAug 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50497183 ↗
  • MalfunctionAug 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50496695 ↗
  • MalfunctionAug 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50490931 ↗
  • MalfunctionAug 27, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 300.00 mg/dL compared to glucose values of 120.00 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was an unknown number of days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50486185 ↗
  • MalfunctionAug 27, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 250.00 mg/dL compared to glucose values of 56.00 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the D Zone, showing the difference in values to be clinically significant. The length of sensor wear was 5 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50469765 ↗
  • MalfunctionAug 17, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 350.00 mg/dL compared to readings of 160.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 5 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50276880 ↗
  • MalfunctionAug 17, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 385 mg/dL compared to readings of 145 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 13 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50275488 ↗
  • InjuryAug 3, 2026Adverse Event Without Identified Device or Use Problem

    An adverse skin reaction was reported with wear of the Abbott Diabetes Care (ADC) device. The customer experienced symptoms described as white pus and self-treated the affected area with triple antibiotic ointment, Betadine, and hydrogen peroxide. There was no report of death or permanent impairment associated with this event.

    Patient problems: Purulent Discharge; Skin Infection

    MAUDE report 50111958 ↗
  • MalfunctionJul 31, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 179.00 mg/dL compared to readings of 97.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 11 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50098322 ↗
  • MalfunctionJul 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50083469 ↗
  • MalfunctionJul 15, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE CUSTOMER RECEIVED SENSOR SCAN RESULTS OF 75.00 MG/DL COMPARED TO READINGS OF 192.00 MG/DL RESPECTIVELY OBTAINED ON A METER AND THE RESULTS, WHEN PLOTTED ON A PARKES ERROR GRID, FALL INTO THE C ZONE, SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THE LENGTH OF SENSOR WEAR WAS UNKNOWN. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25825653 ↗